(322 days)
The Discovery MR750w 3.0T is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750w 3.0T reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The Discovery MR750w 3.0T features a superconducting magnet operating at 3.0 Tesla. The data acquisition system accommodates up to 32 independent receive channels in various increments, and multiple independent coil elements per channel during a single acquisition series. The system uses a combination of time-varying magnetic fields (gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The RF technology of the Discovery MR750w system integrates an RF transmit architecture designed to improve the overall image uniformity. This technology, called Multi-drive, optimizes RF transmit by adjusting the amplitude and phase of the RF output depending on the anatomy being scanned. In order to support Multi-Drive, the RF Transmit (Tx) chain is changed from MR750 and both Tx lines are divided into 2 lines with Dual output Exciter, Dual output RF amp, Dual Transmit/Receive Switch (DTRSW), dual UPM and a 70cm-wide patient bore RF body coil. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences and reconstruction algorithms. The Discovery MR750w 3.0T is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).
Here's a breakdown of the requested information based on the provided 510(k) summary for the Discovery MR750w 3.0T.
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) summary for the Discovery MR750w 3.0T primarily focuses on demonstrating substantial equivalence to predicate devices and adherence to established standards for safety and performance testing for Magnetic Resonance Imaging (MRI) systems. It does not provide specific numerical acceptance criteria with corresponding reported performance values in a direct table format. Instead, it states that various parameters "have been measured and documented through testing to NEMA, IEC or ISO standards" and that these tests were "executed with acceptable results."
Performance Parameters Tested:
| Acceptance Criteria (Implied by NEMA, IEC, ISO Standards) | Reported Device Performance (Stated as "Acceptable Results") |
|---|---|
| Signal-to-noise ratio (SNR) | Met standards |
| Geometric distortion | Met standards |
| Image uniformity | Met standards |
| Slice thickness | Met standards |
| Spatial resolution | Met standards |
Safety Parameters Tested:
| Acceptance Criteria (Implied by NEMA, IEC, ISO Standards) | Reported Device Performance (Stated as "Acceptable Results") |
|---|---|
| Static field strength | Met standards |
| Acoustic noise | Met standards |
| dB/dt | Met standards |
| RF heating (SAR) | Met standards |
| Biocompatibility | Met standards |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a specific sample size for a "test set" in the context of clinical studies where individual cases are evaluated for device performance against a ground truth. Instead, it refers to "clinical images and clinical results summary" used to demonstrate imaging performance.
- Sample Size: Not explicitly stated. The summary refers to "clinical images and clinical results summary" but does not quantify the number of patients or images.
- Data Provenance: The manufacturing entity is GE Healthcare, (GE Healthcare Japan Corporation) and the contact person is in Japan. However, the document does not specify the country of origin of the clinical data (e.g., patient demographics, where the scans were acquired). It does not explicitly state whether the data was retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not detail the process of establishing ground truth for a clinical test set in the way one might expect for an AI/CAD device. It states, "These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis." This implies that the interpretation by trained physicians (plural) constitutes the clinical reference standard or ground truth.
- Number of Experts: The document does not specify the number of experts. It generally refers to interpretation by "a trained physician" (singular in the Indications for Use, but "physicians" is implied for general clinical practice).
- Qualifications of Experts: The document only states "trained physician." No specific qualifications (e.g., specialty, years of experience) are provided.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for a test set. This type of detail is typically found in studies for AI-powered diagnostic devices, where disagreements among readers about ground truth or device performance are resolved. For this MRI system, the primary focus is on technical performance and equivalence to predicates, rather than a diagnostic performance study with specific adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described or reported in this 510(k) summary. The document does not discuss human reader performance, with or without AI assistance. The summary's focus is on the device's technical performance and safety, and its substantial equivalence to predicate MRI systems.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The Discovery MR750w 3.0T is a Magnetic Resonance Imaging System, not an algorithm or AI-powered diagnostic tool in the typical sense that would have "standalone performance" evaluated for diagnostic accuracy. It's a hardware system that generates images. Therefore, the concept of "standalone performance" as it applies to an algorithm without human-in-the-loop is not relevant here. The device's performance is inherently tied to the quality of the images it produces, which are then interpreted by a human physician.
7. The Type of Ground Truth Used
The ground truth for clinical evaluation is implicitly based on expert consensus/interpretation by trained physicians using the images produced by the device. The document states that "images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis." This suggests the clinical utility is assessed through conventional medical diagnostic workflows. There is no mention of pathology, long-term outcomes data, or other objective ground truth methods.
8. The Sample Size for the Training Set
The concept of a "training set" is not applicable in the context of this 510(k) submission. The Discovery MR750w 3.0T is an MRI system, a hardware device, not a machine learning or AI algorithm that requires a training set of data. Its design and performance are based on engineering principles, physics, and established medical imaging standards, rather than learned patterns from a dataset.
9. How the Ground Truth for the Training Set Was Established
Since the device is an MRI system and not an AI algorithm, there is no "training set" or corresponding ground truth establishment process in the machine learning sense. The device's operational parameters and image quality are validated against physical phantoms, engineering specifications, and established industry standards (NEMA, IEC, ISO).
{0}------------------------------------------------
K103327
p1/3
Image /page/0/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a circular design with the letters "GE" in a stylized font in the center. The letters are surrounded by a circular border with four decorative swirls or flourishes evenly spaced around the circle. The logo is black and white.
Discovery MR750w 3.0T 510(k) Premarket Notification
SEP 3 0 2011
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | September 30, 2011 |
|---|---|
| Submitter: | GE Healthcare, (GE Healthcare Japan Corporation)7-127, Asahigaoka 4-chome, Hino-shi,Tokyo 191-8503 JAPAN |
| Primary Contact Person: | Toru ShimizuRegulatory Affairs SpecialistGE Healthcare, (GE Healthcare Japan Corporation)Telephone: +81-42-585-5344Fax: +81-42-585-5075 |
| Secondary ContactPerson: | Glen SabinRegulatory Affairs DirectorGE Healthcare, (GE Medical Systems, LLC)Telephone: (262) 521-6848Fax: (262) 521-6439 |
| Device: Trade Name: | Discovery MR750w 3.0T |
| Common/Usual Name: | Magnetic Resonance Imaging System |
| Classification Names:Product Code: | Magnetic resonance diagnostic deviceLNH |
| Predicate Device(s): | Discovery MR750 (K081028)Optima MR450w (K091536) |
| Device Description: | The Discovery MR750w 3.0T features a superconductingmagnet operating at 3.0 Tesla. The data acquisition systemaccommodates up to 32 independent receive channels invarious increments, and multiple independent coil elementsper channel during a single acquisition series. The system usesa combination of time-varying magnetic fields (gradients) andRF transmissions to obtain information regarding the densityand position of elements exhibiting magnetic resonance.The RF technology of the Discovery MR750w system integratesan RF transmit architecture designed to improve the overallimage uniformity. This technology, called Multi-drive, optimizesRF transmit by adjusting the amplitude and phase of the RFoutput depending on the anatomy being scanned. In order tosupport Multi-Drive, the RF Transmit (Tx) chain is changed from |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. There are four decorative teardrop shapes placed symmetrically around the circle's perimeter. The logo is black and white.
Discovery MR750w 3.0T 510(k) Premarket Notification KI03327
p2/3
MR750 and both Tx lines are divided into 2 lines with Dual output Exciter, Dual output RF amp, Dual Transmit/Receive Switch (DTRSW), dual UPM and a 70cm-wide patient bore RF body coil.
The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences and reconstruction algorithms. The Discovery MR750w 3.0T is desianed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).
The Discovery MR750w 3.0T is a whole body magnetic Intended Use: resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750w 3.0T reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Technology: The Discovery MR750w 3.0T employs the same fundamental scientific technology as its predicate devices of Discovery MR750 and Optima MR450w. Refer to Section 12 for details of the Technical Comparison Table and the Application/Feature Comparison Chart.
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
As stated in the FDA document "Guidance for the Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices" the following parameters have been measured and documented through testing to NEMA, IEC or ISO standards as referenced throughout this submission and listed in Section 9:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script. The letters are enclosed within a circle, and there are three decorative flourishes around the circle's perimeter.
Discovery MR750w 3.0T 510(k) Premarket Notification
Performance:
- Signal-to-noise ratio (SNR) .
- Geometric distortion .
- Image uniformity .
- Slice thickness .
- Spatial resolution-.
Safetv:
- Static field strength .
- Acoustic noise .
- dB/dt .
- . RF heating (SAR)
- Biocompatibility .
The tests outlined above have been executed with acceptable results. Refer to Section 15, 18 of this submission for the above performance and satety testing results.
The Discovery MR750w 3.0T has been designed to comply with applicable IEC standards as reference to Section 9, 17.
The device has been tested by a Nationally Recognized Testing Laboratory and certified to conform to applicable IEC, UL and CSA standards prior to commercialization of the system.
Numerical simulations were conducted to demonstrate the safety of the Multi-Drive RF transmit system.
The following quality assurance measures were applied to the development of the system as reference to Section 11, 16, 18:
- Risk Analysis and control �
- . Requirements Reviews
- Design Reviews .
・
- . Design Verification
- . Performance and Safety testing (Verification)
Summary of Clinical Tests:
Clinical images and clinical results summary demonstrate that the Discovery MR750w 3.0T maintains the same imaging performance results as the predicate systems of Discovery MR750, Optima MR450w. Refer to Section 20 for details of the studies performed.
Conclusion: GE Healthcare considers the Discovery MR750w 3.0T to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three wing-like shapes, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle image.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Toru Shimizu Regulatory Affairs Specialist GE Healthcare Japan Corporation 7-127, Asahigaoka 4-Chrome Hino-Shi, Tokyo, 191-8503 JAPAN
SEP 3 0 2011
Re: K103327
Trade/Device Name: Discovery MR750w 3.0T System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, LNI and MOS Dated: September 2, 2011 Received: September 7, 2011
Dear Mr. Shimizu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 5.10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name:
Discovery MR750w 3.0T
Indications for Use:
The Discovery MR750w 3.0T is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750w 3.0T reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Spatzl
Division Sign Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)
N/A