K Number
K103193
Device Name
DOSEVIEW 3D
Date Cleared
2010-12-27

(59 days)

Product Code
Regulation Number
892.5050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DoseView 3D system is a 3 axis, water phantom scanning system intended to easily and accurately, measure and analyze pulsed photon and electron radiation from a linear acceleratedbased radiation therapy system or continuous radiation from a radioactive source-based radiotherapy system. This data is then often used in support of a radiation treatment planning system. It is comprised of a water tank, electrometer, radiation detector(s), motion controller, controlling software, master pendant and lift/reservoir cart. It is a prescription device intended to be used by a trained medical physicist. The general uses of the DoseView 3D include the following:

  • Acceptance testing and/or commissioning of a radiation therapy or radiotherapy system.
  • Quality assurance measurements to identify and minimize the sources of uncertainty and error in the radiation therapy system, radiotherapy system or radiation treatment planning system.
  • Collection of dose depth data for radiation treatment planning system use.
  • Completion of clinical dosimetry protocols and calibrations.
Device Description

The DoseView 3D system is a 3-axis, water phantom scanning system intended to easily and accurately, measure and analyze pulsed photon and electron radiation from a linear accelerated based radiation therapy system or continuous radiation from a radioactive source-based radiotherapy system. This data is then often used in support of a radiation treatment plasot system. It is comprised of a water tank, electrometer, radiation detector(s), motion controller, controlling software, master pendant and lift/reservoir cart. It is a prescription device intended to be used by a trained medical physicist.

AI/ML Overview

The DoseView 3D is a system designed for measuring and analyzing radiation from linear accelerators and radioactive sources, primarily used in radiation therapy for acceptance testing, quality assurance, dose data collection, and calibration.

Here's an analysis of the provided text regarding its acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly detail a table of acceptance criteria with specific numerical targets. Instead, it states that:

Acceptance Criteria CategoryReported Device Performance
Design Specifications"deemed to meet its predetermined design specifications"
Risk Analysis"deemed to meet its predetermined risk analysis"
Validation Objectives"deemed to meet its predetermined validation objectives"

It implies that the device successfully met the intended design, risk, and validation objectives.

2. Sample Size Used for the Test Set and Data Provenance

The text does not specify a numerical "sample size" in the traditional statistical sense for a test set. Instead, it mentions that the DoseView 3D was "successfully evaluated by the following clinical beta sites":

  • Turville Bay MRI & Radiation Oncology Center, Madison, WI
  • UW Hospitals & Clinics, Madison, WI
  • ATC/Tokyo Metropolitan University, Tokyo, Japan

This indicates a prospective evaluation across three clinical sites located in the United States (Wisconsin) and Japan. The "test set" in this context would likely refer to the data collected and performance observed during its use at these sites.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states that the device "is intended to be used by a trained medical physicist." Therefore, it can be inferred that medical physicists at the clinical beta sites were the users and implicitly, the experts involved in establishing or validating the "ground truth" of the device's performance.

The number of experts or their specific qualifications (e.g., years of experience) is not explicitly stated in the provided text.

4. Adjudication Method for the Test Set

The text does not describe any specific adjudication method (e.g., 2+1, 3+1) for the test set. The evaluation seems to be based on the successful performance observed at the clinical beta sites.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, the provided text does not indicate that a multi-reader multi-case (MRMC) comparative effectiveness study was performed. The focus is on the DoseView 3D's standalone performance and its compliance with standards and predeteremined specifications.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone evaluation was performed. The text indicates that the device was "verified and validated at Standard Imaging" and "successfully evaluated by the following clinical beta sites." This implies that the device's performance was assessed directly, without necessarily comparing it to human performance in a comparative study format. The primary function of the DoseView 3D is to make measurements, so its "standalone" performance refers to its accuracy and reliability in those measurements.

7. The Type of Ground Truth Used

The ground truth used for evaluating the DoseView 3D would likely be based on physical dosimetry standards and established clinical dosimetry protocols and calibrations. The device itself is designed to collect dose depth data and assist in calibration. Therefore, the "ground truth" would be the known and expected dose distributions and measurements based on established physics principles and measurement techniques, against which the DoseView 3D's readings were compared. The text explicitly mentions "Completion of clinical dosimetry protocols and calibrations" as one of its general uses, suggesting that its accuracy against these known standards would be the basis of "ground truth."

8. The Sample Size for the Training Set

The provided text does not mention a training set or its sample size. The DoseView 3D is a measurement system, not a machine learning algorithm that typically requires a training set. Its development and validation would involve engineering verification and calibration rather than machine learning training.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a training set, there is no information on how its ground truth was established.

{0}------------------------------------------------

Image /page/0/Picture/11 description: The image shows the logo for Standard Imaging. The logo consists of the word "STANDARDIMAGING" in a stylized font, with a small dot above the "i". To the right of the word is a circular graphic with a lightning bolt design inside. The logo is black and white.

K103193 Subject: 510(k) Summary of Safety and Effectiveness Information for the Standard Imaging DoseView 3D Standard Imaging DoseView 3D Proprietary Name: DEC 2 7 2010 Device Name Classification(s): Primary: Medical Charged-Particle Radiation Therapy System (accessory to) Secondary: Radiologic Quality Assurance Instrument, Radiation Therapy Simulation System Device Panel: Radiology Device Classification(s): Primary: Class II - 21CFR892.5050 - IYE Secondary: Class II - 21CFR892.1940 - LHO Class II - 21CFR892.5840 - KPQ Predicate Devices: Computerized Medical Systems, Inc., DynaScan Radiation Beam Data Primary: Acquisition System -510(k) K854880 PTW - New York Corp., MP3 Automatic Water Phantom -Secondary: 510(k) Number K954165 Contact Person: Raymond Riddle, PE, RAC, Chief Regulatory Officer

The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.

The DoseView 3D system is a 3-axis, water phantom scanning system intended to easily and accurately, measure and analyze pulsed photon and electron radiation from a linear accelerated based radiation therapy system or continuous radiation from a radioactive source-based radiotherapy system. This data is then often used in support of a radiation treatment plasot system. It is comprised of a water tank, electrometer, radiation detector(s), motion controller, controlling software, master pendant and lift/reservoir cart. It is a prescription device intended to be used by a trained medical physicist. The general uses of the DoseView 3D include:

  • Acceptance testing and/or commissioning of a radiation therapy or . radiotherapy system.
  • Quality assurance measurements to identify and minimize the sources . of uncertainty and error in the radiation therapy system, radiotherapy system or radiation treatment planning system.
  • Collection of dose depth data for radiation treatment planning system use. .
  • Completion of clinical dosimetry protocols and calibrations.

... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ..

STANDARDING, INC. 3120 Dening Way, Midleton, W 53562-1461 600-261-0025 HH 608-831-202 FAX www.standardimaging.com

ADVANCING RADIATION QA"

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Standard Imaging. The logo consists of the words "STANDARD IMAGING" in a stylized font, with the word "STANDARD" in a lighter weight font than "IMAGING". To the right of the words is a circular graphic with a stylized sunburst design inside.

Standard Imaging acquired the product technology, controlling software and labeling for the Dose View 3D System from Computerized Medical Systems, Inc. Previously, Computerized Medical Systems had successfully distributed their DynaScan Radiation Beam Data Acquisition System and further supported their system until the company was acquired by Elekta A. Thus, the new Standard Imaging DoseView 3D is the logical update and evolution of the original Computerized Medical Systems DynaScan Radiation Beam Data Acquisition System. That system was cleared by FDA via a 510(k) premarket notification.

The Standard Imaging DoseView 3D was designed to comply with the applicable portions of the following voluntary product related standards:

  • ISO 13485: Quality Management Systems .
  • ISO 14971: Risk Management
  • IEC 60601-1, plus amendments: Medical Electrical Equipment
  • IEC 60601-1-2, plus amendments: EMC/EMI �
  • IEC 60601-1-4: Programmable Systems ●
  • IEC 60731: Dosimeters with Ionization Chambers as used in Radiotherapy .
  • IEC 61217: Radiotherapy Equipment Coordinates, Movements and Scales .
  • EN 980: Symbols ●
  • . EN 1041: Manuals

The Standard Imaging DoseView 3D has been verified and validated at Standard Imaging. These activities addressed software testing, installation testing, acceptance testing maglify assurance testing, data collection, calibrations and shipment testing related to thing quant Additionally, the DoseView 3D was successfully evaluated by the following clinical heta sites:

  • Turville Bay MRI & Radiation Oncology Center, Madison, WI .
  • · UW Hospitals & Clinics, Madison, WI
  • ATC/Tokyo Metropolitan University, Tokyo, Japan �

The Standard Imaging DoseView 3D has been deemed to meet its predetermined design specifications, risk analysis and validation objectives.

STANDARDING, INC. 3120 Deming Way, Middleton, W153562-1461 600-261-0025 HH 608-831-202 FM 608-831-202 FM www.standardina.goomag.com

Image /page/1/Picture/18 description: The image shows the words "ADVANCING RADIATION QA" in all caps. The words are white and the background is black. The letters are bolded and sans-serif.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the body and wings of the bird, and two curved lines that form the legs and feet.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Raymond T. Riddle, PE, RAC Chief Regulatory Officer Standard Imaging, Inc. 3120 Deming Way MIDDLETON WI 53562-1461

0EC 2 7 2910

Re: K103193

Trade/Device Name: DoseView 3D Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy device Regulatory Class: II Product Code: IYE Dated: October 27, 2010 Received: October 29, 2010

Dear Mr. Riddle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally remarketed predicate device results in a classification for your device and thus, perceits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premy aref notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

signature

David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): | <103193

Device Name:

DoseView 3D

DEC 2 7 2010

Indications For Use:

The DoseView 3D system is a 3 axis, water phantom scanning system intended to easily and accurately, measure and analyze pulsed photon and electron radiation from a linear acceleratedbased radiation therapy system or continuous radiation from a radioactive source-based radiotherapy system. This data is then often used in support of a radiation treatment planning system. It is comprised of a water tank, electrometer, radiation detector(s), motion controller, controlling software, master pendant and lift/reservoir cart. It is a prescription device intended to be used by a trained medical physicist. The general uses of the DoseView 3D include the following:

  • Acceptance testing and/or commissioning of a radiation therapy or radiotherapy system. ◆
  • Quality assurance measurements to identify and minimize the sources of uncertainty and . . error in the radiation therapy system, radiotherapy system or radiation treatment planning system.
  • Collection of dose depth data for radiation treatment planning system use. ●
  • . Completion of clinical dosimetry protocols and calibrations.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

and/or

Over-the-Counter-Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)


(Division Sign-Off)

Division of Radiological D Office of In Vitro Diagnostic Device Evaluat

610K K103/93

Page 1 of 1

12

ﻟﺴﻌ

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.