K Number
K103059
Device Name
TROPHON
Manufacturer
Date Cleared
2011-02-24

(132 days)

Product Code
Regulation Number
892.1570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the Trophon Disinfectant which is intended to be used exclusively with the Trophon EPR device.

The Trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the Trophon EPR for the High-Level Disinfection of ultrasound transducers.

The Trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.

The Trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.

The Trophon Disinfectant should be used with the following contact conditions:

Minimum Operational Cycle Time: 7 minutes
Minimum Concentration: 31.5%
Minimum Disinfectant Dose: 1.0 g
Minimum Chamber Temperature: 56°C

Device Description

The Nanosonics Trophon EPR is a software controlled device which provides High-Level Disinfection of ultrasound transducers. The device consists of a sealed disinfection chamber and operates in conjunction with a multi-dose cartridge of concentrated hydrogen peroxide disinfectant, supplied as an accessory to the device. Pre-cleaned and dried ultrasound transducers are placed within the Trophon EPR chamber and disinfected by means of an automated disinfection and aeration cycle. The disinfected ultrasound transducer is removed from the chamber and is ready for immediate use.

AI/ML Overview

This document describes the Nanosonics Trophon EPR, a device for high-level disinfection of ultrasound transducers. The information provided focuses on its performance testing and claims of equivalence to predicate devices, rather than a typical AI/ML-driven device study with specific acceptance criteria and detailed statistical analysis often seen in more recent submissions.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantifiable "acceptance criteria" in the format of a target metric (e.g., sensitivity > X%, specificity > Y%). Instead, it describes performance through a qualitative statement of efficacy in achieving high-level disinfection.

Acceptance Criteria (Implied)Reported Device Performance
Achieve High-Level Disinfection"High-Level Disinfection was achieved in all cases."
Effective for various materials/transducers"A range of materials and transducers were exposed to the Trophon EPR disinfection cycle and were shown to be compatible."
Disinfectant stability"Stability testing showed that the disinfectant retained effective concentration for the duration of the labeled shelf life."
Meets predicate device standards for high-level disinfection"The Nanosonics Trophon EPR is substantially equivalent... in that it: - achieves validated High-Level Disinfection;"

2. Sample size used for the test set and the data provenance

  • Sample size for test set: The document states "Ultrasound transducers used in a clinical setting were disinfected with the Trophon EPR." However, it does not specify the exact number of transducers or the number of disinfection cycles performed in this simulated use environment.
  • Data provenance: The testing was conducted in a "simulated use environment" using "Ultrasound transducers used in a clinical setting." This suggests a retrospective or mixed prospective/retrospective approach, but the text doesn't explicitly state the country of origin. Given Nanosonics Ltd is an Australian company, it's plausible the testing was conducted there or in a region with similar clinical practices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number or qualifications of experts involved in establishing the ground truth. The evaluation of "High-Level Disinfection" likely relies on microbiological testing rather than expert interpretation of images or other subjective data, which would typically involve human experts.

4. Adjudication method for the test set

The document does not describe an adjudication method as would be relevant for a study involving human interpretation or subjective assessments. The "ground truth" for disinfection efficacy is determined by "Bioburden testing," which is an objective laboratory measure.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is a high-level disinfection system, not an AI-driven diagnostic or interpretative tool that would involve human readers. Therefore, the concept of "human readers improving with AI vs. without AI assistance" does not apply here.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The device is a standalone disinfection system. The "performance testing" described ("Potency testing," "High-Level Disinfection efficacy," "materials and transducers... shown to be compatible," "Stability testing") refers to the performance of the device and its disinfectant mechanism itself, without human intervention in the disinfection cycle once initiated. The "software controlled device" operates autonomously to deliver disinfection. So, in this context, yes, a form of standalone performance was assessed as it relates to the device's functional objective.

7. The type of ground truth used

The ground truth used for evaluating disinfection efficacy was microbiological testing (Bioburden testing), which directly measures the reduction of microorganisms. This is an objective measure of "High-Level Disinfection."

8. The sample size for the training set

The document does not mention a training set. This is not an AI/ML device that requires a distinct training dataset for model development. The described testing is performance validation for a physical device.

9. How the ground truth for the training set was established

As there is no training set for an AI/ML model, this question is not applicable. The device's design and operational parameters (e.g., minimum concentration, temperature, cycle time) would have been established through engineering, chemical, and microbiological research and development, not by training on a dataset with established ground truth in the AI/ML sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K.103059" at the top. Below that is the text "nanosonics" with a circular graphic around the "o". To the right of "nanosonics" is the text "p 1 of 3".

510(k) Summary

FEB 2 4 2011

11 1

510(k) Owner:Nanosonics LtdUnit 24 566, Gardeners RdAlexandria, NSW, 2015AustraliaPh: +61 2 8063 1600Fax: +61 2 9317 5010
Contact Person:Mr. Ron WeinbergerGeneral Manager - Innovation and TechnologyNanosonicsUnit 24, 566 Gardeners RoadAlexandria, NSW 2015Ph: +61 2 8063 1600Fax: +61 2 9317 5010
Brand Name:Nanosonics Trophon EPRNanosonics Trophon Disinfectant
Common Name:Hydrogen Peroxide High-Level Disinfection system for ultrasoundtransducers
Classification Name:21 CFR Sec. 892.1570 Diagnostic ultrasonic transducer (accessory to)
Product Code:FEX OUJ
Regulatory Class:II
Predicate Devices:CS Medical TD-100 (K051305)Reckitt & Colman Sporox™ (K970230)
Date Prepared:February 16, 2011

Description of the Device:

The Nanosonics Trophon EPR is a software controlled device which provides High-Level Disinfection of ultrasound transducers. The device consists of a sealed disinfection chamber and operates in conjunction with a multi-dose cartridge of concentrated hydrogen peroxide disinfectant, supplied as an accessory to the device. Pre-cleaned and dried ultrasound transducers are placed within the Trophon EPR chamber and disinfected by means of an

{1}------------------------------------------------

K103059

p2of3

nanosonics

automated disinfection and aeration cycle. The disinfected ultrasound transducer is removed from the chamber and is ready for immediate use.

Indications for Use:

The Trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the Trophon Disinfectant which is intended to be used exclusively with the Trophon EPR device.

The Trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the Trophon EPR for the High-Level Disinfection of ultrasound transducers.

The Trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.

The Trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.

The Trophon Disinfectant should be used with the following contact conditions:

Minimum Operational Cycle Time:7 minute:
Minimum Concentration:31.5%
Minimum Disinfectant Dose:1.0 g
Minimum Chamber Temperature:56°C

Summary of Equivalence:

The Nanosonics Trophon EPR is substantially equivalent to the CS Medical TD-100 (K051305). The Trophon Disinfectant Cartridge is substantially equivalent to Sporox™ (K970230).

Both the Trophon EPR and predicates are intended to be used in general hospital and health care facilities to achieve High-Level Disinfection of ultrasound transducers.

The Nanosonics Trophon EPR device and the K051305 predicate use a validated and controlled automated cycle to deliver measured doses of disinfectant to a chamber which contains the pre-cleaned and dried ultrasound transducer requiring disinfection. Both the EPR and the K051305 predicate device include a means to verify for each cycle, the correct delivery of the disinfectant.

A risk assessment concluded that there were no significant new safety concerns raised by the design of Nanosonics' Trophon EPR and associated disinfectant

Performance Testing:

Potency testing conducted on the Trophon Disinfectant delivered through the Trophon EPR was performed according to relevant AOAC and EPA methods.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "nanosonics" in a stylized font, with a curved line above and below the word. Above the word "nanosonics" is the text "K103059" and "p 3 of 3" in handwriting. The word "nanosonics" is in all lowercase letters.

High-Level Disinfection efficacy of the Trophon EPR was validated in a simulated use environment.

Ultrasound transducers used in a clinical setting were disinfected with the Trophon EPR. Bioburden testing was conducted pre- and post disinfection cycle and High-Level Disinfection was achieved in all cases.

A range of materials and transducers were exposed to the Trophon EPR disinfection cycle and were shown to be compatible.

Stability testing showed that the disinfectant retained effective concentration for the duration of the labeled shelf life.

Conclusion:

The information summarized above demonstrates that Nanosonics' Trophon EPR is substantially equivalent to the predicate devices in that it:

  • achieves validated High-Level Disinfection; and .
  • provides equivalent or improved safety, in that it uses smaller quantities of disinfectant, . produces no significant residues or hazardous wastes, and has no significant adverse effects on ultrasound transducers.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circle around the eagle. The logo is black and white and appears to be a simple, clean design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MI) 20993-0002

Mr. Ron Weinberger General Manager Nanosonics, Limited Unit 24 566, Gardeners Road Alexandria, NSW 2015 AUSTRALIA

FEB 2 4 2011

Re: K103059

Trade/Device Name: Nanosonics Trophon EPR Regulation Number: 21 CFR 892.7570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: II Product Code: OUJ Dated: February 7, 2011 Received: February 9, 2011

Dear Mr. Weinberger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

510(k) Number (if known):

Device Name:

Nanosonics Trophon EPR

Indications for Use:

The Trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the Trophon Disinfectant which is intended to be used exclusively with the Trophon EPR device.

The Trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the Trophon EPR for the High-Level Disinfection of ultrasound transducers.

The Trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.

The Trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.

The Trophon Disinfectant should be used with the following contact conditions:

Minimum Operational Cycle Time:7 minutes
Minimum Concentration:31.5%
Minimum Disinfectant Dose:1.0 g
Minimum Chamber Temperature:56°C

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth T. Clavie-Will

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K103059

§ 892.1570 Diagnostic ultrasonic transducer.

(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.