(34 days)
Not Found
No
The description focuses on standard oscillometric blood pressure measurement and data storage, with no mention of AI or ML terms or functionalities. The WHO classification is a rule-based lookup, not an AI/ML algorithm.
No
This device is a diagnostic device used for measuring blood pressure and pulse rate, not for treating a condition.
Yes
Explanation: The device measures systolic and diastolic blood pressure and pulse rate, providing data that can be used to assess a person's health status, which is a diagnostic function. It also includes a WHO Blood Pressure Classification indicator, further supporting its role in assessing a condition.
No
The device description explicitly states it is a "battery driven automatic non-invasive blood pressure monitor" and uses an "inflatable cuff," indicating it is a hardware device with integrated software, not a software-only medical device.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
- Device Function: The PG-800B Electronic Blood Pressure Monitor measures blood pressure and pulse rate by applying a cuff to the upper arm and using the oscillometric technique. This is a non-invasive measurement taken directly from the body, not from a sample in vitro (in glass or outside the body).
- Intended Use: The intended use clearly states it measures blood pressure and pulse rate of an adult person via a non-invasive technique.
Therefore, based on the provided information, the PG-800B Electronic Blood Pressure Monitor is a non-invasive medical device used for physiological measurement, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
PG-800B Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm. It can be used at medical facilities or at home. The intended arm circumference is 22-32 cm.
Product codes
DXN
Device Description
The proposed device, PG-800B Electronic Blood Pressure Monitor, is a battery driven automatic non-invasive blood pressure monitor. It can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at arm within its claimed range and accuracy via the oscillometric technique. The device has the data storage function. It has an bar indicating function, which can indicate the WHO (World Health Organization) Blood Pressure Classification of the measured blood pressure by referencing Diastolic Blood Pressure issued at Journal of Hypertension 1999. Vol 17, No.2.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
upper arm
Indicated Patient Age Range
adult
Intended User / Care Setting
medical facilities or at home.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance and Clinical Verification Test per ANSI/AAMI SP10:2002+A1:2003+A2:2006
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
10 2988 1/1
ું તેમ
PANGAO®
f
.
Premarket Notification
Traditional Sec. 510(k) Submission
PG-800B Electronic Blood Pressure Monitor
Sec. III 510(k) Summary
NOV - 9 2010
次;
・・・・・
Sec. III 510(k) Summary
r
This 510(k) Summary is prepared per the request of 21 CFR 807.92. The Assigned 510(k) Number is: | |
---|---|
Date | 20 JUL 2010 |
Sponsor | Shenzhen Pango Electronic Co., Ltd |
No.25, 1st Industrial Park, Fenghuang Road, Xikeng Village, Henggang Town, | |
Longgang District, Shenzhen, Guangdong, 518115, China | |
Contact Person: Ms. Xiaoyun Yang, Vice General Manager | |
T: +86-755-33825988 F: +86-755-33825989 E: sales@pan-go.com | |
Submission | MS. Diana Hong / MR. Lee Fu |
Correspondent | Shanghai Mid-Link Consulting Co., Ltd |
P.O.BOX 237-023, Shanghai, 200237, China | |
Proposed Device | Electronic Blood Pressure Monitor, PG-800B |
Classification | System, Measurement, Blood-pressure, Non-invasive DXN 21 CFR 870.1130 Class II |
Intended Use | PG-800B Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood |
pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an | |
inflatable cuff is wrapped around the upper arm. It can be used at medical facilities or at home. The | |
intended arm circumference is 22-32 cm. | |
Device Description | The proposed device, PG-800B Electronic Blood Pressure Monitor, is a battery driven automatic |
non-invasive blood pressure monitor. It can measure systolic and diastolic blood pressure as well as | |
the pulse rate of adult person at arm within its claimed range and accuracy via the oscillometric | |
technique. The device has the data storage function. It has an bar indicating function, which can | |
indicate the WHO (World Health Organization) Blood Pressure Classification of the measured blood | |
pressure by referencing Diastolic Blood Pressure issued at Journal of Hypertension 1999. Vol 17, No.2. | |
Testing | Electric Safety per IEC 60601-1:1988+A1:1991+A2:1995 / EMC per IEC 60601-1-2:2001+A1:2004 |
Performance and Clinical Verification Test per ANSI/AAMI SP10:2002+A1:2003+A2:2006 | |
Biocompatibility per ISO 10993-5:2006 / ISO 10993-10:2002+AMD:2006 | |
Predicate Device | KD-595 Blood Pressure Monitor as cleared in K070828. |
SE Conclusion | The proposed device, PG-800A Electronic Blood Pressure Monitor, measures the blood pressure via |
same principle, oscillometric, at the same site, upper arm, as the predicate device, Both of | |
them are driven by batteries. The proposed device, PG-800B Electronic Blood Pressure Monitor, is | |
claimed to be Substantially Equivalent (SE) to the predicate device, KD595 Blood Pressure Monitor as | |
cleared in K070828, in aspect of safety and effectiveness. |
Shenzhen Pango Electronic Co., Ltd
Report SN: SHA01420100715-02
III-I
ﺔ ﺗﻌﺪﺩ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ
ਾ ਜਾਂਦਾ ਹੈ।
ਬਹੁਤਾਬ
ット「
1
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus. The logo is simple and recognizable, representing the department's role in promoting health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shenzhen Pango Electronic Co., Ltd. c/o Mr. Ned Devine Third Party reviewer Underwriters Laboratories, Inc. 333 Pfingsten Rd. Northbrook, IL 60062
NOV - 9 2010
Re: K102988
Trade/Device Name: Electronic Blood Pressure Monitor, PG-800B Regulation Number: 21 CFR 870.1130 Regulation Name: Non invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: October 22, 2010 Received: October 28, 2010
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
2
Page 2 - Ned Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Dime R.V. Ines
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
3
1610
PANGAO®
Premarket Notification Traditional Sec. 510(k) Submission PG-800B Electronic Blood Pressure Monitor Sec. II Indication for Use Statement
Indication for Use Statement Sec. II
510(k) Number: K102988 Device Name: PG-800B Electronic Blood Pressure Monitor
2010
Indications for Use:
PG-800B Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.
It can be used at medical facilities or at home.
The intended arm circumference is 22-32 cm.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duma b. V. Limes
Page 1 of _
(novascular Devices
510(k) Number K10 2
Shenzhen Pango Electronic Co., Ltd Report SN: SHA01420100715-02
II-1