(101 days)
The PROLYTE is intended to be a direct replacement for the PROLYTE Electrolyte Analyzer (K070104).
The PROLYTE Electrolyte Analyzer is designed for clinical laboratory use by laboratory professionals to assess the levels of Sodium, Potassium, and Chloride found in whole blood, serum, plasma, and urine of patients. The analysis is performed in-vitro, and neither the analyzer nor any of its components come in contact with the patient.
This analyzer is used by laboratory trained technicians in clinical laboratories to aid in the diagnosis and treatment of palients with electrolyte imbalance. These locations routinely conform to CLIA requlations, and conduct daily quality control programs.
The PROLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, and chloride in whole blood, plasma, serum, and pre-diluted urine samples.
The PROLYTE Sodium Assay is intended to measure sodium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
The PROLYTE Potassium Assay is intended to measure potassium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
The PROLYTE Chloride Assay is intended to measure the level of chloride in whole blood, plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
For In Vitro Diagnostic Use
The PROLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium and chloride in serum, plasma, whole blood and pre-diluted urine samples. The PROLYTE analyzer is designed with the user in mind. It is fully automated with simple 'Yes' or 'No' commands for menu navigation. This simple interface ensures that the analyzer be easy to use for quick analysis, (one minute for most samples), but also that the testing of samples can be done by even non-skilled operators with relative ease. The analyzer can be programmed to selfcalibrate using Mission Diagnostics Fluid Pack Na/K/Cl (510(k) 031159) at set intervals or on request. Sodium, potassium and chloride are commonly measured for use in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders. Mission Controls (510(k) 033063) are the recommended quality control material to be used daily.
Here's an analysis of the acceptance criteria and study details for the PROLYTE Electrolyte Analyzer, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the PROLYTE Electrolyte Analyzer are primarily focused on precision (within-run and total imprecision) and linearity. Another critical criterion for this 510(k) submission is substantial equivalence to its predicate device throughout its claimed measuring range, demonstrated through method comparison studies.
Here's the table summarizing these criteria and the reported performance from the provided document:
| Aspect | Analyte & Matrix | Acceptance Criteria (Performance Goal) | Reported Device Performance (Example Values - Note: Actual tables contain more detailed values, representative ones are presented here) | Pass/Fail |
|---|---|---|---|---|
| Precision | ||||
| Within-Run %CV | Serum/Blood: Na+ | ≤ 1% | Whole Blood: 0.60%, 0.58%, 0.91%Plasma: 0.95%, 0.55%, 0.63%Serum: 0.64%, 0.59%, 0.68% (all P) | Pass |
| Within-Run %CV | Serum/Blood: K+ | ≤ 2% | Whole Blood: 0.79%, 0.78%, 1.20%Plasma: 1.50%, 0.85%, 0.46%Serum: 0.66%, 0.70%, 0.85% (all P) | Pass |
| Within-Run %CV | Serum/Blood: Cl- | ≤ 2% | Whole Blood: 0.80%, 0.64%, 0.72%Plasma: 0.75%, 0.57%, 0.47%Serum: 1.47%, 0.67%, 1.25% (all P) | Pass |
| Within-Run %CV | Urine (1:10 dilution): Na+ | ≤ 2.5% | Urine: 0.54%, 0.68%, 0.55% (all P) | Pass |
| Within-Run %CV | Urine (1:10 dilution): K+ | ≤ 2.5% | Urine: 0.91%, 0.49%, 0.84% (all P) | Pass |
| Within-Run %CV | Urine (1:10 dilution): Cl- | ≤ 2.5% | Urine: 1.72%, 0.62%, 0.56% (all P) | Pass |
| Total %CV | Serum/Blood: Na+ | ≤ 2% | Whole Blood: 0.93%, 0.72%, 0.85%Plasma: 0.71%, 0.70%, 0.59%Serum: 0.76%, 0.88%, 1.03% (all P) | Pass |
| Total %CV | Serum/Blood: K+ | ≤ 2.5% | Whole Blood: 1.23%, 0.59%, 2.15%Plasma: 1.73%, 1.25%, 1.48%Serum: 0.91%, 0.80%, 1.19% (all P) | Pass |
| Total %CV | Serum/Blood: Cl- | ≤ 2.5% | Whole Blood: 1.30%, 1.13%, 1.09%Plasma: 1.25%, 0.83%, 0.81%Serum: 1.80%, 0.83%, 0.87% (all P) | Pass |
| Total %CV | Urine (1:10 dilution): Na+ | ≤ 5% | Urine: 3.82%, 1.02%, 1.48% (all P) | Pass |
| Total %CV | Urine (1:10 dilution): K+ | ≤ 5% | Urine: 1.57%, 1.06%, 0.91% (all P) | Pass |
| Total %CV | Urine (1:10 dilution): Cl- | ≤ 5% | Urine: 2.53%, 1.69%, 1.76% (all P) | Pass |
| Linearity | Na+, K+, Cl- across all matrices | R² > ~0.99 (Implicit in strong linear fit required for medical devices) | R² values for all analytes and matrices are very high, ranging from 0.9965 to 0.9997. (All P) | Pass |
| Method Comparison | Na+, K+, Cl- across all matrices | Strong correlation (R² > ~0.98) and slope near 1, intercept near 0, when compared to predicate device within reportable range to demonstrate substantial equivalence. | All R² values are between 0.9844 and 0.9995. Slopes are close to 1, and intercepts are close to 0, indicating strong agreement with the predicate. (All P) | Pass |
2. Sample Size and Data Provenance for the Test Set
- Precision Test Set Sample Sizes:
- Within-Run Imprecision: 30 replicates for each sample (low, mid, high concentrations) per analyte and matrix.
- Total Imprecision: 40 replicates (2 measurements daily for 10 days for serum, plasma, urine; 40 replicates on 2 instruments for whole blood).
- Linearity Test Set Sample Sizes:
- Whole Blood: 42 (Na+), 36 (K+), 42 (Cl-)
- Plasma: 42 (Na+), 54 (K+), 42 (Cl-)
- Serum: 33 (Na+), 33 (K+), 27 (Cl-)
- Urine: 40 (Na+), 36 (K+), 34 (Cl-)
- Method Comparison Test Set Sample Sizes:
- Whole Blood: 110 (Na+), 111 (K+), 108 (Cl-)
- Plasma: 101 (Na+), 104 (K+), 101 (Cl-)
- Serum: 103 (Na+), 100 (K+), 106 (Cl-)
- Urine: 102 (Na+), 105 (K+), 101 (Cl-)
- Data Provenance: The document does not specify the country of origin. It indicates that these are "non-clinical tests" and "clinical tests" referring to method comparisons with patient samples. The samples were either patient samples or spiked/diluted samples to cover the entire measuring range. This implies a prospective collection and testing approach for portions of the method comparison and possibly for precision and linearity.
3. Number of Experts and Qualifications for Ground Truth
- This device is an in vitro diagnostic (IVD) device (analyzing chemical analytes in body fluids) for which the "ground truth" is typically established by reference methods, calibrated standards, or a predicate device, not by human expert interpretation like in imaging.
- For the method comparison studies, the predicate device (original PROLYTE Electrolyte Analyzer K070104) served as the reference for ground truth. Therefore, the "experts" in this context would be the predicate device itself, which has been previously cleared by the FDA based on its own validation.
4. Adjudication Method for the Test Set
- No adjudication method (like 2+1 or 3+1 consensus) is applicable or mentioned. As an IVD device, measurements are quantitative and compared directly to either itself (for precision and linearity) or a predicate device (for method comparison), not to subjective expert interpretations requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done or is applicable. This is an IVD device providing quantitative measurements, not an AI imaging or diagnostic algorithm that assists human readers.
6. Standalone Algorithm Performance
- Yes, the entire study focuses on standalone performance of the PROLYTE Electrolyte Analyzer. The "algorithm" here refers to the instrument's internal measurement and processing capabilities. The non-clinical (precision and linearity) and clinical (method comparison) tests assess the device's inherent performance characteristics directly.
7. Type of Ground Truth Used
- For precision and linearity, the ground truth is established by the known concentrations of the analytes in the control materials and diluted stock solutions. These are internal reference points.
- For method comparison, the ground truth is established by the measurements obtained from the predicate device (original PROLYTE Electrolyte Analyzer K070104). This approach aims to demonstrate substantial equivalence, meaning the new device performs comparably to a device already cleared for market.
8. Sample Size for the Training Set
- The document does not explicitly mention a "training set" in the context of machine learning. The device is an analyzer that uses ion-selective electrodes, and its "software" was rewritten, but its underlying measurement principles are the same as the predicate. The performance studies described are for validation/testing of the implemented system, not for training a model in the machine learning sense. The device is likely calibrated using standardized solutions, which could be considered an ongoing "training" or calibration process, but this is distinct from a one-time dataset for algorithm development.
9. How the Ground Truth for the Training Set Was Established
- As noted above, the concept of a training set as typically understood in machine learning is not directly applicable here. The device functions based on established electrochemical principles. The "calibration" of the device likely uses certified reference materials with known concentrations, ensuring accuracy. This would be part of the device's operational procedure rather than a study-specific training set.
{0}------------------------------------------------
Diamond Diagnostics 510(k) Submission for PROLYTE (New Software)
1. 510(k) Summary- PROL YTE Electrolyte Analyzer
(1) Submitted by:
Diamond Diagnostics 333 Fiske St. Holliston, MA 01746
(2) Contact Person:
Liann Voo, Director, QA and Scientific Development 508-429-0450 (x351) Phone: Fax: 508-429-0452 E-mail: lvoo@diamonddiagnostics.com
(3) Summary Prepared:
January 07, 2011
(4) Device Trade Name:
PROLYTE Electrolyte Analyzer Na*/K*/Cl
(5) Regulatory Information:
| Description | CFR Section | Device Class | Product Code |
|---|---|---|---|
| Sodium Test System | 862.1665 | Class II | JGS |
| Potassium Test System | 862.1600 | Class II | CEM |
| Chloride Test System | 862.1170 | Class II | CGZ |
Panel: Chemistry, 75
(6) Predicate Devices:
| Description | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) | Analytes |
|---|---|---|
| PROLYTE Electrolyte Analyzer | K070104 | Potassium. ChlorideSodium. |
Statement of Technology Characteristics of the Device Compared to Predicate Device:
| Operating Principle | Predicate Device | PROLYTE |
|---|---|---|
| Potentiometric Na+, K+, Cl- Electrodes | K070104 | SAME |
(7) Device Description:
The PROLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium and chloride in serum, plasma, whole blood and pre-diluted urine samples. The PROLYTE analyzer is designed with the user in mind. It is fully automated with simple 'Yes' or 'No' commands for menu navigation. This simple interface ensures that the analyzer be easy to use for quick analysis, (one minute for most samples), but also that the testing of samples can be done by even non-skilled operators with relative ease. The analyzer can be programmed to selfcalibrate using Mission Diagnostics Fluid Pack Na/K/Cl (510(k) 031159) at set intervals or on request. Sodium, potassium and chloride are commonly measured for use in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders. Mission Controls (510(k) 033063) are the recommended quality control material to be used daily.
17 4 2011
{1}------------------------------------------------
(8) Intended Use:
a. Intended use(s):
The PROLYTE is intended to be a direct replacement for the PROLYTE Electrolyte Analyzer (K070104).
The PROLYTE Electrolyte Analyzer is designed for clinical laboratory use by laboratory professionals to assess the levels of Sodium, Potassium, and Chloride found in whole blood, serum, plasma, and urine of patients. The analysis is performed in-vitro, and neither the analyzer nor any of its components come in contact with the patient.
This analyzer is used by laboratory trained technicians in clinical laboratories to aid in the diagnosis and treatment of palients with electrolyte imbalance. These locations routinely conform to CLIA requlations, and conduct daily quality control programs.
b. Indication(s) for use:
The PROLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, and chloride in whole blood, plasma, serum, and pre-diluted urine samples.
The PROLYTE Sodium Assay is intended to measure sodium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
The PROLYTE Potassium Assay is intended to measure potassium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
The PROLYTE Chloride Assay is intended to measure the level of chloride in whole blood, plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
For In Vitro Diagnostic Use
(9) Technological Characteristics of the Device:
Measurement Principles:
The principles of measurement used in the PROLYTE Electrolyte Analyzer are identical to those principles existing in the Predicate electrolyte analyzer K070104 PROLYTE.
The PROLYTE measures sodium, potassium and chloride in whole blood, serum, plasma, and urine, using ion selective electrode technology. The sodium electrode contains a glass tube, specially formulated to be sensitive to sodium ions. The potassium incorporates a neutral carrier ionophore membrane. The chloride contains an ionophore covalently bound to a substrate which is sensitive to negatively charged ions. The potential of each electrode is measured relative to a fixed, stable reference established by a silver/silver chloride electrode in concentrated salt solution. The measured potential varies with the concentration of the ion sensed by the electrode.
Comparison to Predicate Device:
The software for the PROLYTE has been rewritten. However, the functionality has remained primarily unchanged from the Predicate. The major change is in QC storage which allows for 3 control levels instead of 2 to be stored. In addition, the reportable range for the PROLYTE is wider for Na and K in Urine. Blood analysis time is slightly longer in the PROLYTE than the Predicate. Urine analysis time is shorter.
{2}------------------------------------------------
Comparison Table below shows primary similarities and differences.
| SIMILARITIES | ||
|---|---|---|
| Item | Device | Predicate |
| Measurement Method | Ion Selective Electrodes | Same |
| Analytes Measured | Sodium, Potassium, Chloride | Same |
| Sample Matrix | Whole Blood, Plasma, Serum,Urine | Same |
| Calibration | Automatic and On-Demand | Same |
| Measuring Range Whole Blood,Plasma, Serum | Na: 45 - 205 mEq/LK: 1.5 - 11 mEq/LCl: 45 - 205 mEq/L | Same |
| Measuring Range Urine | Cl: 25 - 505 mEq/L | Same |
| DIFFERENCES | ||
|---|---|---|
| Item | Device | Predicate |
| Measuring Range Urine | Na: 25 - 1020 mEq/LK: 10 - 505 mEq/L | Na: 30 - 1020 mEq/LK: 20 - 505 mEq/L |
| Blood Analysis Time | 59 seconds | 57 seconds |
| Urine Analysis Time | 69 seconds | 93 seconds |
| QC Storage | Low, Normal, High, 30 each | Normal, Abnormal, 20 each |
(10) Summary of non-clinical tests:
Precision -
Within run precision was calculated with results from three whole blood, serum, plasma, and urine samples for each analyte. The sample concentrations were at the low and high end of reference ranges and near the mid point range. The protocol called for running 30 replicates of each sample without calibration between measurements. The replicates were run consecutively in one day. Sodium, potassium and chloride all demonstrated precision within the limits defined below.
| Serum/Blood | Urine (1:10 dilution) |
|---|---|
| Na+ C.V. ≤ 1% | C.V. ≤ 2.5% |
| K+ C.V. ≤ 2% | C.V. ≤ 2.5% |
| Cl- C.V. ≤ 2% | C.V. ≤ 2.5% |
Typical Within Run Imprecision Results:
Whole Blood, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 114.27 | 2.678 | 70.79 |
| SD | 0.69 | 0.021 | 0.57 |
| %CV | 0.60 | 0.79 | 0.80 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 136.11 | 4.103 | 96.02 |
| SD | 0.80 | 0.032 | 0.62 |
| %CV | 0.58 | 0.78 | 0.64 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 160.57 | 6.887 | 125.42 |
| SD | 1.47 | 0.083 | 0.91 |
| %CV | 0.91 | 1.20 | 0.72 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
Plasma, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 93.07 | 2.376 | 61.27 |
| SD | 0.89 | 0.036 | 0.46 |
| %CV | 0.95 | 1.50 | 0.75 |
| N | 30 | 30 | 30 |
| PASS/FAIL | P | P | P |
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 142.95 | 4.354 | 103.69 |
| SD | 0.78 | 0.037 | 0.59 |
| %CV | 0.55 | 0.85 | 0.57 |
| N | 32 | 32 | 32 |
| PASS/FAIL | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 165.80 | 6.622 | 123.57 |
| SD | 1.04 | 0.031 | 0.58 |
| %CV | 0.63 | 0.46 | 0.47 |
| N | 30 | 30 | 30 |
| PASS/FAIL | P | P | P |
{3}------------------------------------------------
Serum, mEq/L
| Na+ | K+ | CI- | |
|---|---|---|---|
| Mean | 120.10 | 3.026 | 70.30 |
| SD | 0.77 | 0.020 | 1.03 |
| %CV | 0.64 | 0.66 | 1.47 |
| N | 30 | 30 | 30 |
| PASS/FAIL | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 145.16 | 4.755 | 97.02 |
| SD | 0.86 | 0.034 | 0.65 |
| %CV | 0.59 | 0.70 | 0.67 |
| N | 30 | 30 | 30 |
| PASS/FAIL | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 163.21 | 6.536 | 118.80 |
| SD | 1.12 | 0.056 | 1.48 |
| %CV | 0.68 | 0.85 | 1.25 |
| N | 30 | 30 | 30 |
| PASS/FAIL | P | P | P |
Urine, mEq/L
| Na+ | K+ | CI- | Na+ | K+ | CI- | Na+ | K+ | CI- | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | 33.26 | 31.967 | 52.89 | Mean | 121.83 | 78.548 | 184.34 | Mean | 202.84 | 112.432 | 274.38 |
| SD | 0.18 | 0.290 | 0.91 | SD | 0.83 | 0.384 | 1.15 | SD | 1.11 | 0.945 | 1.52 |
| %CV | 0.54 | 0.91 | 1.72 | %CV | 0.68 | 0.49 | 0.62 | %CV | 0.55 | 0.84 | 0.56 |
| N | 30 | 30 | 30 | N | 30 | 30 | 30 | N | 30 | 30 | 30 |
| PASS/FAIL | P | P | P | PASS/FAIL | P | P | P | PASS/FAIL | P | P | P |
Within Run Imprecision at Instrument Reporting Limits
Whole Blood, mEq/L
| Na+ | K+ | Cl⁻ | |
|---|---|---|---|
| Mean | 53.15 | 1.678 | 47.90 |
| SD | 0.53 | 0.025 | 0.88 |
| %CV | 1.0 | 1.5 | 1.8 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 193.33 | 10.363 | 196.26 |
| SD | 1.48 | 0.056 | 2.00 |
| %CV | 0.76 | 0.54 | 1.0 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
Plasma, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 53.10 | 1.753 | 49.64 |
| SD | 0.65 | 0.025 | 0.83 |
| %CV | 1.23 | 1.41 | 1.68 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 189.21 | 9.6407 | 197.15 |
| SD | 1.10 | 0.0934 | 1.26 |
| %CV | 0.58 | 0.97 | 0.64 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
Serum, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 54.01 | 1.693 | 53.38 |
| SD | 0.46 | 0.024 | 0.88 |
| %CV | 0.85 | 1.41 | 1.65 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 187.37 | 9.914 | 180.09 |
| SD | 0.94 | 0.070 | 1.42 |
| %CV | 0.50 | 0.71 | 0.79 |
| N | 30 | 30 | 30 |
| Pass/Fall | P | P | P |
Urine, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 28.52 | 11.63 | 39.91 |
| SD | 0.56 | 0.14 | 0.99 |
| %CV | 1.95 | 1.17 | 2.47 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 908.9 | 493.1 | 420.9 |
| SD | 6.4 | 3.1 | 1.8 |
| %CV | 0.70 | 0.62 | 0.42 |
| N | 30 | 30 | 30 |
| Pass/Fail | P | P | P |
Total precision was calculated with results from three whole blood, serum, plasma, and urine samples for each analyte. The samples concentrations were at the low and high end of reference ranges and near the mid point range. The serum, plasma and urine samples were measured twice each morning and twice each afternoon for ten consecutive days resulting in 40 replicates. Whole blood total precision was conducted on 2 different instruments with calibrations every consecutive replicates of 10 to total 40 replicates for each sample. Sodium, potassium and chloride all demonstrated total precision within the limits defined below.
{4}------------------------------------------------
| Serum/Blood | Urine (1:10 dilution) | |
|---|---|---|
| Na+ | C.V. ≤ 2% | C.V. ≤ 5% |
| K+ | C.V. ≤ 2.5% | C.V. ≤ 5% |
| Cl- | C.V. ≤ 2.5% | C.V. ≤ 5% |
Typical Total Imprecision Results:
Whole Blood, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 113.69 | 2.649 | 71.24 |
| SD | 1.06 | 0.032 | 0.92 |
| %CV | 0.93 | 1.23 | 1.30 |
| N | 40 | 40 | 40 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 136.36 | 4.103 | 96.51 |
| SD | 0.99 | 0.024 | 1.09 |
| %CV | 0.72 | 0.59 | 1.13 |
| N | 39 | 40 | 39 |
| Pass/Fail | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 161.22 | 6.924 | 126.41 |
| SD | 1.37 | 0.149 | 1.37 |
| %CV | 0.85 | 2.15 | 1.09 |
| N | 40 | 40 | 40 |
| Pass/Fail | P | P | P |
Plasma, mEq/L
| Na+ | K+ | Cl- | Na+ | K+ | Cl- | Na+ | K+ | Cl- | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | 118.98 | 2.902 | 74.01 | Mean | 143.89 | 4.502 | 99.58 | Mean | 160.66 | 6.230 | 121.81 |
| SD | 0.85 | 0.050 | 0.93 | SD | 1.00 | 0.056 | 0.83 | SD | 0.94 | 0.092 | 0.99 |
| %CV | 0.71 | 1.73 | 1.25 | %CV | 0.70 | 1.25 | 0.83 | %CV | 0.59 | 1.48 | 0.81 |
| N | 40 | 40 | 40 | N | 40 | 40 | 40 | N | 40 | 40 | 40 |
| PASS/FAIL | P | P | P | PASS/FAIL | P | P | P | PASS/FAIL | P | P | P |
Serum, mEq/L
| Na+ | K+ | Cl- | Na+ | K+ | Cl- | Na+ | K+ | Cl- | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | 119.04 | 2.970 | 69.02 | Mean | 144.03 | 4.732 | 100.13 | Mean | 160.71 | 6.502 | 119.93 |
| SD | 0.91 | 0.027 | 1.24 | SD | 1.26 | 0.038 | 0.83 | SD | 1.66 | 0.078 | 1.05 |
| %CV | 0.76 | 0.91 | 1.80 | %CV | 0.88 | 0.80 | 0.83 | %CV | 1.03 | 1.19 | 0.87 |
| N | 40 | 40 | 40 | N | 40 | 40 | 40 | N | 40 | 40 | 40 |
| PASS/FAIL | P | P | P | PASS/FAIL | P | P | P | PASS/FAIL | P | P | P |
Urine, mEq/L
| Na+ | K+ | Cl- | |
|---|---|---|---|
| Mean | 39.32 | 25.452 | 109.67 |
| SD | 1.50 | 0.398 | 2.77 |
| %CV | 3.82 | 1.57 | 2.53 |
| N | 40 | 40 | 40 |
| PASS/FAIL | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 116.87 | 79.485 | 183.51 |
| SD | 1.19 | 0.845 | 3.11 |
| %CV | 1.02 | 1.06 | 1.69 |
| N | 40 | 40 | 40 |
| PASS/FAIL | P | P | P |
| Na+ | K+ | Cl- | |
| Mean | 227.06 | 129.180 | 253.35 |
| SD | 3.35 | 1.178 | 4.45 |
| %CV | 1.48 | 0.91 | 1.76 |
| N | 40 | 40 | 40 |
| PASS/FAIL | P | P | P |
Total Imprecision at Instrument Reporting Limits
Whole Blood, mEq/L
| Na+ | K+ | Cl- | Na+ | K+ | Cl- | ||
|---|---|---|---|---|---|---|---|
| Mean | 53.46 | 1.667 | 48.93 | Mean | 195.85 | 10.337 | 190.47 |
| SD | 0.71 | 0.030 | 0.83 | SD | 1.91 | 0.109 | 3.05 |
| %CV | 1.33 | 1.82 | 1.70 | %CV | 0.98 | 1.05 | 1.60 |
| N | 40 | 40 | 40 | N | 40 | 18 | 40 |
| Pass/Fail | P | P | P | Pass/Fail | P | P | P |
Plasma, mEq/L
| Na+ | K+ | Cl- | Na+ | K+ | Cl- | ||
|---|---|---|---|---|---|---|---|
| Mean | 55.43 | 1.7795 | 50.07 | Mean | 189.28 | 9.709 | 198.86 |
| SD | 0.83 | 0.0297 | 0.87 | SD | 1.39 | 0.065 | 1.97 |
| %CV | 1.49 | 1.67 | 1.73 | %CV | 0.74 | 0.67 | 0.99 |
| N | 40 | 40 | 40 | N | 40 | 40 | 40 |
| Pass/Fail | P | P | P | Pass/Fail | P | P | P |
{5}------------------------------------------------
Serum, mEq/L
| Na+ | K+ | Cl- | Na+ | K+ | Cl- | ||
|---|---|---|---|---|---|---|---|
| Mean | 55.26 | 1.700 | 53.56 | Mean | 187.10 | 9.9300 | 181.39 |
| SD | 0.75 | 0.028 | 0.93 | SD | 1.70 | 0.1050 | 1.68 |
| %CV | 1.36 | 1.66 | 1.73 | %CV | 0.91 | 1.06 | 0.93 |
| N | 40 | 40 | 40 | N | 40 | 40 | 40 |
| Pass/Fail | P | P | P | Pass/Fail | P | P | P |
Urine, mEq/L
| Na+ | K+ | Cl- | Na+ | K+ | Cl- | ||
|---|---|---|---|---|---|---|---|
| Mean | 28.67 | 11.48 | 35.52 | Mean | 907.82 | 501.53 | 447.05 |
| SD | 0.66 | 0.17 | 1.01 | SD | 6.26 | 6.06 | 2.89 |
| %CV | 2.30 | 1.44 | 2.83 | %CV | 0.69 | 1.21 | 0.65 |
| N | 40 | 40 | 40 | N | 40 | 40 | 40 |
| Pass/Fail | P | P | P | Pass/Fail | P | P | P |
Linearity
Linearity was evaluated by preparing stock solutions with high concentrations of Na+, K+, and CI- in whole blood, plasma, serum, and urine. These stocks were diluted to concentrations across the measuring ranges of each analyte and matrix. Linear regression was performed on the results using expected values based on the stock sample dilution. The results are shown below.
Whole blood Linearity, Measured compared to Expected Values
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 1.0050 | -3.8 | 0.9993 | 23 - 204 | 42 |
| Potassium | 1.0112 | -0.24 | 0.9994 | 0.7 - 11.6 | 36 |
| Chloride | 1.0005 | 3.4 | 0.9973 | 33 - 211 | 42 |
Plasma Linearity, Measured compared to Expected Values
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 0.9806 | 1.666 | 0.9993 | 30 - 205 | 42 |
| Potassium | 0.999 | 0.0555 | 0.9995 | 0.9 - 11.2 | 54 |
| Chloride | 0.9777 | 1.555 | 0.9985 | 31 - 209 | 42 |
Serum Linearity, Measured Concentration compared to Expected Values
| Parameter | Slope | Intercept | R² | Range | n |
|---|---|---|---|---|---|
| Sodium | 1.0021 | 0.3748 | 0.9997 | 20 - 219 | 33 |
| Potassium | 0.9989 | 0.0184 | 0.9995 | 1.3 - 13.1 | 33 |
| Chloride | 1.0264 | -1.0423 | 0.9965 | 15 - 238 | 27 |
Urine, Measured Concentration compared to Expected Values
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 1.0002 | -2.1049 | 0.9996 | 16 - 1085 | 40 |
| Potassium | 1.0082 | 3.1433 | 0.9991 | 9 - 547 | 36 |
| Chloride | 0.9860 | 1.965 | 0.9991 | 22 - 504 | 34 |
The linearity studies support the following reportable range.
| Measuring Range Whole Blood,Plasma, Serum | Na: 45 - 205 mEq/L |
|---|---|
| K: 1.5 - 11 mEq/L | |
| Cl: 45 - 205 mEq/L | |
| Measuring Range Urine | Na: 20 - 1020 mEq/L |
| K: 10 - 505 mEq/L | |
| Cl: 25 - 505 mEq/L |
{6}------------------------------------------------
(11) Summary of clinical tests submitted with the pre-market notification for the device.
Method comparisons to predicate device were performed with whole blood, plasma, serum and urine patient samples. A small number of samples were spiked or diluted to fully span the claimed measuring ranges. The results are summarized below.
Whole Blood Comparison, New PROLYTE versus Predicate, Original PROLYTE (K070104)
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 1.0054 | -2.9043 | 0.9875 | 48 - 205 | 110 |
| Potassium | 1.0095 | -0.0675 | 0.9891 | 1.59 - 10.2 | 111 |
| Chloride | 0.9987 | -2.1711 | 0.9844 | 45 - 203 | 108 |
Plasma Comparison of New PROLYTE and Predicate, Original PROLYTE (K070104)
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 1.0243 | -4.1003 | 0.9971 | 49 - 205 | 101 |
| Potassium | 0.9875 | 0.1004 | 0.9961 | 1.5 - 11.0 | 104 |
| Chloride | 1.0176 | -7.9389 | 0.9965 | 45 - 204 | 101 |
Serum Comparison, of New PROLYTE and Predicate, Original PROLYTE (K170104)
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 1.0119 | -4.2748 | 0.9936 | 46 - 204 | 103 |
| Potassium | 0.9800 | 0.106 | 0.9950 | 1.5 - 10.8 | 100 |
| Chloride | 1.0144 | -4.8408 | 0.9922 | 49 - 203 | 106 |
Urine Comparison, of New PROLYTE and Predicate, Original PROLYTE (K070104)
| Parameter | Slope | Intercept | R2 | Range | n |
|---|---|---|---|---|---|
| Sodium | 0.9798 | 4.5946 | 0.9995 | 25 - 1020 | 102 |
| Potassium | 0.9879 | 0.0762 | 0.9991 | 11 - 499 | 105 |
| Chloride | 1.0179 | -1.809 | 0.9933 | 26 - 504 | 101 |
The method comparison studies support correlation in the reportable range.
| Measuring Range Whole Blood,Plasma, Serum | Na: 45 - 205 mEq/L |
|---|---|
| K: 1.5 - 11 mEq/L | |
| Cl: 45 - 205 mEq/L | |
| Measuring Range Urine | Na: 25 - 1020 mEq/L |
| K: 10 - 505 mEq/L | |
| Cl: 25 - 505 mEq/L |
(12) Conclusions drawn from the clinical and non-clinical testing.
Analysis of the comparative measurement presented in the 510(k) for this device, together with the linearity and precision data collected during these clinical trails demonstrates that the Diamond Diagnostics PROLYTE Electrolyte Analyzer with new software is safe, effective and substantially equivalent to its predicate device, the original PROLYTE (K070104).
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Diamond Diagnostic, Inc. c/o Liann Voo Regulatory Affairs Specialist 333 Fiske Street Holliston, MA 01746
K102959 Re:
Trade Name: Prolyte Electrolyte Analyzer Regulation Number: 21 CFR §862.1665 Regulation Name: Sodium Test System. Regulatory Class: Class II Product Codes: JGS, CEM, CGZ Dated: December 23, 2010 Received: December 23, 2010
JAN 1 4 201
Dear Liann Voo:
We have reviewed your Section 510(k) premarket notification of intent to market the We liave levice your Sector 310(t) promised the device is substantially equivalent (for device felefenecod above and have actore) to legally marketed predicate devices
the indications for use stated in the enclosure) to legally marketed predicate of the the multations for use stated in the viace in the enactment date of the marketed in mersiale commerce prior to may 2016 to have been reclassified in accordance Medical Device Amendinents, of to devices management Act (Act (Act) that do not require
with the provisions of the Federal Food, Drug, and Cosmetic Act (Accefere, morket the with the provisions of the Federal Food, Bruggest A. You may, therefore, market the approval of a premarket approval upprovalissions of the Act. The general controls device, subject to the general controls pro connual registration, listing of devices, and provisions of the Act merded requirements on prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III II your device is classified (300 additional controls. Existing major regulations (CE (PMA), it may be subject to such additional connects. Encludes Regulations (CFR), Parts
affecting your device can be found in ' like for consements concerners concerning you affecting your device can be round in This 22, 80 and 11 and 11 and 11 and 11 and device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does Please be advised that FDA > Issuance of a sacterming the complies with other not mean that FDA has made a determinations and regulations administered by other
requirements of the Act or any Federal statutes and regulations and not requirements of the Act of ally Proclar statute Act 3 requirements, including, but not
Federal agencies. You must comply with all the Act ' S requirements 801 and Federal agencies. You must compry with and 007); labeling (21 CFR Parts 801 and limited to: registration and ilsting (21 CF it an ocal device-related adverse events) (21
809); medical device reporting (reporting of medical device as sortorial in the qual 809); medical device reporting (cporting or mouter as set forth in the quality
CFR 803); and good manufacturing practice requirements as set forth in the quality
CFR 803); an systems (QS) regulation (21 CFR Part 820).
මෙම ප්රධාන ක්රි.
{8}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollofire number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
4. Indications for Use
510 (k) Number (if known):
Device Name: PROLYTE Electrolyte Analyzer
Indication For Use:
The PROLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potessium, and chloride undryzer which unders in the electived unine samples.
The PROLYTE Sodium Assay is intended to measure sodium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the horme aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydrate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
The PROLYTE Potassium Assay is intended to measure potassium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood odession levels.
The PROLYTE Chloride Assay is intended to measure the level of chloride in whole blood, plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Prescription Use × (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Richard Stone, D.D.S.
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102959
§ 862.1665 Sodium test system.
(a)
Identification. A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.(b)
Classification. Class II.