(142 days)
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphyqmanometers. The operational principle is based on oscillometric and silicon integrates pressure sensor technology. it can calculate the systolic and diastolic blood pressure, the measurements results can also be classified by the function of blood pressure classification indicator. If any irreqular heartbeat is detected, it can be shown to the user. More over, it also obtains the function of averaging the measurement results. iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock achieves its function by integrate the device with an iPhone, ipod or ipad. For it does not contain an LCD or other display components, so It's necessary for the new device to connect to an iPhone, ipod or ipad containing a support software to constitute a complete blood pressure measurement system.
Here's an analysis of the provided text regarding the acceptance criteria and study for the iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the iHealth BP3 are based on compliance with the ANSI/AAMI SP10 standard for manual, electronic, or automated sphygmomanometers. While the document doesn't provide a specific numerical table of acceptance criteria and the device's numerical performance against each, it states a blanket compliance.
| Acceptance Criteria (from ANSI/AAMI SP10) | Reported Device Performance |
|---|---|
| All applicable requirements of ANSI/AAMI SP10 | Device met all applicable requirements of the standard. |
(Note: The document doesn't provide the explicit numerical performance values as would be typical for accuracy studies, but rather asserts compliance with the standard's requirements.)
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated. The document mentions "a new clinical test is done in accordance with ANSI/AAMI SP10," but does not specify the number of subjects or measurements included in this test.
- Data Provenance: Not explicitly stated. The document indicates the submitting company is based in Tianjin, P.R. China, but does not confirm where the clinical test was conducted or if the data was retrospective or prospective. Given it was a "new clinical test," it was likely prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not explicitly stated. The ground truth for blood pressure measurements in accordance with ANSI/AAMI SP10 typically involves simultaneous measurements by trained personnel using a reference sphygmomanometer. However, the number and qualifications of these "experts" (e.g., clinicians, trained observers) are not mentioned in this document.
4. Adjudication Method for the Test Set
Not explicitly stated. For blood pressure clinical trials, adjudication often involves averaging multiple expert readings or using a standard auscultatory method as reference. This document does not detail such a method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an automatic blood pressure monitor, not an AI-assisted diagnostic imaging or analysis tool that would typically involve human readers or MRMC studies. Its primary function is to directly measure blood pressure.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, in principle. The clinical test described is likely a standalone performance validation of the device's algorithm and hardware. The device itself (when connected to an iPhone/iPod/iPad) automatically measures and reports blood pressure; it doesn't require human interpretation of complex data for a diagnosis, but rather provides a direct measurement. The "algorithm" for calculating systolic and diastolic pressure has changed from the predicate device (inflation detection vs. deflation detection), which necessitated this new clinical test.
7. The Type of Ground Truth Used
The ground truth used would have been a reference measurement from a validated sphygmomanometer performed by trained observers, as prescribed by the ANSI/AAMI SP10 standard. This is a form of expert reference measurement, but not "pathology" or "outcomes data" in the typical sense for diagnostic devices.
8. The Sample Size for the Training Set
Not applicable. This device is a measurement system, not a machine learning model that requires a "training set" in the conventional sense. The "arithmetic" (algorithm) for inflation detection would have been developed and refined through engineering and calibration, not through a data-driven training process in the same way an AI algorithm for image recognition would be.
9. How the Ground Truth for the Training Set was Established
Not applicable for the reasons stated above. The "ground truth" for developing the measurement algorithm would come from physical principles of blood pressure measurement and engineering standards, not a labeled dataset for training a statistical model.
{0}------------------------------------------------
510(k) Summary
FEB 2 3 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
1.0 submitter's information
| Name: | Andon Health Co., Ltd. |
|---|---|
| Address: | No 3, Jinping Street Ya An Road, Nankai District,Tianjin, P.R. China |
| Phone number: | 86-22-6052 6161 |
| Fax number: | 86-22-6052 6162 |
| Contact: | Liu Yi |
| Date of Application: | 09/27/2010 |
2.0 Device information
- iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Trade name: Pressure Dock
Classification name: Noninvasive blood pressure measurement system
3.0 Classification
Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: ll Panel: Cardiovascular
4.0 Predicate device information
Andon Health Co., Ltd. Manufacturer: Device: KD-930 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K101950
{1}------------------------------------------------
5.0 Device description
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphyqmanometers.
The operational principle is based on oscillometric and silicon integrates pressure sensor technology. it can calculate the systolic and diastolic blood pressure, the measurements results can also be classified by the function of blood pressure classification indicator. If any irreqular heartbeat is detected, it can be shown to the user. More over, it also obtains the function of averaging the measurement results.
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock achieves its function by integrate the device with an iPhone, ipod or ipad. For it does not contain an LCD or other display components, so It's necessary for the new device to connect to an iPhone, ipod or ipad containing a support software to constitute a complete blood pressure measurement system.
6.0 Intended use
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
The intended use and the indication for use of the iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock, as described in its labeling are the same as the predicate device KD-930.
{2}------------------------------------------------
Pg 3 of 4
7.0 Summary comparing technological characteristics with predicate device
| Technological Characteristics | Comparison result |
|---|---|
| Design principle | Identical |
| Appearance | Similar |
| Patients contact Materials | Identical |
| Performance | Similar |
| Biocompatibility | Identical |
| Mechanical safety | Identical |
| Energy source | Identical |
| Standards met | Identical |
| Electrical safety | Identical |
| EMC | Identical |
| Function | Similar |
8.0 Discussion of non-clinical and clinical test performed
Non-clinical Tests have been done as follows:
a. Electromagnetic compatibility test according to IEC 60601-1-2;
b: Electrical safety according test to IEC 60601-1 and IEC 60601-1-1
c. FCC test according to FCC part 15 (2009)
d. Safety and performance characteristics of the test according to SP10
None of the test demonstrates that iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock brings new questions of safety and effectiveness.
Clinical Test Concerning the Compliance of ANSI/AAMI SP10
Compared to deflation detection of its predicate device KD-930, iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is an inflation detection device, so the arithmetic is changed. As a result, a new clinical test is done in accordance with ANSI/AAMI SP10, and the device met all applicable requirements of the standard.
{3}------------------------------------------------
9.0 Performance summary
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock conforms to the following standards:
- 60601-1, Medical Electrical ● IEC Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- 60601-1, Medical Electrical Equipment Part 1: • UL General Requirements for Safety, 2003.
-
- IEC 60601-1-1, Medical Electrical Equipment Part 1: General Requirements for Safety – 1. Collateral standard: Safety Requirements for Medical Electrical Systems, 2000.
- EN 60601-1-2, Medical Electrical Equipment Part 1-2: General . Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests, 2007.
- AAMI SP10:2002, Manual, electronic or automated sphygmomanometers. e
- AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI . SP10:2002 Manual, electronic, or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
10.0 Comparison to the predicate device and the conclusion
Our device iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-930 whose 510(k) number is K101950.
The two devices are very similar in the intended use, the design principle, the material, the performance and the applicable standards. Only their appearance, the memory time, and the user interface are different. The measure process is also changed, that is the new device will get the measurement results when the device is inflating, while KD-930 gets the result during the deflating period. What's more, iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock can achieve its function with an iphone, ipod or ipad, while KD-930 can only connect an iphone to achieve its function.
However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. To the left of the bird symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Andon Health Co., Ltd. C/O Mr. Liu Yi, President No. 3 Jinping Street Ya'an Road Nankai District, Tianjin 300190 China
FEB 2 3 2011
Re: K102939
Trade Name: iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: Not Dated Received: February 15, 2011
Dear Mr. Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Mr. Liu Yi
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
O. Macf.
CBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Pressure Dock
Statement of Indications for Use
| 510(k) Number : | K102939 |
|---|---|
| Device name: | iHealth BP3 Fully Automatic Arm Cuff Electronic Blood |
Indications for use:
iHealth BP3 Fully Automatic Arm Cuff Electronic Blood Pressure Dock is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Prescription use Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use YES (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number (K102939)
Page I of I
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).