(402 days)
The NexGen Peripheral Expandable Catheter is indicated for percutaneous access to the perioheral vascular system and is designed to assist in the placement and removal of devices. This device is not intended for use in the coronary or cerebral vasculature.
The NexGen Peripheral Expandable Catheter is a sterile single use catheter intended to facilitate the placement or removal of vascular catheters and devices including the NexGen Mechanical Retrieval Device (MRD).
The Device features a Pebax/HDPE catheter shaft attached Nitinol tip that expands as the device is deployed from the lumen of a guide catheter. The Nitinol braid passively expands to the contour shape of the vessel lumen and allows a separate medical device such as a catheter to be retracted into the Nitinol Braid. Both the Nitinol Braid and the other device are then retracted into the Guide Catheter. The retracted Expandable Catheter with the encapsulated device are then removed together via removal of the Guide Catheter.
The Expandable Catheter comes in multiple expanded profiles of 5mm, 7mm, 9mm and 11mm, each having a working length of 43cm and 116cm. The Expandable Catheter is packaged in a Tear Away Sheath (B. Braun 510(k) K000313) that is used to facilitate loading the catheter into a Guide Catheter.
Proximally located on the catheter shaft is a female luer connection for the attachment to a Hemostasis Valve Y Connector, which provides a means of introducing contrast media or other fluids. The Hemostasis Valve Y Connector is separate or a prepackaged component.
The provided text describes a medical device, the NexGen Peripheral Expandable Catheter, and its 510(k) submission to the FDA. However, the document does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria in the way typically applied to AI/ML medical devices.
Instead, this document details the regulatory clearance process for a physical medical device (catheter) based on its substantial equivalence to predicate devices. The "testing" mentioned refers to bench-top non-clinical tests to verify the device met design specifications and functional/structural integrity, not clinical performance acceptance criteria in the context of diagnostic accuracy or reader improvement that would be relevant for an AI study.
Therefore, many of the requested categories for AI/ML device studies (like sample sizes, ground truth establishment, expert qualifications, MRMC studies) are not applicable or present in this document.
Below is a table summarizing the information available in the provided text, while explicitly noting where the requested AI-relevant information is not present.
Acceptance Criteria and Device Performance (as presented for a physical medical device)
| Acceptance Criteria Category (for physical device) | Reported Device Performance / Evaluation Method |
|---|---|
| Visual / Dimension Inspection | Met design specifications |
| Catheter Pushability / Trackability Testing | Met design specifications |
| Catheter Tensile Testing | Met design specifications |
| Catheter Torque Testing | Met design specifications |
| Catheter Burst Testing | Met design specifications |
| Catheter Leak Testing | Met design specifications |
| Catheter Kink Diameter Testing | Met design specifications |
| Catheter Flow Rate Testing | Met design specifications |
| Catheter Radial Force Testing | Met design specifications |
| Catheter Corrosion Resistance Testing | Met design specifications |
| Biocompatibility Tests | Demonstrated material safety |
| Functional Safety | Demonstrated functional safety |
| Structural Integrity | Demonstrated structural integrity |
| Equivalence to Predicate Devices (K963886 & K070865) | Found "substantially equivalent" by FDA |
Study Details (as applicable to a physical medical device submission, not AI)
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable / Not Provided for AI-style "test set." The testing described is non-clinical bench testing of the physical device. The documents do not mention clinical studies with human participants that would have a "test set" in the context of an AI/ML device evaluation.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable / Not Provided. This is a physical device submission, not an AI/ML device requiring expert-established ground truth for performance evaluation in the diagnostic sense.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable / Not Provided.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This document describes a physical medical device, not an AI/ML algorithm that would assist human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is not an algorithm.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Not Applicable. For the bench-top tests, "ground truth" was likely defined by engineering specifications and measurements (e.g., a specific tensile strength requirement, a maximum leak rate).
-
The sample size for the training set
- Not Applicable / Not Provided. This is a physical device, not an AI/ML algorithm requiring a training set.
-
How the ground truth for the training set was established
- Not Applicable / Not Provided.
Conclusion:
The provided 510(k) summary is for a physical medical catheter and details non-clinical bench testing to demonstrate compliance with design specifications and substantially equivalence to predicate devices, as required for FDA clearance of such devices. It does not provide information relevant to the typical acceptance criteria and study designs used for AI/ML-driven medical devices, as it is not an AI/ML device.
{0}------------------------------------------------
$p^{1/2}$
NOV 1 0 2011
SECTION 3 - 510(k) SUMMARY
| 510(k) Number: | K102925 |
|---|---|
| Submitted by: | NexGen Medical Systems, Inc.10471 Double R Blvd, Suite AReno, NV 89521 US |
| Contact Person: | John Kucharczyk. CEOPhone (775) 849-4165 |
| Date Prepared: | March 18, 2011 |
| Name of the Device: | NexGen Peripheral Expandable Catheter |
| Proprietary Name: | Undetermined at this time |
| Common Name: | Catheter Introducer |
| Classification: | Catheter Introducer (870.1340) |
SUMMARY STATEMENT:
Predicate Device: The NexGen Peripheral Expandable Catheter is substantially equivalent to the Applied Medical Expandable Vascular Sheath cleared under 510(k) K070865 and K963886.
Device Description:
The NexGen Peripheral Expandable Catheter is a sterile single use catheter intended to facilitate the placement or removal of vascular catheters and devices including the NexGen Mechanical Retrieval Device (MRD).
The Device features a Pebax/HDPE catheter shaft attached Nitinol tip that expands as the device is deployed from the lumen of a guide catheter. The Nitinol braid passively expands to the contour shape of the vessel lumen and allows a separate medical device such as a catheter to be retracted into the Nitinol Braid. Both the Nitinol Braid and the other device are then retracted into the Guide Catheter. The retracted Expandable Catheter with the encapsulated device are then removed together via removal of the Guide Catheter.
The Expandable Catheter comes in multiple expanded profiles of 5mm, 7mm, 9mm and 11mm, each having a working length of 43cm and 116cm. The Expandable Catheter is packaged in a Tear Away Sheath (B. Braun 510(k) K000313) that is used to facilitate loading the catheter into a Guide Catheter.
Proximally located on the catheter shaft is a female luer connection for the attachment to a Hemostasis Valve Y Connector, which provides a means of introducing contrast media or other fluids. The Hemostasis Valve Y Connector is separate or a prepackaged component.
{1}------------------------------------------------
K/o 2925 p 2/2
Intended Use: The NexGen Peripheral Expandable Catheter is indicated for percutaneous access to the peripheral vascular system and is designed to assist in the placement and removal of devices. This device is not intended for use in the coronary or cerebral vasculature.
Non-Clinical Testing: The following bench top testing was conducted to verify the device met design specifications:
- Visual / Dimension Inspection .
- Catheter Pushability / Trackability Testing .
- Catheter Tensile Testing .
- Catheter Torque Testing .
- . Catheter Burst Testing Catheter Leak Testing
- Catheter Kink Diameter Testing .
- Catheter Flow Rate Testing .
- Catheter Radial Force Testing ●
- Catheter Corrosion Resistance Testing ●
The NexGen Peripheral Expandable Catheter is a sterile Technological Characteristics: single use catheter intended for percutaneous access to the peripheral vascular system. The Expandable Catheter is designed to assist in the placement and removal of vascular catheters and devices. The device consists of 2 major components, a catheter shaft and a deployable braided nitinol funnel. The NexGen Expandable Catheter is designed for percutaneous insertion via a standard 8F Guide Catheter. The braided funnel passively expands to the contour shape of the vessel lumen and allows a separated medical device such as a catheter to be inserted and Both the braided funnel and the other device are then retracted into the Guide retracted. Catheter. The retracted Expandable Catheter with the encapsulated device is then removed together via removal of the Guide Catheter. The Expandable Catheter comes in multiple expanded profiles of 5mm. 7mm, 9mm and 11mm, each having a working length of 43cm and l 16cm.
The NexGen Expandable Catheter is substantially equivalent to the Applied Medical Expandable Vascular Sheath (K963886 and K070865).
Mechanical, safety and biocompatibility tests were performed to verify the functional safety, structural integrity and material safety. All testing demonstrated that the NexGen Expandable Catheter is comparable to the predicate devices and introduces no new safety and effectiveness issues when used as indicated.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle symbol with three curved lines representing its wings and tail. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 1 0 2011
NexGen Medical Systems, Inc. c/o Craig Pagan Regulatory Affairs 1050 W NASA Blvd. Suite 136 Melbourne, FL 32901
Re: K102925
Trade/Device Name: NexGen Peripheral Expandable Catheter Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: October 25, 2011 Received: October 26, 2011
Dear Mr. Pagan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. Craig Pagan
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
K L
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
SECTION 4 - INDICATIONS FOR USE STATEMENT
K162925 510(k) Number:
NexGen Peripheral Expandable Catheter Device Name:
INDICATIONS:
The NexGen Peripheral Expandable Catheter is indicated for percutaneous access to the perioheral vascular system and is designed to assist in the placement and removal of devices. This device is not intended for use in the coronary or cerebral vasculature.
CONTRAINDICATIONS:
No known contraindications.
Prescription Use _ V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
166
vision Sign-Off) Division of Cardiovascular Devices
510(k) Numb
Page 1 of 1
§ 870.1340 Catheter introducer.
(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).