K Number
K102841
Device Name
PANTEX AM/PM SALIVARY CORTISOL ENZYME IMMUNOASSAY
Date Cleared
2012-05-08

(587 days)

Product Code
Regulation Number
862.1205
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
For the in-vitro diagnostic quantitative determination of free and protein bound salivary cortisol in human saliva as an aid in the assessment of Cushing Syndrome and Addison's Disease. Measurements of cortisol in saliva are used in the diagnosis and treatment of disorders of the adrenal gland.
Device Description
The kit consists of a 96 well GARGG (Goat Anti-Rabbit Gamma Globulin) coated microplate (12x8 breakable strip wells), seven ready-to-use calibrators (range 0.1-30 ng/ml) of gravimetrically prepared cortisol from a commercial source (Steraloids) and compared and traced to NIST cortisol, low and high controls, anti-Cortisol (rabbit), 10X concentrated Cortisol (analog)-peroxidase, substrate solution, stop reaction solution and 10X concentrated wash solution.
More Information

Not Found

No
The device description and performance studies describe a standard immunoassay kit for measuring cortisol levels, with no mention of AI or ML algorithms for data analysis or interpretation.

No.
This device is an in-vitro diagnostic test kit used to measure cortisol levels in saliva, aiding in the diagnosis of Cushing Syndrome and Addison's Disease. It does not provide treatment or therapy.

Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "For the in-vitro diagnostic quantitative determination of free and protein bound salivary cortisol in human saliva as an aid in the assessment of Cushing Syndrome and Addison's Disease." This directly indicates its purpose as a diagnostic tool.

No

The device description clearly outlines a physical kit containing a microplate, calibrators, controls, antibodies, and solutions, indicating it is a hardware-based in-vitro diagnostic device, not software only.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The "Intended Use / Indications for Use" section explicitly states "For the in-vitro diagnostic quantitative determination of free and protein bound salivary cortisol in human saliva as an aid in the assessment of Cushing Syndrome and Addison's Disease." The phrase "in-vitro diagnostic" is a direct indicator.
  • Purpose: The device is designed to measure a substance (cortisol) in a sample taken from the human body (saliva) to provide information for the diagnosis and treatment of specific medical conditions (disorders of the adrenal gland, Cushing Syndrome, and Addison's Disease). This aligns perfectly with the definition of an in vitro diagnostic device.
  • Device Description: The description details a kit containing reagents and components used to perform a laboratory test on a biological sample.
  • Performance Studies: The document includes details of analytical performance studies (precision, linearity, recovery, method comparison, interference) which are typical for IVD devices to demonstrate their analytical validity.

N/A

Intended Use / Indications for Use

For the in-vitro diagnostic quantitative determination of free and protein bound salivary cortisol in human saliva as an aid in the assessment of Cushing Syndrome and Addison's Disease. Measurements of cortisol in saliva are used in the diagnosis and treatment of disorders of the adrenal gland.

Product codes (comma separated list FDA assigned to the subject device)

NHG

Device Description

The basis of the Cortisol Enzyme Immunoassay (EIA) is the quantitative relation between ligand concentration and the proportion of Cortisol (analog) enzyme conjugate bound to the antiserum. For example: Cortisol in the calibrators and unknowns compete with Cortisol coupled to peroxidase for antibody binding sites. After incubation, unbound components are washed away. The reaction between Cortisol peroxidase with the substrate (TMB) produces a blue color. The pre-determined time of incubation the reaction is stopped and a yellow color is formed. The optical density (read at 450 nm) is inversely proportional to the cortisol of calibrators, saliva samples and saliva controls.

The kit consists of a 96 well GARGG (Goat Anti-Rabbit Gamma Globulin) coated microplate (12x8 breakable strip wells), seven ready-to-use calibrators (range 0.1-30 ng/ml) of gravimetrically prepared cortisol from a commercial source (Steraloids) and compared and traced to NIST cortisol, low and high controls, anti-Cortisol (rabbit), 10X concentrated Cortisol (analog)-peroxidase, substrate solution, stop reaction solution and 10X concentrated wash solution.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Saliva (human)

Indicated Patient Age Range

The expected reference values were obtained from subjects aged 23-68 years.

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

I. Analytical Performance

  • a. Precision/Reproducibility

    • (i) Intra-assay precision: Determined from 20 replicates of low, medium and high samples.
      • Low: Mean 0.627 ng/mL, SD 0.034 ng/mL, %CV 5.4
      • Medium: Mean 3.995 ng/mL, SD 0.266 ng/mL, %CV 6.7
      • High: Mean 25.232 ng/mL, SD 1.579 ng/mL, %CV 6.3
    • (ii) Inter-assay precision: Determined from the mean of average duplicates for twelve (12) separate assays.
      • Low: Mean 0.587 ng/mL, SD 0.037 ng/mL, %CV 6.3
      • Medium: Mean 4.163 ng/mL, SD 0.301 ng/mL, %CV 7.2
      • High: Mean 25.126 ng/mL, SD 0.712 ng/mL, %CV 2.8
    • (iii) Inter-lot or between-lot variation: Determined by duplicate measurements of five (5) pools of saliva samples and three (3) saliva controls using three (3) different lots. Concluded a %CV of ≤10% for each sample tested.
    • (iv) Repeatability: Conducted over 20 days of testing with 2 assays performed daily. Used 3 different reagent lots and 3 saliva pools (low, medium, high concentration).
      • Low Concentration: Concentration 0.600 ng/ml, SD 0.0141. Total Device Precision: SD 0.0538, %CV 9.09.
      • Medium Concentration: Concentration 4.0 ng/ml, SD 0.0894. Total Device Precision: SD 0.1869, %CV 4.56.
      • High Concentration: Concentration 25 ng/ml, SD 0.5477. Total Device Precision: SD 0.8185, %CV 3.24.
  • b. Linearity

    • Ten (10) sample concentrations spanning the assay measuring range were performed per EP6-A. Recoveries ranged from 93.0% to 101.4%.
  • c. Recovery

    • Ten (10) saliva samples with endogenous cortisol were spiked with known quantities of cortisol and assayed. Recoveries ranged from 93.7% to 103.9%.
  • d. Traceability/Reagent Stability/Sample Stability/Expected Values

    • Calibrators and controls traced to NIST cortisol.
    • Reagents stored at 2-8°C are stable for 9 months. Opened vial and working Cortisol-HRP Conjugate solution are stable for 31 days at 2-8°C.
    • Sample stability:
      • Room Temperature (20-30°C): Up to 7 days
      • 37°C: Up to 7 days
      • 2-8°C: Up to 7 days

§ 862.1205 Cortisol (hydrocortisone and hydroxycorticosterone) test system.

(a)
Identification. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.(b)
Classification. Class II.

0

K102841 510(k) Summary

| Submitted By: | Pantex, Division of Bio-Analysis, Inc.
1701 Berkeley Street
Santa Monica, CA 90404
USA
(310) 828-7423 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Romulo Garza, Ph.D., President/Senior Scientist |
| Date Summary Prepared: | 4-3-12 |
| Trade Name: | Pantex AM/PM Salivary Cortisol EIA Kit |
| Common Name: | Enzyme immunoassay, cortisol, salivary |
| Regulation Number and Panel: | 862.1205-Clinical Chemistry |
| Classification Product Code: | NHG |
| Classification: | Class II |
| Substantially Equivalent Device: | K011323 Salimetrics HS Salivary Cortisol EIA
Kit Item No. 1-3102 (single) 96 well kit |
| Device Description: | |

A. Test principle.

The basis of the Cortisol Enzyme Immunoassay (EIA) is the quantitative relation between ligand concentration and the proportion of Cortisol (analog) enzyme conjugate bound to the antiserum. For example: Cortisol in the calibrators and unknowns compete with Cortisol coupled to peroxidase for antibody binding sites. After incubation, unbound components are washed away. The reaction between Cortisol peroxidase with the substrate (TMB) produces a blue color. The pre-determined time of incubation the reaction is stopped and a yellow color is formed. The optical density (read at 450 nm) is inversely proportional to the cortisol of calibrators, saliva samples and saliva controls.

B. Kit Description.

The kit consists of a 96 well GARGG (Goat Anti-Rabbit Gamma Globulin) coated microplate (12x8 breakable strip wells), seven ready-to-use calibrators (range 0.1-30 ng/ml) of gravimetrically prepared cortisol from a commercial source (Steraloids) and compared and traced to NIST cortisol, low and high controls, anti-Cortisol (rabbit), 10X concentrated Cortisol (analog)-peroxidase, substrate solution, stop reaction solution and 10X concentrated wash solution.

C. Intended Use/Indications for Use:

For the in-vitro diagnostic quantitative determination of free and protein bound salivary cortisol in human saliva as an aid in the assessment of Cushing Syndrome and Addison's Disease. Measurements of cortisol in saliva are used in the diagnosis and treatment of disorders of the adrenal gland.

Predicate Device:

The predicate device for substantial equivalence in this submission is:

1

Device NameSalimetrics Salivary Cortisol HS EIA Kit
CompanySalimetrics
510(k) referenceK011323

Technology Comparison:

Predicate Device:New Device:
Salimetrics Salivary Cortisol HS
EIA Kit
(K011323)Pantex AM/PM Salivary
Cortisol EIA Kit
(K102841) pending
Indications for useFor the in-vitro diagnostic
quantitative determination of free
and protein bound salivary
cortisol in human saliva as an aid
in the assessment of Cushing
Syndrome and Addison's Disease.
Measurement s of cortisol in
saliva are used in the diagnosis
and treatment of disorders of the
adrenal gland.For the in-vitro diagnostic
quantitative determination of
free and protein bound
salivary cortisol in human
saliva as an aid in the
assessment of Cushing
Syndrome and Addison's
Disease.
Measurement s of cortisol in
saliva are used in the diagnosis
and treatment of disorders of
the adrenal gland.
AnalyteFree and Protein-bound
CortisolFree and Protein-bound
Cortisol
Sample TypeSalivaSaliva
MethodEnzyme immunoassayEnzyme immunoassay
Detection MethodColormetric microplate readerColormetric microplate reader
Test PrincipleCortisol in the sample competes
with Cortisol-enzyme conjugate
for binding sites to antibody
bound to a microwell. Unbound
components are washed away and
enzyme is measured by a colored
reaction with the TMB substrate.Cortisol in the sample
competes with cortisol-
enzyme conjugate for binding
sites to the antibody (rabbit
anti cortisol) bound to a
GARGG microplate.
Unbound components are
washed away and enzyme is
measured by a colored
reaction with the TMB
substrate.
CalculationsQuantitative determination
with standard curveQuantitative determination
with standard curve
Quality ControlUse of reference controls is
recommendedUse of reference controls is
recommended
Analytical Measuring
Range (AMR)0.12 ng/ml - 30.0 ng/ml0.1 ng/ml - 30.0 ng/ml
Expected Values
(Normal range)N=192
Ages: 18-70
AM range: 0.8 - 15.5 ng/ml
PM range: II)No
Claim Accepted (I Yes

4

Precision Low Concentration Pool
Standard Deviation, SD)(% Coefficient of Variation CV)
Within Run0.02243.79
Between Run0.04627.80
Repeatability0.01622.73
Total Device Precision0.05389.09

.

and the comments of the comments of the comments of

Repeatability Medium Concentration
Concentration (ng/ml)4.0 ng/ml
Standard Deviation0.0894
(I) (User Variance/Claim Variance) x R63.073
(II) Critical Chi-square65.171
Claim Rejected (I>II)No
Claim Accepted (I Yes
Precision Medium Concentration Pool
Standard Deviation, SD)(% Coefficient of Variation CV)
Within Run0.14753.60
Between Run0.05141.26
Repeatability0.10252.50
Total Device Precision0.18694.56
Repeatability High Concentration
Concentration (ng/ml)25 ng/ml
Standard Deviation0.5477
(I) (User Variance/Claim Variance) x R62.035
(II) Critical Chi-square65.171
Claim Rejected (I>II)No
Claim Accepted (I Yes
Precision High Concentration Pool
Standard Deviation, SD(% Coefficient of Variation CV)
Within Run0.44421.176
Between Run0.29151.15
Repeatability0.622762.46
Total Device Precision0.81853.24

.

5

b. Linearity

Ten (10) sample concentrations that span the assay measuring range were performed Per EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures.

S=10 samples (dilutions) Concentration = (C1V1+C10V10)/(V1+V10)

| | C1
ng/ml | V1
ng/ml | C10
ng/ml | V10
ng/ml | Calculated
Concentration
ng/ml | Obtained
Concentration
ng/ml | Recovery
% |
|----|-------------|-------------|--------------|--------------|--------------------------------------|------------------------------------|---------------|
| 1 | 0.093 | * | | | 0.100 | 0.093 | 93.0 |
| 2 | 0.093 | 0.889 | 33.788 | 0.111 | 3.833 | 3.729 | 97.3 |
| 3 | 0.093 | 0.778 | 33.788 | 0.222 | 7.573 | 7.620 | 100.6 |
| 4 | 0.093 | 0.667 | 33.788 | 0.333 | 11.313 | 10.842 | 95.8 |
| 5 | 0.093 | 0.556 | 33.788 | 0.444 | 15.054 | 14.350 | 95.3 |
| 6 | 0.093 | 0.444 | 33.788 | 0.556 | 18.827 | 18.313 | 97.3 |
| 7 | 0.093 | 0.333 | 33.788 | 0.667 | 22.568 | 21.547 | 95.5 |
| 8 | 0.093 | 0.222 | 33.788 | 0.778 | 26.308 | 24.694 | 93.9 |
| 9 | 0.093 | 0.111 | 33.788 | 0.889 | 30.048 | 30.459 | 101.4 |
| 10 | | | | * | 35.000 | 33.788 | 96.5 |

*Targets of low and high sample concentrations

c. Recovery:

Ten (10) saliva samples containing different levels of endogenous cortisol were spiked with known quantities of cortisol and assayed.

| Sample | Endogenous
(ng/ml) | Added
(ng/ml) | Expected
(ng/ml) | Observed
(ng/ml) | Recovery
(%) |
|--------|-----------------------|------------------|---------------------|---------------------|-----------------|
| 1 | 0.493 | 0.250 | 0.743 | 0.739 | 99.5 |
| 2 | 0.878 | 0.500 | 1.378 | 1.291 | 93.7 |
| 3 | 1.551 | 1.000 | 2.551 | 2.641 | 103.5 |
| 4 | 1.850 | 2.000 | 3.850 | 3.958 | 102.8 |
| 5 | 0.936 | 4.000 | 4.936 | 4.951 | 100.3 |
| 6 | 1.042 | 8.000 | 9.042 | 9.394 | 103.9 |
| 7 | 0.691 | 16.000 | 16.691 | 17.165 | 102.8 |
| 8 | 0.622 | 20.000 | 20.622 | 19.997 | 97.0 |
| 9 | 2.057 | 24.000 | 26.057 | 24.938 | 95.7 |
| 10 | 0.348 | 28.000 | 28.348 | 28.943 | 102.1 |

d. Traceability/Reagent Stability/Sample Stability/Expected Values

The calibrators and controls are prepared from stock cortisol (Steraloids) and are gravimetrically weighed and prepared. Concentration of stock cortisol (Steraloids) concentrations were confirmed by comparison to NIST cortisol (y=1.029 x-0.2195. R20.9924)

6

Summary of the stability results

Real time stability studies are conducted to determine the reagent and kit shelf life (expiration date). Expiration date of the Pantex AM/PM Salivary Cortisol EIA Kit. Cat 631, is determined by results of shelf life studies and is based on the reagent that has the shorter assigned expiration day.

    1. The reagents stored at 2-8°C are stable for 9 months; therefore, the expiration date of the kit components is established at 9 months from the manufacturing date. We are basing the 9 month stability claim based on the results obtained with the real time stability study when stored at 2-8°C, supported by the reagents stored at room temperature (20 -28°C) and in reference to the stress chart that predicts reagents year stability at +5℃, by Kennon, L. "Use of models in determining chemical pharmaceutical stability".

Sample stability

1 ปี 1 ปี 2017 คนนิสเทคร กิน ที่ 11 ของแล้ว ค.ศ. 2017 ค.ศ. 2010 พ.ศ. 25 เมตร พ.ศ. 2010 พ.ศ. 25
StorageRoom37 °C2-8 °C10% cross reaction prednisolone is a potential interfering substance.

8

  • II. Method Comparison Studies
    Tests were conducted for comparison between the Pantex AM/PM Salivary Cortisol EIA Kit, Cat and the predicate assay , Salimetrics HS Salivary Cortisol results of 160 samples were compared . Comparison of the Pantex AM/PM Salivary Cortisol EIA Kit (new device) and the Salimetrics HS Salivary Cortisol EIA (predicate) demonstrated acceptable regression and correlation statistics and appears to be substantially equivalent to the FDA cleared predicate device.
Pantex AM/PM Salivary Cortisol EIA versus Salimetrics HS Salivary Cortisol EIA
Linear Regression equation$Y = 1.0269x + 0.0994$
Correlation (r²)0.9797

III. Interferences Studies.

An in-vitro experiment was performed by spiking three (3) levels of Cortisol (low, medium and high) with high concentrations of five (5) potentially interfering substances: alcohol, coffee (as caffeine), cigarette (as nicotine) and food and gum extracts. The results obtained appear to demonstrate no significant interference of the substances tested in this study with the measurement of Cortisol in saliva using the Pantex AM/PM Salivary Cortisol EIA Kit, Cat #631.

In-vitro experiment results:

PoolsPotential Interferent Caffeine Added (ug/mL)Obtained Value (ng/mL)Recovery from Control (%)
Low Pool01.420100
Low Pool8001.444101.7
Low Pool4001.438101.3
Low Pool2001.477104.0
Middle Pool04.932100
Middle Pool8005.143104.3
Middle Pool4005.063102.7
Middle Pool2004.53992.0
High Pool028.302100
High Pool80026.01091.9
High Pool40026.81194.7
High Pool20027.56397.4

9

| Pools | Potential Interferent
Food Added
(mg/mL) | Obtained Value
(ng/mL) | Recovery from
Control
(%) |
|-------------|------------------------------------------------|---------------------------|---------------------------------|
| Low Pool | 0.000 | 1.345 | 100 |
| | 426 | 1.340 | 99.6 |
| | 213 | 1.332 | 99.0 |
| | 106.5 | 1.291 | 96.0 |
| Middle Pool | 0.000 | 4.739 | 100 |
| | 426 | 4.730 | 99.8 |
| | 213 | 4.834 | 102.0 |
| | 106.5 | 4.835 | 102.0 |
| High Pool | 0.000 | 26.447 | 100 |
| | 426 | 27.495 | 104.0 |
| | 213 | 26.283 | 99.4 |
| | 106.5 | 28.055 | 106.1 |

PoolsPotential Interferent NicotineAdded (ug/mL)Obtained Value (ng/mL)Recovery from Control (%)
Low Pool01.362100
8001.436101.7
6001.437101.3
4001.4345104.0
2001.31796.7
Middle Pool04.871100
8005.243107.6
6005.258107.9
4005.087104.4
2005.155105.8
High Pool025.503100
80027.033106.0
60026.397103.5
40025.642100.6
20024.92897.8

.

.

·

·

10

| Pools | Potential Interferent
Gum Added
(mg/mL) | Obtained Value
(ng/mL) | Recovery from
Control
(%) |
|-------------|-----------------------------------------------|---------------------------|---------------------------------|
| Low Pool | 0 | 1.289 | 100 |
| | 270 | 1.346 | 104.4 |
| | 135 | 1.286 | 99.8 |
| | 67.5 | 1.248 | 96.8 |
| | 33.75 | 1.280 | 99.3 |
| Middle Pool | 0 | 4.843 | 100 |
| | 270 | 4.957 | 102.4 |
| | 135 | 4.845 | 100 |
| | 67.5 | 4.712 | 97.3 |
| | 33.75 | 4.764 | 98.4 |
| High Pool | 0 | 28.367 | 100 |
| | 270 | 29.419 | 103.7 |
| | 135 | 29.247 | 103.1 |
| | 67.5 | 27.870 | 98.3 |
| | 33.75 | 28.216 | 99.5 |

| Pools | Potential Interferent
Ethanol Added
(%) | Obtained Value
(ng/mL) | Recovery from
Control
(%) |
|-------------|-----------------------------------------------|---------------------------|---------------------------------|
| Low Pool | 0 | 1.268 | 100 |
| | 0.025 | 1.267 | 104.4 |
| | 0.050 | 1.230 | 99.8 |
| | 0.100 | 1.213 | 96.8 |
| Middle Pool | 0 | 4.539 | 100 |
| | 0.025 | 4.403 | 102.4 |
| | 0.050 | 4.320 | 100 |
| | 0.100 | 4.490 | 97.3 |
| High Pool | 0 | 26.272 | 100 |
| | 0.025 | 28.362 | 108.0 |
| | 0.050 | 28.744 | 109.4 |
| | 0.100 | 28.750 | 109.4 |

Concluding Statement:

Taken together, the performance characteristics, comparison studies with a predicate device and acceptable statistical performance studies in this 510(k) submission demonstrates that the Pantex AM/PM Salivary Cortisol EIA Kit, Cat #631, is safe and effective for its intended use and is substantially equivalent to the predicate device.

11

Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an image of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

Pantex, Division of Bio-Analysis, Inc c/o Romulo Garza 1701 Berkeley Street Santa Monica, CA 90404

MAY - 8 2012

Re: K102841 ·

Trade Name: Pantex AM/PM Salivary Cortisol Enzyme Immunoassay Regulation Number: 21 CFR §862.1205 Regulation Name: Cortisol (hydrocortisone and hydroxycorticosterone) test system Regulatory Class: Class II Product Codes: NHG Dated: April 3, 2012 Received: May 2, 2012

Dear Dr. Garza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III -(PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr 1), it can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

12

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 Jou ability of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

N.

Steven H. Liao, Ph.D.

Courney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

13

Indications for Use

K102841 510(k) Number (if known):

Device Name: Pantex AM/PM Salivary Cortisol Enzyme Immunoassay

Indications For Use:

For the in-vitto diagnostic quantitative determination of free and protein bound salivary cortisol in human saliva as an aid in the assessment of Cushing Syndrome and Addison's Disease. Measurements of cortisol in saliva are used in the diagnosis and treatment of disorders of the adrenal gland.

× Prescription Use (21 CFR Part 801 Subpart D) .

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Ruto Chem

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 102841

Page 1 of 1