K Number
K102643
Date Cleared
2011-07-15

(304 days)

Product Code
Regulation Number
862.1315
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SPOTCHECK Neonatal GALT Microplate Reagent Kit is for the quantitative determination of galactose-1-phosphate uridyltransferase, EC 2.7.7.12 (GALT), activity in whole blood saturated filter paper disks, using a microplate absorbance reader. Measurements of GALT enzyme activity are used primarily in the diagnosis and treatment of the hereditary disease galactosemia. This method is intended for in vitro diagnostic use as an aid in neonatal screening for decreased levels of GALT enzyme activity, and not for monitoring purposes.

The SPOTCHECK Pro is used for automated sample processing in the application of in vitro diagnostic assays. Specimens containing patient bodily substances are introduced and analyzed in microtiter plates using qualitative/quantitative determination through absorbance measurements.

These devices are intended for use by trained, qualified laboratory personnel.

Device Description

SPOTCHECK Neonatal GALT Microplate Reagent Kit - 60 Plate: Four enzyme mediated reactions are employed in the determination of GALT activity. GALT activity is determined by measuring the colored formazan produced by the addition of the color reagent to the incubated blood/substrate mixture. Patient samples of whole blood collected on standardized filter paper are placed into the wells of a standard 96 well microplate. A buffered enzyme mixture is added to each well and the plate is incubated at 37 °C for 120 minutes on a plate shaker/incubator. Following incubation, an aliquot of the mixture from each well is transferred to the corresponding wells on a clean 96 well microplate. Color reagent is added to each well, the color is developed over the course of 10 minutes, and the absorbance of each sample is determined on the plate reader. A blank absorbance reading is made prior to the addition of the color reagent to correct for endogenous sample color. The color developed is proportional (1:1) to the GALT activity in the sample. A standard curve prepared from a stock NADH solution is used to quantitate the results. Results are expressed as units of GALT enzyme activity per gram of hemoglobin or U/g Hb.

SPOTCHECK Pro: INSTRUMENT COMPONENTS: Tecan Freedom EVO and accessories necessary for assay.

AI/ML Overview

This is a 510(k) premarket notification for a medical device, not an AI/ML device, therefore, some of the requested information (e.g., number of experts, adjudication method, MRMC study, sample size for training set) is not applicable or cannot be extracted from the provided text. The document describes the device's technical characteristics, performance studies performed to demonstrate substantial equivalence to a predicate device, and its intended use.

Here's an analysis of the provided information, tailored to what is available in the regulatory submission format:

1. Table of Acceptance Criteria (Performance Goals) and Reported Device Performance

The acceptance criteria are generally implied by the comparative studies to the predicate device and established CLSI guidelines for analytical performance. The studies aim to show the SPOTCHECK Kit performs comparably or better than the predicate device across various metrics.

Performance MetricAcceptance Criteria (Implied/Standard)Reported Device Performance (SPOTCHECK Neonatal GALT Microplate Reagent Kit)
LinearityAdherence to CLSI EP6-A; 2nd order regression from 0 to 15 U/g Hb.Non-linear (2nd order regression) in the range of 0.25 to 15 U/g Hb. Confirmed by adherence to CLSI EP6-A. Calibration curve (NADH standards) conforms to a 2nd order regression from 0 to 15 U/g Hb. Results > 15 U/g Hb are reported as such.
Analytical Sensitivity (LoD)Consistent with CLSI EP17-A; α < 0.3%, β < 0.3%Limit of Detection (LoD) = 0.2 U/g Hb. Determined consistent with CLSI EP17-A protocol, with α < 0.3% and β < 0.3%, based on 300 measurements (60 blank, 240 low-level samples); LoB = 0.08 U/g Hb. Total error (3xSD) < 0.2 U/g Hb. LoD = LoQ.
Analytical Sensitivity (LoQ)< 20% total imprecision at LoQ (functional LoQ).Functional LoQ = 0.3 U/g Hb. Established using a criterion of < 20% total imprecision at the LoQ. Neonatal specimens < 0.3 U/g Hb reported as such (presumed positive for galactosemia).
Automated vs. Manual PerformanceManual and automated processing should provide similar results. Screening equivalence should be confirmed if manual is backup.Linear Regression (Automated vs. Manual): - Multiple R: 0.946- R²: 0.896- Adjusted R²: 0.895- Standard Error: 0.804- Observations: 204- Intercept: 0.393 (SE 0.168, 95% CI 0.063-0.724)- X Variable: 0.963 (SE 0.023, 95% CI 0.917-1.01)Study confirms similar results can be expected.
Clinical Classification (Automated)Screening results should correlate to the predicate device.Comparison to Predicate Device (Automated):- Total N = 1752- Positive agreement: (57/58) = 98.3% (calculated from table)- Negative agreement: (1746/1747) = 99.9% (calculated from table)- Overall agreement: (57+1746)/1752 = 99.8% (calculated from table)(Note: There are conflicting tables for automated classification, using one where SPOTCHECK Positive/Negative vs Predicate Positive/Negative sum to 1752 cases (the "57" table)).
Clinical Classification (Manual)High degree of correlation to predicate device classification. Correct classification of known deficient samples.Comparison to Predicate Device (Manual):- Total N = 292- Positive agreement: (58/62) = 93.5%- Negative agreement: (214/230) = 93.0%- 10 known GALT deficient samples correctly classified.
PrecisionComparable to predicate device; adherence to CLSI EP5-A2.Within-Run Precision (Manual, n=80):- Deficient (0.45 U/g Hb): CV 8.1%- Partial (1.3 U/g Hb): CV 6.5%- Near cutoff (2.4 U/g Hb): CV 8.3%- Normal (6.7 U/g Hb): CV 8.2%Within-Run Precision (Automated, n=80):- Deficient (0.60 U/g Hb): CV 7.8%- Partial (1.3 U/g Hb): CV 6.2%- Near cutoff (2.4 U/g Hb): CV 6.7%- Normal (6.1 U/g Hb): CV 6.6%Total Precision (Manual, n=80):- Deficient (0.45 U/g Hb): CV 11%- Partial (1.3 U/g Hb): CV 9.2%- Near cutoff (2.4 U/g Hb): CV 9.2%- Normal (6.7 U/g Hb): CV 9.7%Total Precision (Automated, n=80):- Deficient (0.60 U/g Hb): CV 11%- Partial (1.3 U/g Hb): CV 7.3%- Near cutoff (2.4 U/g Hb): CV 7.1%- Normal (6.1 U/g Hb): CV 6.7%Demonstrates comparable precision to predicate device.
Analytical Specificity (Interference)No statistically or clinically significant interference from common substances. Adherence to CLSI EP7-A2.γ globulin (6000 mg/dL): No significant interference. Minor GALT increases near cutoff, but not enough to misclassify deficient. (Predicate: no interference up to 2500 mg/dL).Albumin (6000 mg/dL): No significant interference or misclassification. (Predicate: not evaluated).Bilirubin, conjugated (28.8 mg/dL): No significant interference. (Predicate: no interference up to 40 mg/dL).Bilirubin, unconjugated (20 mg/dL): No significant interference. (Predicate: no interference up to 40 mg/dL).Hemoglobin (200 mg/dL): Statistically significant decrease in GALT activity, could lead to false positive near cutoff. (Predicate: not evaluated).Triglycerides (3270 mg/dL): No significant interference. (Predicate: no interference up to 1000 mg/dL).Sulfamethoxazole (400 ug/mL): No significant interference. (Predicate: not evaluated).Trimethoprim (40 ug/mL): No significant interference. (Predicate: not evaluated).
Contributions from HematocritVarying hematocrit should not lead to false negatives.45% Hct: Deficient 0.43 U/g Hb, Near Cutoff 2.4 U/g Hb, Normal 5.4 U/g Hb55% Hct: Deficient 0.55 U/g Hb, Near Cutoff 2.7 U/g Hb, Normal 6.9 U/g Hb65% Hct: Deficient 0.72 U/g Hb, Near Cutoff 3.3 U/g Hb, Normal 8.3 U/g HbDifferences in hematocrit had a statistically significant effect on samples with low GALT activity, but no indication of misclassifying a deficient sample as normal (false negative).

2. Sample Size Used for the Test Set and Data Provenance

  • Automated and Manual Performance Comparison:

    • Test Set Sample Size: 128 newborn patient dried blood spot samples, plus dried blood spot controls (manufactured to mimic newborn specimens) and dried specimens of mixed adult blood. A total of 216 samples were analyzed.
    • Data Provenance: Not explicitly stated (e.g., country of origin, specific state). The samples are referred to as "newborn patient dried blood spot samples" and "mixed adult blood," suggesting they are from human subjects, but the geographical origin is not specified. The study was retrospective, as existing dried blood spots were analyzed.
  • Expected Values and Clinical Cutoff Determination (Automated Processing):

    • Test Set Sample Size: 1752 routine samples and 53 known GALT deficient samples (controls, retrospective galactosemic neonates, and galactosemic non-neonates).
    • Data Provenance: From a "state screening laboratory," indicating US origin and retrospective collection.
  • Screening Using Manual Processing:

    • Test Set Sample Size: 247 newborn patient dried blood spot samples, plus dried blood spot controls (with newborn hematocrit) and dried specimens of mixed adult blood. A total of 292 samples were analyzed.
    • Data Provenance: Not explicitly stated, likely similar to the automated comparison study (retrospective human samples, likely US state screening program).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not applicable as this is a chemical reagent kit for GALT activity measurement, not an AI/ML device relying on human expert image interpretation or similar.
  • The "ground truth" for GALT deficiency or normal activity was established by:
    • Analysis using the predicate device.
    • "Known GALT deficient samples" which included "retrospective galactosemic neonates and galactosemic non-neonates," meaning their clinical status was previously established.
    • "Routine neonatal screening by the state DOH classified all but one of these specimens (as well as the specimens classified as negative) to be presumptive negative for galactosemia." For some samples, the reference came from a state Department of Health (DOH) screening classification.

4. Adjudication Method for the Test Set

  • Not applicable. The "ground truth" was established based on predicate device results and previously classified clinical samples, not through a consensus or adjudication process among multiple human readers for a primary diagnostic task.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/ML device, and no human reader study with or without AI assistance was performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This device is a standalone diagnostic kit (the "algorithm" here being the chemical assay and spectrophotometric measurement, interpreted by trained lab personnel). The studies described (linearity, sensitivity, precision, comparison to predicate) are effectively standalone performance evaluations. The SPOTCHECK Pro (instrument) automates the process, meaning the "algorithm" (assay) performs without manual intervention during the assay process itself, aside from initial setup and final interpretation.

7. The Type of Ground Truth Used

  • Comparative Reference Standard: The primary ground truth for evaluating the SPOTCHECK Kit was the performance of a legally-marketed predicate device (Bio-Rad Quantase Neonatal GALT Test). The studies explicitly compare the SPOTCHECK Kit's results to the predicate device's results.
  • Clinical Diagnosis/Known Status: For validation of clinical classification, "known GALT deficient samples" (retrospective galactosemic neonates and non-neonates) were used. This indicates a form of outcomes data or pre-established clinical diagnosis.
  • State DOH Screening Classification: In some instances, the "routine neonatal screening by the state DOH" provided a classification (presumptive positive/negative) which served as a reference for comparison.

8. The Sample Size for the Training Set

  • Not applicable in the context of AI/ML. The device is a chemical reagent kit. There wasn't a "training set" in the machine learning sense. The device's calibration curve is established using NADH standards, which are analytical controls, not a training dataset for an algorithm.

9. How the Ground Truth for the Training Set was Established

  • Not applicable as there is no "training set" in the AI/ML context. The calibration curve is established analytically using NADH standards, which have known concentrations. These standards are foundational to quantifying the GALT activity, not for training a predictive model.

{0}------------------------------------------------

ASTORIA. PACIFIC

Neonatal GALT Microplate Reagent Kit

Name, Address of Contact Person 1.

Applicant's name and address

Astoria-Pacific, Inc. FDA Establishment No. 3050015 15130 SE 82nd Drive P.O. Box 830 Clackamas, OR 97015-0830

Tel 1-503-657-3010 Fax 1-503-655-7367

Charles A. Peterson President

Jason Reynolds Director of R & D, Official Correspondent

.

Name of the Device 2.

Product Classification

Device Classification

Reagent Kit

Regulation Number21 CFR 862.1315
510(k) NumberK102643
Classification PanelClinical Chemistry
Product CodeKQP
Device ClassificationClass II
Product Nomenclature
Common NameGALT (Galactose-1-phosphate uridyl transferase)Screening Test
Classification NameGalactose-1-phosphate uridyl transferase test system
Proprietary NameAstoria-Pacific SPOTCHECK® Neonatal GALT MicroplateReagent Kit - 60 Plate
Model NumberAstoria-Pacific Part No. 81-4000-60K
Instrument
Regulation Number21 CFR 862.2300
510(k) NumberK102643
Classification PanelClinical Chemistry
Product CodeJJO

Class I

510(k) Summary

K102643

JUL 15 2011

{1}------------------------------------------------

510(k) Summary

Product Nomenclature

Common NamePhotometer for clinical use
Classification NameColorimeter, photometer, spectrophotometer for clinical use
Proprietary NameSPOTCHECK ProTM
Model NumberAstoria-Pacific Part No. 910-0500-00

3. Identification of the legally-marketed device for which substantial equivalence is claimed.

Product Classification

Reagent Kit

Regulation Number510(k) NumberClassification PanelProduct CodeDevice Classification21 CFR 862.1315K990827Clinical ChemistryKOPClass II
Product Nomenclature
Common NameGALT (Galactose-1-phosphate uridyl transferase)Screening Test
Classification NameGalactose-1-phosphate uridy1 transferase test system
Proprietary NameBio-Rad Quantase Neonatal GALT Test
Model Number(s)Bio-Rad Part No. 532-6000, 192 Test Kit
Bio-Rad Part No. 532-6001, 960 Test Kit
Instrument
Regulation Number21 CFR 862.2300
510(k) NumberK953710
Classification PanelClinical Chemistry
Product CodeJJQ
Device ClassificationClass I
Product Nomenclature
Common NamePhotometer for clinical use
Classification NameColorimeter, photometer, spectrophotometer for clinicaluse
Proprietary NameModel Number(s)Bio-Tek ELX808 Automated Microplate ReadersPart No. ELX808IUAP
Page 2 of 12

{2}------------------------------------------------

510(k) Summary

Description of the Device 4.

SPOTCHECK Neonatal GALT Microplate Reagent Kit - 60 Plate

Astoria-Pacific Part No. 81-4000-60K Galactose-1-phosphate uridyl transferase test system

KIT CONTENTS:

Substrate Color Reagent Stock Standard Tris Buffer

Four enzyme mediated reactions are employed in the determination of GALT activity. The GALT enzyme catalyzes the conversion of galactose-1-phosphate to glucose-1phosphate and concurrently the conversion of UDP-glucose to UDP-galactose. Then, glucose-1-phosphate is converted to glucose-6-phosphate catalyzed by the enzyme phosphoglucomutase. Next, glucose-6-phosphate is oxidized to 6-phosphogluconate with the concurrent reduction of NADP to NADPH, catalyzed by glucose-6-phosphate dehydrogenase. Finally, a tetrazolium salt, catalyzed by diaphorase, reacts with the NADPH to form the product that is measured to yield GALT activity.

Image /page/2/Figure/8 description: This image shows a series of four chemical reactions. In the first reaction, Gal-1-P and UDP-Glu are converted to Glu-1-P and UDP-Gal, with GALT as the catalyst. In the second reaction, Glu-1-P is converted to Glu-6-P, with PGluM as the catalyst and Mg as a cofactor. In the third reaction, Glu-6-P and NADP are converted to 6-PG and NADPH, with G6PD as the catalyst and Mg as a cofactor. In the fourth reaction, NADPH and MTT are converted to Colored Formazan and NADP, with Diaphorase as the catalyst.

GALT activity is determined by measuring the colored formazan produced by the addition of the color reagent to the incubated blood/substrate mixture.

Patient samples of whole blood collected on standardized filter paper are placed into the wells of a standard 96 well microplate. A buffered enzyme mixture is added to each well and the plate is incubated at 37 °C for 120 minutes on a plate shaker/incubator. Following incubation, an aliquot of the mixture from each well is transferred to the corresponding wells on a clean 96 well microplate. Color reagent is added to each well, the color is developed over the course of 10 minutes, and the absorbance of each sample is

Page 3 of 12

{3}------------------------------------------------

determined on the plate reader. A blank absorbance reading is made prior to the addition of the color reagent to correct for endogenous sample color.

The color developed is proportional (1:1) to the GALT activity in the sample. A standard curve prepared from a stock NADH solution is used to quantitate the results. Results are expressed as units of GALT enzyme activity per gram of hemoglobin or U/g Hb. A unit is defined as the quantity of GALT enzyme that catalyzes the formation of one micromole of UDP-galactose per minute at 37 °C.

SPOTCHECK Pro

Astoria-Pacific Part No. 910-0500-00 Galactose-1-phosphate uridyl transferase test system

INSTRUMENT COMPONENTS:

Tecan Freedom EVO® and accessories necessary for assay

Statement of Intended Use 5.

The SPOTCHECK Neonatal GALT Microplate Reagent Kit is for the quantitative determination of galactose-1-phosphate uridyl transferase, EC 2.7.7.12 (GALT), activity in whole blood saturated filter paper disks, using a microplate absorbance reader. Measurements of galactose-1-phosphate uridyltransferase are used primarily in the diagnosis and treatment of the hereditary disease galactosemia. This method is intended for in vitro diagnostic use as an aid in newborn screening for decreased levels of GALT enzyme activity, and not for monitoring purposes.

The SPOTCHECK Pro is used for automated sample processing in the application of in vitro diagnostic assays. Specimens containing patient bodily substances are introduced and analyzed in microtiter plates using qualitative/quantitative determination through absorbance measurements.

These devices are intended for use by trained, qualified laboratory personnel.

Summary of the Technological Characteristics of the Device 6.

DEVICE COMPARISON

The most significant difference between the SPOTCHECK Neonatal GALT Microplate Reagent Kit and the predicate device is the use of a calibration curve with the SPOTCHECK Kit to quantify results. Additionally, the SPOTCHECK Kit is also intended for use on automated platforms. Both the proposed and predicate devices use approximately the same reagent formulation and both use the same technology (spectrophotometric microplate reader) to determine GALT activity.

{4}------------------------------------------------

Neonatal patient dried blood specimens are punched into microplate wells, eluted and incubated with approximately the same substrate and buffer system on the SPOTCHECK Kit as on the predicate device. On the SPOTCHECK Kit, the tetrazolium salt (MTT) is not introduced into the reagent scheme until the final chemical reaction, whereas on the predicate device the MTT is included in the incubation substrate at the start of the procedure. The final step in the reaction, the formation of the colored formazan, is the same in both devices.

The SPOTCHECK Kit has two additional enzymes in the incubation substrate that are not included on the predicate device. Galactose-6-phosphate dehydrogenase (G6PD) and phosphoglucomutase (PGluM) are normally present in clinical specimens, but may become damaged or destroyed if samples are exposed to excess heat or moisture during handling or if samples are stored at room temperature or above for extended periods of time. The two additional enzymes are included so that samples with degradation of G6PD or PGluM are not incorrectly classified as having deficient GALT activity.

SPOTCHECK NeonatalGALT Microplate KitPredicate Device
Specimen collection,handling and storageUse standardized blood spotcollection cards; followprotocol in CLSI LA4-A5Same collection, handlingand storage
Specimen1 x 1/8" punched dried bloodspot (DBS)Same sample size
IncubationIn microplate, on combinationincubator/shakerIn covered or sealedmicroplate, onincubator/shaker
Incubation temperature37 °C37 °C
Incubation time2 hours (120 minutes)3 hours (180 minutes)
Substrate reagentBuffered NADP + Gal-1-P +UDP-Glu + G6PD + PGluMBuffered NADP +tetrazolium salt + Gal-1-P +UDP-Glu + glucose-1,6-diphosphate
Color reagentBuffered MTT + diaphoraseBuffered diaphorase
Absorbance measurementson microplate reader600 nm (750 nm reference)550 or 570 nm
Reporting unitsU/g HbU/g Hb
Limit of quantitation0.3 U/g Hb0.64 U/g Hb
Range0.3 U/g Hb to 15 U/g HbUpper range not stated
CalibrationLiquid NADH standardsFactor multiplication
Clinical classificationPresumptive positive andnegative (normal)Presumptive positive andnegative
Quality control materialDBS normal, deficientDBS normal, deficient

Summary of SPOTCHECK Neonatal GALT Microplate Kit and Predicate Device Comparison of Technological Characteristics

{5}------------------------------------------------

510(k) Summary

LINEARITY

The assay is non-linear (200 order regression) in the range of 0.25 to 15 U/g Hb. This correlation was confirmed by adherence to CLSI EP6-A: Evaluation of the Linearity of Ouantitative Measurement Procedures: A Statistical Approach; Approved Guideline. The calibration curve, established by the use of NADH standards, conforms to a 200 order regression from 0 to 15 U/g Hb. Results > 15 U/g Hb are reported as such.

ANALYTICAL SENSITIVITY

An overall analytical sensitivity of the assay was determined by adhering to CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline. The limit of detection (LoD), defined as the lowest amount of analyte in a sample that can be detected (although perhaps not accurately quantifiable), is 0.2 U/g Hb and the limit of quantitation (LoQ) is 0.3 U/g Hb. The LoD is determined consistent with the guidelines in CLSI EP17-A protocol and with proportions of false positives (a) less than 0.3% and false negatives (B) less than 0.3%, based on 300 measurements, consisting of 60 blank and 240 low-level samples; limit of blank (LoB) = 0.08 U/g Hb. The total error (TE, 3xSD) is less than the goal of 0.2 U/g Hb, assuming that there is no bias due to the unavailability of standard reference materials. Therefore, according to the guidelines in CLSI EP17-A, the LoD = LoQ. However, a functional LoQ of 0.3 U/g Hb was established using a criterion of < 20% total imprecision at the LoQ. Neonatal specimens with reported concentrations < 0.3 U/g Hb are reported as such, and can be presumed positive for galactosemia.

AUTOMATED and MANUAL PERFORMANCE COMPARISON

A study was performed to compare the SPOTCHECK GALT Reagent Kit processed using the SPOTCHECK Pro automated platform (configured to precisely automate the manual steps) against the manual method. Newborn patient dried blood spot samples (n = 128), dried blood spot controls manufactured to mimic newborn specimens, as well as dried specimens consisting of mixed adult blood were analyzed using both manual processing and the SPOTCHECK Pro. A total of 216 samples were analyzed using singlicate measurements for both devices. To reduce sources of error not attributed to the different approaches, the same reagent preparations were used, and individual samples were punched and analyzed using each process on the same day. The study was performed over three days.

Linear Regression Statistics
Multiple R0.946
R'0.896
Adjusted R20.895
Standard Error0.804
Observations-204

Regression Results

CoefficientsStandard ErrorLower 95%Upper 95%
Intercept0.3930.1680.0630.724
X Variable0.9630.0230.9171.01

Statistics include results within claim measuring range only.

{6}------------------------------------------------

This study confirms that manual and automated processing can be expected to provide similar results in screening patients for GALT deficiency.

If manual processing of the reagent kit is intended for use as backup for automated analysis, screening equivalence between the processing methods should be confirmed.

EXPECTED VALUES and DETERMINATION of CLINICAL CUTOFF

An exemplary normal range was established by analyzing 1752 routine samples and 53 known GALT deficient samples (controls, retrospective galactosemic neonates, and galactosemic non-neonates) at a state screening laboratory using the SPOTCHECK Kit (automated on the SPOTCHECK Pro). The same specimens were analyzed using the predicate device.

Data Summary

GALT (Routine)Predicate DeviceSPOTCHECK
Number of Observations17481748
Mean Value Observed9.4 U/g Hb7.9 U/g Hb
Standard Deviation2.22.1
Range of the Data1.4 to 18.5 U/g Hb2.5 to 14.5 U/g Hb
0.5 Percentile3.53.2
0.25 Percentile3.22.9

Statistics include only results within measuring range of both devices.

Specimens known deficient in GALT yielded results typically below one or both devices' limits of quantitation (49 of 53).

Specimens with results equal to or below the 0.5 and 0.25 percentiles were classified as deficient in GALT activity, or presumptive positive for galactosemia, and require followup testing according to institutional, local, state, regional, and/or national guidelines or regulations. Using the same criteria, cutoffs were very similar to the values for presumptive positive and normal GALT activity using the predicate device.

Each laboratory must determine its range of normal and deficient levels of GALT activity, based on its patient population and analytical variables.

CLASSIFICATION of SAMPLES

The performance of the SPOTCHECK Neonatal GALT Microplate Reagent Kit was evaluated against the performance of the predicate device according to CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples: Approved Guideline -Second Edition. Patient samples as well as dried blood spot control samples were used in this study.

{7}------------------------------------------------

Screening Results Comparison

Reference Test (Predicate Device)
PositiveNegativeTotal
SPOTCHECKPositive81210
Negative417381743
Total1217401752
Reference Test (Predicate Device)
PositiveNegativeTotal
SPOTCHECKPositive61263
Negative17381743
Totalર્ણ રેસ17401805
Reference Test (Predicate Device)
PositiveNegativeTotal
SPOTCHECKPositive$4^2$15
SPOTCHECKNegative117461747
SPOTCHECKTotal517471752
Reference Test (Predicate Device)
PositiveNegativeTotal
SPOTCHECKPositive57158
Negative117461747
Total5817471805

&quot;Routine neonatal screening by the state DOH classified all but one of these specimens (as well as the specimens classified as negative) to be presumptive negative for galactosemia. The laboratory classified the one specimen as having partial GALT activity.

2 Routine neonatal screening by the state DOH classified all but one of these specimens (as well as the specimens classified as negative) to be presumptive negative for galactory classified the one specimen as having partial GALT activity.

{8}------------------------------------------------

Screening Using Manual Processing

An internal study was performed for manual comparison to the predicate device. Newborn patient dried blood spot samples (n = 247), dried blood spot controls with newborn hematocrit, and dried specimens consisting of mixed adult blood were analyzed using both the SPOTCHECK Kit and the predicate device. A total of 292 samples were analyzed using singlicate measurements for both devices. A population of specimens that provided a large number of low GALT activity results was analyzed to ensure a high number of relevant screening classification comparisons.

GALT {All Specimens}Predicate DeviceSPOTCHECK
No. of Observations (in range)265265
Mean Value4.8 U/g Hb5.2 U/g Hb
Standard Deviation2.42.5
Range of the Data0.6 to 12.6 U/g Hb1.1 to 14.3 U/g Hb

Statistics include only results within measuring range of both devices.

Deficient GALT Activity(Presumptive Positive)$\u2264$ 2.3 U/gHb$\u2264$ 3.2 U/g Hb (0.5th percentile)and$\u2264$ 2.9 U/g Hb (0.25th percentile)
Normal GALT Activity (Negative)> 2.3 U/gHb> 3.2 U/g Hb (0.5th)and> 2.9 U/g Hb (0.25th)
Predicate Device
PositiveNegativeTotal
SPOTCHECKPositive602787
Negative2203205
Total62230292
Percent positive agreement: (60/62) = 96.7%
Predicate Device
PositiveNegativeTotal
SPOTCHECKPositive581674
Negative4214218
Total62230292
Percent positive agreement: (58/62) = 93.5%
Percent negative agreement: (214/230) = 93.0%

The SPOTCHECK GALT in vitro diagnostic kit demonstrated a high degree of correlation to specimens classified positive by the predicate device. Additionally, 10 specimens known to be deficient in GALT activity (analyzed in an unbiased manner) were correctly classified.

This evaluation complements the study demonstrating equivalent performance between manual and automated processing. Combined, they demonstrate safety and effectiveness of the new GALT screening method, using manual or automated specimen analysis.

Page 9 of 12

{9}------------------------------------------------

PRECISION PERFORMANCE

Within-run and total precision for the SPOTCHECK Kit were determined according to CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition. Samples at 4 different levels of GALT activity were analyzed using 16 replicates on 1 run per day for 5 days (2 distinct calibration curves on 2 plates); a total of 80 data points at each level were collected. Within-run and total precision using the predicate device for normal GALT activity levels was determined by analyzing samples in duplicate during 10 separate runs (data reported below was copied from product insert). The method used to determine precision of deficient and partial activity GALT samples in the predicate device was not stated.

Within-Run Precision, Sr SPOTCHECK Neonatal GALT Kit

(Manual, n = 80)
Activity(U/g Hb)DeficientPartialNearcutoffNormal
Mean0.451.32.46.7
S.D.0.0370.0870.200.55
C.V.8.1%6.5%8.3%8.2%
Activity(U/g Hb)DeficientPartialNearcutoffNormal
Mean0.601.32.46.1
S.D.0.0460.0810.160.40
C.V.7.8%6.2%6.7%6.6%

(Automated on the SPOTCHECK Pro, n = 80)

Predicate Device

(Manual, n = 20)
Activity(U/g Hb)NormalNormalNormal
Mean3.53.65.0
S.D.0.150.270.37
C.V.4.3%12%7.4%

Total Precision, ST

SPOTCHECK Neonatal GALT Kit

(Manual, n = 80)

Activity(U/g Hb)DeficientDeficientNearcutoffNormal
Mean0.451.32.46.7
S.D.0.050.120.220.65
C.V.11%9.2%9.2%9.7%

{10}------------------------------------------------

Automated on the SPOTCHECK Pro, n = 80)
Activity(U/g Hb)DeficientPartialNearcutoffNormal
Mean0.601 .3246.1
S.D.0.0630.0950.170.41
C.V.11%7.3%7.1%6.7%

Total Precision, Sr (continued)

Predicate Device

(Manual, n = 20)

Activity(U/g Hb)Partial"NormalNormalNormal
Mean1.83.53.65.0
S.D.0.450.370.430.40
C.V.23%11%12%8.0%

"(Number of samples (n) is not reported.)

The results of the precision study demonstrate that the SPOTCHECK Neonatal GALT Microplate Reagent Kit, at a minimum, exhibits comparable precision performance to that reported in the predicate device insert. Additionally, performance is similar whether the SPOTCHECK kit is processed manually or with automation.

ANALYTICAL SPECIFICITY

The study of potential interfering substances when using the SPOTCHECK Neonatal GALT Microplate Reagent Kit was carried out according to CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition.

Interference EvaluatedSPOTCHECK Neonatal GALTMicroplate KitPredicate Device
γ globulin (protein)Up to 6000 mg/dL showed nostatistically significant interference;minor increases in GALT wereobserved near the cutoff, but adeficient neonate would not beclassified as normalup to 2500 mg/dLshowed no significantinterference
Albumin (protein)Up to 6000 mg/dL showed nostatistically significant interference;minor increases in GALT wereobserved near the cutoff, but adeficient neonate would not beclassified as normalnot evaluated
Bilirubin, conjugatedUp to 28.8 mg/dL showed nostatistically or clinically significantinterferenceup to 40 mg/dL showedno significantinterference
Bilirubin, unconjugatedUp to 20 mg/dL showed noup to 40 mg/dL showed

Page 11 of 12

{11}------------------------------------------------

statistically or clinically significantno significant
interferenceinterference
Hemoglobin (Hb)Up to 200 mg/dL showed anot evaluated
statistically significant decrease in
GALT activity, which could result
in a false positive near the cutoff
TriglyceridesUp to 3270 mg/dL showed noup to 1000 mg/dL
statistically or clinically significantshowed no significant
interferenceinterference
SulfamethoxazoleUp to 400 ug/mL showed nonot evaluated
(SMX)statistically or clinically significant
interference
Trimethoprim (TMP)Up to 40 ug/mL showed nonot evaluated
statistically or clinically significant
interference

Contributions from Hematocrit

To determine any possible effects on the performance of the Astoria-Pacific, Inc. SPOTCHECK Neonatal GALT Microplate Reagent due to varying blood hematocrit, three blood spot samples were manufactured, one with deficient GALT, one near the clinical cutoff, and one with normal GALT. Within each of the three blood spot samples, three different levels of hematocrit (45%, 55% and 65%) were achieved by adjusting the quantity of red blood cells prior to spotting on blood spot collection paper.

Hematocrit %Deficient (n = 6)Near Cutoff (n = 6)Normal (n = 6)
450.43 U/g Hb2.4 U/g Hb5.4 U/g Hb
550.55 U/g Hb2.7 U/g Hb6.9 U/g Hb
650.72 U/g Hb3.3 U/g Hb8.3 U/g Hb

GALT activity resides in the red blood cells, so it is expected that varying hematocrit will lead to varying GALT response. Samples with lower hematocrit levels had lower GALT activity and samples with higher hematocrit levels had GALT activity. Differences in hematocrit had a statistically significant effect on samples with low GALT activity, however there is no indication that a sample deficient in GALT activity would be misclassified as normal (false negative) due to varying hematocrit levels.

7. Determination of Substantial Equivalency

Based on the performance characteristics and comparison data, the SPOTCHECK Kit is safe, effective, and substantially equivalent to the legally-marketed predicate device. The indications for use are essentially the same for the SPOTCHECK Neonatal GALT Microplate Reagent Kit and the predicate device. Technological characteristics are very similar to the predicate device and there is sufficient evidence that demonstrates that the differences do not adversely affect the safety and effectiveness of the SPOTCHECK Kit.

{12}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Jul. 15 2011

Astoria-Pacific, Inc. c/o Mr. Jason C. Reynolds Director of Research & Development 15130 SE 82nd Dr. Clackamas. OR 97015

રિભ: K102643 Trade Name: SPOTCHECK Neonatal GALT Microplate Reagent Kit, SPOTCHECK Pro Regulation Number: 21 CFR §862.1315 Regulation Name: Galactose-1-Phosphate Uridyl Transferase test system Regulatory Class: Class II Product Codes: KQP, JJQ Dated: July 8. 2011 Received: July 11, 2011

Dear Mr. Reynolds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{13}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-fremember (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

CJC.

Courtney Harper. Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

: 评论:

Enclosure

{14}------------------------------------------------

Indications for Use Form

510(k) Number: K102643

Device Name:

Indications for Use:

The SPOTCHECK Neonatal GALT Microplate Reagent Kit is for the quantitative determination of galactose-1-phosphate uridyItransferase, EC 2.7.7.12 (GALT), activity in whole blood saturated filter paper disks, using a microplate absorbance reader. Measurements of GALT enzyme activity are used primarily in the diagnosis and treatment of the hereditary disease galactosemia. This method is intended for in vitro diagnostic use as an aid in neonatal screening for decreased levels of GALT enzyme activity, and not for monitoring purposes.

This device is intended for use by trained, qualified laboratory personnel.

Prescription UseX(Part 21 CFR 801 Subpart D)AND/OROver-The-Counter Use(21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C. Benam

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102643

Page 1 of 2

{15}------------------------------------------------

Indications for Use Form

510(k) Number: K102643

Device Name: SPOTCHECK Pro™

Indications for Use:

The SPOTCHECK Pro is used for automated sample processing in the application of in vitro diagnostic assays. Specimens containing patient bodily substances are introduced and analyzed in microtiter plates using qualitative/guantitative determination through absorbance measurements.

This device and assays are intended for use by trained, qualified laboratory personnel.

Prescription UseX
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C. Benam

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

4-102643 510(k)________________________________________________________________________________________________________________________________________________________________________

Page 2 of 2

§ 862.1315 Galactose-1-phosphate uridyl transferase test system.

(a)
Identification. A galactose-1-phosphate uridyl transferase test system is a device intended to measure the activity of the enzyme galactose-1-phosphate uridyl transferase in erythrocytes (red blood cells). Measurements of galactose-1-phosphate uridyl transferase are used in the diagnosis and treatment of the hereditary disease galactosemia (disorder of galactose metabolism) in infants.(b)
Classification. Class II.