K Number
K102547

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-02-07

(157 days)

Product Code
Regulation Number
888.3080
Age Range
12 - 120
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CoRoent® Small Interlock™ System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CoRoent Small Interlock System. The CoRoent Small Interlock System is intended for use with autograft.

Device Description

The NuVasive CoRoent Small Interlock System is a standalone anterior cervical interbody device consisting of a PEEK (polyetheretherkeytone) implant cage with titanium alloy Sevious markers and washers, and three (3) titanium alloy bone fixation screws. The de roos alloy conforming to ASTM F136. The implants are available in a variety of sizes The CoRoent Small Interlock System is a to accommodate anatomical conditions. standalone system intended to be used with the bone screws provided, and when used as such requires no additional supplementary fixation systems.

AI/ML Overview

The NuVasive® CoRoent® Small Interlock™ System is an intervertebral body fusion device. The acceptance criteria and the study proving the device meets these criteria are detailed in the "Performance Data" section of the 510(k) summary.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Test Standard)Device Performance (Results)
Static and dynamic torsion (ASTM F2077)Meets or exceeds the performance of the predicate device.
Static and dynamic compression (ASTM F2077)Meets or exceeds the performance of the predicate device.
Subsidence (ASTM F2267)Meets or exceeds the performance of the predicate device.
Wear Debris (ASTM F2077 & ASTM F1877)Meets or exceeds the performance of the predicate device.

Based on these results, the device was found to be substantially equivalent to legally marketed predicate devices, indicating it is safe and effective for its intended use.

2. Sample size used for the test set and the data provenance:

The document states that "Nonclinical testing and engineering analysis were performed." However, it does not provide specific sample sizes for each test (e.g., number of implants tested). The data provenance is also not explicitly stated beyond being "nonclinical testing and engineering analysis." It does not specify country of origin or whether the data was retrospective or prospective, as these are typically not applicable to nonclinical engineering bench testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable to this submission. The "ground truth" for this device's performance is established through adherence to recognized ASTM standards for mechanical testing, not through expert consensus on medical images or clinical data. The performance is objectively measured against physical and mechanical properties defined by these standards.

4. Adjudication method for the test set:

This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies or image interpretation where human judgment is involved in establishing a ground truth. For mechanical bench testing, results are quantitative and directly measured against predefined criteria.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. This document describes a mechanical medical device, an intervertebral body fusion system, not an AI-assisted diagnostic or imaging system. Therefore, MRMC studies and the concept of "human readers improve with AI" are irrelevant to this submission.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. This device is a passive implant; it does not involve algorithms or human-in-the-loop performance in the context of an AI system. The term "standalone system" in the device description refers to the fact that it does not require additional supplementary fixation systems when used with the provided bone screws.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the "performance data" presented, the "ground truth" is defined by the established parameters and pass/fail criteria within the referenced ASTM international standards (ASTM F2077, ASTM F2267, ASTM F1877). These standards specify the test methods and acceptable performance limits for intervertebral body fusion devices.

8. The sample size for the training set:

This information is not applicable. As this is a mechanical device, there is no "training set" in the context of machine learning or AI algorithms. The design and manufacturing process are subject to quality control and design verification, but not machine learning training.

9. How the ground truth for the training set was established:

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized graphic on the left and the word "NUVASIVE" in a bold, sans-serif font on the right. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller font. The graphic on the left appears to be an abstract representation of a curved shape.

510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations. Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

Submitted by: A.

Sheila Bruschi Associate Manager, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 320-4515 Fax: (858) 320-4615

FEB - 7 2011

Date Prepared: August 30, 2010

Device Name B.

Common or Usual Name: Classification Name: Device Class: Classification: Product Code:

Trade or Proprietary Name: NuVasive® CoRoent® Small Interlock™ System Intervertebral Body Fusion Device Intervertebral Body Fusion Device Class II 8888.3080 ODP

Predicate Devices C.

The subject CoRoent Small Interlock System is substantially equivalent to the following devices previously cleared by FDA:

  • · K072415 Surgicraft Limited STALIFTM C
  • · K072981- Synthes Spine Synthes Zero-P
  • · K083389 Globus Medical Coalition™ Spacer
  • · K094042- Medtronic Sofamor Danek PEEK PREVAIL™ Cervical Interbody Device
  • · K092521- SeaSpine, Inc. Zuma-CTM
  • · K100043- NuVasive, Inc. CoRoent XLR Standalone System

Device Description D.

The NuVasive CoRoent Small Interlock System is a standalone anterior cervical interbody device consisting of a PEEK (polyetheretherkeytone) implant cage with titanium alloy Sevious markers and washers, and three (3) titanium alloy bone fixation screws. The de roos alloy conforming to ASTM F136. The implants are available in a variety of sizes The CoRoent Small Interlock System is a to accommodate anatomical conditions. standalone system intended to be used with the bone screws provided, and when used as such requires no additional supplementary fixation systems.

Page 1 of 2

{1}------------------------------------------------

K102547

Image /page/1/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized abstract shape on the left, followed by the word "NUVASIVE" in a blocky, outlined font. Below the word "NUVASIVE" is the tagline "speed of innovation" in a smaller, sans-serif font.

510(k) Premarket Notification K102547 NuVasive® CoRoent® Small Interlock™ System

E. Intended Use

The CoRoent® Small Interlock System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DD) at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CoRoent Small Interlock System. The CoRoent Small Interlock System is intended for use with autograft.

F. Technological Characteristics

As was established in this submission, the subject CoRoent Small Interlock System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes.

G. Performance Data

Nonclinical testing and engineering analysis were performed to demonstrate that the subject CoRoent Small Interlock System is substantially equivalent to other predicate devices. The following testing was performed:

  • Static and dynamic torsion per ASTM F2077 .
  • Static and dynamic compression per ASTM F2077 .
  • . Subsidence per ASTM F2267
  • Wear Debris per ASTM F2077 & ASTM F1877 .

The results of these studies showed that the subject CoRoent Small Interlock System meets or exceeds the performance of the predicate device, and the device was therefore found to be substantially equivalent.

Conclusions H.

Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject CoRoent Small Interlock System has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use.

Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus. The seal is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NuVasive, Inc. % Ms. Sheila Bruschi Associate Manager, Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121

.............................................................................................................................................................................. SEF

Re: K102547

Trade/Device Name: NuVasive® CoRoent® Small Interlock™ System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: January 10, 2011 Received: January 11, 2011

Dear Ms. Bruschi:

This letter corrects our substantially equivalent letter of February 7, 2011.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{3}------------------------------------------------

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A Millberson

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K102547

Device Name: NuVasive® CoRoent® Small Interlock™ System

Indications For Use:

The CoRoent Small Interlock™ System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease TODD at one level from C2-T1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CoRoent Small Interlock System. The CoRoent Small Interlock System is intended for use with autograft.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

.

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorauve ذنت:د:ندن

ל X102547 510(k) Number_

Page 1 of 1

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.