K Number
K102470
Date Cleared
2011-02-07

(161 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LigaSure™ Curved, Small Jav, Open Sealer/Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and lymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc.

The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Device Description

The LigaSure™ Curved, Small Jaw, Open Sealer/Divider is a sterile, single-use, hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad™ energy platform (generator) to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics during open general surgical procedures (as indicated). The ForceTriad's tissue-fusion (LigaSure) mode delivers precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion, and has been designed to produce minimal sticking, charring, or thermal spread to adjacent tissue. LigaSure instruments attach to the generator with a ten-foot containing a "smart" connector. This connector functions as a unique product identifier for device-specific recognition by the generator. The Curved, Small Jaw, Open Sealer/Divider instrument is designed to be both ergonomic and intuitive for the user. Its hemostat-style body and symmetrically placed controls facilitate handling by both left and righthanded users, and its small, curved jaws maximize visibility and access when the instrument is used in confined surgical spaces. Ring handles function to allow the user to grasp tissue by opening and closing the jaws of the instrument. The interior surfaces of the jaws contain the electrodes, which serve to ligate by delivering energy to the grasped tissue. RF energy can be activated by the user in two ways: (1) through the use of a single button incorporated into the handle body or (2) through the use of a footswitch attached to the generator. A cutting mechanism functions to mechanically divide tissue following tissue fusion. It consists of a stainless steel blade and is controlled by the user through a trigger located on the handle body.

AI/ML Overview

The submission describes the device's technological and performance characteristics compared to predicate devices, drawing upon non-clinical (electrical/mechanical/functional) performance testing, preclinical (bench tissue/animal) evaluations, and usability studies to demonstrate substantial equivalence. It does not provide explicit "acceptance criteria" for novel performance metrics nor a "study that proves the device meets the acceptance criteria" in the format of a clinical trial with pre-defined endpoints and statistical analysis. Instead, it relies on comparative testing to known predicate devices and established safety standards.

Here's a breakdown of the requested information based on the provided text, indicating where information is not explicitly stated or is not applicable in this type of submission:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present explicit "acceptance criteria" as pass/fail thresholds for novel performance metrics in a tabular format. Instead, it demonstrates substantial equivalence through comparative performance testing to predicate devices and compliance with relevant safety standards. The "reported device performance" is qualitative and comparative for most aspects.

Acceptance Criteria (Implied)Reported Device Performance
Non-Clinical:
Compliance with IEC 60601-1, 60601-2-2, 60601-1-2Performed in accordance with these standards.
Mechanical/Functional performance demonstrationVerification and comparison studies conducted. (Specific results not detailed but implied to be acceptable for equivalence)
Cutting mechanism/blade retraction performanceTest results provided. (Implied to be acceptable)
Jaw temperature performanceTest results provided. (Implied to be acceptable, minimal thermal spread claims)
Jaw force performance (compared to predicates)Jaw pressure of LF1212 shown to be lower than LS1200, but comparable to LS1520. Justified by larger jaw electrode surface area for adequate tissue fusion and division.
Button force performanceTest results provided. (Implied to be acceptable)
Cutter trigger force performanceTest results provided. (Implied to be acceptable)
Power curve performanceTest results provided. (Implied to be acceptable for energy delivery)
Preclinical (in vitro):
Isolated vessel burst pressures/hemostasis (porcine renal/pulmonary)Evaluated, demonstrating substantial equivalence to predicate devices in grasping, ligating, and dividing tissue. (Specific values not provided)
Ligation (sealing) speedEvaluated, demonstrating substantial equivalence. (Specific values not provided)
Division (cutting) performanceEvaluated, demonstrating substantial equivalence. (Specific values not provided)
Preclinical (in vivo - acute):
Overall hemostasis (isolated arteries, veins, tissue bundles)Evaluated using porcine and ovine models, demonstrating substantial equivalence. (Specific values not provided)
Overall activation timeEvaluated, demonstrating substantial equivalence. (Specific values not provided)
Overall grasping performanceEvaluated, demonstrating substantial equivalence. (Specific values not provided)
Lateral thermal damage to adjacent tissueEvaluated on sealed isolated vessels, designed to produce "minimal sticking, charring, or thermal spread to adjacent tissue." (Specific values not provided, but implicitly acceptable if "minimal")
Preclinical (in vivo - chronic):
Overall chronic hemostasis and healing (21-day survival)Evaluated using a porcine model, demonstrating substantial equivalence during acute procedures and following the survival period for isolated arteries, veins, and tissue bundles. (Specific outcomes not provided but implied to be positive and equivalent)
Usability:
Adequate assurance of safety and performance (human factors/usability)Evaluated with users in simulated operating environments through formative studies, demonstrating adequate assurance for patient and operator. (Specific metrics or pass/fail not provided but implied to be met).

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a single "test set" in the context of a clinical study. The evaluations involved various bench testing samples, excised porcine renal and pulmonary arteries, porcine and ovine models (for in vivo acute studies), and a porcine model (for in vivo chronic studies). The specific number of each type of sample/animal used is not detailed.
  • Data Provenance:
    • In vitro: Excised porcine renal and pulmonary arteries.
    • In vivo: Porcine and ovine models.
    • Usability: "Simulated operating environments" with "users."
    • The origin (e.g., country) of these animal models or where the usability studies were conducted is not specified. All studies are preclinical/non-clinical, not human clinical data, so the retrospective/prospective distinction for human subjects is not directly applicable.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable in the context of this 510(k) submission. Ground truth for performance (e.g., burst pressure, thermal spread, healing) was established through objective measurements in bench and animal studies, not through expert human interpretation of images or other data typically requiring adjudicated ground truth. Usability studies involved "users," but these are for evaluating human factors, not for establishing diagnostic ground truth.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used for establishing ground truth in clinical imaging studies where multiple human readers interpret data. The evaluations here are based on objective physical measurements and animal model outcomes.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This device is a surgical instrument, not an AI-powered diagnostic or assistive tool for human readers of medical images. No MRMC study was conducted.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a physical electrosurgical instrument. It operates with a human user in the loop. The "standalone" concept refers to algorithms performing diagnostic tasks without human intervention, which is not relevant here.

7. The Type of Ground Truth Used

  • Objective Measurements: For electrical, mechanical, and functional performance (e.g., jaw temperature, force, power curves).
  • Biological Endpoints in Animal Models: For preclinical evaluations, the ground truth was based on measurable biological outcomes such as:
    • Isolated vessel burst pressures/hemostasis
    • Ligation (sealing) speed
    • Division (cutting) performance
    • Overall hemostasis
    • Overall activation time
    • Overall grasping performance
    • Lateral thermal damage
    • Chronic hemostasis and healing over a 21-day survival period.

8. The Sample Size for the Training Set

Not applicable. This device did not involve machine learning or AI that would require a "training set" for an algorithm. The development and testing focused on engineering, materials science, and preclinical biological performance.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set for an algorithm was used.

{0}------------------------------------------------

FEB = 7 2011

K10247

Image /page/0/Picture/1 description: The image shows the word "COVIDIEN" in black letters. To the left of the word is a black square with a white rectangle inside. The word is written in a bold, sans-serif font. The image is simple and clear.

510(k) Summary

Date summary prepared: 2/4/2011

510(k) Submitter/Holder

Covidien, formerly Valleylab, a division of Tyco Healthcare 5920 Longbow Drive Boulder, CO 8030 I

Contact

David M. Horton Project Manager, Regulatory Affairs Telephone: 303-530-6391 Fax: 303-516-8307 Email: david.m.horton@covidien.com

Name of Device

Trade Name: LigaSure™ Curved, Small Jaw, Open Sealer/Divider Common Name: Bipolar electrosurgical instrument Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, class II, GEI).

Predicate Devices

The LigaSure™ Curved, Small Jaw, Open Sealer/Divider was compared and found to be substantially equivalent to the following products of comparable type in commercial distribution:

Device Common Name: Bipolar electrosurgical instrument Trade Name: LigaSure™ Precise Instrument Vessel Sealing System Catalog Number: LS1200 510(k) Number: K010010 (cleared 4/2/2001) Manufacturer: Covidien, formerly Valleylab, a division of Tyco Healthcare

Device Common Name: Bipolar electrosurgical instrument Trade Name: LigaSure™ 5mm Laparoscopic Sealer-Divider Catalog Number: LS1520 510(k) Number: K031011 (cleared 5/29/2003) Manufacturer: Covidien, formerly Valleylab, a division of Tyco Healthcare

Device Description

The LigaSure™ Curved, Small Jaw, Open Sealer/Divider is a sterile, single-use, hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad™ energy platform (generator) to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics during open general

{1}------------------------------------------------

surgical procedures (as indicated). The ForceTriad's tissue-fusion (LigaSure) mode delivers precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion, and has been designed to produce minimal sticking, charring, or thermal spread to adjacent tissue,

LigaSure instruments attach to the generator with a ten-foot containing a "smart" connector. This connector functions as a unique product identifier for device-specific recognition by the generator. The Curved, Small Jaw, Open Sealer/Divider instrument is designed to be both ergonomic and intuitive for the user. Its hemostat-style body and symmetrically placed controls facilitate handling by both left and righthanded users, and its small, curved jaws maximize visibility and access when the instrument is used in confined surgical spaces.

Ring handles function to allow the user to grasp tissue by opening and closing the jaws of the instrument. The interior surfaces of the jaws contain the electrodes, which serve to ligate by delivering energy to the grasped tissue. RF energy can be activated by the user in two ways: (1) through the use of a single button incorporated into the handle body or (2) through the use of a footswitch attached to the generator. A cutting mechanism functions to mechanically divide tissue following tissue fusion. It consists of a stainless steel blade and is controlled by the user through a trigger located on the handle body.

Intended Use

The LigaSure™ Curved, Small Jaw, Open Sealer/Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and lymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc.

The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Technological and Performance Characteristics

A detailed comparison of the LigaSure Curved, Small Jaw, Open Sealer/Divider LigaSure LS1200 and LS1520 predicate instruments found several technological and performance similarities and very few relevant differences. The following six similarities were identified and discussed in the submission: (1) basic RF tissue fusion technology and its fundamental variables, (2) basic user (manual) control mechanisms, (3) system compatibility, (4) functional materials, (5) dimensions, shape, and weight, and (6) non-clinical and preclinical performance characteristics.

Three differences were found to be relevant when comparing the LigaSure LF1212 instrument to the predicate devices. These differences, identified below, were found to not pose additional or different risks when compared to the instruments already in commercial distribution.

(1) Energy activation mechanisms: The hand-activated button of the LigaSure LF1212 instrument is a single two-stage button design activated through the use of the hemostat-style handles, whereas the hand activation button of the LS1520 instrument is a single-stage design actuated independent of the handle body. The placement of the hand-activated button in this location reduces the number of finger-operated controls. In addition, the ForceTriad's LigaSure touchscreen displays a "hand-activation-on" button (icon). The use of the hand-activated button can be disabled by the user by touching this icon when helshe desires to only activate energy using a footswitch.

{2}------------------------------------------------

(2) Latching/Ratcheting mechanism: Unlike the predicate devices, the LigaSure LF1212 instrument is not designed with a latching/ratcheting mechanism. The non-latching design of the LF1212 instrument is influenced by the placement of the hand-activated button. The location of the hand-activated button requires the instrument handles to be completely closed in order to activate energy, thereby not requiring a latching mechanism in order to assure adequate jaw pressure during activation.

(3) Jaw force/pressure: The jaw pressure of the LF 1212 instrument was shown to be lower than the jaw pressure of the LS1200. This lower pressure is due primarily to the 35 percent larger jaw electrode surface area when compared to the LS1200 instrument. A larger electrode surface area is necessary for the instrument to be able to divide the sealed tissue while achieving adequate tissue fusion performance. Notably, the jaw pressure of the LS1520 instrument was found to be comparable to the jaw pressure of the LF1212 instrument.

Non-Clinical/Preclinical Performance

Evidence of safety and effectiveness was obtained from three primary areas: (1) non-clinical (electrical/mechanical/functional) performance testing, (2) preclinical (bench tissue/animal) evaluations and testing, and (3) usability studies.

Non-clinical: Basic safety and performance testing was performed in accordance with IEC 60601-1. IEC 60601-2-2, and 60601-1-2. In addition, verification and comparison studies were conducted to evaluate the mechanical and functional performance. Specifically, test results in the following areas were provided: mechanical/functional, cutting mechanism/blade retraction, jaw temperature, jaw force, button force, cutter trigger force, and power curve performance.

Preclinical: Evidence obtained from preclinical bench tissue (in vitro) and animal (in vivo acute and chronic) studies using porcine and ovine models (and in a variety of tissue types and conditions) demonstrate that the LigaSure Curved, Small Jaw, Open Sealer/Divider performs substantially equivalent to the predicate devices in relevant aspects associated with grasping, ligating, and dividing tissue; namely,

Evaluated in vitro using excised porcine renal and pulmonary arteries -isolated vessel burst pressures/hemostasis

-ligation (sealing) speed

-division (cutting) performance

Evaluated in vivo/acute using porcine and ovine models '

-overall hemostasis (isolated arteries, veins, tissue bundles)

-overall activation time

-overall grasping performance

-lateral thermal damage to adjacent tissue (measured on sealed isolated vessels)

Evaluated in vivo/chronic using a porcine model1 -overall chronic hemostasis and healing during acute procedures and following a 21-day survival period (isolated arteries, veins, tissue bundles)

'evaluation in vivo was necessary because the characteristics requiring evaluation could not be adequately studied in vitro.

Usability: In addition to non-clinical testing, usability was evaluated with users in simulated operating environments. These studies consisted of several formative study,

{3}------------------------------------------------

which demonstrate the instrument provides adequate assurance of safety and performance (in regards to human factors/usability aspects) for the patient and operator.

Clinical Performance

This premarket notification report does not rely on the assessment of human clinical trial data to demonstrate substantial equivalence.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird-like figure. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

FEB 7 2011

Covidien % Mr. David Horton 5920 Longbow Drive Boulder, CO 80301

Re: K102470

Trade/Device Name: LigaSure Curved, Small Jaw, Open Sealer/Divider Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: January 28, 2011 Received: January 31, 2011

Dear Mr. Horton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially-equivalent-(for the-indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2) i CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

:

: :

{5}------------------------------------------------

Page 2 – Mr. David Horton

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Aling B. 12 h

Mark N. Melkerson Director Division of Surgical, Orthonedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K 102470

Device Name: LigaSure™ Curved, Small Jaw, Open Sealer/Divider

The LigaSure™ Curved, Small Jav, Open Sealer/Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and lymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc.

The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Prescription Use ✔ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

K102470 510(k) Number

K102470: LigaSure™ Curved, Small Jaw, Open Sealer/Divider

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.