K Number
K102323
Manufacturer
Date Cleared
2010-12-16

(121 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cervical Standalone Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Cervical Standalone Intervertebral Body Fusion Device implants are to be used with autogenous bone graft and implanted via an anterior approach. The cervical device is to be used in patients who have had six weeks of non-operative treatment. The Cervical Standalone Intervertebral Body Fusion Device is to be used with two titanium alloy screws which accompany the implant.

Device Description

The Cervical Standalone Intervertebral Body Fusion Device is comprised of PEEK-OPTIMA® LT cages which are available in a variety of sizes. The cages include toothed spikes on the top and bottom surfaces to engage with the superior and inferior end plates of neighboring vertebral bodies to resist rotation and migration. The cage shape and open center allow for bony in-growth in and around the implant. A single cage is sufficient to be used at each intervertebral level. Screws are inserted through the anterior titanium face and screwed into the vertebral bodies for bony fixation.

AI/ML Overview

The acceptance criteria for the Cervical Standalone Intervertebral Body Fusion Device are based on demonstrating substantial equivalence to a predicate device, the COALITION™ Spacer Intervertebral Body Fusion Device (K083389). The study proving the device meets these criteria involved mechanical testing.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Technological EquivalenceEquivalent Indications for UseMeets: Identical indications for use.
Equivalent Design ConceptsMeets: Same basic design.
Equivalent Feature ComparisonsMeets: Similar features.
Equivalent Biocompatible MaterialsMeets: Uses established, known materials (PEEK-OPTIMA LT, Tantalum, Titanium 6AI-4V ELI), which are equivalent to the predicate.
Equivalent Operating PrincipleMeets: Same operating principle.
Equivalent Manufacturing EnvironmentMeets: Same manufacturing environment.
Equivalent Sterilization ProcessMeets: Same sterilization process.
Equivalent Packaging ConfigurationsMeets: Same packaging configurations.
Mechanical PerformanceStatic Compression per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Static Compression Shear per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Static Torsion per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Static Load Induced Subsidence per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Static Expulsion per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Dynamic Compression per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Dynamic Compression Shear per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Dynamic Torsion per ASTM F-2077Performed as designed and met or exceeded all product specifications.
Wear TestingCharacteristics of particulate wear debrisThe result from the wear testing was found to be substantially equivalent to legally marketed devices.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document does not specify a distinct "test set" in terms of patient data or
    clinical samples. The testing described is primarily mechanical and wear testing of the device itself.
  • Data Provenance: Not applicable as the testing is non-clinical, mechanical testing of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. The ground truth for this device's acceptance is based on adherence to engineering standards (ASTM F-2077 and F-2267) and comparison to a predicate device, not on expert clinical review of test data.

4. Adjudication method for the test set

  • Not applicable. This was a non-clinical, mechanical testing study against established ASTM standards and
    predicate device equivalence.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a
    physical implant, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a physical implant, not an algorithm. The "standalone" in the device name refers to its design as an intervertebral fusion device that does not require an anterior plate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the mechanical testing, the ground truth was defined by established engineering standards (ASTM F-2077 and F-2267) and the performance specifications of the device itself, with the aim of demonstrating that the device performed "as designed" and met/exceeded these specifications.
  • For substantial equivalence, the ground truth was based on the characteristics and performance of the legally marketed predicate device (K083389).

8. The sample size for the training set

  • Not applicable. This is a physical medical device undergoing mechanical validation, not a machine
    learning model requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable, as there was no training set for a machine learning model.

{0}------------------------------------------------

510(k) Summary (21 CFR Part 807.92)

Submitter Information A.

Submitter's Name: Theken Spine, LLC DEC 16 2010 Address: 1800 Triplett Blvd. Akron, Ohio 44306 Telephone Number: 330-475-8600 Fax Number: 330-773-7697 Contact Person: Dale Davison Date Prepared: 08/16/2010 B. Device Information Trade Name: Cervical Standalone Intervertebral Body Fusion Device Common Name: Cervical Standalone Intervertebral Body Fusion Device Classification Name: Spinal Intervertebral Body Fusion Device (per 21 CFR 888.3080) Device Classification: Class 11 (per 21 CFR 888.3080) Panel: Orthopedic, Product Code: ODP Medtronic Sofamor Danek, PEEK PREVAIL™ Cervical Interbody Predicate Device: Device (K073285) Globus Medical Inc. COALITIONTM Spacer Intervertebral Body Fusion Device (K083389) Material Composition: Polyetherctherketone (PEEK-OPTIMA LT) per ASTM F-2026 Tantalum per ASTM F-560, and Titanium 6AI-4V ELI per ASTM F-136 Subject Device Description: The Cervical Standalone Intervertebral Body Fusion Device is comprised of PEEK-OPTIMA® LT cages which are available in a variety of sizes. The cages include toothed spikes on the top and bottom surfaces to engage with the superior and inferior end plates of neighboring vertebral bodies to resist rotation and migration. The cage shape and open center allow for bony in-growth in and around the implant. A single cage is sufficient to be used at each intervertebral level. Screws are inserted through the anterior titanium face and screwed into the vertebral bodies for bony fixation. Intended Use: The Cervical Standalone Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Cervical Standalone Intervertebral Body Fusion Device implants are to be used with autogenous bone graft and implanted via an anterior approach. The cervical device is to be used in patients who have had six weeks of non-operative treatment. The Cervical Standalone Intervertebral Body Fusion Device is to be used with two titanium alloy screws which accompany the implant.

{1}------------------------------------------------

The technological characteristics of the Cervical Technological Characteristics: Standalone Intervertebral Body Fusion Device are equivalent to the predicate device COALITION Spacer Intervertebral Body Fusion Device (K083389) manufactured by Globus Medical Inc. These technological characteristics include the same indications for use, design concepts, feature comparisons, and known biocompatible materials. Summary of test data: Mechanical testing of the subject device consisted of static compression, static compression shear, static torsion, static load

induced subsidence, static expulsion, dynamic compression, dynamic compression shear and dynamic torsion. All testing was conducted per ASTM F-2077 and ASTM F-2267 guidelines and the device performed as designed and met or exceeded all product specifications. In addition, wear testing was performed in order to determine the characteristics of any particulate wear debris generated during the tests. The result from the wear testing was found to be substantially equivalent to legally marketed devices.

C. Substantial Equivalence

Theken Spine believes that sufficient evidence exists to reasonably conclude that the Cervical Standalone Intervertebral Body Fusion Device is substantially equivalent to the predicate device COALITION™ Spacer Intervertebral Body Fusion Device (K083389) manufactured by Globus Medical Inc. This is based on the design concept, the use of established, known materials, feature comparisons, mechanical testing, indications for use, pre-production quality assurance planning analysis. All implants o represent a basic design concept in terms of safety and effectiveness, and differ only in minor details.

Substantial equivalence include:

  • The same Indications for use ●
  • The same basic design ●
  • The same operating principle ●
  • . The same materials
  • The same manufacturing environment
  • The same sterilization process .
  • The same packaging configurations .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, with three human profiles incorporated into the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Theken Spine, LLC % Mr. Dale Davison 1800 Triplett Boulevard Akron, OH 44306

Re: K102323

Trade/Device Name: Cervical Standalone Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: November 12, 2010 Received: November 15, 2010

SEP 1 2 201

Dear Mr. Davison:

This letter corrects our substantially equivalent letter of December 16, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not

{3}------------------------------------------------

limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark A. Wilkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

DEC 1 6 2010

Device Name: Cervical Standalone Intervertebral Body Fusion Device

Indications For Use: .

The Cervical Standalone Intervertebral Body Fusion Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Cervical Standalone Intervertebral Body Fusion Device implants are to be used with autogenous bone graft and implanted via an anterior approach. The cervical device is to be used in patients who have had six weeks of non-operative treatment. The Cervical Standalone Intervertebral Body Fusion Device is to be used with two titanium alloy screws which accompany the implant.

Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

$\overline{R.i.m}$ B.nh In nol)

(Division Sign Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number K102323

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.