(261 days)
GENOSS Collagen Membrane is intended to be used during guided bone and guided tissue regeneration procedures as a biodegradable barrier membrane into bony defects surrounding teeth and on alveolar ridges.
- Periodontal/infrabony defects
- Ridge augmentation
- Extraction sites (implant preparation/placement)
- Guided bone regeneration (GBR) procedures
- Sinus lifts
GENOSS Collagen Membrane is a resorbable membrane to aid the regenerative healing of bone or bone/periodontal ligament defects of the oral cavity. GENOSS Collagen Membrane is designed to be a passive barrier which excludes epithelial and gingival connective tissue from the defect site so that only the desirable cells repopulate the space, allowing regeneration to occur. Its resorbability eliminates the need for a second surgery. GENOSS Collagen Membrane is made from bovine tendon, the Type 1 collagen and is sterilized in gamma-radiation.
The provided document, a 510(k) summary for the "GENOSS Collagen Membrane," describes the device, its intended use, and a series of performance tests to demonstrate its substantial equivalence to predicate devices. However, this document does not contain acceptance criteria for specific performance metrics in a quantitative sense, nor does it detail a study that proves the device meets such criteria within the typical framework of an AI/software device evaluation.
Instead, the document focuses on biocompatibility testing and in-vitro performance bench testing for a medical device (collagen membrane) used in dental procedures. The "acceptance criteria" are implied to be the successful completion of these tests with results indicating non-toxicity, non-sensitization, and meeting "established specifications necessary for consistent performance according to its intended use."
The information requested for AI device evaluations, such as sample sizes for test and training sets, data provenance, ground truth establishment by experts, and MRMC studies, are not applicable to this type of medical device (collagen membrane) and the information provided in this 510(k) submission. This is a premarket notification for a physical resorbable barrier membrane, not an AI or software-as-a-medical-device (SaMD) product.
Therefore, the following table and explanations are adapted to the context of the provided document, highlighting what is reported rather than what would be expected for an AI device.
1. Table of Acceptance Criteria and Reported Device Performance
As explicit, quantitative acceptance criteria are not stated for this physical medical device in the document, the table below summarizes the types of tests performed and the general performance outcome reported. The acceptance criteria are broadly implied as "meeting established specifications" or "non-toxic and non-sensitizing."
| Acceptance Criteria Category (Implied) | Reported Device Performance |
|---|---|
| Biocompatibility | - ISO Cytotoxicity: Completed, non-toxic |
| - ISO Systemic toxicity: Completed, non-toxic | |
| - ISO Pyrogenicity: Completed, non-pyrogenic | |
| - ISO Sensitization: Completed, non-sensitizing | |
| - ISO Intracutaneous reactivity: Completed, non-reactive | |
| - ISO Implantation: Completed, biocompatible to tissue | |
| - ISO Genotoxicity (Ames): Completed, non-genotoxic | |
| - ISO Genotoxicity (Micro-nucleus): Completed, non-genotoxic | |
| - ISO Haemocompatibility: Completed, haemocompatible | |
| - ISO Elution assay: Completed, non-toxic eluates | |
| - ISO Genotoxicity (Sister chromatid): Completed, non-genotoxic | |
| In-vitro Performance | - Shrink temperature (DSC): Evaluated, met specifications |
| - Tensile strength: Evaluated, met specifications | |
| - Decomposition time: Evaluated, met specifications | |
| - Swelling: Evaluated, met specifications | |
| Raw Material Safety | - Virus inactivation: Evaluated in accordance with ASTM F2212-08, confirmed safety |
| - Endotoxin: Evaluated in accordance with ASTM F2212-08, confirmed safety | |
| - Amino acid analysis: Evaluated in accordance with ASTM F2212-08, confirmed safety | |
| - SDS-PAGE: Evaluated in accordance with ASTM F2212-08, confirmed safety | |
| Overall Conclusion | "All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the GENOSS Collagen Membrane met the established specifications necessary for consistent performance according to its intended use." |
Study Details (as applicable to a physical medical device submission):
-
Sample size used for the test set and the data provenance:
- The document does not specify sample sizes for the biocompatibility or in-vitro performance tests. These tests are typically performed on a statistically relevant number of samples to ensure reproducibility and reliability, but the exact number isn't reported in the summary.
- Data provenance: Not explicitly stated, meaning no country of origin or whether it's retrospective/prospective data as these concepts are generally confined to clinical studies or data-driven AI models rather than benchtop testing of material properties. The tests were likely conducted in a controlled laboratory environment by the manufacturer or a contract research organization.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. "Ground truth" in the context of expert consensus is typically relevant for diagnostic AI algorithms. For a physical medical device like a collagen membrane, the "truth" is established through standardized laboratory tests (e.g., ISO standards for biocompatibility, ASTM for material properties) which have defined methods and measurable outcomes, rather than expert interpretation of a diagnostic finding.
-
Adjudication method for the test set:
- Not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among multiple human readers/experts when establishing ground truth for diagnostic classifications. This is not relevant for the bench and biocompatibility testing of a physical membrane.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This submission is for a physical medical device (collagen membrane), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed or is relevant.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This submission is for a physical medical device, not an algorithm.
-
The type of ground truth used:
- The "ground truth" for this device's performance is based on objective, standardized laboratory measurements and observations as defined by ISO and ASTM standards (e.g., cell viability in cytotoxicity tests, temperature readings for shrink temperature, chemical composition analysis, etc.). It is not based on expert consensus, pathology, or outcomes data in the sense typically discussed for AI devices.
-
The sample size for the training set:
- Not applicable. This is a physical medical device, not an AI algorithm, so there is no "training set."
-
How the ground truth for the training set was established:
- Not applicable. As there is no training set, this question is not relevant.
{0}------------------------------------------------
Genoss Co., Ltd. 1F Gyeonggi R&DB center, 906-5 Iui-dong, Suwon-si, Yeongtong-gu, Gyeonggi-do, 433-270, Korea Tel. 82-31-888-5100, Fax. 82-31-888-5105, Web: www.genoss.com
Image /page/0/Picture/1 description: The image shows the text "MAY - 4 2011" on the top line. Below that is the word "GENOSS" in a bold, stylized font. The text appears to be a date and a company or product name.
510(k) Summary
1. Company
| Submitter | |
|---|---|
| Name | GENOSS Co., Ltd. |
| Address | 1F Gyeonggi R&DB center, 906-5 lui-dong, Suwon-si, Yeongtong-gu,Gyeonggi-do, 433-270, Korea |
| Phone/Fax | +82-31-888-5100/ +82-31-888-5105 |
| Contact Person | Hoseog Cha / QA Team leaderhscha@genoss.com |
| Summary Date | March 8, 2011 |
2. Device Name
| Proprietary name: | GENOSS Collagen Membrane |
|---|---|
| Common name: | Resorbable dental barrier membrane, animal source |
| Classification name: | Barrier, animal source, dental |
3. Predicated Device
RCM6 membrane (K003339) Fortgen Oral Membrane (K071555) OSSIX Plus (K053260)
4. Description
GENOSS Collagen Membrane is a resorbable membrane to aid the regenerative healing of bone or bone/periodontal ligament defects of the oral cavity.
GENOSS Collagen Membrane is designed to be a passive barrier which excludes epithelial and gingival connective tissue from the defect site so that only the desirable cells repopulate the space, allowing regeneration to occur. Its resorbability eliminates the need for a second surgery.
GENOSS Collagen Membrane is made from bovine tendon, the Type 1 collagen and is sterilized in gamma-radiation.
{1}------------------------------------------------
GENOSS
| Model name | Size | ||
|---|---|---|---|
| Width (mm) | Length (mm) | Thickness (mm) | |
| GCM1020 | 10 | 20 | |
| GCM1515 | 15 | 15 | |
| GCM1520 | 15 | 20 | |
| GCM2030 | 20 | 30 | 0.3 |
| GCM2530 | 25 | 30 | |
| GCM3040 | 30 | 40 |
5. Indication for use
GENOSS Collagen Membrane is intended to be used during guided bone and guided tissue regeneration procedures as a biodegradable barrier membrane into bony defects surrounding teeth and on alveolar ridges.
- Periodontal/infrabony defects
- Ridge augmentation
- Extraction sites (implant preparation/placement)
- Guided bone regeneration (GBR) procedures
- Sinus lifts
6. Technological Characteristics
GENOSS Collagen membrane has the similar technological characteristics as the predicate device; main material, indication for use and design. Technological characteristics of GENOSS Collagen membrane, RCM6 membrane, FORTAGEN ORAL membrane, OSSIX -PLUS are as following
| GENOSS Collagen Membrane | |
|---|---|
| Indication for use | GENOSS Collagen Membrane is intended to be used during guided bone andguided tissue regeneration procedures as a biodegradable barrier membrane intobony defects surrounding teeth and on alveolar ridges.- Periodontal/infrabony defects- Ridge augmentation- Extraction sites (implant preparation/placement)- Guided bone regeneration (GBR) procedures- Sinus lifts |
| Material |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "GENOSS" in bold, black letters. The letters are sans-serif and appear to be slightly slanted. The word is horizontally oriented and takes up most of the frame. The background is white.
| Children and country and consider of the consistedCrosslinking | Carbodiimide (EDC) |
|---|---|
| Sterilization | Gamma Irradiation------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| RCM6 membrane | |
|---|---|
| Indication for use | RESORBABLE COLLAGEN MEMBRANE is intended for use in oral surgicalprocedures as a resorbable material for placement to aid in wound healing in theareas of dental implants, bone defects, periodontal defects and ridgeaugmentation.- Ridge Augmentation- Bone Regeneration around Implants- Sinus Window- Extraction Sites- Periodontal Defects |
| Material | Type I bovine collagen |
| Crosslinking | Formaldehyde (FA) |
| Sterilization | Gamma Irradiation |
| FORTAGEN ORAL MEMBRANE | |
|---|---|
| Indication for use | FortaGen Oral Membrane is intended to be used during guided bone and guided |
| tissue regeneration procedures as a biodegradable barrier membrane for: | |
| - Ridge augmentation for later implant insertions; | |
| - Simultaneous ridge augmentation and implant insertions; | |
| - Ridge augmentation around implants inserted in delayed extraction sites; | |
| - Ridge augmentation around implants inserted in immediate extraction sites; | |
| - Alveolar ridge preservation consequent to tooth (teeth) extraction(s); | |
| - Over the window in lateral window sinus elevation procedures; | |
| - In implants with vertical bone loss due to infection, | |
| - In implants with vertical bone loss due to infection, only in cases where | |
| satisfactory debridement and implant surface disinfection can be achieved; | |
| - In intra bony defects around teeth | |
| - For treatment of recession defects, together with coronally positioned flap; | |
| - In furcation defects in multi-rooted teeth. |
.
{3}------------------------------------------------
102307
Image /page/3/Picture/1 description: The image shows the word "GENOSS" in a bold, sans-serif font. The letters are black and the background is white. The letters are slightly distorted, with some of the edges appearing jagged.
| Material | Type I porcine collagen |
|---|---|
| Crosslinking | Carbodiimide (EDC) |
| Sterilization | Gamma Irradiation |
| OSSIX -PLUS | |
|---|---|
| Indication for use | OssixHt-PLUS biodegradable collagen membrane is intended for use during theprocess of guided bone regeneration (GBR) and guided tissueregeneration (GTR) as a biodegradable barrier for:- Ridge augmentation for later implant insertions.- Simultaneous ridge augmentation and implant insertions.- Ridge augmentation around implants inserted in delayed extraction sites.- Ridge augmentation around implants inserted in immediate extraction sites.- Alveolar ridge preservation consequent to tooth (teeth) extraction(s).- Over the window in lateral window sinus elevation procedures.- In implants with vertical bone loss due to infection, only in cases wheresatisfactory debridment and implant surface disinfection can be achieved.- In intra bony defects around teeth.- For treatment of recession defects, together with coronally positioned flap- In furcation defects in multi rooted teeth. |
| Material | |
| Crosslinking | |
| Sterilization |
7. Performance Data
Biocompatibility testing on the proposed GENOSS Collagen Membrane has been completed. Requirements for biological evaluation of the proposed device were based on the Blue Book Memorandum G95-1 issued on May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices, Part I: Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed:
-ISO Cytotoxicity
-ISO Systemic toxicity
-ISO Pyrogencity -ISO Sensitization
510(k) Summary: K102307
{4}------------------------------------------------
102307
Image /page/4/Picture/1 description: The image contains the word "GENOSS" in all caps. The letters are bold and black. The font is sans-serif.
-ISO Intracutaneous reactivity -ISO Implantation -ISO Genotoxicity(Ames)
-ISO Genotoxicity(Micro-nucleus) -ISO Haemocompatibility -ISO Elution assay
-ISO Genotoxicity(Sister chromatid)
The proposed GENOSS Collagen Membrane was evaluated using the following in-vitro performance bench testing to confirm the performance characteristics:
-Shrink temperature (DSC) -Tensile strength
-Decomposition time -Swelling
Also, raw material was evaluated in accordance with ASTM F2212-08 to confirm the safety of collagen raw material:
| -Virus inactivation | -Amino acid analysis |
|---|---|
| -Endotoxin | -SDS-PAGE |
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the GENOSS Collagen Membrane met the established specifications necessary for consistent performance according to its intended use.
8. Conclusion
Based on the information provided in this premarket notification GENOSS Co., Ltd. concludes that GENOSS Collagen Membrane is substantially equivalent to predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Genoss Company, Limited C/O Dr. Eunkyung Son Dentium USA 11075 Knott Avenue, Suite A Cypress, California 90630
- 4 2011 MAY
Re: K102307
Trade/Device Name: GENOSS Collagen Membrane Regulation Number: 21 CFR 872,3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: April 26, 2011 Received: May 2, 2011
Dear Dr. Son:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Dr. Son
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Ph for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
GENOSS
Indication for use
510(k) Number:
长102307
Device Name:
GENOSS Collagen Membrane
Indication for use:
GENOSS Collagen Membrane is intended to be used during guided bone and guided tissue regeneration procedures as a biodegradable barrier membrane into bony defects surrounding teeth and on alveolar ridges.
- Periodontal/infrabony defects
- Ridge augmentation
- Extraction sites (implant preparation/placement)
- Guided bone regeneration (GBR) procedures
- Sinus lifts
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runne
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Indication for use
510(k) Number:
Page 1 of 1
§ 872.3930 Bone grafting material.
(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.