(109 days)
HASS Zirtooth is indicated for use as a substructure for porcelain fused ceramic fixed dental restorations; namely crown, bridges, inlays, and onlays.
HASS Zirtooth is a zirconia porcelain product to be used to make the core of the all porcelain crown. It consists of ceramic blocks and corresponds to ISO 6872 Type 2 Class 1. This ceramic product provides the supporting structure for mounting the ceramic restoration. The shelf life of the ceramic block is semi-permanent.
The provided text describes a 510(k) submission for a dental device, "HASS Zirtooth," and focuses on its substantial equivalence to a predicate device based on physical characteristics. However, it does not contain information about a clinical study involving human subjects or AI performance. The summary details engineering and materials testing, not a clinical trial with specific performance metrics like sensitivity, specificity, or reader improvement.
Therefore, many of the requested sections cannot be filled from the provided text.
Here's a breakdown of what can be extracted and what cannot, based on your request:
Acceptance Criteria and Study Details for HASS Zirtooth (HASS Corp.)
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Standards) | Reported Device Performance (HASS Zirtooth) |
|---|---|
| ISO 6872 | Met the test criteria |
| ISO 9693 | Met the test criteria |
Note: The document states: "The following properties were tested for the device according to ISO 6872 and 9693 and all the results met the test criteria." It lists some of the properties under each standard (e.g., Uniformity, Extraneous Materials, Chemical Solubility, Flexural Strength, Radioactivity of dental ceramic, Flexural strength after the low-temperature deterioration for ISO 6872; Coefficient of thermal expansion and Bonding strength with veneer ceramic for ISO 9693). Specific numerical thresholds for acceptance criteria and the exact numerical performance values are not provided in this summary.
2. Sample size used for the test set and the data provenance
- Sample Size for test set: Not specified. The study was a materials engineering test, not a clinical study on a patient population.
- Data Provenance: Not applicable in the context of clinical data. It's an in-vitro materials test.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This question is not applicable as the study was based on material property testing according to established ISO standards, not expert clinical assessment for ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. This relates to clinical interpretation and adjudication, which was not part of this materials standard testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This submission is for a dental material, not an AI-powered diagnostic device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a ceramic material, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this device's performance was adherence to material property specifications defined by ISO 6872 and ISO 9693 standards.
8. The sample size for the training set
- Not applicable. There was no "training set" as this is a physical material, not a machine learning model.
9. How the ground truth for the training set was established
- Not applicable. There was no "training set" or corresponding ground truth establishment in the context of machine learning.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for HASS Corporation. The logo is in black and features a bold, sans-serif font. The letters are closely spaced and have a modern, industrial look. The word "CORPORATION" is written in a smaller font size below the main logo.
510(k) Summary (K102305) ,
This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 10/28/2010
-
- Submission Sponsor
| Submitter | |
|---|---|
| Name | HASS CORP. |
| Address | Rm.301 & 306, Venture Plaza II, |
| Block Ga-2-3, Gangneung Science Park, Daejeon-dong, | |
| Gangneung, Gangwon-do, Korea | |
| Phone | Tel: +82-70-7712-1300 |
| Fax | Fax: +82-33-644-1231 |
-
- Submission Correspondent
LK Consulting Group 2341 W. Crescent Ave. #3, Anaheim, CA 92801 Priscilla Chung Phone: 714-844-2612 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
- Submission Correspondent
- Device
Trade Name: HASS Zirtooth Common Name: Dental Frame Material for Dental Prosthesis Classification Name: Porcelain Powder for Clinical Us Classification regulation: 21 CFR 872.6660
4. Predicate Device:
Vita In-Ceram YZ® Cubes for Cerec® (K022996), VITA Zahnfabrik GmbH & Co. KG
-
- Description:
HASS Zirtooth is a zirconia porcelain product to be used to make the core of the all porcelain crown. It consists of ceramic blocks and corresponds to ISO 6872 Type 2 Class 1. This ceramic product provides the supporting structure for mounting the
- Description:
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for HASS Corporation. The logo is in black and white. The word "HASS" is in large, bold letters. Below the word "HASS" is the word "CORPORATION" in smaller letters.
ceramic restoration. The shelf life of the ceramic block is semi-permanent.
-
- Indication for use:
HASS Zirtooth is indicated for use as a substructure for porcelain fused ceramic fixed dental restorations: namely crown, bridges, inlays, and onlays.
- Indication for use:
-
- Safety and Effectiveness:
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device as safe and effective as the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
- Safety and Effectiveness:
It has been shown in this 5109(K) submission that the differences between the HASS Zirtooth and the predicate device do not raise any questions regarding its safety and effectiveness.
-
- Physical Characteristics
The following properties were tested for the device according to ISO 6872 and 9693 and all the results met the test criteria.
- Physical Characteristics
-
ISO 6872 Uniformity, Extraneous Materials, Chemical Solubility, Flexural Strength, Radioactivity of dental ceramic and Flexural strength after the low temperature deterioration
-
· ISO 9693 Coefficient of thermal expansion and Bonding strength with veneer ceramic .
-
- Conclusion
Based on the information provided in this premarket notification, HASS Zirtooth is safe, effective and substantially equivalent to the predicate devices as described herein.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Hass Corporation C/O Ms. Priscilla Chung LK Consulting Group 2341 W. Crescent Avenue #3 Anaheim, California 92801
Re: K102305
Trade/Device Name: Hass Zirtooth Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: October 13, 2010 Received: November 12, 2010
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
DEC - 3 2010
{3}------------------------------------------------
Page 2- Ms. Chung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), i please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance-and Biometrics/Division-of-Postmarket-Surveillance:
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh. for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
DEC - 3 2010
510(k) Number (if known): K102305
Device Name: Hass Zirtooth
Indications for Use:
HASS Zirtooth is indicated for use as a substructure for porcelain fused ceramic fixed dental restorations; namely crown, bridges, inlays, and onlays.
Prescription Use (Per 21 CFR 801 Subpart D)
AND
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEE DED)
Concurrence of CDRH, Office of Device Evaluation
וויוצועיו בוקר Division of Anesthesicilogy, General Hue, .... Infection Control, Dental Devices
Number: K102305
§ 872.6660 Porcelain powder for clinical use.
(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.