K Number
K102251
Device Name
DASH KNEE
Manufacturer
Date Cleared
2011-05-17

(281 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DASH knee is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to the anatomy. The system aids the surgeon to accurately navigate a knee prosthesis to the intraoperatively planned position. Ligament balancing and measurements of bone alignment are provided by DASH knee.

Example orthopedic surgical procedures include but are not limited to:

  • · Total Knee Replacement
  • Ligament Balancing
Device Description

Dash is an image guided surgery system for total knee replacement surgery based on landmark based visualization of the femur and tibia. It is intended to enable operational navigation in orthopedic surgery. It links a surgical instrument, tracked by passive markers to virtual computer image space on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. DASH knee uses the registered landmarks to navigate the femoral and tibial cutting guides to the optimally position.

DASH knee software registers the patient data needed for navigating the surgery intraoperatively. No preoperative CT-scanning is necessary.

AI/ML Overview

The provided text describes the K102251 510(k) Summary for the BrainLAB DASH knee system. This document outlines the device, its intended use, comparisons to predicate devices, and verification/validation activities. However, it does not contain specific acceptance criteria, reported device performance metrics in a quantitative manner, or detailed study results that would typically be presented in a table format with numerical values.

Therefore, I cannot directly extract and fill the table for acceptance criteria and reported device performance, nor can I provide answers for all the requested study design details as the information is not present in the given text.

Here's what can be inferred or stated based on the provided text, along with the limitations:

1. Table of Acceptance Criteria and Reported Device Performance

  • Acceptance Criteria: The document implies that the system must function correctly according to its specifications and ensure safety. It mentions "correct system functionality as it has been specified" and "correct behavior of the system for all possible procedures," but no specific numerical targets or thresholds are given.
  • Reported Device Performance: The document states that "All tests have been successfully completed" and that the device was found "substantially equivalent" to predicate devices. This indicates that the device met its internal performance benchmarks, but these benchmarks or the numerical results are not disclosed.
Acceptance Criteria (Implied)Reported Device Performance (Implied)
Correct system functionality during registration/navigationAll tests successfully completed; system functions as specified.
Correct software algorithm behaviorAll tests successfully completed; software functions as specified.
Correct behavior for all possible workflowsAll tests successfully completed; system functions as specified for all procedures.
Safety measures against defined risks are effectiveSafety measures tested and found effective.
Substantial equivalence to predicate devicesDevice found substantially equivalent to K073615 and K014256.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified. The document mentions "usability workshops (use labs)" with "plastic bones (sawbones)" and "cadaver lab" testing, but no specific number of instruments, tests, or cadavers is provided.
  • Data Provenance: Not explicitly stated. The testing appears to be primarily in a lab setting ("usability workshops," "cadaver lab") and therefore would be considered prospective in nature, generating new simulated or ex vivo data. There is no mention of patient data or country of origin for the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of Experts & Qualifications: Not specified. The document mentions "testing persons went through same procedure like for the non clinical use lab sessions" during cadaver testing, implying the involvement of individuals simulating surgical procedures. However, their qualification, number, or their role in establishing a formal "ground truth" (beyond following specified procedures) is not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not specified. The document describes a series of verification and validation steps, including comparing "registration values... to external measured reference values" and testing for "correct behavior." This implies a comparison to a predefined standard or ideal outcome, but no formal adjudication process for discrepancies is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study involving human readers and AI assistance is described. The DASH knee is an image-guided surgery system, not an AI diagnostic tool that assists human readers in interpreting images. Its purpose is to guide surgical instruments, not to enhance human interpretation of images.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, a form of standalone testing was performed for the software algorithms. The document states: "After the verification of the instruments in combination with the software the verification of the software algorithms itself has been performed." This implies that the core algorithms were tested independently of full human interaction, likely against defined computational outputs or expected results.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The ground truth appears to be based on external measured reference values for registration accuracy and predefined specifications/expected behavior for software functionality and workflow correctness. In the cadaver and sawbone labs, the ground truth would likely be the anatomically correct positions or optimal surgical trajectories as determined by established surgical principles and perhaps simulated by experienced users. It is not expert consensus from reviewing patient data, nor is it pathology or long-term outcomes data as these are pre-market non-clinical validations.

8. The sample size for the training set

  • Not applicable as this is not an AI/machine learning device in the contemporary sense that requires a "training set" for model development. The system uses "landmark based visualization" and "acquiring multiple landmarks on the bone surface" to generate a 3D model. Its functionality is based on established algorithms in image guidance and tracking, not on learning from a large dataset.

9. How the ground truth for the training set was established

  • Not applicable, for the same reason as point 8.

{0}------------------------------------------------

MAY 1 7 2011

KIO2251 510 (k) Summary of Safety and Effectiveness for DASH knee

Manufacturer:

Address:BrainLAB AG
Kapellenstrasse 12
85622 Feldkirchen
Germany
Phone:+49 89 99 15 68 0
Fax:+49 89 99 15 68 33
Contact Person:Mr. Alexander Schwiersch
Summary Date:July 12, 2010

Device Name:

Trade name:

DASH knee

Common/Classification Name:

BrainLAB DASH, BrainLAB Image Guided Surgery System / Instrument, Stereotaxic

Predicate Device:

BrainLAB Knee (K073615) Kolibri Image Guided Surgery System (K014256)

Device Classification Name: Instrument, Stereotaxic Regulatory Class: Class II Regulation Number: 21 CFR 882.4560 Product Code: OLO

Device Description:

Dash is an image guided surgery system for total knee replacement surgery based on landmark based visualization of the femur and tibia. It is intended to enable operational navigation in orthopedic surgery. It links a surgical instrument, tracked by passive markers to virtual computer image space on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. DASH knee uses the registered landmarks to navigate the femoral and tibial cutting guides to the optimally position.

DASH knee software registers the patient data needed for navigating the surgery intraoperatively. No preoperative CT-scanning is necessary.

Intended Use:

DASH knee is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical

{1}------------------------------------------------

structure, such as the skull, a long bone, or vertebra, can be identified relative to the anatomy. The system aids the surgeon to accurately navigate a knee prosthesis to the intraoperatively planned position. Ligament balancing and measurements of bone alignment are provided by DASH knee.

Changes made to the predicate device:

Reduced SW complexity: The spectrum of possible workflows has been reduced to one already contained universal express workflow. Within that express workflow registration steps have been combined with the help of an additional instrument.

Additional instrument: Device for multiple point acquisition on femur condyles. Compared to the predicate device, where femoral landmarks have been calculated out of two surface acquisition steps, regarding landmarks are here acquired directly in one step. In combination with a regarding femoral acquisition page the amount of registration steps has been reduced for the user.

Reduced Platform: Compared to predicate device, the main calculation unit is part of the camera stand. A smaller separated display, represented by an iPod touch, acts an embedded display in the instrumentation for navigation.

Wireless communication between embedded display (iPod touch) and main calculation unit (integrated in camera stand).

Completed verification activities:

Following design verification activities have been per formed to ensure correct system functionality as it has been specified:

The first part of the verification covered the instrument and system accuracy during registration and navigation. The registration values have been compared to external measured reference values.

After the verification of the instruments in combination with the software the verification of the software algorithms itself has been performed.

Part three of the verification includes the testing of all possible workflows to ensure the correct behavior of the system for all possible procedures.

With the knowledge of the above named points the current device has been compared to the predicate device.

The next step was the detailed verification of the signed specifications covering the detailed functionality of buttons for example.

At last the measures aqainst the defined risks of the risk analysis have been tested.

This strategy ensures the verification of basic software algorithms up to specific detailed functionality, the comparison to the predicate device and the safety of the defined measures of the risk analysis. All tests have been successfully completed.

{2}------------------------------------------------

Completed validation activities

Following design validation activities have been performed:

A literature search has been performed to prove safety and effectiveness of BrainLAB computer assisted total knee replacement software. This applies directly to DASH knee, since DASH knee is derived from the previous marketed device BrainLAB knee (K073615),

Non-clinical Validation has been performed to prove the system targets and supplement requirement specifications if necessary. With the help of usability workshops (use labs) OR setups and surgery proceedings have been simulated with plastic bones (sawbones).

Pre-clinical Validation has been performed to confirm/ complete detailed specification for each requirement. Here OR setups and surgery proceedings have been simulated in a cadaver lab. Testing persons went through same procedure like for the non clinical use lab sessions.

To prove that all validation issues are addressed, a final validation has been performed under non clinical conditions.

Substantial equivalence:

Dash knee has been verified and validated according to BrainLAB's procedures for product design and development. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate device BrainLAB Knee (K073615) and Kolibri Image Guided Surgery System (K014256).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

BrainLAB AG % Mr. Alexander Schwiersch Kapellenstrasse 12 85622 Feldkirchen, Germany

Re: K102251

MAY 1 7 2011

Trade/Device Name: DASH Knee Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: April 29, 2011 Received: May 06, 2011

Dear Mr. Schwiersch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Mr. Alexander Schwiersch

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

erely yours,

Kof

Petro

N.C.P D.I

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number: K102251

Device Name: DASH knee

Indications For Use:

DASH knee is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to the anatomy. The system aids the surgeon to accurately navigate a knee prosthesis to the intraoperatively planned position. Ligament balancing and measurements of bone alignment are provided by DASH knee.

Example orthopedic surgical procedures include but are not limited to:

  • · Total Knee Replacement
  • Ligament Balancing

Prescription Use X
(Per 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nilfo Dyden for mxm
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) NumberK102251
------------------------
Page1 of 1
--------------

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).