(175 days)
The TYRO™-97 (hofocon A) and ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.
The lenses may be disinfected only by using chemical disinfection
The TYRO™ 97 (hofocon A) Rigid Gas Permeable lens material, hofocon A, is a polymer of trifluoroethyl methacrylate and silicone methacrylate, with no methyl methacrylate. The blue tinted lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and Cl Solvent Yellow 18; the blue-UV lenses contain D&C Green No 6 and a UV absorber, [ 2-(2'-hydroxy-5'-methacryloxyethylphenyl)-2H-benzotriazole. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
The ONSI™-56 (onsifocon A) Rigid Gas Permeable lens material, onsifocon A,is a polymer of trifluoroethyl methacrylate polymer with tris (trimethylsiloxy) methacryloxypropylsilane 3trimethoxysilylpropylmethacrylate methacrylic acid 1,3-bis(3methacryloxypropyl)tetrakis(trimethylsiloxane ethylene glycol dimethacrylate 2-hydroxyethyl methacrylate N-vinylpyrrolidone.
The blue tinted onsifocon A lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and Cl Solvent Yellow 18; the blue-UV lenses contain D&C Green No 6 and a UV absorber, { 2-(2'-hydroxy-5'methacryloxyethylphenyl)-2H-benzotriazole. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
Here's a breakdown of the acceptance criteria and the study information for the TYRO™-97 and ONSI™-56 Rigid Gas Permeable Contact Lenses, based on the provided text:
Acceptance Criteria and Device Performance
The acceptance criteria for these devices are based on demonstrating substantial equivalence to existing legally marketed predicate devices. This is achieved by comparing several key physical properties of the new contact lens materials (hofocon A and onsifocon A) to those of the predicate devices (BOSTON ES, BOSTON EO, and BOSTON XO).
Table of Acceptance Criteria and Reported Device Performance:
| Property | Acceptance Criteria (Predicate Devices: BOSTON ES, BOSTON EO, BOSTON XO) | TYRO™-97 (hofocon A) Performance | ONSI™-56 (onsifocon A) Performance |
|---|---|---|---|
| Water Content | <1% | <0.2% | <1.0% |
| O₂ Permeability (Fatt Method) | 18 to 141 x 10⁻¹¹ (cm³O₂)(cm)/[(sec)(cm²)(mmHg)] @ 35°C (ranges of predicates) | 97 x 10⁻¹¹ ANSI units | 56.5 x 10⁻¹¹ ANSI units |
| Refractive Index | 1.415 to 1.443 (ranges of predicates) | 1.440 | 1.452 |
| Hardness | Not explicitly stated for predicates, but D/82 and D/85 are reported | D/82 (Shore) | D/85 (Shore) |
| Specific Gravity | 1.22 to 1.27 (ranges of predicates) | 1.087 | 1.206 |
| Wetting Angle | 49° to 52° (ranges of predicates) | estimate <25° | 7.25° ± 1.55° (sessile drop method) |
| Mechanical (flexural) Strength | Not explicitly stated for predicates | 3952 psi | 2698 psi |
| Light Transmittance | >85% to >92% (ranges of predicates) | Clear >95%T, Blue >70% T, Green >70% T | Clear >95% T, Blue >70% T, Green >70% T, Grey >70% T, Blue-violet >70% T, Blue-UV >70%T(400-780nm) |
Study Information:
The document describes a 510(k) Premarket Notification process, which relies on demonstrating substantial equivalence to predicate devices rather than independent clinical trials for new materials per se.
-
Sample size used for the test set and the data provenance:
- The document does not explicitly mention a "test set" in the traditional sense of a clinical trial for evaluating the performance of the new devices against specific acceptance criteria. Instead, it relies on bench testing of the material properties to justify substantial equivalence.
- Therefore, the sample sizes for the material property tests (e.g., number of lenses tested for oxygen permeability, hardness) are not specified in the provided text.
- The data provenance for the reported performance values (e.g., Oxygen Permeability of 97 ANSI units for hofocon A) is from laboratory testing of the new materials (hofocon A and onsifocon A). The location or context of these labs is not given, nor is it specified if this was retrospective or prospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable in this context. The "ground truth" for the material properties is established through standardized physical and chemical testing methods (e.g., Fatt Method for oxygen permeability, sessile drop method for wetting angle, Shore Hardness for hardness). These are objective measurements, not subjective expert assessments.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This is not applicable. There is no "adjudication" in the sense of resolving discrepancies between expert opinions, as the evaluation is based on objective material property measurements.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is not applicable. This submission is for rigid gas permeable contact lenses, which are physical medical devices, not AI-powered diagnostic tools. Therefore, MRMC studies involving human readers and AI assistance are irrelevant to this submission.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This is not applicable. This refers to a medical device's algorithm performance, which is not relevant for contact lenses.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- As mentioned above, the "ground truth" for the performance claims (which are material properties) is established through standardized physical and chemical laboratory testing measurements. The document explicitly lists methods like "Fatt Method" for oxygen permeability and "sessile drop method" for wetting angle.
-
The sample size for the training set:
- This is not applicable. This submission is for physical medical devices (contact lenses), not a machine learning algorithm that requires a training set.
-
How the ground truth for the training set was established:
- This is not applicable for the same reason as point 7.
{0}------------------------------------------------
-K 102154.
JAN 2 1 2011
December 14, 2010
510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS
TYRO™ -97 (hofocon A) and ONSI™ -56 (onsifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear
1. SUBMITTER INFORMATION:
The Lagado Corporation 2890 South Tejon St Englewood, CO 80110
Contact:
John M. Szabocsik, Ph.D.
Official agent Szabocsik and Associates 203 N. Wabash, Ste 2208 Chicago, IL 60601 . (312) 553-0828 .
-
- December 13, 2010 DATE PREPARED:
3. DEVICE NAME:
rigid gas permeable (hydrophilic) contact lenses Classification Name:
TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric Proprietary Names: and Bifocal Contact Lens
ONSI™ -56 (onsifocon A) Rigid Gas Permeable Spherical. Aspheric. Toric and Bifocal Contact Lens
CLASSIFICATION: 4 .
Class II (Performance Standards) Classification: HQD 21 CFR 886.5916 (b) (1) Rigid Gas Permeable Contact Lens for Daily Wear
5. DESCRIPTION OF DEVICES:
The TYRO™ 97 (hofocon A) Rigid Gas Permeable lens material, hofocon A, is a polymer of trifluoroethyl methacrylate and silicone methacrylate, with no methyl methacrylate. The blue tinted lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and Cl Solvent Yellow 18; the blue-UV lenses contain D&C Green No 6 and a UV absorber, [ 2-(2'-hydroxy-5'-methacryloxyethylphenyl)-2H-benzotriazole. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
The physical properties of the hofocon A lenses are as follows:
Typical Property
Test Value
{1}------------------------------------------------
| Hardness | D/82 (Shore HardnessUnits) |
|---|---|
| Water Content | <0.2% |
| Wetting Angle | 23.3E (sessile drop method) |
| Oxygen Permeability | 97 ANSI units |
| Dimensional Stability | · Stable |
| Refractive Index | 1.440 |
| Specific Gravity | 1.087 |
| Flexural Strength | 3952 psi |
The ONSI™-56 (onsifocon A) Rigid Gas Permeable lens material, onsifocon A,is a polymer of trifluoroethyl methacrylate polymer with tris (trimethylsiloxy) methacryloxypropylsilane 3trimethoxysilylpropylmethacrylate methacrylic acid 1,3-bis(3methacryloxypropyl)tetrakis(trimethylsiloxane ethylene glycol dimethacrylate 2-hydroxyethyl methacrylate N-vinylpyrrolidone.
The blue tinted onsifocon A lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and Cl Solvent Yellow 18; the blue-UV lenses contain D&C Green No 6 and a UV absorber, { 2-(2'-hydroxy-5'methacryloxyethylphenyl)-2H-benzotriazole. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
The physical properties of onsifocon A lenses are as follows:
| Typical Property | Test Value |
|---|---|
| Hardness | D/85 (Shore HardnessUnits) |
| Water Content | <1.0% |
| Wetting Angle | 7.25° ±1.55 (sessile drop method) |
| Oxygen Permeability | 56.2 ANSI units |
| Dimensional Stability | Stable |
| Refractive Index | 1.452 |
| Specific Gravity | 1.206 |
| Flexural Strength | 3952 psi |
INDICATIONS FOR USE: 6.
The TYRO™-97 (hofocon A) and ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit comeal astigmatism. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.
The lenses may be disinfected only by using chemical disinfection
SAFETY AND SUBSTANTIAL EQUIVALENCE: 7.
The safety and efficacy of the TYRO™ -97 (hofocon A) and ONSI™ -56 (onsifocon A) Rigid Gas Permeable Contact Lenses were demonstrated in 510(k) Premarket Notifications as follows: K052507 - SE Decision Date 02-NOV-05 and K033599 - SE Decision Date 16-SEP-04, respectively.
{2}------------------------------------------------
December 14, 2010
The TYRO™-97 (hofocon A) and ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses
are substantially equivalent to the BOSTON ES (enflufocon A), BOSTON EO (enflufocon B) XO (hexafocon A) Rigid Gas Permeable Contact Lenses indicated for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery K053124 - SE Decision Date 30-JAN-06.
{3}------------------------------------------------
December 14, 2010
SUBSTANTIAL EQUIVALENCE
| Hofocon A | Onsifocon A | Hexafocon B (Boston XO) | Enflufocon A (Boston ES) | Enflufocon B (Boston EO) | |
|---|---|---|---|---|---|
| Water Content | <0.2% | <1.0% | <1% | <1% | <1% |
| O₂ Permeability | Fatt Method: D units=97 x 10-11 (cm³O₂)(cm)/[(sec)(cm²)(mmHg)] @ 35°C | Fatt Method: D units=56.5 x 10-11 (cm³O₂)(cm)/[(sec)(cm²)(mmHg)] @ 35°C | Fatt Method: D units=141 x 10-11 (cm³O₂)(cm)/[(sec)(cm²)(mmHg)] @ 35°C | Fatt Method: D units= 18 x 10-11 (cm³O₂)(cm)/[(sec)(cm²)(mmHg)] @ 35°C | Fatt Method: D units= 58 x 10-11 (cm³O₂)(cm)/[(sec)(cm²)(mmHg)] @ 35°C |
| Refractive Index | 1.440 | 1.452 | 1.415 | 1.443 | 1.420 |
| Hardness | D/82 (Shore) | D/85 (Shore) | |||
| Specific Gravity | 1.087 | 1.206 | 1.27 | 1.22 | 1.23 |
| Wetting Angle | estimate <25° | 7.25 ± 1.55° | 49° | 52° | 49° |
| Mechanical (flexural) Strength | 3952 psi | 2698 psi | |||
| Light Transmittance | Clear >95%TBlue >70% TGreen>70% T | Clear>95% TBlue>70% TGreen>70% TGrey>70% TBlue-violet>70% TBlue-UV>70%T(400-780nm) | >92% | >85% | >85% |
| 0%T |
な
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings and body. The text "DEPARTMENT OF HEALTH &" is vertically oriented on the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Szabocsik and Associates c/o Dr. John M. Szabocsik 203 North Wabash Avenue Suite 2208 Chicago, Illinois 60601
JAN 2 1 201
Re: K102154
Trade/Device Name: TYRO-97 (hofocon A), ONSI™ - 56 (onsifocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens for Daily Wear Regulatory Class: Class II Product Code: HQD Dated: December 20, 2010 Received: December 23, 2010
Dear Dr. Szabocsik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Dr. John M. Szabocsik
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and liasting (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11178001,1 profox the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFF), Paul 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/5/Picture/7 description: The image shows the handwritten word "for" in cursive. The letter "f" has a large, elongated loop extending below the baseline, and the "o" and "r" are connected with a smooth, flowing line. The writing appears to be done with a pen or marker, and the ink is black. The background is plain white.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
December 14, 2010
Page 1 of 1
510(k) NUMBER (IF KNOWN)
DEVICE NAME TYRO™-97 (hofocon A) and ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear
INDICATIONS FOR USE
The TYRO™-97 (hofocon A) and ONSI™-56 (onsifocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.
The lenses may be disinfected only by using chemical disinfection
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ X Over-The-Counter-Use (Per 21 CFR 801.109) OR (Optional Format 1-2-96)
Page 1 of 1
Jana Postell
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K102154
§ 886.5916 Rigid gas permeable contact lens.
(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.