(290 days)
No
The device description details a homogeneous enzyme immunoassay system based on recombinant DNA technology and spectrophotometric measurement. There is no mention of AI or ML in the intended use, device description, or performance studies.
No
This device is an in vitro diagnostic (IVD) device used to detect opiate in human oral fluid, not to treat a medical condition.
Yes
The "Intended Use / Indications for Use" section explicitly states, "This in vitro diagnostic device is intended for clinical laboratory use only." The device performs a qualitative determination of opiate in human oral fluid, which is a diagnostic purpose.
No
The device description clearly outlines a chemical assay system involving reagents, an enzyme, and a collection system with physical components (absorbent pad, plastic handle, collection tube). This is an in vitro diagnostic device with significant hardware and chemical components, not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Explicit Statement: The "Intended Use / Indications for Use" section clearly states: "This in vitro diagnostic device is intended for clinical laboratory use only."
- Nature of the Test: The device performs a test on a biological sample (human oral fluid) to determine the presence of a substance (opiate) for diagnostic purposes (preliminary analytical test result for drug of abuse).
- Clinical Laboratory Use: The intended user and setting are specified as "clinical laboratory use only," which is a common characteristic of IVDs.
Therefore, based on the provided text, the CEDIA® Opiate OFT Assay is an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The CEDIA® Opiate OFT Assay is intended for use in the qualitative determination of opiate in human oral fluid at a cutoff concentration of 30 ng/mL in neat oral fluid. The specimen must be collected exclusively with the Oral-Eze™ Saliva Collection System. The assay is calibrated against morphine and performed on the MGC240. This in vitro diagnostic device is intended for clinical laboratory use only.
The CEDIA Opiate OFT Assay provides only a preliminary analytical test result. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result particularly when preliminary positive results are used.
Product codes
DJG
Device Description
Microgenics CEDIA® Opiate OFT Assay uses recombinant DNA technology to produce a unique homogeneous enzyme immunoassay system. The assay is based on the bacterial enzyme ß-galactosidase, which has been genetically engineered into two inactive fragments. These fragments spontaneously re-associate to form fully active enzyme that, in the assay format, cleave a substrate, generating a color change that can be measured spectrophotometrically.
In the assay, analyte in the sample competes with analyte conjugated to one inactive fragment (enzyme donor) of β-galactosidase for antibody binding site. If analyte is praesent in the sample, it binds to antibody, leaving the inactive enzyment free to form active enzyme. If the analyte is not present in the sample, antibody binds to analyte conijugated on the inactive fragment, inhibiting the re-association of inactive B-galectosidase fragments, and no active enzyme is formed. The amount of active enzyme formed and fightently, absorbance change are directly proportional to the amount of analyte present it the sample.
The Oral-Eze™ Saliva Collection System consists of Oral-Eze™ saliva collector and collection tube with preservative buffer. Oral-Eze™ saliva collector consists of an absorbent pad attached to a plastic handle. The saliva collector is provided with a volume adecruacy indicator. The plastic handle has a round window where blue color will appear when sufficient volume of oral fluid is collected. Samples are collected by placing the collector pad and plastic shield between lower cheek and gum with the plastic shield facing the cheek. Oral fluid collection is done when blue color appears in the window of the handle. The pad is ejected in to the collection tube by placing thumb on the ridges on the handle and pushing the thumb forward. The collection tube is capped and sent to the laboratory for processing and testing.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
oral fluid
Indicated Patient Age Range
Not Found
Intended User / Care Setting
prescription professional use only in clinical laboratories
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Qualitative Precision: All samples tested recovered accurately. Samples at levels below the cutoff read as negative and samples at levels above the cutoff read as positive.
Qualitative Cutoff Characterization: All samples tested recovered accurately, low control as negative and high control level as positive.
Interferences: Results demonstrated that there was no significant interference from endogenous and exogenous substances tested at clinically relevant concentrations and in samples adjusted to pH range of 5 to 9.
Specificity and Cross-Reactivity: Cross-reactivity to metabolites and structurally related compounds was tested in the assay. No significant cross-reactivity was observed with other structurally unrelated compounds.
Method Comparison: The overall concordance between the CEDIA® Opiate OFT Assay and GC/MS is 97.6%. The comparison of sample results by the CEDIA® Opiate OFT Assay to GCMS showed 100.0 % sensitivity and 95.2 % specificity.
Key Metrics
Method Comparison:
Sensitivity: 100.0 %
Specificity: 95.2 %
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
0
APR - 8 2011
510K SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in summary of STO(K) salety and enectronics of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is: K101754
Company/Contact person
Lisa Charter Manager, Regulatory Affairs Thermo Fisher Scientific, Clinical Diagnostic Division 46360 Fremont Blvd Fremont, CA 94538 Phone: (510) 979-5142 Facsimile: (510) 979-5422 Email: Lisa.Charter@ThermoFisher.com
Date Prepared
January 6, 2011
Requiatory Declarations
Common / Usual Name | CEDIA® Opiate OFT Assay |
---|---|
Trade/ Proprietary Name | Thermo Scientific CEDIA® Opiate OFT Assay |
Classification Regulation | 21 CFR 862.3650 |
Device Class | Class II |
Device Regulation Panel | Toxicology |
Product Code | DJG |
Intended use
The CEDIA® Opiate OFT Assay is intended for use in the qualitative determination of The CEDIA "Oplate Of T Assay is Intended for 30 ng/mL in neat oral fluid. The oplate in numan oral liud at a caton contonitution of 50 mg Saliva Collection System. The specimen must be collected exclusively with the era on the MGC240. This in vitro assay is oulibrator againical laboratory use only.
The CEDIA Opiate OFT Assay provides only a preliminary analytical test result. A more The CEDIA Oplate OF TASSY provides only and to obtain a confirmed analytical result. Gas specific alternative method must be used (b Chromatography-Tandem Mass
Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography-Tandemical in ChromalographyMass Spectrometry (GOATS) and internations . Olinical consideration Spectrometly (CC-MS/NG) are the profess somments) of abuse test result particularly when preliminary positive results are used.
Conditions for use
The CEDIA® Opiate OFT Assay is for prescription professional use only in clinical The - OEDIA - Oplate - It is not for use in Point of Care settings.
Legally marketed device to which equivalency is claimed
1
CEDIA® Opiate OFT Assay is substantially equivalent to the previously cleared STC Opiates MICRO-PLATE EIA, K981341 (At present OTI, OraSure Technologies Inc.)
DESCRIPTION OF DEVICE
Principle of the CEDIA® Opiate OFT Assay
Microgenics CEDIA® Opiate OFT Assay uses recombinant DNA technology to produce a unique homogeneous enzyme immunoassay system. The assay is based on the bacterial enzyme ß-galactosidase, which has been genetically engineered into two inactive fragments. These fragments spontaneously re-associate to form fully active enzyme that, in the assay format, cleave a substrate, generating a color change that can be measured spectrophotometrically.
In the assay, analyte in the sample competes with analyte conjugated to one inactive fragment (enzyme donor) of β-galactosidase for antibody binding site. If analyte is praesent in the sample, it binds to antibody, leaving the inactive enzyment free to form active enzyme. If the analyte is not present in the sample, antibody binds to analyte conijugated on the inactive fragment, inhibiting the re-association of inactive B-galectosidase fragments, and no active enzyme is formed. The amount of active enzyme formed and fightently, absorbance change are directly proportional to the amount of analyte present it the sample.
Principle of Oral-Eze™ Saliva Collection System
The Oral-Eze™ Saliva Collection System consists of Oral-Eze™ saliva collector and collection tube with preservative buffer. Oral-Eze™ saliva collector consists of an absorbent pad attached to a plastic handle. The saliva collector is provided with a volume adecruacy indicator. The plastic handle has a round window where blue color will appear when sufficient volume of oral fluid is collected. Samples are collected by placing the collector pad and plastic shield between lower cheek and gum with the plastic shield facing the cheek. Oral fluid collection is done when blue color appears in the window of the handle. The pad is ejected in to the collection tube by placing thumb on the ridges on the handle and pushing the thumb forward. The collection tube is capped and sent to the laboratory for processing and testing.
2
Comparison of Technological Characteristics
The CEDIA® Opiate OFT Assay is substantially equivalent to the OTI Opiates MICRO-PLATE EIA. (K981341)
| Comparison | Subject Device
CEDIA® Opiate OFT Assay | Predicate Device
OTI Opiates MICRO-PLATE EIA
K981341 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The CEDIA® Opiate OFT Assay is
intended for use in the qualitative
determination of opiate in human
oral fluid at a cutoff concentration of
30 ng/mL in neat oral fluid. The
specimen must be collected
exclusively with the Oral-Eze™
Saliva Collection System. The assay
is calibrated against morphine and
performed on the MGC240. This in
vitro diagnostic device is intended
for clinical laboratory use only.
The CEDIA Opiate OFT Assay
provides only a preliminary
analytical test result. A more specific
alternative method must be used to
obtain a confirmed analytical result.
Gas Chromatography/Mass
Spectrometry (GC/MS) and Liquid
Chromatography-Tandem Mass
Spectrometry (LC-MS/MS) are the
preferred confirmatory methods.
Clinical consideration and
professional judgment should be
applied to any drug of abuse test
result particularly when preliminary
positive results are used. | The OTI Opiates MICRO-PLATE
EIA is intended for use in the
qualitative determination of opiates
in oral fluid collected with the
OraSure® Oral Collection Device.
For In Vitro Diagnostic Use. |
| Test
Principle | Microgenics CEDIA® Opiate OFT
Assay uses recombinant DNA
technology to produce a unique
enzyme
homogeneous
immunoassay system. The assay is
based on the bacterial enzyme ß-
galactosidase, which has been
genetically engineered into two
inactive fragments. These fragments
spontaneously re-associate to form
fully active enzyme that, in the
assay format, cleave a substrate,
generating a color change that can
be measured
spectrophotometrically.
In the assay, analyte in the sample | The OTI Opiates is a competitive
micro-plate immunoassay for the
detection of opiates in oral fluid
collected with the OraSure® Oral
Specimen Collection Device.
Specimen or standard is added to
an EIA well in combination with an
enzyme-labeled hapten derivative.
In an EIA well containing an
OraSure® specimen positive for
opiates, there is a competition
between the drug and the enzyme-
labeled hapten to bind the antibody
fixed onto the EIA well. ElA wells
are then washed, substrate is
added, and color is produced. The
absorbance measured for each well
at 450nm is inversely proportional to |
| one inactive fragment (enzyme donor) of β-galactosidase for antibody binding site. If analyte is present in the sample, it binds to antibody, leaving the inactive enzyme fragment free to form active enzyme. If the analyte is not present in the sample, antibody binds to analyte conjugated on the inactive fragment, inhibiting the re-association of inactive β-galactosidase fragments, and no active enzyme is formed. The amount of active enzyme formed and resultant absorbance change are directly proportional to the amount of analyte present in the sample. | the amount of opiates present in the specimen or calibrator/control. | |
| Sample
Matrix | Oral Fluid | Oral Fluid |
| Calibrator
levels . | 0, 10.0, 80.0 ng/mL | 0, 10 ng/mL |
| Cutoff level | 30.0 ng/mL in neat oral fluid | 10.0 ng/mL |
| Unassayed
Control
levels | 5.0, 15.0 ng/mL | 5.0, 20.0 ng/mL |
3
. . . . . .
:
:
and the contraction of the comments of the comments of the comments of
: :
ﺍ
4
SUMMARY OF CLINICAL TESTING
Qualitative Precision
All samples tested recovered accurately. Samples at levels below the cutoff read as negative All samples tested recoverou assembly
and samples at levels above the cutoff read as positive.
o
Qualitative Cutoff Characterization
All samples tested recovered accurately, low control as negative and high control level as positive.
Interferences
Results demonstrated that there was no significant interference from endogenous and Results demonstrated that there was The Sighlindine interested oncentrations and in samples adjusted to pH range of 5 to 9.
Specificity and Cross-Reactivity
Cross-reactivity to metabolites and structurally related compounds was tested in the assay. No Cross-reactivity to metabolites and Structurally related of the structurally unrelated compounds.
significant cross-reactivity was observed with other structurally unrelated
Method Comparison
The overall concordance between the CEDIA® Opiate OFT Assay and GC/MS is 97.6%. The over 100.0 % The overall concordance between the CEDIA "Opiate OFT Assay to GCMS showed 100.0 %
comparison of sample results by the CEDIA® Opiate OFT Assay to GCMS showed 100.0 % sensitivity and 95.2 % specificity.
Conclusion
As summarized, the CEDIA® Opiate OFT Assay is substantially equivalent to the OTI Opiates As summarized, the CEDIA Opliale Of 1 Assay Jobsentation through performance
MICRO-PLATE EIA. Substantial equivalence has been demonstrations have been MICRO-PLATE EIA. Substantial equivalence Nas boom democracy in that design specifications have been satisfied.
5
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is depicted in a simple, black and white design.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Microgenies Corp. ThermoFisher Scientific, Clinical Diagnostic Division c/o Ms. Lisa Charter Manager, Regulatory Affairs 46360 Fremont Blvd. Fremont, CA 94538-6406
APR 0 8 2011
Re: K101754
Trade Name: Thermo Scientific CEDIA Opiate OFT Assay Regulation Number: 21 CFR §862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Codes: DJG Dated: March 10, 2011 Received: March 14, 2011
Dear Ms. Charter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
6
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
7
Indications for Use
510(k) Number (if known): K101754
Device Name: CEDIA® Opiate OFT Assay
Indications for Use:
The CEDIA® Opiate OFT Assay is intended for use in the qualitative determination of opiate in human oral fluid at a cutoff concentration of 30 ng/mL in neat oral fluid. The specimen must be collected exclusively with the Oral-Eze™ Saliva Collection System. The assay is calibrated against morphine and performed on the MGC240. This in vitro diagnostic device is intended for clinical laboratory use only.
The CEDIA Opiate OFT Assay provides only a preliminary analytical test result. A more specific alternative method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result particularly when preliminary positive results are used.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k/01754