(26 days)
Not Found
No
The summary mentions an "Arrhythmia Detection Algorithm" but provides no details suggesting it utilizes AI or ML. The "Mentions AI, DNN, or ML" section is explicitly marked as "Not Found".
No
The device is intended for monitoring and detection of cardiac arrhythmias, not for treating them. It provides data for medical professionals to evaluate, but does not administer therapy itself.
Yes
The device is described as an "Arrhythmia Detector" which continuously monitors patient ECG and "automatically generates an alarm triggered by an arrhythmia detection algorithm." The recorded data is then transmitted to a monitoring center for evaluation by a medical practitioner, indicating its role in identifying and characterizing a medical condition.
No
The device description explicitly mentions a "chest-worn sensor" with electrodes, a battery, a Bluetooth transceiver, and a buzzer, which are hardware components. It also mentions "Hardware Verification and Validation" in the performance studies summary.
Based on the provided information, this device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs analyze samples taken from the body. IVDs typically involve testing biological samples like blood, urine, tissue, etc., outside of the body to diagnose or monitor a condition.
- This device monitors physiological signals from the body. The CG-6108 ACT-3L directly acquires and analyzes the electrical activity of the heart (ECG) from the patient's chest. This is a physiological measurement, not an analysis of a biological sample.
- The intended use and device description focus on ECG monitoring and arrhythmia detection. The text clearly describes the device's function as continuously monitoring ECG, detecting arrhythmias, and transmitting this data. There is no mention of analyzing any biological samples.
Therefore, the CG-6108 ACT-3L falls under the category of a physiological monitoring device, not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
Product codes
DSI, DXH
Device Description
CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is designed for self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center.
The chest-worn sensor is used for the acquisition, recording, and transmission of the ECG signal. The device is equipped with 4 electrodes on a harness and it houses a 3.6V AA battery, a Bluetooth transceiver and a buzzer.
The ECG signals are transmitted via Bluetooth to a handheld device with a proprietary interactive application, configured to process and transmit the ECG recordings. The handheld device is a mobile computing device with a display and a touch input such as a cell-phone. It has sufficient memory and processing capability to run the proprietary application.
When an arrhythmia event is detected the handheld device transmits the recorded ECG information automatically via cellular link, to the Monitoring Center or professional analysis. When cellular service is unavailable the patient has an option to transmit via a landline telephone.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
The modified version CG-6108 has been subjected to extensive verification / validation testing. Final testing of the system included various performance tests and software validation tests designed to ensure that the device meet all of its functional and performance requirements and is fit for its intended use. The following list summarizes the testing performed on the device;
- Software Verification and Validation.
- Software Functional Unit Verification
- System Level Software Validation
- Arrhythmia Detection Algorithm Performance Validation
- Hardware Verification and Validation
- Verification of Conformance with Special Controls Guidance Document.
Key Metrics
Not Found
Predicate Device(s)
Unmodified version of CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector, cleared in K081257 on May 29, 2008.
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
0
Special 510(k) for CG-6108 ACT-3L Continuous ECG monitor and Arrhythmia Detector Section 7: 510(k) Summary
510(k) Summary: Modified CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector
Introduction
This document contains the 510(k) summary for the modified CG-6108 1L Continuous ECG Monitor and Arrhythmia Detector. The content of this summary is based on the requirements of 21 CFR Section 807 92(c).
Submitter | Card Guard Scientific Survival Ltd., | |
---|---|---|
Establishment | ||
Registration | ||
Number | 9681879 | |
Address | 2 Pekeris St., P.O.B. 527, Rehovot, 76100, Israel | |
Contact person: | Asher Kassel, Director of RA & QA, Card Guard Scientific Survival Ltd. | |
Phone: | 972-8-9484010 | |
(direct) | Fax: 972-8-9484044 | |
E-mail: | asherk@cardguard.com | |
Date Prepared: | June 8, 2010 | |
Predicate device | Unmodified version of CG-6108 ACT-3L Continuous ECG Monitor and | |
Arrhythmia Detector, cleared in K081257 on May 29, 2008. | ||
Trade Name: | CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector | |
Classification: | Detector and alarm, arrhythmia / | |
Transmitters and receivers, electrocardiograph, telephone | ||
Product Code: | DSI, DXH | |
Regulation No: | 870.1025, 870.2920 | |
Class: | II |
Device Description
CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is designed for self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center.
The chest-worn sensor is used for the acquisition, recording, and transmission of the ECG signal. The device is equipped with 4 electrodes on a harness and it houses a 3.6V AA battery, a Bluetooth transceiver and a buzzer.
The ECG signals are transmitted via Bluetooth to a handheld device with a proprietary interactive application, configured to process and transmit the ECG recordings. The handheld device is a mobile computing device with a display and a touch input such as a cell-phone. It has sufficient memory and processing capability to run the proprietary application.
When an arrhythmia event is detected the handheld device transmits the recorded ECG information automatically via cellular link, to the Monitoring Center or professional analysis. When cellular service is unavailable the patient has an option to transmit via a landline telephone.
Indications for Use:
The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
1/2
1
$\frac{2}{2}$
Image /page/1/Picture/1 description: The image shows the words "CARD GUARD" in bold, black letters. Below that, in a smaller font, are the words "Scientific Survival Kit". To the right of the text is a jagged line that resembles a heartbeat monitor.
Special 510(k) for CG-6108 ACT-3L Continuous ECG monitor and Arrhythmia Detector Section 7: 510(k) Summary
Summary of the Technological Characteristics / Principles of Operation
The technological characteristics and principles of operation of the modified device are the same as the predicate device. The chest-worn ECG sensor transmits signals via Bluetooth to the handheld device equipped with the Medical Application, which incorporates an algorithm for detection of cardiac events: Atrial Fibrillation, Tachycardia and Pause. A detected artifact triggers transmission of the signal to the Monitoring Center for analysis.
Non-clinical performance data for the CG-6108 ACT-3L:
The modified version CG-6108 has been subjected to extensive verification / validation testing. Final testing of the system included various performance tests and software validation tests designed to ensure that the device meet all of its functional and performance requirements and is fit for its intended use. The following list summarizes the testing performed on the device;
- Software Verification and Validation .
- Software Functional Unit Verification o
- System Level Software Validation 0
- Arrhythmia Detection Algorithm Performance Validation 0
- . Hardware Verification and Validation
- Verification of Conformance with Special Controls Guidance Document .
Performance Standards:
This 510(k) submission was written in accordance with the FDA Guidance document "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm, October 28, 2003" and the device conforms to the applicable performance requirements contained in and referenced in this document. In addition, this submission was prepared in accordance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005", The design of the CG-6108 ACT-1L conforms to the following voluntary standards:
- ANSI/AAMI/ISO EC57:1998 (R) 2008: Testing and Reporting Performance Results of Cardiac . Rhythm and ST Seament Measurement Algorithms
- . ANSI/AAMI EC38:1998 Ambulatory Electrocardiograph
- ISO 14971:2007: Medical devices application of risk management to medical devices; .
- IEC 60601-1:1988, 20d edition, Part 1, plus A1:1991 and A2:1995: Medical electrical . equipment; Part 1: General requirements for safety
- IEC 60601-1-2: 2001, plus am. 1:2004, Part 1: Medical electrical equipment, Part 1-2; . Electromagnetic compatibility - Requirements and tests
- . IEC 60601-1-4:2000, plus Amendment 1:2004: Medical electrical equipment; Part 1: 4. Collateral Std: Programmable electric medical systems
- IEC 62304:2006: Medical device software Software life cycle processes .
- ISO 15223:2007: Medical devices Symbols to be used with medical device labels, labeling . and information to be supplied
Substantial Equivalence:
The modified CG-6108 ACT-3L device is substantially equivalent with respect to the indications for use, technological characteristics and performance characteristics to the identified legally marketed predicate devices.
2
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services (HHS). The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. In the center of the seal is an abstract image of a stylized eagle.
Food and Druq Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
JUL 1 3 2010
Card Guard Scientific Survival Ltd.,
c/o Clay Anselmo
President and CEO,
Reglera LLC.
· 555 Zang Street Suite 100 Lakewood, Colorado 80228
Re: K101703
Trade/Device Name: CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm. Regulatory Class: Class II (two) Product Code: DSI, DXH Dated: June 14, 2010 Received: June 17, 2010
Dear Mr. Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device. referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 - Mr. Anselmo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W.M.D.
-Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Indications for Use
长101703 510(k) Number (if known): _
Device Name: CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector
Indications for Use:
The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by paiients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) |
---|
(Division Sign-Off) |
Division of Cardiovascular Devices
Page 1 of 1
-11-
510(k) Number K101703
Page 19 of 791