K Number
K101659

Validate with FDA (Live)

Date Cleared
2010-12-16

(188 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SuperCross microcatheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents.

Device Description

The SuperCross™ microcatheter (SuperCross) is a single-lumen catheter designed for use in the coronary and peripheral vasculature. The SuperCross tubing consists of a braided shaft that tapers distally to create a low profile. SuperCross is compatible with a 0.014 inch diameter guidewire with a minimum length of 180 cm. The SuperCross has a single gold marker band located 0.035 inch (0.89 mm) from the distal tip of the catheter. The distal 40 cm of the SuperCross has a hydrophilic coating to enhance deliverability to the target vasculature. The device has positioning marks located at 95 cm and 105 cm from the distal tip, respectively. The proximal end of the catheter incorporates a strain relief and a luer-lock adapter for flushing, wire insertion, and accessory attachment.

AI/ML Overview

The provided text describes a 510(k) summary for the SuperCross™ microcatheter. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with specific acceptance criteria and performance metrics in the format requested.

Here's an attempt to extract and interpret the information based on the provided text, while acknowledging that the direct format for acceptance criteria and a study proving those are not fully detailed:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states: "Results of the verification testing and biomaterial assessments met the specified acceptance criteria and did not raise new safety or performance questions." However, the specific numerical or quantitative acceptance criteria for each test are not provided in the document. The performance is generally stated as "met the specified acceptance criteria."

Acceptance Criteria CategoryReported Device Performance
Mechanical Performance
Simulated anatomy/concomitant device useMet specified acceptance criteria
Liquid leakMet specified acceptance criteria
KinkMet specified acceptance criteria
Static pressureMet specified acceptance criteria
Air aspirationMet specified acceptance criteria
TensileMet specified acceptance criteria
TorqueMet specified acceptance criteria
Dimensional verificationMet specified acceptance criteria
RadiopacityMet specified acceptance criteria
Material/Coating Performance
Hydrophilic coating integrity and particulatesMet specified acceptance criteria
Biocompatibility
CytotoxicityMet specified acceptance criteria
SensitizationMet specified acceptance criteria
Irritation/intracutaneous reactivityMet specified acceptance criteria
Acute systemic toxicityMet specified acceptance criteria
Material mediated pyrogensMet specified acceptance criteria
Hemocompatibility (Hemolysis, Coagulation, Prothrombin time, Hematological parameters, Complement activation, Thrombogenicity)Met specified acceptance criteria

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not specify the sample sizes used for any of the verification tests.
  • Data Provenance: The tests conducted appear to be retrospective engineering/laboratory bench tests performed by the manufacturer (Vascular Solutions, Inc.) to qualify the device design. Country of origin for data is not explicitly stated but implied to be internal testing by a US-based company.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not Applicable. This type of information (number of experts, qualifications, ground truth establishment for a medical image or diagnostic device) is relevant for studies involving human interpretation or clinical endpoints. The SuperCross microcatheter is an interventional device, and its approval is based on engineering and biocompatibility testing against predefined specifications, not human interpretation of data where "ground truth" would be established by experts.

4. Adjudication Method for the Test Set:

  • Not Applicable. As explained above, this concept typically applies to studies involving human reading or interpretation of data, which is not the nature of the tests described for the microcatheter. The "adjudication" in this context would likely be internal quality control and regulatory review to ensure test results meet specifications.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No. The document does not mention a multi-reader, multi-case comparative effectiveness study. This type of study is more common for diagnostic imaging AI devices, assessing the impact of AI assistance on human performance (e.g., radiologists' accuracy). The SuperCross microcatheter is an interventional device.

6. Standalone Performance Study:

  • Yes, implicitly. The verification tests listed (e.g., tensile, torque, liquid leak, biocompatibility) represent standalone performance evaluations of the device's physical and biological characteristics. The device's performance was assessed independently against a set of predetermined criteria. The studies are described as "verification testing and biomaterial assessments."

7. Type of Ground Truth Used:

  • The "ground truth" in this context refers to predefined engineering specifications, international standards, and internal quality control criteria for device performance and safety (e.g., specific tensile strength values, leak rates, biocompatibility assay results). These are objective, measurable parameters rather than expert consensus, pathology, or outcomes data, which are typically used for diagnostic or predictive devices.

8. Sample Size for the Training Set:

  • Not Applicable. This device is a traditional medical device (catheter), not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How Ground Truth for the Training Set Was Established:

  • Not Applicable. Since there is no AI/machine learning component and thus no "training set," this question is not relevant.

{0}------------------------------------------------

DEC 16 2010

2 510(k) Summary

[As required by 21 CFR 807.92]

510(k) Number: K101659

Submitter's Information / Contact Person

Manufacturer

Vascular Solutions, Inc. 6464 Sycamore Court Minneapolis, MN 55369 USA Establishment Registration # 2134812

Date Prepared: December 13, 2010

Contact Person

Matt Nienstedt Regulatory Affairs Associate Tel: 763.656.4317 (direct) Fax: 763.656.4253 Email: mnienstedt@vascularsolutions.com

Alternate Contact Person

Stacy Quellette Senior Regulatory Affairs Operations Associate Tel: 763.656.4217 (direct) Fax: 763.656.4253 Email: souellette@vascularsolutions.com

General Information

Trade NameSuperCross™ microcatheter
Common / Usual Namemicrocatheter
Classification Name21 CFR 870.1250, Catheter, Percutaneous
Predicate DeviceGopher™ Gold catheter (K091345 - Vascular Solutions, Inc.)

Device Description

The SuperCross™ microcatheter (SuperCross) is a single-lumen catheter designed for use in the coronary and peripheral vasculature. The SuperCross tubing consists of a braided shaft that tapers distally to create a low profile. SuperCross is compatible with a 0.014 inch diameter guidewire with a minimum length of 180 cm. The SuperCross has a single gold marker band located 0.035 inch (0.89 mm) from the distal tip of the catheter. The distal 40 cm of the SuperCross has a hydrophilic coating to enhance deliverability to the target vasculature. The device has positioning marks located at 95 cm and 105 cm from the distal tip, respectively. The proximal end of the catheter incorporates a strain relief and a luer-lock adapter for flushing, wire insertion, and accessory attachment.

{1}------------------------------------------------

Intended Use / Indications

The SuperCross microcatheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents.

Technological Characteristics

The SuperCross microcatheter is substantially equivalent in device design and performance to the predicate device. The SuperCross and predicate Gopher Gold devices have the following characteristics in common:

  • . Compatibility with 0.014 inch guidewires
  • . Shaft configuration - The shafts of the SuperCross and Gopher Gold are both constructed from a braided stainless steel backbone and several thermoplastics bonded together that taper from larger, proximal inner and outer diameters to smaller, distal inner and outer diameters.
  • Radiopaque distal tips .
  • . Positioning marks
  • . Similar inner and outer diameters
  • Tapered distal tips The SuperCross and predicate Gopher Gold are both designed with . tapered distal tips to aid in navigating small, tortuous anatomy.
  • . Hub configuration - Both the predicate and subject device have proximal hubs (with strain reliefs) that are compatible with ISO 594-1 and ISO 594-2 compliant luer fittings.
  • . Rated to 300 psi injection pressure
  • . Sterilized by ethylene oxide
  • Packaged in identical sterile pouch and retail box .

The SuperCross and Gopher Gold devices differ in the following:

  • Specific catheter shaft and hub materials and construction .
  • "Hydrophilic coating SuperCross has a hydrophilic coating and the predicate Gopher Gold . does not.
  • Torque device and threaded tip The predicate Gopher Gold device features a threaded distal tip and is supplied with a torque device while the SuperCross does not contain these features

{2}------------------------------------------------

Substantial Equivalence and Summary of Studies

The SuperCross microcatheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design was qualified through the following tests:

  • Simulated anatomy/concomitant device use .
  • Hydrophilic coating integrity and particulates .
  • Liquid leak .
  • . Kink
  • Static pressure .
  • Air aspiration .
  • . Tensile
  • Torque ●
  • Dimensional verification .
  • Radiopacity .
  • . Biocompatibility
    • o Cytotoxicity
    • Sensitization O
    • o Irritation/intracutaneous reactivity
    • Acute systemic toxicity o
    • Material mediated pyrogens o
    • Hemocompatibility o
      • Hemolysis ■
      • Coagulation
      • . Prothrombin time
      • Hemotological parameters
      • Complement activation 트
      • 피 Thrombogenicity

Results of the verification testing and biomaterial assessments met the specified acceptance criteria and did not raise new safety or performance questions.

{3}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

DEC 16 2010

Vascular Solutions, Inc. c/o Mr. Matt Nienstedt Regulatory Affairs Associate 6464 Sycamore Court Minneapolis, MN 55369

Re: K101659

Trade/Device Name: SuperCross™ Microcatheter Common Name: Catheter, Percutaneous Regulation Number: 21 CFR 870.1250 Regulatory Class: II Product Code: DOY Dated: December 3, 2010 Received: December 6, 2010

Dear Mr. Nienstedt:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed four becaused the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sured in the encreate) to tegary coment date of the Medical Device Amendments, or to commerce pror to May 20, 1977, are eccordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicierore, manol the Act include requirements for annual registration, listing of general oonly of over in the practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Trease note: CDTCF Good not over be truthful and not misleading.

{4}------------------------------------------------

Page 2 - Mr. Matt Nienstedt

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

Dina R. Vahner

Image /page/4/Picture/8 description: The image shows a black and white drawing of a symbol. The symbol is a stylized letter, possibly a cursive 'f' or 's', with a loop at the bottom and a curved line extending to the right. The drawing is simple and lacks any shading or texture.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

DEC 1 6 2010

510(k) Number (if known): K | 0 | 659

Device Name: SuperCross microcatheter

Indications for Use:

. .

.

The SuperCross microcatheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R. dehner

ാ ടിറ്റ്ന-Oli)
സി Cardiovascular Devices

510(k) Numb

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).