K Number
K101639
Date Cleared
2010-06-25

(15 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector are intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECC, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center.

The monitoring center provides the ECG data to the medical practitioner for evaluation.

Device Description

CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector is designed for self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center.

The chest-worn sensor is used for the acquisition, recording, and transmission of the ECG signal. The device is equipped with 3 electrodes on a harness and it houses a 3.6V AA battery, a Bluetooth transceiver and a buzzer.

The ECG signals are transmitted via Bluetooth to a handheld device with a proprietary interactive application, configured to process and transmit the ECG recordings. The handheld device is a mobile computing device with a display and a touch input such as a cell-phone. It has sufficient memory and processing capability to run the proprietary application.

When an arrhythmia event is detected the handheld device transmits the recorded ECG information automatically via cellular link, to the Monitoring Center or professional analysis. When cellular service is unavailable the patient has an option to transmit via landline telephone.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector:

The provided text focuses on a 510(k) summary for a modified device (CG-6108 ACT-1L) and states that the technological characteristics and principles of operation are the same as the predicate device. This often means that the original predicate device's performance data and acceptance criteria are largely being referenced or assumed. The document does not contain a detailed de novo study demonstrating the performance of this modified device against specific, new acceptance criteria, but rather references validation against existing guidance and standards.

Therefore, many of the requested items (sample size, expert qualifications, etc.) are not explicitly provided for the modified device's performance study within these documents. The provided information focuses on conformance to standards and validation of software/hardware, rather than a clinical effectiveness study with an independent test set and ground truth.


1. Table of Acceptance Criteria and Reported Device Performance

The document references existing performance standards and guidance. It does not provide a specific table of acceptance criteria for this particular submission or a direct report of device performance against such. Instead, it states conformance to:

  • "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm, October 28, 2003"
  • "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005"
  • ANSI/AAMI/ISO EC57:1998 (R) 2008: Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms
  • ANSI/AAMI EC38:1998 Ambulatory Electrocardiograph

The document states: "Final testing of the system included various performance tests and software validation tests designed to ensure that the device meet all of its functional and performance requirements and is fit for its intended use."

Specific performance metrics of the device (sensitivity, specificity, accuracy for arrhythmia detection) are NOT provided in this document.


2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified in the provided document.
  • Data Provenance: Not specified in the provided document. The device is manufactured in Israel, but the origin of any test data is not mentioned. It is unclear if it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

4. Adjudication method for the test set

  • Adjudication Method: Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study is not mentioned in the provided document. This device is primarily a detector and transmitter of ECG data for medical professionals to evaluate, not a system assisting human readers directly in interpretation for a comparative effectiveness outcome as typically measured in an MRMC study. The device's algorithm detects events to trigger transmission, it doesn't provide a diagnostic interpretation for human readers to then improve upon.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: The document states: "Arrhythmia Detection Algorithm Performance Validation". This indicates that the algorithm's performance was validated, which would typically be a standalone assessment. However, specific metrics (e.g., sensitivity, specificity, PPV, NPV for Atrial Fibrillation, Tachycardia, and Pause) are not provided in this summary.

7. The type of ground truth used

  • Type of Ground Truth: Not explicitly stated. For "Arrhythmia Detection Algorithm Performance Validation" and conformance to ANSI/AAMI/ISO EC57, the ground truth would typically be established by expert cardiologists reviewing the ECG tracings and labeling arrhythmias. The document does not specify if pathology or outcomes data were used, which is less common for arrhythmia detection algorithms unless used for long-term prognosis.

8. The sample size for the training set

  • Training Set Sample Size: Not specified. The document focuses on validation rather than the development or training process.

9. How the ground truth for the training set was established

  • Training Set Ground Truth Establishment: Not specified.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the words "CARD GUARD" in bold, black letters. Below the words "CARD GUARD" are the words "Scientific Survival LTD" in a smaller font. The image is a close-up of the text, and the background is white.

510(k) Summary: Modified CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector

Introduction

JUN 2 5 2010

This document contains the 510(k) summary for the modified CG-6108 1L Continuous ECG Monitor and Arrhythmia Detector. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

SubmitterCard Guard Scientific Survival Ltd.,
EstablishmentRegistrationNumber9681879
Address2 Pekeris St., P.O.B. 527, Rehovot, 76100, Israel
Contact person:Asher Kassel, Director of RA & QA, Card Guard Scientific Survival Ltd.
Phone:972-8-9484010 Fax: 972-8-9484044(direct)
E-mail:asherk@cardguard.com
Date Prepared:June 8, 2010
Predicate deviceUnmodified version of CG-6108 ACT-1L Continuous ECG Monitor andArrhythmia Detector, cleared in K071995 on December 18, 2007.
Trade Name:CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector
Classification:Detector and alarm, arrhythmia /Transmitters and receivers, electrocardiograph, telephone
Product Code:DSI DXH
Regulation No:870.1025, 870.2920
Class:II

Device Description

CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector is designed for self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center.

The chest-worn sensor is used for the acquisition, recording, and transmission of the ECG signal. The device is equipped with 3 electrodes on a harness and it houses a 3.6V AA battery, a Bluetooth transceiver and a buzzer.

The ECG signals are transmitted via Bluetooth to a handheld device with a proprietary interactive application, configured to process and transmit the ECG recordings. The handheld device is a mobile computing device with a display and a touch input such as a cell-phone. It has sufficient memory and processing capability to run the proprietary application.

When an arrhythmia event is detected the handheld device transmits the recorded ECG information automatically via cellular link, to the Monitoring Center or professional analysis. When cellular service is unavailable the patient has an option to transmit via landline telephone.

Indications for Use:

The CG-6108 Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors a one lead ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the words "CARD GUARD" in bold, black letters. Below that, in a smaller font, are the words "Scientific Survival LTD". The text appears to be part of a logo or label. The background is not clear, but it seems to be part of a larger document or image.

Special 510(k) for CG-6108 Continuous ECG monitor and Arrhythmia Detector Section 7: 510(k) Summary .

Summary of the Technological Characteristics / Principles of Operation

The technological characteristics and principles of operation of the modified device are the same as the predicate device. The chest-worn ECG sensor transmits signals via Bluetooth to the handheld device equipped with the Medical Application. which incorporates an algorithm for detection of cardiac events: Atrial Fibrillation, Tachycardia and Pause. A detected artifact triggers transmission of the signal to the Monitoring Center for analysis.

Non-clinical performance data for the CG-6108 ACT-1L:

The modified version CG-6108 has been subjected to extensive verification / validation testing. Final testing of the system included various performance tests and software validation tests designed to ensure that the device meet all of its functional and performance requirements and is fit for its intended use. The following list summarizes the testing performed on the device;

  • . Software Verification and Validation
    • Software Functional Unit Verification o
    • System Level Software Validation o
    • Arrhythmia Detection Algorithm Performance Validation o
  • Hardware Verification and Validation .
  • Verification of Conformance with Special Controls Guidance Document .

Performance Standards:

This 510(k) submission was written in accordance with the FDA Guidance document "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm, October 28, 2003" and the device conforms to the applicable performance requirements contained in this document. In addition, this submission was prepared in accordance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005". The design of the CG-6108 ACT-1L conforms to the following voluntary standards:

  • . ANSI/AAMI/ISO EC57:1998 (R) 2008: Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms
  • ANSI/AAMI EC38:1998 Ambulatory Electrocardiograph ●
  • ISO 14971:2007: Medical devices application of risk management to medical devices; .
  • IEC 60601-1:1988, 2nd edition, Part 1, plus A1:1991 and A2:1995: Medical electrical . equipment; Part 1: General requirements for safety
  • IEC 60601-1-2: 2001, plus am. 1:2004, Part 1: Medical electrical equipment, Part 1-2; . Electromagnetic compatibility - Requirements and tests
  • IEC 60601-1-4:2000, plus Amendment 1:2004: Medical electrical equipment; Part 1: 4. Collateral . Std: Programmable electric medical systems
  • IEC 62304:2006: Medical device software Software life cycle processes .
  • ISO 15223:2007: Medical devices Symbols to be used with medical device labels, labeling . and information to be supplied

Substantial Equivalence:

The modified CG-6108 ACT-1L device is substantially equivalent with respect to the indications for use, technological characteristics and performance characteristics to the identified legally marketed predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Card Guard Scientific Survival Ltd., c/o Clay Anselmo President and CEO, Reglera LLC. 555 Zang Street Suite 100 Lakewood, Colorado 80228

Re: K101639

Trade/Device Name: CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector Regulation Number: 21 CFR 870.1025, 21 CFR 870.2920 Regulation Name: Arrhythmia Detector and Alarm, Telephone Electrocardiograph Transmitter and Receiver. Regulatory Class: Class II (two) Product Code: DSI, DXH Dated: June 8, 2010 Received: June10, 2010

JUN 2 5 2010

Dear Mr. Anselmo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{3}------------------------------------------------

Page 2 - Mr. Anselmo

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm142641.htm

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): _ K10 1639

Device Name: CG-6108 ACT-11 Continuous ECG Monitor and Arrhythmia Detector

Indications for Use:

The CG-6108 ACT-1L Continuous ECG Monitor and Arrhythmia Detector are intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECC, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center.

The monitoring center provides the ECG data to the medical practitioner for evaluation.

Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular
510(k) NumberK10-1639

Page 1 of 1

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.