K Number
K101592

Validate with FDA (Live)

Date Cleared
2011-10-06

(486 days)

Product Code
Regulation Number
882.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Erase Pen (RF Controller) is only used with the Erase Tip (Single Use Sterile Detachable Electrode Cartridge system) for thermal coagulation of soft tissue and peripheral nerves by creating radiofrequency (RF) lesions in nerve tissue. It is intended for use only by trained clinicians in hospital or clinical setting.

Device Description

The Erase Pen (hand-held Bi-Polar Radiofrequency battery operated controller) and the Erase Tip (single use, sterile detachable electrode cartridge system) is a hand-held Electrosurgical Device to locate and ablate nerves. The depth of the electrodes, the amount of RF energy. delivered, and the duration of energy applied control the size of the ablation.

AI/ML Overview

The provided 510(k) summary for the Cheng Medical Corporation Erase Pen and Erase Tip System for Nerve Ablation (K101592) does not provide explicit acceptance criteria or details of a specific study proving the device meets them.

Instead, the document focuses on demonstrating substantial equivalence to a predicate device (Radionic's RFG-3C, K982489) based on shared technological characteristics and similar operating parameters. The "Device Testing Results and Conclusion" section broadly states that "All necessary testing was performed... to ensure that the [device] is substantially equivalent... and to ensure that the new device does not raise new issues regarding safety and effectiveness." This indicates that general safety and performance testing was conducted, but the specifics of what was tested, what the acceptance criteria were, and what the results were are not detailed in this summary.

Therefore, many of the requested items cannot be answered directly from the provided text.

Here is a breakdown of what can be inferred or directly stated based on the given information, and where information is missing:


1. A table of acceptance criteria and the reported device performance

  • No explicit acceptance criteria are provided.
  • The document presents a table comparing "Erase Pen Operating Characteristics" with the predicate device (Radionics RFG-3C), which serves as a comparison of performance parameters rather than predefined acceptance criteria. The implied "performance" is that these parameters are sufficiently similar to the predicate device for substantial equivalence.
ParameterErase PenRadionics RFG-3C
a. Maximum voltage deliverable, Vp-p56 into 2000Ω and 3000Ω load200 into 2000Ω load*
b. Maximum current deliverable, amps0.18 into 100Ω load0.70 into 100Ω load
c. Maximum power deliverable, watts3.5 into 100Ω load50 into 100Ω load
d. Maximum voltage accessible, Vp-p71161
e. Maximum current accessible, amps0.201.17
f. Maximum power accessible, watts568
g. Frequency of RF energy, kHz460+/-5%500+/-10%
h. Shape of the waveform during RF ablationsinusoidalsinusoidal
i. Shape of the waveform during Tissue Stimulation200µs one-shot pulse (monophasic square wave)200µs one-shot pulse (monophasic square wave)
j. Impedance cutoffs, ohms20 to 22050 to 5000
k. Maximum current limit, amps0.180.70

Note: The document states "The set output level Erase Pen and adjustable output level Radionics RF generators would have an output power equivalence when the Radionics RFG-3C is at a setting of 3 to 5, depending on the tissue impedance." This implies that despite some numerical differences, the functional output can be considered equivalent under certain conditions.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not provided. The summary mentions "All necessary testing was performed," but does not specify the nature of the test set (e.g., in-vitro, ex-vivo, animal, human clinical). Therefore, details like sample size, provenance, or retrospective/prospective nature are absent.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not provided. Since no specific test set involving human interpretation or subjective assessment is described, there's no mention of experts or ground truth establishment in this context. The testing described appears to be engineering/performance testing of the device itself.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not provided. As above, no such test design is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This device is a radiofrequency ablation system, not an AI-assisted diagnostic or interpretative device. Therefore, an MRMC study comparing human reader performance with and without AI assistance is irrelevant and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This device is a physical electrosurgical instrument, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not provided in detail. For a device like this, the "ground truth" for performance testing typically refers to objective measurements of physical parameters (e.g., power output, lesion size/temperature in a tissue phantom or animal model) and adherence to engineering specifications and safety standards. The document states "to ensure that the product... does not raise new issues regarding safety and effectiveness," implying such testing was done, but specific ground truth types are not detailed.

8. The sample size for the training set

  • Not applicable / Not provided. As this is a physical medical device and not an AI/ML algorithm requiring a training set, this information is not relevant and not present in the document.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. See point 8.

{0}------------------------------------------------

K101592

Cheng Medical Corporation

510(k) Notification

Erase Pen and Erase Tip System for Nerve Ablation (Models HC-0001 and CS-0001)

APPENDIX A: 510(k) SUMMARY

510(k) SUMMARY

OCT - 6 2011

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

Name, Address, Phone and Fax number of the Applicant A.

Cheng Medical Corporation. 12945 Saratoga Avenue Saratoga, CA 95070 Telephone: (408) 255-3223 Fax: (408) 255-3957

в. Contact Person

Mary P. Gallup Regulatory Affairs Consultant Hantel Technologies, Inc. Telephone: (510) 441-4017

C. Date Prepared

06/04/2010

D. Device Name

Trade Name: Model HC-0001 Erase Pen Model CS-0001 Erase Tip Regulation Number: 21 CFR 882.4400 and 21 CFR 882.4725 Regulation Number: Radiofrequency Lesion Generator and Lesion Probe Regulation Class: II Product Code: GXD, GXI

แ่ Device Description

The Erase Pen (hand-held Bi-Polar Radiofrequency battery operated controller) and the Erase Tip (single use, sterile detachable electrode cartridge system) is a hand-held Electrosurgical Device to locate and ablate nerves. The depth of the electrodes, the amount of RF energy. delivered, and the duration of energy applied control the size of the ablation.

F. Intended Use

The Erase Pen and Erase Tip (Single use Sterile Detachable Electrode Cartridge system) is intended to create radiofrequency (RF) heat lesion in nerve tissue. It is intended for use only by trained clinicians in hospital or clinical setting.

The Erase Pen and Erase Tip (Single use Sterile Detachable Electrode Cartridge system) ™ Last update: 10/03/2011

{1}------------------------------------------------

STATES AND

Cheng Medical Corporation

Erase Pen and Erase Tip System for Nerve Ablation (Models HC-0001 and CS-0001)

Substantial Equivalence G.

The Erase Pen Model HC-0001 and Erase Tip Model CS-0001 is substantially equivalent in the following technological ways to the identified predicate device: Radionic's RFG-3C (K982489) system.

  • � Indication For Use
  • RF Energy Type Bi Polar �
  • Functional (stimulation and ablation) ●
  • Probe Type (single use) .

Erase Pen Operating Characteristics compared with predicate device

ParameterErase PenRadionics RFG-3C
a. Maximum voltage deliverable, Vp-p56 into 2000Ω and 3000Ω load200 into 2000Ω load*
b. Maximum current deliverable, amps0.18 into 100Ω load0.70 into 100Ω load
c. Maximum power deliverable, watts3.5 into 100Ω load50 into 100Ω load
d. Maximum voltage accessible, Vp-p71161
e. Maximum current accessible, amps0.201.17
f. Maximum power accessible, watts568
g. Frequency of RF energy, kHz460+/-5%500+/-10%
h. Shape of the waveform during RF ablationsinusoidalsinusoidal
i. Shape of the waveform during Tissue Stimulation200µs one-shot pulse (monophasic square wave)200µs one-shot pulse (monophasic square wave)
j. Impedance cutoffs, ohms20 to 22050 to 5000
k. Maximum current limit, amps0.180.70

Note: The set output level Erase Pen and adjustable output level Radionics RF generators would have an output power equivalence when the Radionics RFG-3C is at a setting of 3 to 5, depending on the tissue impedance.

  • Information not available for the maximum voltage deliverable of the Radionics RFG-3C into a 3000Ω impedance.

H. Device Testing Results and Conclusion

All necessary testing was performed on the Erase Pen Model HC-0001 and Erase Tip (Single use Sterile Detachable on the Life Life Life Life Fer Model 110-0001 and Erate The (Single
is substantially ogying and the Cartridge system) Model CS-0001 to ensure that the is substantially equivalent to the overlies devices and to ensure that the product
raise new issues recording of the sub-cess and to ensure that the new device does not raise new issues regarding safety and effectiveness

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services. The seal is circular and contains the department's logo in the center. The logo is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the perimeter of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WQ66-G609 Silver Spring, MD 20993-0002

Cheng Medical Corporation c/o Mary Pascual Gallup Vice President of Regulatory Affairs, Consultant Hantel Technologies, Inc. 12945 Saratoga Avenue Saratoga, California 95070

  • 6 2011

Re: K101592

Trade/Device Name: Erase Hand-held Controller and Erase Single Use Electrode Cartridge System Regulation Number: 21 CFR 882,4400 Regulation Name: Radiofrequency lesion generator Regulatory Class: Class II Product Code: GXD Dated: September 28, 2011 Received: September 30, 2011

Dear Ms. Pascual Gallup:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practive requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Alexander

$\int_a^b$

Malvina B. Evdelman, M.D. Director Division of Ophthalmic. Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Cheng Medical Corporation

510(k) Notification

Erase Pen and Erase Tip System for Nerve Ablation (Models HC-0001 and CS-0001)

SECTION 4: INDICATION FOR USE STATEMENT

Indications for Use

510(k) Number (if known): {{ 0 (5922

Device Name:

Model HC-0001 The Erase Pen Model CS-0001 The Erase Tip

Indications for Use:

Page 1 of 1_

The Erase Pen (RF Controller) is only used with the Erase Tip (Single Use Sterile Detachable Electrode Cartridge system) for thermal coagulation of soft tissue and peripheral nerves by creating radiofrequency (RF) lesions in nerve tissue. It is intended for use only by trained clinicians in hospital or clinical setting.

Prescription Use _Yes_and/or Over-The-Counter Use _ No

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)


(Division Sign-Off)

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K101592

The Erase Pen and Erase Tip (Single use Sterile Detachable Electrode Cartridge system) ™

Page 12 Last update: 10/03/2011

§ 882.4400 Radiofrequency lesion generator.

(a)
Identification. A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.(b)
Classification. Class II (performance standards).