(299 days)
Not Found
No
The description focuses on the physical characteristics of the stent and delivery system, and the performance studies are bench and clinical tests related to mechanical properties and equivalence to predicate devices. There is no mention of AI, ML, image processing, or data-driven algorithms.
Yes
The device is described as being used for the "palliative treatment of duodenal or gastric outlet obstruction and duodenal strictures caused by malignant neoplasms," which indicates a therapeutic purpose rather than diagnostic or prognostic.
No
The device is described as a stent used for the "palliative treatment of duodenal or gastric outlet obstruction and duodenal strictures caused by malignant neoplasms," indicating a therapeutic rather than diagnostic function. It is designed to physically treat blockages, not to identify or characterize medical conditions.
No
The device description clearly details a physical stent made of nitinol wire and an introducer system with catheters and a handle, indicating it is a hardware medical device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for the palliative treatment of duodenal or gastric outlet obstruction and duodenal strictures caused by malignant neoplasms. This is a therapeutic intervention performed directly on the patient's body.
- Device Description: The device is a stent and an introducer system designed to be implanted within the body.
- Lack of IVD Characteristics: An IVD is a medical device used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health. This device does not perform any such tests on samples.
Therefore, this device falls under the category of an implantable medical device used for treatment, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
This device is used for palliative treatment of duodenal or gastric outlet obstruction and duodenal strictures caused by malignant neoplasms.
Product codes
MUM
Device Description
Stent Description: This flexible, self-expanding stent is constructed of nitinol wire. The total length of the stent is indicated by radiopaque markers on the inner catheter, indicating the actual length of the stent at nominal stent diameter. The stent has flanges at both stent ends. Introducer System Description: The stent is mounted on an inner catheter, which accepts a 0.035 inch wire guide and is constrained by an outer catheter. A pistol-grip delivery handle allows stent deployment or recapture.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
duodenal, gastric outlet
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Performance (bench and clinical) testing was carried out to determine the equivalence of the Evolution™ Duodenal Stent System to the predicate devices and to verify the safety and effectiveness of the device. The following bench tests were carried out: deployment force testing, expansion force testing, compression force testing, dimensional testing, corrosion testing, tensile strength testing and MRI testing.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 878.3610 Esophageal prosthesis.
(a)
Identification. An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
0
Section 5.0 510(k) Summary
MAR 2 9 2011
Name: | Cook Ireland Ltd |
---|---|
Address: | O'Halloran Road |
National Technology Park | |
Limerick, Ireland | |
Phone: | 353 61 334440 |
Fax: | 353 61 239293 |
Contact Persons: | Jacinta Kilmartin, Regulatory Affairs Specialist |
Emmett Devereux, Director Quality & | |
Regulatory Affairs | |
Phone: | 353 61 334440 |
Fax: | 353 61 239 293 |
Date: | March 24, 2011 |
Trade Name: | Evolution™ Duodenal Stent System |
Common Name: | Stent, Metallic, Expandable, Duodenal |
Classification Name: | Stent, Metallic, Expandable, Duodenal (21 CFR |
878.3610, Product Code: MUM) | |
Predicate Devices: | Boston Scientific Wallstent® Enteral |
Endoprosthesis with Unistep™ Delivery System | |
(K991056) | |
Boston Scientific WallFlex™ Enteral Duodenal | |
Stent System with Anchor Lock Delivery System | |
(K062750) | |
Description of the Device: | Stent Description: |
This flexible, self-expanding stent is constructed | |
of nitinol wire. The total length of the stent is | |
indicated by radiopaque markers on the inner | |
catheter, indicating the actual length of the stent at | |
nominal stent diameter. The stent has flanges at | |
both stent ends. | |
Introducer System Description: | |
The stent is mounted on an inner catheter, which | |
accepts a 0.035 inch wire guide and is constrained | |
by an outer catheter. A pistol-grip delivery | |
handle allows stent deployment or recapture. | |
Indications for use: | This device is used for palliative treatment of |
duodenal or gastric outlet obstruction and | |
duodenal strictures caused by malignant | |
neoplasms. | |
Comparison of Characteristics: | The Evolution™ Duodenal Stent System is |
substantially equivalent to the currently marketed | |
predicate devices, the Boston Scientific | |
Wallstent® Enteral Endoprosthesis with | |
Unistep™ Delivery System (K991056), and the | |
Boston Scientific WallFlex™ Enteral Duodenal | |
Stent System with Anchor Lock Delivery System | |
(K062750). | |
Performance Data: | Performance (bench and clinical) testing was |
carried out to determine the equivalence of the | |
Evolution™ Duodenal Stent System to the | |
predicate devices and to verify the safety and | |
effectiveness of the device. The following bench | |
tests were carried out: deployment force testing, | |
expansion force testing, compression force | |
testing, dimensional testing, corrosion testing, | |
tensile strength testing and MRI testing. |
1
.
Image /page/1/Picture/2 description: The image contains a sequence of alphanumeric characters, seemingly handwritten. The sequence reads 'KI01530', with each character distinct and clearly formed. The characters are arranged linearly from left to right, with consistent spacing between them.
:
:
2
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". The text is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Jacinta Kilmartin Regulatory Affairs Specialist Cook Ireland Ltd. National Technology Park O'Halloran Road Limerick IRELAND
MAR 2 9 201
. Re: K101530
Trade/Device Name: Evolution™ Duodenal Stent System Regulation Number: 21 CFR 8878,3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: MUM Dated: February 14. 2011 Received: February 16, 2011
Dear Ms. Kilmartin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability v warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
3
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Hubert Lemon MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Section 4.0 Indications for Use
| 510(k) Number (if known): | ﺳﺎﺗﮫ ﺍﻭﺭ ﺍﺱ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾ
C
( |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Device Name: Evolution™ Duodenal Stent System
Indications for Use:
This device is used for palliative treatment of duodenal or gastric outlet obstruction and duodenal strictures caused by malignant neoplasms.
Prescription Use V AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helen Klemm
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K101530
Page 1 of _1