K Number
K101512

Validate with FDA (Live)

Date Cleared
2010-07-01

(30 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) is a device intended to measure the systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.

Device Description

Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Tarqet (BP3MD1-4) is designed to measure systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an individual by using a noninvasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses a resistive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4), based on the provided 510(k) summary:

This device did not undergo a clinical accuracy study for its blood pressure measurement technology as part of this 510(k) submission. The manufacturer claimed substantial equivalence to a predicate device (Model BP3BT0-1) on the basis that the differences between the new device and the predicate did not affect the blood pressure measurement technology. Therefore, they concluded that repeated clinical testing in accordance with ANSI/AAMI SP10 was not required.

However, the submission does detail non-clinical tests that were performed.


1. Table of Acceptance Criteria and Reported Device Performance

Since no new clinical accuracy study for blood pressure measurement was performed for this specific device model, the table below refers to the non-clinical tests mentioned. The acceptance criteria for these would generally be adherence to the relevant standards and guidance documents.

Acceptance Criteria (Non-Clinical)Reported Device Performance
Reliability Test Requirements:Met all relevant requirements.
a. Storage testPassed
b. Operating testPassed
c. Vibration testPassed
d. Drop testPassed
e. Life testPassed
EMC Test RequirementsMet all relevant requirements.
f. EMC TestPassed
Software Validation Requirements (Moderate level of concern)Met all relevant requirements.
Validation conducted in accordance with FDA November 2005 guidance document.
Overall Safety and EffectivenessDemonstrated substantial equivalence with predicate devices.
No significant differences in terms of safety and effectiveness (electrical, mechanical, environmental) per FDA DCRND November 1993 Draft "Reviewer Guidance" and ANSI/AAMI SP10:2008.

2. Sample Size Used for the Test Set and Data Provenance

  • Non-Clinical Tests: The documentation does not specify the sample sizes (e.g., number of devices tested) for the reliability, EMC, or software validation tests.
  • Clinical Accuracy (for blood pressure measurement): No new clinical test set was used for this 510(k) submission. The manufacturer relied on the clinical accuracy of its predicate device, Model BP3BT0-1, which presumably underwent clinical testing prior to its own clearance.
  • Data Provenance: Not applicable for new clinical data for this submission. The non-clinical tests would have been performed by the manufacturer (Microlife Intellectual Property GmbH, Switzerland).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Not applicable for this 510(k) submission. Since no new clinical accuracy testing was performed, there was no need to establish ground truth for a new clinical test set. The manufacturer's argument for substantial equivalence was based on engineering analysis and non-clinical testing, plus the prior clearance of a clinically-tested predicate device.

4. Adjudication Method for the Test Set

  • Not applicable for this 510(k) submission. No clinical test set requiring adjudication was used.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically for evaluating devices where human interpretation of medical images or data is involved, often with AI assistance. This device is a standalone blood pressure monitor.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • No standalone clinical performance study was performed for this specific device model as part of this 510(k) submission. The device (Microlife Upper Arm Automatic Digital Blood Pressure Monitor) inherently operates in a "standalone" manner (algorithm-only for blood pressure measurement, without human intervention in the interpretive phase), but the clinical accuracy validation was based on substantial equivalence to a predicate device, rather than a de novo clinical trial for this specific model.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

  • For the non-clinical tests, the ground truth was derived from adherence to engineering standards and regulatory guidance documents.
  • For the blood pressure measurement accuracy (which was deemed unchanged from the predicate), the ground truth for the predicate device's clinical testing would have been reference measurements from trained clinicians using a validated gold-standard method (e.g., auscultation by multiple observers, or intra-arterial measurement, as per ANSI/AAMI SP10). However, this was not directly assessed again for the subject device.

8. The Sample Size for the Training Set

  • Not applicable. This device uses a well-known "oscillometric method" algorithm, not a machine learning model that requires a "training set" in the typical sense. The algorithm is based on established physical principles.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As the device uses an established oscillometric algorithm, there isn't a "training set" with associated ground truth in the context of AI/ML development. The algorithm's principles and parameters would have been developed and refined through engineering and physiological understanding, and validated against clinical data during the development of earlier predicate devices.

{0}------------------------------------------------

510(k) SUMMARY

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K101572

1. Submitter's Identification:

Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland

JUL -1 2010

Date Summary Prepared: May 25, 2010 .

Contact: Mr. Gerhard Frick Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard.frick@microlife.ch

2. Name of the Device:

Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4)

Requlation Number: 21 CFR Part 870.1130 Requlation Name: Non-Invasive Blood Pressure Measurement System Requiatory Class: II Product Code: DXN

3. Information for the 510(k) Cleared Device (Predicate Device):

  • a. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3BT0-K013485. Microlife Corporation.
  • b. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (Twin200), K082357, Microlife Intellectual Property GmbH.

4. Device Description:

Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Tarqet (BP3MD1-4) is designed to measure systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an individual by using a noninvasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses a resistive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

{1}------------------------------------------------

The target blood pressure value 140/90 mmHg is used for hypertensive patient. The target blood pressure value 130/80 mmHq is used for hypertension patients with diabetes, coronary heart disease or chronic kidney disease. If both the measured systolic and diastolic blood pressure values are below the set target values (140/90 or 130/80), the screen will light in green with two <> signs and show the pulse pressure (PP). If either the measured systolic result is above the set target value (140/90 or 130/80), the screen will light in red with the difference to the goal value shown in display as number.

5. Intended Use:

The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) is a device intended to measure the systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.

Comparison to the 510(k) Cleared Devices (Predicate Devices): 6.

a) Subject Modified Device Compared to BP3BT0-1 Predicate, K013485:

The subject modified device, Model WatchBP Office Target (BP3MD1-4) and the predicate device model BP3BT0-1, use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. Upper arm cuff is inflated automatically, deflation rate is controlled by one factory set exhaust valve and the deflation pressures are transferred to one sensor.

The differences between the two models are the sensor type and additional features such as blood pressure targets setting function, red/green backlight function and PP automatically calculation function. However, the differences do not affect the accuracy and normal use of this device based on the internal clinical tests comparing different sensors and cuff bladder materials.

  • b) Subject Modified Device Compared to WatchBP Office (Twin200) Predicate, K082357:
    The subject modified device. WatchBP Office Target (BP3MD1-4) uses the same oscillometric method as the predicate WatchBP Office (Twin200) to determine the svstolic and diastolic blood pressure, pulse rate and pulse pressure (PP). Upper arm cuff is inflated automatically by pump. the deflation rate is controlled by factory set exhaust valve and the deflation pressures are transferred via tubing to a sensor in these two units.

They differ by the intended use, measuring location, measuring mode, power source type. MAP automatically calculation function etc., however, the differences do not affect the accuracy and normal use of this device because they use the same fundamental scientific technology. Based upon the aforementioned information, the two devices are substantially equivalent.

The intended use of the subject modified device, the WatchBP Office Target (BP3MD1-4), has not changed from the 510(k) cleared predicate device labeling

{2}------------------------------------------------

since it is a combination of the two (2) predicate devices stated in this submission. All comparison parameters stated in our comparison chart (included in the submission) come from either one of the predicate devices cited.

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Testing information demonstrating safety and effectiveness of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements.

The following testing was conducted:

  • a. Reliability Test Storage test
  • b. Reliability Test Operating test
  • c. Reliability Test Vibration test
  • d. Reliability Test Drop test
  • e. Reliability Test Life test
  • EMC Test نه

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) tested met all relevant requirements of the aforementioned tests.

8. Discussion of Clinical Tests Performed:

The subject modified device Model WatchBP Office Target (BP3MD1-4) is from the technical point of view, identical to the blood pressure monitor Model BP3BT0-1. The differences between them do not related to blood pressure measurement technology; therefore the clinical accuracy in terms of blood pressure detection will not be affected. Based on Microlife's risk analysis, repeated clinical testing in accordance with ANSI/AAMI SP10 is therefore not required.

9. Software information:

Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

10. Conclusions:

We have demonstrated that there are no significant differences between the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) and the predicate devices, in terms of safety and effectiveness based on electrical, mechanical and environmental test results per the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and the ANSI/AAMI Voluntary Standard, SP10: 2008.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings, rendered in black lines. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Microlife Intellectual Property GmbH Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021

JUL - 1 2010

Re: K101512

Device Name: Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (Two) Product Code: DXN Dated: May 26, 2010 Received: June 1, 2010

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 - Ms. Susan D. Goldstein-Falk

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Exhibit B

Indications for Use

510(k) Number (if known):

K101512

Device Name: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4)

Indications For Use:

The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) is a device intended to measure the systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

J.W.C.

Page 1 of 1

ivision Sign-C diovascular Devices Division of Car

510(k) Number K101502

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).