K Number
K013485

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-01-17

(90 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 with Optional Thermal Printer, Model PR 1KA1 is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. This device can be used in connection with the Microlife Thermal Printer.

Device Description

The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method". An optional printer is included as an accessory which can be sold separately in a separate box with the printer user manual.

AI/ML Overview

The provided text describes the Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1.

Here's an analysis of the acceptance criteria and the study information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
ANSI/AAMI SP10-1992 "National Standard for Electronic or Automated Sphygmomanometers"The BP-3BTO-1 met all relevant requirements of this standard.
FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND (Electrical, Mechanical and Environmental Performance Requirements)The device met all relevant requirements of the aforementioned tests (General Functions, Reliability (Operation Conditions, Drop, Storage, Vibrating), EMC, IEC 60601-1 Safety, Printer Port Function, Thermal Printer Software Validation, EMC with Printer).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the sample size (number of subjects) used for the clinical testing in accordance with ANSI/AAMI SP10-1992.
  • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective/prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document refers to the ANSI/AAMI SP10-1992 standard for clinical testing. This standard typically involves comparison of the automated device readings against auscultatory measurements performed by trained observers (experts). However, the specific number of experts and their qualifications (e.g., "radiologist with 10 years of experience") are not explicitly stated in the provided 510(k) summary.

4. Adjudication Method for the Test Set

The document does not explicitly describe the adjudication method (e.g., 2+1, 3+1, none) for the test set. For blood pressure measurement, the ground truth is typically established by trained observers taking multiple readings.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

A Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. This type of study is more common for diagnostic imaging devices where human readers interpret results, and the AI's effect on their performance is evaluated. This device is an automated blood pressure monitor, and its performance is assessed against a standard, not in terms of improving human reader effectiveness.

6. Standalone Performance Study

Yes, a standalone study was performed. The entire clinical testing section (Discussion of Clinical Tests Performed) is dedicated to evaluating the device's performance (i.e., algorithm only without human-in-the-loop performance) against the ANSI/AAMI SP10-1992 standard. The statement "The BP-3BTO-1 met all relevant requirements of this standard" confirms the standalone performance evaluation.

7. Type of Ground Truth Used

The ground truth for the clinical testing (per ANSI/AAMI SP10-1992) would typically involve expert auscultatory measurements. These are considered the "expert consensus" or "reference method" in blood pressure measurement testing. The device states its method is "similar to the auscultatory method" in its description.

8. Sample Size for the Training Set

The document does not provide information regarding a specific "training set" sample size. This device relies on an "oscillometric method" and a software algorithm. While the algorithm was developed, the submission focuses on validation against a standard, not on a machine learning training paradigm with a distinct training set. The "fuzzy logic inflation" and "split slope resolution" are algorithmic improvements, but information about a formal "training set" as understood in modern AI development is absent.

9. How the Ground Truth for the Training Set Was Established

Since there is no explicit mention of a "training set" in the context of modern machine learning, there is no information on how ground truth for such a set was established. The device uses a well-known "oscillometric method" and algorithmic enhancements, implying that the underlying principles and parameters of the algorithm were likely developed based on established physiological principles and empirical data, rather than a separate, labeled "training set" in the AI sense.

{0}------------------------------------------------

EXHIBIT # 1

510(K) SUMMARY

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 5l0(k) number is:

1. Submitter's Identification:

Microlife Corporation 9F, 431 Rui Guang Road Nei Hu. Taipei 114 Taiwan, Republic of China

Contact:

Mr. Laurence Yanq

Date Summary Prepared: October 17, 2001

2. Name of the Device:

Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1.

3. Predicate Device Information:

The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 is substantially equivalent to the Microlife Automatic Blood Pressure Monitor, Model BP-2BHO, K# 970211.

4. Device Description:

The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse

{1}------------------------------------------------

rate, which is a well - known technique in the market called the "oscillometric method".

An optional printer is included as an accessory which can be sold separately in a separate box with the printer user manual.

Intended Use: 5.

The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. This device can be used in connection with the Microlife Thermal Printer.

Comparison to Predicate Devices: 6.

Both devices use the well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. An upper arm cuff is inflated automatically; deflate rate is controlled by a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. Each device uses a similar capacitance-type pressure sensor to translate the pressure variations to electrical signals that can be interpreted by an integrating circuit. Once the reading is determined each unit operates a solenoid valve to release the pressure to zero. Our Blood Pressure Monitor, Model BP-3BTO-1, differs from the predicate device in the fuzzy logic inflation.

The interface between the sensor and the microprocessor determines the system's accuracy. For our Pressure Monitor, Model BP-3BTO-1, the software is capable of a split slope resolution to improve accuracy over the entire range. Since the range is "split" into the three sections (0 to 100mmHg) (100 to 200mmHg) (200 to 300mmHg) error due to nonlinearity is reduced by the ability to adjunct the slope to best fit the output curve. A nonlinearity of 1% is reduced to 0.33% by splitting the span into three separated linear relations. This way the sensor is matched to the software by using a series of jumpers that profile the slopes to the output of the sensor.

The Microlife Automatic Blood Pressure Monitor, Model BP-3BTO-1, has field calibration access; this model is initiated by the following steps: (a) Remove one battery from compartment first, (b) When inserting back the battery, press and hold "I/O" key, and don't release it until the battery is inserted well.

Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:

{2}------------------------------------------------

Testing information demonstrating safety and effectiveness of the Microlife Automatic Blood Pressure Monitor, Model BP-3BTO-1 in the intended Automatio Drosa is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.

The following testing was conducted:

  • General Functions Test a.
  • Reliability Test Operation Conditions b.
  • Reliability Test Drop Testing C.
  • Reliability Test Storage ರ.
  • Reliability Test Vibrating Testing e.
  • f. EMC Test
  • IEC 60601-1 Safety Test g.
  • Blood Pressure Monitor Printer Port Function Test Report h.
  • Thermal Printer Software Validation Report ....
  • EMC Test Report Blood Pressure Monitor Connected with j. Thermal Printer

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 tested met all relevant requirements of the aforementioned tests.

Discussion of Clinical Tests Performed: 8.

ANSI/AAMI SP10-1992 "National Standard for Electronic or Automated Sphygmomanometers" testing was performed. All relevant sections were addressed and testing conducted. The BP-3BTO-1 met all relevant requirements of this standard.

9. Conclusions:

We have demonstrated that the Microlife Automatic Blood Pressure Monitor, Model BP-3BTO-1, is as safe and effective as the predicate, the Microlife Automatic Blood Pressure Monitor, Model BP-2BHO based on electrical, mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and, the ANSI/AAMI Voluntary Standard, SP10-1992 testing results.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized eagle-like symbol, which is the official emblem of the HHS. The seal is black and white.

JAN 1 7 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Microlife Corporation c/o Ms. Susan D. Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021

Re: K013485

Trade Name: Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1 Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: October 17, 2001 Received: October 19, 2001

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Susan D. Goldstein-Falk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Kala Teler

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Exhibit B

Page __ 1_____________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): __ KO13485

Device Name: Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1

Indications For Use:

The Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1 with Optional Thermal Printer, Model PR 1KA1 is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. This device can be used in connection with the Microlife Thermal Printer.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
5'10(k) Number K01348

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use X
(Optional Format 1-2-96)

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).