K Number
K101435
Device Name
IMIX PANORAD AND SOMARAD
Date Cleared
2010-09-15

(117 days)

Product Code
Regulation Number
892.1680
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
IMIX PanoRad and SomaRad X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. (Not for mammography.)
Device Description
The PanoRad consists of: A high resolution CCD detector with 14" x 17" or 17" x 17" image area. An X-Ray Tube & Collimator. A fully motorized positioned. A 65 kW high frequency generator. A 4 way floating table. An Image Acquisition workstation with DICOM 3 compliance. The SomaRad consists of: A high resolution CCD detector with 14" x 17" or 17" x 17" image area. An X-Ray Tube & Collimator. A fully motorized positioned. A 50 kW high frequency generator. A table. An Image Acquisition workstation with DICOM 3 compliance.
More Information

Not Found

No
The summary describes a standard digital X-ray system with image acquisition and DICOM compliance, but there is no mention of AI or ML in the intended use, device description, or performance studies. The "Mentions AI, DNN, or ML" field is explicitly marked as "Not Found".

No.
The device is used for diagnostic imaging purposes (generating radiographic images) and does not treat or cure any medical condition.

Yes

The 'Intended Use / Indications for Use' section explicitly states that the device is "intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures."

No

The device description explicitly lists multiple hardware components including CCD detectors, X-Ray Tubes, collimators, motorized positioners, generators, and tables. While it includes an Image Acquisition workstation with DICOM compliance (which would involve software), the device is clearly a system incorporating significant hardware.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to "generat[e] radiographic images of human anatomy." This is a diagnostic imaging device that works by capturing images of the inside of the body using X-rays.
  • Device Description: The components listed (X-Ray Tube, Detector, Generator, Table, Workstation) are all typical of an X-ray imaging system.
  • Lack of IVD Characteristics: IVD devices are used to examine specimens (like blood, urine, or tissue) outside of the body to provide information about a person's health. This device does not interact with biological specimens in this way.

Therefore, the IMIX PanoRad and SomaRad X-Ray Systems are diagnostic imaging devices, not IVDs.

N/A

Intended Use / Indications for Use

IMIX PanoRad and SomaRad X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. Not for mammography.

Product codes

KPR, MOB, LLZ

Device Description

The PanoRad consists of:

  • A high resolution CCD detector with 14" x 17" or 17" x 17" image area.
  • An X-Ray Tube & Collimator.
  • A fully motorized positioned.
  • A 65 kW high frequency generator.
  • A 4 way floating table.
  • An Image Acquisition workstation with DICOM 3 compliance.

The SomaRad consists of:

  • A high resolution CCD detector with 14" x 17" or 17" x 17" image area.
  • An X-Ray Tube & Collimator.
  • A fully motorized positioned.
  • A 50 kW high frequency generator.
  • A table.
  • An Image Acquisition workstation with DICOM 3 compliance.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray

Anatomical Site

human anatomy

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

The results of bench, user, and standards testing indicates that the new device is as safe and effective as the predicate devices. The changed components are in fact identical to the predicate devices.

Key Metrics

Not Found

Predicate Device(s)

K073114, K083645, K070618

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

0

Image /page/0/Picture/0 description: The image shows the text "510(k) Summary K10j" at the top. Below this text is a graphic of a wave pattern with two black dots above it. Underneath the wave pattern is the text "IMIX Immediate Method of Imaging X-rays".

K101435

SEP 1 5 2010

IMIX ADR Finland OY Peltokatu 16 D3 33100 Tampere Finland Telephone: +358 (3) 2129 850 Fax: +358 (3) 2129 852 E-mail: emea@imixadr.com

Contact: Sigrid Smitt-Jeppesen President and CEO IMIX Americas Date Prepared: May 13, 2010

    1. Identification of the Device: Proprietary-Trade Name: IMIX PanoRad and SomaRad X-Ray Systems Classification Name: Stationary X-ray system, Product Codes Product Codes KPR, MOB, and LLZ Common/Usual Name: General purpose diagnostic X-ray Unit.
    1. Equivalent legally marketed devices: K073114, IMIX Insight; K083645 Radstar Digital Imaging System made by 5-Star, and K070618, DICOM Pacs made by O&R. The IMIX PanoRad and SomaRad X-Ray Systems use the identical panel to K083645 and the identical software to K070618.
    1. Indications for Use (intended use) IMIX PanoRad and SomaRad X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. Not for mammography.

4. Description of the Device:

  • The PanoRad consists of:
    • A high resolution CCD detector with 14" x 17" or 17" x 17" image area. t
    • An X-Ray Tube & Collimator. .
    • A fully motorized positioned .
    • A 65 kW high frequency generator .
    • A 4 way floating table .
    • An Image Acquisition workstation with DICOM 3 compliance.

The SomaRad consists of:

  • A high resolution CCD detector with 14" x 17" or 17" x 17" image area. .
  • An X-Ray Tube & Collimator. .
  • A fully motorized positioned .
  • A 50 kW high frequency generator .
  • A table .
  • An Image Acquisition workstation with DICOM 3 compliance.

1

    1. Safety and Effectiveness, comparison to predicate device. The results of bench, user, and standards testing indicates that the new device is as safe and effective as the predicate devices. The changed components are in fact identical to the predicate devices.

| Characteristic | K073114, IMIX Insight | K083645 Radstar
Digital Imaging System
made by 5-Star | IMIX PanoRad and
SomaRad X-Ray
Systems, Combines
cleared devices |
|---------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Intended Use: | General purpose diagnostic
X-ray unit | Acquisition of x-ray images
when incorporated into an
x-ray system, a film
substitute. | SAME |
| User Interface | Software Driven Touch
Panel LCD | Windows computer | Software Driven Touch
Panel LCD |
| Generator | Stadler | Not supplied | CPI or Stadler |
| Maximum output | 40, 50, and 65 kW | Not supplied | 50 and 65 kW |
| Stand | Supplied by Sedecal | Not supplied | Supplied by Shinyoung For
M Co Ltd |
| Image Acquisition | Digital: IMIX Digital
Radiographic Detector
K974863 | Varian | IMIX or Varian Digital
Radiographic Detectors, 9
or 16 mp. |
| Digital Panel Size | Active image size: 16 x 16
inches (40cm x 40cm) | 17" x 17", 14" x 17", 12" x
16" and 8" x 10" panel | 14" x 17" (4336R) OR
17" x 17" (4343R) |
| Digital Panel
Supplier | IMIX | Varian 4336 and others.
(not specified) | Original IMIX panels, same
as in K073114 OR:
Varian 4343R OR 4336R |
| Digital Resolution | 160 Micron.
3056 x 3056 (9 megapixels)
or 120 Micron
4096 x 4096 (16
megapixels) | 7.9 megapixel or 9.4
megapixel. 139 micron. | Same as original 510(k), 16
megapixels OR
7.9 megapixel or 9.4
megapixel. 139 micron. |
| DICOM | Yes | Optional | Yes, via O&R software
cleared in K091364 |
| Method of Control | Touch Panel LCD | Windows computer | Touch Panel LCD |
| Collimator | Ralco R302L/A DHHS | Not supplied | Ralco R302L/A DHHS |
| Safety | UL listed | UL listed | UL listed |

    1. Substantial Equivalence Chart, IMIX PanoRad and SomaRad X-Ray Systems
  1. Conclusion: After analyzing bench, user (clinical image pairs), and standards testing data, it is the conclusion of IMIX ADR that the IMIX PanoRad and SomaRad X-Ray Systems are as safe and effective as the predicate devices, have few technological differences, and have no new indications for use, thus rendering them substantially equivalent to the predicate devices.

2

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three wavy lines representing the branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

IMIX ADR Finland OY % Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

SEP 1 5 2010

Re: K101435

Trade/Device Name: IMIX PanoRad and SomaRad X-Ray Systems Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: July 32, 2010 Received: July 29, 2010

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

3

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket. notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Donald J. Trump

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

4

510(k) Number (if known):K101435
------------------------------------

KIDI435 SEP 1 5 2010

Device Name: IMIX PanoRad and SomaRad X-Ray Systems

Indications For Use:

IMIX PanoRad and SomaRad X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. (Not for mammography.)

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page 1 of 1

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K101435