(35 days)
The ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated. .
- Failure of anticoagulant therapy for thromboembolic disease. .
- Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced.
- Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed . or is contraindicated.
ECLIPSE™ Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
The ECLIPSE™ Filter consists of twelve electropolished shape-memory nitinol wires emanating from a central electropolished nitinol filter hook. These 12 wires form two levels of embolic filtration: the legs provide the lower level of filtration and the arms provide the upper level of filtration. The ECLIPSE™ Filter is intended to be used in the inferior vena cava with diameters less than or equal to 28 mm.
The ECLIPSE™ Filter System - Femoral Delivery Kit consists of a 7 French inner diameter (I.D.) introducer catheter and dilator set and a storage tube preloaded with the ECLIPSE™ Filter and pusher system. The dilator is fitted with 2 radiopaque marker bands spaced 28 mm apart for caval sizing. The introducer sheath contains a radiopaque tip and hemostasis valve with a side port for injecting contrast medium via a syringe. The storage tube and pusher system attach to the introducer and allow for delivery and deployment of the ECLIPSE™ Filter.
The ECLIPSE™ Filter System – Jugular/Subclavian Delivery Kit consists of a 10 French I.D. introducer catheter and dilator set and a delivery device preloaded with the ECLIPSE™ Filter. The dilator is fitted with 2 radiopaque marker bands spaced 28 mm apart for caval sizing. The introducer sheath contains a radiopaque tip and hemostasis valve with a side port for injecting contrast medium via a syringe. The delivery device fits within the introducer sheath and has a delivery mechanism to deploy the ECLIPSE™ Filter.
The provided text describes the ECLIPSE™ Filter System, an intravascular filter, and its FDA 510(k) submission (K101431). However, it does not contain information related to acceptance criteria or a study proving device performance using clinical data, AI, or human-in-the-loop performance.
The submission focuses on demonstrating substantial equivalence to a predicate device (K093659) based on modifications primarily related to labeling and minor physical changes (addition of a kangaroo pouch). The performance testing described is limited to in vitro testing for packaging, sterilization, and latex, which are not directly about the clinical efficacy or diagnostic performance of the filter itself.
Therefore, most of the requested information cannot be extracted from this document.
Here's what can be gathered:
1. A table of acceptance criteria and the reported device performance
- NotFound: The document does not provide specific acceptance criteria or clinical performance metrics related to the filter's efficacy (e.g., emboli capture rate, complication rates). It only mentions that in vitro tests (packaging, sterilization, latex) demonstrated comparability to the predicate device.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- NotFound: No clinical test set or data provenance is mentioned. The testing was in vitro.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- NotFound: As no clinical test set is described, there's no mention of experts establishing ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- NotFound: Not applicable as there's no clinical test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No: No MRMC study was done, and the device is a physical filter, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No: This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable: For the in vitro tests mentioned (packaging, sterilization, latex), "ground truth" would relate to meeting pre-defined physical or chemical specifications, not clinical outcomes or expert consensus on a diagnosis.
8. The sample size for the training set
- NotFound: No training set is mentioned as this is a physical device, not an AI model.
9. How the ground truth for the training set was established
- Not Applicable: No training set is mentioned.
{0}------------------------------------------------
Response to FDA Questions (K101431) ECLIPSE™ Filter System
Page 11
ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery Kits 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
Submitter Information:
- Bard Peripheral Vascular, Inc Applicant: 1625 West 3rd Street Tempe, Arizona 85281
480-638-2906 Phone:
480-449-2546 Fax:
Contact: Joni Creal, Regulatory Affairs Associate
Date: June 21, 2010
Subject Device Name:
| Device Trade Name: | ECLIPSE™ Filter System - FemoralDelivery Kit (EC500F) and ECLIPSE™ FilterSystem – Jugular/Subclavian Delivery Kit(EC500J) |
|---|---|
| Common or Usual Name: | Filter, Intravascular, Cardiovascular |
| Classification: | Class II |
| Classification Panel: | Cardiovascular Devices |
| Product Code: | DTK |
| Predicate Devices: | ECLIPSE™ Filter System - Femoral andJugular/Subclavian Delivery Kits (K003659) |
Summary of Change:
The primary modification from the predicate device, the ECLIPSE™ Filter System -Femoral and Jugular/Subclavian Delivery System (K093659), to the subject device, the ECLIPSE™ Femoral Filter System and Jugular/Subclavian Delivery System, was an addition of a Patient Brochure and Implant Card to the device labeling. In addition, a
Bard Peripheral Vascular, Inc.
{1}------------------------------------------------
kangaroo pouch was added to accommodate the Patient Brochure and Implant Card, and minor labeling modifications were made.
Device Description:
The ECLIPSE™ Filter consists of twelve electropolished shape-memory nitinol wires emanating from a central electropolished nitinol filter hook. These 12 wires form two levels of embolic filtration: the legs provide the lower level of filtration and the arms provide the upper level of filtration. The ECLIPSE™ Filter is intended to be used in the inferior vena cava with diameters less than or equal to 28 mm.
The ECLIPSE™ Filter System - Femoral Delivery Kit consists of a 7 French inner diameter (I.D.) introducer catheter and dilator set and a storage tube preloaded with the ECLIPSE™ Filter and pusher system. The dilator is fitted with 2 radiopaque marker bands spaced 28 mm apart for caval sizing. The introducer sheath contains a radiopaque tip and hemostasis valve with a side port for injecting contrast medium via a syringe. The storage tube and pusher system attach to the introducer and allow for delivery and deployment of the ECLIPSE™ Filter.
The ECLIPSE™ Filter System – Jugular/Subclavian Delivery Kit consists of a 10 French I.D. introducer catheter and dilator set and a delivery device preloaded with the ECLIPSE™ Filter. The dilator is fitted with 2 radiopaque marker bands spaced 28 mm apart for caval sizing. The introducer sheath contains a radiopaque tip and hemostasis valve with a side port for injecting contrast medium via a syringe. The delivery device fits within the introducer sheath and has a delivery mechanism to deploy the ECLIPSE™ Filter.
Indications for Use of Device:
The subject device, the ECLIPSE™ Filter Systems - Femoral and Jugular/Subclavian Delivery Kits, are indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated. .
- Failure of anticoagulant therapy for thromboembolic disease. .
- Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced.
{2}------------------------------------------------
Response to FDA Questions (K101431) ECLIPSE™ Filter System
- Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed . or is contraindicated.
ECLIPSE™ Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Technological Comparison to Predicate Devices:
The technological characteristics of the subject device, the ECLIPSE™ Filter System -Femoral and Jugular/Subclavian Delivery Kits, are substantially equivalent to those of the predicate device, the ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery System (K093659), in terms of intended use, indications for use, application, user population, same operating principle, filter design, delivery system design, fundamental scientific technology, performance, and sterilization method.
Performance Testing Summary:
To demonstrate substantial equivalence of the subject device, the ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery Kits to the predicate device, the technological characteristics and performance criterion were evaluated using in vitro testing performed as outlined below:
- Packaging Testing .
- Sterilization Testing .
- Latex Testing .
The results from these tests demonstrate that the technological characteristics and performance criteria of the ECLIPSE™ Filter System – Femoral and Jugular/Subclavian Delivery Kits are comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
Conclusions:
The ECLIPSE™ Filter Systems - Femoral and Jugular/Subclavian Delivery Kits are y substantially equivalent to the legally marketed predicate device, the ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery System (K093659).
Bard Peripheral Vascular, Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird in flight. The profiles are stacked on top of each other, creating a sense of depth and movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
JUN 2 5 2010
Bard Peripheral Vascular, Inc. c/o Joni Creal Regulatory Affairs Associate 1625 West 3rd Street Tempe, AZ 85281
Re: K101431
Trade Name: ECLISPE Filter System - Femoral Delivery Kit and ECLISPE Filter System - Jugular/Subclavian Delivery Kit Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular intravascular filter Regulatory Class: Class II Product Code: DTK Dated: June 21, 2010 Received: June 22, 2010
Dear Ms. Creal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Creal
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Duna R. Vahner
ABram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K101431
Device Name: ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery Kits
Indications for Use:
The ECLIPSE™ Filter System - Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations:
- . Pulmonary thromboembolism when anticoagulants are contraindicated.
- . Failure of anticoagulant therapy for thromboembolic disease.
- Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced.
- Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed . or is contraindicated.
ECLIPSE™ Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Prescription Use_X (Part21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K101431 |
|---|---|
| --------------- | --------- |
Bard Peripheral Vascular, Inc.
§ 870.3375 Cardiovascular intravascular filter.
(a)
Identification. A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.(b)
Classification. Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”