(30 days)
The Coaxial Introducer Set is indicated for percutaneous introduction of up to a 0.038 inch guidewire or catheter into the vascular system following a small 21 gauge needle stick. The Coaxial Introducer Set is not intended for use in the coronary or cerebral vasculature.
The Coaxial Introducer Set is designed to provide dilation of the initial entry point and provide a channel for attaining vascular access. The coaxial introducer consists of an inner dilator within a slightly shorter outer sheath. The kit contains a 21 gauge introducer needle, a .018" diameter guidewire, and a 4F or 5F coaxial introducer. The coaxial introducer is offered in a standard and stiff version. The Coaxial Introducer Set is provided in sterile and non-sterile configurations.
The provided document is a 510(k) summary for a medical device (Coaxial Introducer Set) and does not describe acceptance criteria or a study proving that the device meets those criteria in the context of an AI/ML medical device. Instead, it details in-vitro bench testing conducted to assure reliable design and performance in accordance with ISO standards and/or internal procedures for a physical medical device. It also includes information about biocompatibility testing.
Therefore, many of the requested categories for AI/ML device studies cannot be populated from the given text.
Here's a breakdown of what can be extracted and what cannot:
1. Table of Acceptance Criteria and Reported Device Performance:
The document mentions in-vitro testing but does not provide specific acceptance criteria or quantitative performance metrics for each test. It only lists the types of tests performed.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Bench/Performance Data: | |
| Guidewire Passage | Assured reliable design and performance (qualitative statement) |
| Air Leakage | Assured reliable design and performance (qualitative statement) |
| Liquid Leakage | Assured reliable design and performance (qualitative statement) |
| Force at Break | Assured reliable design and performance (qualitative statement) |
| Simulated Use | Assured reliable design and performance (qualitative statement) |
| Biocompatibility: | |
| ISO 10993 requirements | Materials meet the requirements (qualitative statement) |
2. Sample size used for the test set and the data provenance: Not applicable. This refers to physical device testing, not data for an AI/ML model.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This pertains to AI/ML model evaluation, not physical device testing.
4. Adjudication method for the test set: Not applicable. This pertains to AI/ML model evaluation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a physical medical device, not an AI-powered system.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This document describes a physical medical device, not an AI algorithm.
7. The type of ground truth used: For the bench testing, the "ground truth" would be objective measurements and observations against established engineering specifications and ISO standards. For biocompatibility, it's compliance with ISO 10993 standards.
8. The sample size for the training set: Not applicable. This pertains to AI/ML model training.
9. How the ground truth for the training set was established: Not applicable. This pertains to AI/ML model training.
Summary based on the provided text:
The Coaxial Introducer Set underwent in-vitro bench testing for:
- Guidewire Passage
- Air Leakage
- Liquid Leakage
- Force at Break
- Simulated Use
These tests were performed "to assure reliable design and performance in accordance with ISO standards and/or internal procedures." The specific acceptance criteria and the quantitative results (e.g., "passed 100/100 guidewire passages," or "leakage rate was X") are not provided in this summary.
Additionally, biocompatibility testing was performed, and the results demonstrated that the materials used meet the requirements of ISO 10993.
The conclusion states that the proposed devices meet the performance criteria of design verification as specified by ISO standards and test protocols, and are substantially equivalent to the legally marketed predicate devices.
{0}------------------------------------------------
K1.01399
JUN 1 7 2010
510(k) SUMMARY
Submitter Information: A.
Submitter:
Contact:
Date Prepared:
MARTECH MEDICAL PRODUCTS 1500 Delp Drive Harleysville, PA 19438 (215) 256-8833 Telephone (215) 256-9191 Fax Susan Pileggi May 17, 2010
Coaxial Introducer Set B. Device Name:
Common Name: Classification Name: C.F.R. Section: Class:
Introducer Set Catheter Introducer (74 DYB) 880.1340
K091954: Medcomp, Micro-Stick Set Predicate Devices:
II
D. Device Description:
C.
The Coaxial Introducer Set is designed to provide dilation of the initial entry point and provide a channel for attaining vascular access. The coaxial introducer consists of an inner dilator within a slightly shorter outer sheath.
The kit contains a 21 gauge introducer needle, a .018" diameter guidewire, and a 4F or 5F coaxial introducer. The coaxial introducer is offered in a standard and stiff version.
The Coaxial Introducer Set is provided in sterile and non-sterile configurations.
E. Intended Use:
The Coaxial Introducer Set is indicated for percutaneous introduction of up to a 0.038 inch guidewire or catheter into the vascular system following a small 21 gauge needle stick. The Coaxial Introducer Set is not intended for use in the coronary or cerebral vasculature.
F. Comparison to Predicate Devices:
The Coaxial Introducer Set is substantially equivalent to the predicate devices in terms of intended use, anatomical location, general design, and method of sterilization.
{1}------------------------------------------------
G. Bench / Performance Data:
The following in-vitro testing was performed on the Coaxial Introducer Set to assure reliable design and performance in accordance with ISO standards and/or internal procedures.
- Guidewire Passage ●
- . Air Leakage
- � Liquid Leakage
- Force at Break .
- Simulated Use .
H. Biocompatibility:
Results for all biocompatibility testing demonstrate the materials used meet the requirements of ISO 10993.
I. Conclusion:
The proposed devices meet the performance criteria of design verification as specified by ISO standards and test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed devices are substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three thick, curved lines.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 7 2010
Martech Medical c/o Ms. Susan Pileggi Regulatory Specialist 1500 Delp Drive Harleysville, PA 19438
Re: K101399
Trade/Device Name: Coaxial Introducer Set Regulation Number: 21 CFR 870.1340 Regulation Name: Introducer Catheter Regulatory Class: Class II (two) Product Code: DYB Dated: May 17, 2010 Received: May 26, 2010
Dear Ms. Pileggi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Susan Pileggi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutiFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Duna R. Vichner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): Kjol 399
Device Name: Coaxial Introducer Set
Indications for Use:
The Coaxial Introducer Set is indicated for percutaneous introduction of up to a 0.038 inch guidewire or catheter into the vascular system following a small 21 gauge needle stick. The Coaxial Introducer Set is not intended for use in the coronary or cerebral vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Viclun
(Division Sign-Off) (Division Sign-On)
Division of Cardiovascular Devices
510(k) Number_K10739
§ 870.1340 Catheter introducer.
(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).