(64 days)
Not Found
No
The document does not mention AI, ML, or related terms, and the device description and performance studies do not suggest the use of such technologies.
No
Explanation: The device is explicitly stated to be a "diagnostic tool" and is used for mapping and displaying the position of catheters, not for treating any condition.
Yes
The "Intended Use / Indications for Use" section explicitly states that the EnSite Velocity™ Cardiac Mapping System "is a suggested diagnostic tool".
No
The device description explicitly lists hardware components (Display Workstation, Amplifier) as part of the system, indicating it is not software-only.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The EnSite Velocity Cardiac Mapping System is described as a "suggested diagnostic tool" used in patients during electrophysiology studies. It works by displaying the position of catheters within the heart and potentially helping to identify complex arrhythmias. This is an in vivo (within the living body) diagnostic process, not an in vitro (outside the living body) one.
- Lack of IVD Indicators: The description does not mention analyzing biological samples, using reagents, or performing tests on specimens outside the body, which are hallmarks of IVD devices.
Therefore, the EnSite Velocity Cardiac Mapping System is a medical device used for in vivo diagnosis and mapping within the heart, not an IVD.
N/A
Intended Use / Indications for Use
The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.
- When used with the EnSite Array™ Catheter, the EnSite System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
OR - When used with the EnSite NavX Surface Electrode Kit, the EnSite System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.
Product codes
DQK
Device Description
The EnSite Velocity System consists of the following:
- Display Workstation .
- Amplifier
Consisting of:
EnSite Multi-electrode Diagnostic Catheter (EnSite Array™ - Model EC1000) EnSite Electrophysiology Workstation - Model EE3000 EnSite NavX Surface Electrode Kit - Model EN0010
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
right atrium, heart
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1425 Programmable diagnostic computer.
(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).
0
12 510(k) Summary for Public Disclosure
JUL 14 2010
12.1 Submitter's Name/Contact Person
Donna R. Lunak St. Jude Medical One St. Jude Medical Drive St. Paul, MN 55117 USA
The Establishment Registration Number is 2184149.
12.2 Predicate Device
EnSite Velocity System v.1.3 (K093942)
12.3 Common or Usual Name
Electrophysiology Mapping System with console and catheter
12.4 Proprietary Name
EnSite Velocity System
Consisting of:
EnSite Multi-electrode Diagnostic Catheter (EnSite Array™ - Model EC1000) EnSite Electrophysiology Workstation - Model EE3000 EnSite NavX Surface Electrode Kit - Model EN0010
12.5 Classification Name
DQK, Programmable diagnostic computer (21 CFR 870.1425), Class II, Circulatory Systems Device Panel
12.6 Hardware Description
The EnSite Velocity System consists of the following:
- Display Workstation .
- Amplifier
1
12.7 Indications for Use
The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.
- When used with the EnSite Array™ Catheter, the EnSite System is ● intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
- OR
- When used with the EnSite NavX Surface Electrode Kit, the EnSite . System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.
12.8 Device Comparison to the Predicate Device
The EnSite Velocity System v.2.0 has the same intended use and fundamental scientific technology as the predicate device. All technological characteristics of the EnSite Velocity System v.2.0 are substantially equivalent to the predicate device.
Summary of Non-Clinical Testing 12.9
Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product.
12.10 Summary of Design Control Activities
The development of the EnSite Velocity System v.2.0 was performed in accordance with St. Jude Medical's Quality System requirements, and in compliance with Quality System Regulation design controls requirements documented in 21 CFR 820.30. A Declaration of Conformity with Design Controls follows in section 13.4.
12.11 Conclusion
The EnSite Velocity v.2.0 has the same indications for use, intended use and fundamental scientific technology as the predicate device. All technological characteristics of the EnSite Velocity v.2.0 are substantially equivalent to the predicate device.
2
Where operational and performance differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness.
Therefore, St. Jude Medical considers the EnSite Velocity v.2.0 to be substantially equivalent to the predicate device.
3
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circumference of the circle. The text is in all capital letters and is arranged to follow the curve of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 1 4 2010
St. Jude Medical c/o Donna R. Lunak Regulatory Specialist One St. Jude Medical Drive Saint Paul MN 55117
Re: K101328
Trade/Device Name: EnSite Velocity System V.2,0 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: June 23, 2010 Received: June 24, 2010
Dear Ms. Lunak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
4
Page 2 - Ms. Donna Lunak
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical or ice related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
7 Indications for Use
Device Name: EnSite Velocity System
Indications for Use:
The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.
-
When used with the EnSite Array™ Catheter, the EnSite System is intended to . be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
OR -
When used with the EnSite NavX Surface Electrode Kit, the EnSite System is t intended to display the position of conventional electrophysiology (EP) catheters in the heart.
AND/OR Over-The-Counter Use
(21 CER 801 Subnart C Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W. Wood.
510(k) Number k01328