(190 days)
The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation.
This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.
The IOLMaster 500 is a non-contact biometry instrument for measurements of the eye required for preoperative computation of intraocular lens (IOL) type and power. As with the IOLMaster predicate device, the IOLMaster 500 provides measurements of axial length, corneal radius (keratometry), and anterior chamber depth. In the new model of the IOLMaster, the optical measurement technology has remained the same and the measurements of axial length, corneal radius and anterior chamber depth are achieved in the same fashion. In addition to these three parameters, using the same optical techniques with advances in software, the modified IOLMaster also has the capability to measure a fourth parameter, i.e., the "white-to-white" distance (WTW). Incorporation of this additional information extends the physician's choice of computational formulas.
Here's a breakdown of the acceptance criteria and the studies conducted for the IOLMaster 500, based on the provided 510(k) summary:
1. Table of Acceptance Criteria (Implied) and Reported Device Performance
The submission does not explicitly state acceptance criteria in terms of specific thresholds for agreement or variability. Instead, it presents the results of studies and implies that these results demonstrate substantial equivalence and acceptable performance. Therefore, the "acceptance criteria" listed here are inferred from the demonstrated performance that led to 510(k) clearance.
| Measurement Parameter | Implied Acceptance Criterion (from demonstrated performance) | Reported Device Performance (Mean Difference / Standard Deviation / Repeatability / Reproducibility) |
|---|---|---|
| Agreement with Other Instruments | ||
| Axial Length (vs. GBS) | Agreement with predicate/reference device (Grieshaber Biometric System) indicating minimal mean difference and low standard deviation. | -0.01 mm ±0.19 mm |
| Corneal Curvature (vs. Manual Keratometer) | Agreement with predicate/reference device (manual keratometer) indicating minimal mean difference and low standard deviation. | -0.01 mm ±0.05 mm |
| Anterior Chamber Depth (vs. GBS) | Agreement with predicate/reference device (Grieshaber Biometric System) indicating minimal mean difference and low standard deviation. | +0.03 mm ±0.18 mm |
| White-to-White (vs. Lenstar LS900) | Agreement with predicate/reference device (Lenstar LS900) indicating minimal mean difference and low standard deviation. | +0.06 mm ±0.33 mm |
| Repeatability and Reproducibility | ||
| Axial Length | High repeatability (low SD, low limit) within a single instrument/operator and high reproducibility (low SD, low limit) across instruments/operators. | Repeatability SD: 0.0206 mm / Limit: 0.0577 mmReproducibility SD: 0.0222 mm / Limit: 0.0623 mm |
| Corneal Curvature | High repeatability (low SD, low limit) within a single instrument/operator and high reproducibility (low SD, low limit) across instruments/operators. | Repeatability SD: 0.0162 mm / Limit: 0.0455 mmReproducibility SD: 0.0167 mm / Limit: 0.0468 mm |
| Anterior Chamber Depth | High repeatability (low SD, low limit) within a single instrument/operator and high reproducibility (low SD, low limit) across instruments/operators. | Repeatability SD: 0.0347 mm / Limit: 0.0972 mmReproducibility SD: 0.0494 mm / Limit: 0.1383 mm |
| White-to-White | High repeatability (low SD, low limit) within a single instrument/operator and high reproducibility (low SD, low limit) across instruments/operators. | Repeatability SD: 0.0558 mm / Limit: 0.1562 mmReproducibility SD: 0.0647 mm / Limit: 0.1811 mm |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Agreement Studies (Test Set):
- Axial Length: 146 phakic eyes (from an initial 189 consecutive eyes).
- Corneal Curvature: 154 eyes.
- Anterior Chamber Depth: 151 eyes.
- White-to-White: 112 cataract surgery eyes.
- Data Provenance: The studies referenced are:
- Haigis W, Lege B (2000) - This implies a German origin given the journal title.
- Buckhurst PJ et al. (2009) - This is a British Journal of Ophthalmology paper, suggesting UK or international data.
- All referenced studies compared measurements from eyes evaluated for cataract surgery. The studies appear to be retrospective literature reviews of previously published comparisons, rather than a new prospective study specifically for this 510(k).
-
Repeatability and Reproducibility Study (Test Set):
- Sample Size: 30 subjects.
- Data Provenance: "Carl Zeiss Meditec, Inc., Clinical Study: Repeatability of IOLMaster 500; 2010." This suggests a prospective study conducted specifically for this device by the manufacturer, likely in the USA where Carl Zeiss Meditec, Inc. is located. Subjects were 18 years or older with healthy eyes and no ocular opacities.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
No explicit ground truth experts or their qualifications are mentioned for the agreement studies, as these studies compared the IOLMaster 500 to established reference devices (Grieshaber Biometric System for AL/ACD, manual keratometer for corneal curvature, Lenstar LS900 for WTW). These reference devices are themselves considered "ground truth" or highly accurate measurement methods in their respective fields.
For the repeatability and reproducibility study, there is no "ground truth" in the sense of expert annotation, but rather statistical analysis of repeated measurements. The study involved "one operator" for inter-device variability and "each of three operators" for inter-operator variability, but their specific qualifications are not detailed beyond "operator."
4. Adjudication Method for the Test Set
No adjudication method is described for the test sets. The agreement studies directly compared measurements between devices, and the repeatability/reproducibility study analyzed variations in measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No MRMC comparative effectiveness study was done. This device is a measurement instrument, not an AI-assisted diagnostic tool that requires human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the device operates in a standalone capacity for measurement. The performance tables (Table 1 and Table 2) reflect the inherent accuracy and precision of the device's measurements, independent of human interpretation or intervention beyond performing the measurement itself.
7. The Type of Ground Truth Used
- Agreement Studies: The ground truth was effectively established by comparison with established, high-precision, and clinically accepted reference devices for ocular measurements:
- High precision immersion ultrasound system (Grieshaber Biometric System) for axial length and anterior chamber depth.
- Handheld keratometer (Renaissance, Alcon) for corneal curvature.
- Optical low coherence reflectometry (Lenstar LS900, Haig-Streit) for white-to-white measurements.
- Repeatability and Reproducibility Study: The "ground truth" here is the true underlying measurement of the subject, with the study assessing the variability around that true value by repeated measurements from the device itself.
8. The Sample Size for the Training Set
No explicit training set is mentioned in the provided document. The IOLMaster 500's measurement technology (partial coherence interferometry and traditional ophthalmic biometry) remained the same as the predicate device, with software advances primarily improving the Graphical User Interface (GUI) and adding the white-to-white measurement. It's possible the core algorithms for AL, corneal radius, and ACD were "trained" or validated during the development of the original IOLMaster (K993357), but this document doesn't detail it. The IOLMaster 500's white-to-white measurement uses "the same optical techniques with advances in software," implying it leverages existing optical principles rather than requiring a large separate training set for a new AI model.
9. How the Ground Truth for the Training Set Was Established
As no specific training set for a machine learning model is described, there's no information on how its ground truth was established. The device relies on established biophysical principles and optical measurements.
{0}------------------------------------------------
4.0 510(K) SUMMARY
NOV - 3 2010 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
- Submitter's Name, address, telephone number, contact person, and date summary 4.1 PREPARED
| a. | Applicant: | Carl Zeiss Meditec AGGoeschwitzer Strasse 51-5207745 JenaGermany |
|---|---|---|
| b. | Contact Person: | Judith A. Brimacombe, MADirector, Clinical/Regulatory AffairsCarl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, CA 94568USAj.brimacombe@meditec.zeiss.comTel: (925) 557-4616Fax: (925) 557-4259 |
- October 7, 2010 Summary Prepared: C.
Name of device, including trade name and classification name 4.2
| a. | Trade/Proprietary Name: | IOLMaster® 500 |
|---|---|---|
| b. | Common/Usual Name: | IOLMaster® |
| c. | Classification Name: | AC-powered slit lamp biomicroscope |
| d. | Product Code and Class: | HJO - Class II |
| e. | Classification Number: | 886.1850 |
{1}------------------------------------------------
PREDICATE DEVICE 4.3
The IOLMaster 500, the subject of this 510(k), is substantially equivalent to the predicate IOLMaster (Carl Zeiss Meditec, AG) cleared for marketing under 510(k) number K993357 and the Lenstar Model LS 900 (Haag-Streit AG) cleared for marketing under 510(k) number K082891.
DEVICE DESCRIPTION 4.4
The IOLMaster 500 is a non-contact biometry instrument for measurements of the eye required for preoperative computation of intraocular lens (IOL) type and power. As with the IOLMaster predicate device, the IOLMaster 500 provides measurements of axial length, corneal radius (keratometry), and anterior chamber depth. In the new model of the IOLMaster, the optical measurement technology has remained the same and the measurements of axial length, corneal radius and anterior chamber depth are achieved in the same fashion. In addition to these three parameters, using the same optical techniques with advances in software, the modified IOLMaster also has the capability to measure a fourth parameter, i.e., the "white-to-white" distance (WTW). Incorporation of this additional information extends the physician's choice of computational formulas.
4.5 STATEMENT OF INTENDED USE
The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation.
This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.
Comparison of Technological Characteristics 4.6
The IOLMaster 500 has the same indications for use and operating characteristics as the predicate IOLMaster with the exception that the white-to-white measurement has been added to the indications for use statement.
The IOLMaster utilizes partial coherence interferometry and traditional ophthalmic biometry techniques to obtain measurements for axial length, corneal radius, anterior chamber depth and white-to white distance. Integrated algorithms in the software allow for the use of clinically recognized formulas for the calculation of IOL type and power prior to
{2}------------------------------------------------
cataract surgery. These functions are the same for the IOLMaster 500 and the predicate device.
Software enhancements using the Windows XP platform have improved the Graphic User Interface (GUI) and the usability of the IOLMaster. Additionally, modifications have been made to the IOLMaster computer hardware to incorporate USB ports for importing and exporting data to and from the IOLMaster 500.
Since the measurements are achieved in the same fashion using the identical optical technology, the IOLMaster 500 is therefore substantially equivalent to the predicate IOLMaster (K993357).
4.7 CLINICAL EVALUATION
Clinical data reported in the literature was reviewed to determine agreement between the IOL Master, Grieshaber Biometric System, and Lenstar Model LS900. The results are provided in Table 1 on the following page.
Axial length, anterior chamber depth and corneal radius
IOLMaster axial length (AL) and anterior chamber depth (ACD) measurements were compared to measurements obtained with a high precision immersion ultrasound system (Grieshaber Biometric System (GBS)). A total of 189 consecutive eyes presented for cataract evaluation were measured with the IOLMaster. Excluded from the study were 32 eyes not measured with GBS. Mean difference for axial length was calculated based on 146 phakic eyes comparative axial length measurements between IOLMaster and GBS (10 pseudophakic eyes and 1 aphakic eye were excluded). Mean difference for ACD was calculated based on measurements on 151 eyes comparing IOLMaster and GBS (in 6 pseudophakic eves. ACD could not be measured). Mean difference for the average corneal radius was calculated based on keratometry measurements on 154 eyes comparing IOLMaster and a handheld keratometer ((Renaissance, Alcon). Two eyes with no available keratometry results and one eye with unreliable readings due to dry eye were excluded).2
White-to-white measurement
IOLMaster white-to-white measurements were compared to measurements obtained with a Lenstar LS 900 (Haag-Streit) in 112 cataract surgery eyes.2
1 Haigis W, Lege B: Akustische und optische Biometrie im klinischen Einsatz; in Wenzel M, Kohnen B. (eds): 14. Kongress der Deutschsprachigen Gesellschaft für Intraokulariinsen-Implantation und refractive Chirurgie, Jahresband der DGII, Luzern, February 2000. Koln, Biermann, 2000, pp 73-78. [ Haigis W, Lege B: Acoustical and Optical Biometry in Clinical Application; in Wenzel M, Kohnen T, Blumer B. (eds): 14. Congress of the german-speaking society for implantation of intraocular lenses and refractive surgery (DGII), Annual book of the DGII, Luzern, February 2000. Koln, Biermann, 2000, pp 73-78.] 4 Buckhurst PJ, Wolffson JS, Shah S, Naroo SA, Davies LN, Berrow EJ. A new optical low coherence reflectometry device for ocular biometry in cataract patients. Br J Ophthalmol 2009; 93 (7): 949-953.
{3}------------------------------------------------
Table 1. Agreement with Other Instruments
| N | Mean Difference | Standard Deviation | |
|---|---|---|---|
| Axial length* | 146 | -0.01 mm | ±0.19 mm |
| Corneal curvature** | 154 | -0.01 mm | ±0.05 mm |
| Anterior chamber depth* | 151 | + 0.03 mm | ±0.18 mm |
| White-to-white*** | 112 | + 0.06 mm | ±0.33 mm |
All referenced studies compared measurements from eyes evaluated for cataract surgery.
- In comparison to high precision immersion ultrasound instrument (Grieshaber Biometric Systems, GBS) *
- ** In comparison to manual keratometer (Renaissance, Alcon)*
- *** In comparison to optical low coherence reflectometry (Lenstar LS900, Haig-Streit)*
Repeatability and Reproducibility
A study was conducted to determine repeatability and reproducibility of the IOLMaster 500 measurements of axial length (AL), corneal radius (corneal curvature), anterior chamber depth (ACD) and white-to-white (WTW). Phase I of the study enrolled 30 subjects and was designed to determine inter-device variability, wherein each subject was imaged 3 times during a single visit on each of three IOLMaster 500 instruments by one operator. Phase II enrolled the same 30 subjects and was designed to determine inter-operator variability, wherein each subject was imaged three times during a single visit by each of three operators. Subjects in the study were 18 years or older with healthy eyes and no ocular opacities. The precision of axial length measurements and anterior chamber depth measurements may be different in eyes with cataract or corneal abnormalities.
The repeatability and reproducibility results are shown in Table 2 on the following page.
3 Carl Zeiss Meditec, Inc., Clinical Study: Repeatability of IOLMaster 500; 2010.
{4}------------------------------------------------
| Repeatability | Reproducibility | |||
|---|---|---|---|---|
| RepeatabilitySD(mm) | RepeatabilityLimit(mm) | ReproducibilitySD(mm) | ReproducibilityLimit(mm) | |
| Axial length | 0.0206 | 0.0577 | 0.0222 | 0.0623 |
| Corneal curvature | 0.0162 | 0.0455 | 0.0167 | 0.0468 |
| Anterior chamber depth | 0.0347 | 0.0972 | 0.0494 | 0.1383 |
| White-to-white | 0.0558 | 0.1562 | 0.0647 | 0.1811 |
Table 2. Repeatability and Reproducibility of the IOLMaster 500 Measurements
Repeatability is defined as the standard deviation of measurements taken by a single operator/single instrument. The repeatability standard deviation was estimated by the squared error. Reproducibility is the standard deviation of measurements that include between operator and between instrument variability. The reproducibility standard deviation was estimated by the square root of the sum of random measurement variability, inter-device variability, inter-operator variability, subject and device interaction variability, and subject and operator interaction variability.
The repeatability limit is the upper 95% limit for the differences between repeated results. Per ISO 5725-6, the repeatability limit = 2.8 x repeatability SD.
The reproducibility limit is the upper 95% limit calculated for the difference between individual measurements using different operators and different instruments. Per ISO 5725-6, the reproducibility limit = 2.8 x the reproducibility SD.
BRIEF SUMMARY OF PERFORMANCE TESTS AND RESULTS 4.8
The IOLMaster has been designed and tested to the applicable safety standards. The performance data supporting safety and substantial equivalence of the IOLMaster 500 to the predicate device demonstrates that:
- the axial length, corneal radius, anterior chamber depth measurements of the . modified IOLMaster are substantially equivalent to those of the predicate device;
- white-to-white measurements using the IOLMaster 500 are substantially equivalent . to those of the predicate device;
- the IOLMaster meets the requirements for electrical and laser safety; .
- usability of the IOLMaster has been verified in accordance with appropriate . standards.
4.9 CONCLUSION
The IOLMaster 500 is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, and three parallel lines extending from the eagle's body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Carl Zeiss Meditec Incorporated c/o Judith A. Brimacombe, M.A. Director, Clinical and Regulatory Affairs 5160 Hacienda Drive Dublin, CA 94568
Re: K101182
Trade Name: IOLMaster 500 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: September 13, 2010 Received: September 14, 2010
Dear Ms. Brimacombe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
NOV - 3 2010
{6}------------------------------------------------
Page 2 - Ms. Judith A. Brimacombe, M.A.
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Debra Falls
Malvina B. Eydelman, M. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K101182
Device Name(s): IOLMaster 500 NOV - 3 2010
Indications for Use:
The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation.
This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
DemR
Page 1 1 of of 1 1
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number K101182
§ 886.1850 AC-powered slitlamp biomicroscope.
(a)
Identification. An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.(b)
Classification. Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.