(224 days)
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
This premarket notification covers line extensions for the PONTiS™ product line. The ferroFibre™ stainless steel (SS) sutures were cleared for marketing in 2008, and are marketed in the PONTiS ™ product group. The PONTiS™ product family has been expanded to include stainless steel suture anchors geometrically similar to the reNOVO™ titanium product family (cleared in 2007), in combination with ferroFibre™ SS Sutures and optional suture crimps.
The PONTiS [1] implants incorporate non-absorbable ferroFibre [1] 316 L Stainless Steel Surgical sutures and are available in United States Pharmacopoeia (USP) sizes 4-0 and 3-0 in various lengths.
316L Stainless steel is used in a wide variety of medical devices including previously approved implants of this type. The ferroFibre™ Stainless Steel Sutures may be secured by either knot tying or crimping with a collar.
The ferroFibre™ multifilament sutures are supplied sterile, armed with cutting needles.
The implant components also include stainless steel collars specific to each suture size. The collars are used to secure multiple suture strands together simultaneously by crimping. For example, crimps are sized to accept between four (4) and eight (8) strands of each specific size of ferroFibre™ suture for crimping. A crimping instrument is supplied to crimp the collars to the suture strands. An adjustable tendon holder with suture tensioning slots is provided to facilitate suture cable tightening and tensioning. Excess suture is cut and removed.
The stainless steel PONTiS™ Suture Anchors are 3.0mm in diameter, are selftapping, and fully threaded. The anchors are each available with 4-0 and 3-0 multistrand ferroFibre™ stainless steel sutures. A disposable anchor driver holds the excess suture and delivers the anchor directly into the bone. The strands are then used to secure the soft tissue to bone.
Prior to installation, a hole is created in the bone using k-wire, drill or punch. The anchor is threaded into this hole. The anchor has self-tapping flutes.
Unlike the previously cleared titanium reNOVO™ Suture Anchor, PONTiS™ Suture Anchors are fully threaded allowing thread fixation in both cancellous and cortical bone.
The shorter overall length of the 3.0mm anchors, when compared to the self-drilling titanium reNOVO™ anchors, allows use in smaller/thinner bones where a lower profile is needed to achieve results without subsequent tissue injury.
The PONTiS™ threaded suture anchors are stainless steel to assure compatibility with the stainless steel in the ferroFibre™ stainless steel multi-strand sutures and the stainless steel crimp collars.
The size of the nonabsorbable stainless steel ferroFibre™ sutures used in the PONTiS™ Anchors is USP 4-0 and 3-0.
A single use driver and hand piece holds the excess suture and delivers the preloaded anchor into the bone. The suture strands are used to reapproximate and secure the soft tissue to bone.
The PONTiS™ Suture Anchors are provided sterile for single use applications.
The sizes and materials are designed to address the indications cited.
The provided text is a 510(k) summary for a medical device (PONTiS™ Sutures and Suture Anchors with Optional Crimps). This type of document is heavily focused on demonstrating substantial equivalence to previously approved devices rather than conducting novel clinical or performance studies with acceptance criteria in the way one might see for a new drug or a highly innovative device.
Therefore, the specific information requested in the prompt (acceptance criteria, details about a "study that proves the device meets the acceptance criteria," sample sizes, ground truth establishment, expert qualifications, MRMC studies, standalone performance, etc.) is not explicitly present in the level of detail typically associated with those terms in a rigorous scientific study. The document describes performance testing and biocompatibility reviews, but not in a format that lends itself to a direct mapping to the requested table and study details.
Here's an attempt to extract and interpret the information based on the provided text, acknowledging that many fields will be "Not Applicable" or "Not Specified" in the context of a 510(k) for substantial equivalence:
Acceptance Criteria and Device Performance (Based on Substantial Equivalence)
The PONTiS™ Sutures and Suture Anchors with Optional Crimps are not evaluated against specific quantitative acceptance criteria in this 510(k) summary in the same way a new, novel device might be. Instead, their performance is deemed acceptable if they are substantially equivalent in materials, operational principles, indications, and function to predicate devices already on the market.
The "studies" conducted primarily involve comparison to predicate devices and established standards, ensuring material safety and adequate mechanical properties for the stated indications.
Table 1: Acceptance Criteria and Reported Device Performance (Interpretation for Substantial Equivalence)
| Acceptance Criteria Category | Specific Criteria (Interpreted from Text) | Reported Device Performance (Summary) |
|---|---|---|
| Biocompatibility | Compliance with ISO 10993 standards for cytotoxicity, sensitization, and irritation. | All materials subjected to biocompatibility reviews or tests; deemed safe for intended use. |
| Material Properties | Compliance with USP 32-NF 26 Monographs <861>, <871>, <881> for suture diameter and tensile strength. | Compliant. |
| Suture Loop Pull-Strength (Crimped) | Sufficient strength to secure a tendon repair. Equivalent to knot-secured suture products currently marketed for hand tendon repair. Sufficient to support active motion without resistance. | Achieved sufficient strength and exhibited equivalent results to knotted sutures and literature benchmarks. |
| Suture Anchor Pull-Out Strength | Sufficient to support active motion without resistance following tendon reattachment in the hand. | Found to be sufficient to support active motion without resistance. |
| Suture Anchor Insertion Torque | Integrity of the anchor and anchor driver interface. | Quantified in hard bone (bovine cortical bone) to assure integrity. |
| Functional Equivalence (Anchors) | Equivalent intended use, materials, operational principles, and indications to predicate suture anchors (reNOVO™, Mitek Micro Anchor). | Deemed substantially equivalent to predicate devices. Performance tests showed equivalent results to currently marketed products for the same indications. |
| Functional Equivalence (Sutures+Crimps) | Equivalent intended use, materials, operational principles, and indications to predicate metallic sutures/cables secured with crimps (Core Essence PONTiS™ ferroFibre, Pioneer Surgical SDB Cerclage, Ortheon Medical Tenofix). | Deemed substantially equivalent to predicate devices. |
Study Details for Demonstrating Substantial Equivalence
1. Sample sized used for the test set and the data provenance:
- Suture Loop Pull-Strength Test: Not specified.
- Anchor Pull-Out Performance (foam bone, bovine cortical bone): Not specified.
- Biocompatibility Tests: Not specified for sample sizes of tested materials or specimens.
- Data Provenance: The mechanical testing appears to be lab-based developmental and comparative testing, not necessarily clinical patient data. The biocompatibility tests are lab-based.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This is not a study requiring human expert assessment for a "ground truth" in the diagnostic or clinical outcome sense.
- The document mentions "surgeons and engineers" functionally evaluating suture anchors in animal and human cadaver bone, but this is described as an "evaluation" rather than a formal "ground truth" establishment for a test set. No specific number or qualifications are given beyond "surgeons and engineers."
3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. This concept is not relevant to the type of testing described (mechanical and biocompatibility).
4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This type of study (MRMC, AI-assisted reading) is entirely irrelevant to this device (sutures and anchors) and the nature of the 510(k) submission.
5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This is not an algorithm-based device.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not Applicable / Industry Standards and Predicate Device Performance. The "ground truth" here is compliance with established material standards (e.g., ISO, USP) and demonstrating equivalent mechanical properties and biocompatibility to predicate devices that are already deemed safe and effective. For mechanical tests, the "ground truth" is derived from physical measurements (e.g., tensile strength, pull-out force).
7. The sample size for the training set:
- Not Applicable. This is not a machine learning or algorithm-based device that requires a training set.
8. How the ground truth for the training set was established:
- Not Applicable. See above.
{0}------------------------------------------------
Subject 510(k) Number: K101126
Sponsor
Core Essence Orthopaedics, Inc 575A Virginia Drive Fort Washington, PA 19034
FDA Establishment Registration Number 3004613836
Official Contact
Jeff Miller Vice President Operations Core Essence Orthopaedics, Inc. 575A Virginia Drive Fort Washington, PA 19034 Phone - (215) 310-9534 Fax - (215) 660-5015
Proprietary Name
1. PONTiSTM Sutures and Suture Anchors with Optional Crimps
Common Name
-
- Suture Anchor, Stainless Steel Suture
Regulatory Class & Device Product Codes, Suture Anchors
Class II 21 CFR § 888.3040 Smooth or threaded metallic bone fixation fastener Class II 21 CFR § 878.4495 Suture, Nonabsorbable, Steel, Monofilament and Multifilament, Sterile
Product Codes (Panel 87, Orthopedic):
PONTiS™ Suture Anchor with optional crimp
- MBI (fastener, fixation, nondegradable, soft tissue) .
Date Prepared: November 26, 2010
{1}------------------------------------------------
PONTiS™ Sutures with optional crimp
· GAQ
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the Core Essence PONTiS™ ferroFibre stainless steel suture (K081060), the Pioneer Surgical SDB Cerclage System (K992616), the Ortheon Medical Tenofix tendon repair devices (K023594).
The substantial equivalence of the PONTiS™ 3.0mm suture anchors are based on the equivalence in intended use, materials, operational principals, and indications to reNOVO™ and other suture anchors including the Core Essence Orthopaedics reNOVO™ (K071520) and the J&J DePuy Mitek Micro Anchor (K962511).
Brief Description of the Device, Sutures with Crimp Collars
This premarket notification covers line extensions for the PONTiS™ product line. The ferroFibre™ stainless steel (SS) sutures were cleared for marketing in 2008, and are marketed in the PONTiS ™ product group. The PONTiS™ product family has been expanded to include stainless steel suture anchors geometrically similar to the reNOVO™ titanium product family (cleared in 2007), in combination with ferroFibre™ SS Sutures and optional suture crimps.
The PONTiS [1] implants incorporate non-absorbable ferroFibre [1] 316 L Stainless Steel Surgical sutures and are available in United States Pharmacopoeia (USP) sizes 4-0 and 3-0 in various lengths.
316L Stainless steel is used in a wide variety of medical devices including previously approved implants of this type. The ferroFibre™ Stainless Steel Sutures may be secured by either knot tying or crimping with a collar.
The ferroFibre™ multifilament sutures are supplied sterile, armed with cutting needles.
The implant components also include stainless steel collars specific to each suture size. The collars are used to secure multiple suture strands together simultaneously by crimping. For example, crimps are sized to accept between four (4) and eight (8) strands of each specific size of ferroFibre™ suture for crimping. A crimping instrument is supplied to crimp the collars to the suture strands. An adjustable tendon holder with suture tensioning slots is provided to facilitate suture cable tightening and tensioning. Excess suture is cut and removed.
{2}------------------------------------------------
Indications for Use
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
Safety and Performance, PONTiSTM Anchors with ferroFibre™ Sutures and Crimp Collar Securing Method
A collection of tests was conducted to characterize biocompatibility, diameter and tensile strength of the sutures in accordance with:
- 0 · ISO 10993 standards
- 0 USP 32-NF 26 Monographs <861>, <871>, <881>
- Class II Special Control Guidance, Surgical Suture; Guidance for Industry and � FDA, June 3, 2003
In addition, a crimped suture loop pull-strength test was developed and used to verify that multiple sutures secured together using a crimped multi-strand collar had sufficient strength to secure a tendon repair. In the performance tests, the PONTiSTM Sutures secured with Crimp Collars had sufficient strength and exhibited equivalent results to knot-secured suture products currently marketed for hand tendon repair as confirmed by a comparison to knotted sutures and to the literature.
The Crimp-Collar-secured multi-strand suture tensile strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. All tests were performed in accordance with Part-10993 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices). The performance studies and biocompatibility reviews indicate that the PONTISTM Sutures, secured with crimped Collars are safe for their intended use.
Basis for Substantial Equivalence, Sutures with alternate Crimp Collar securing method:
Core Essence Orthopaedics PONTiS™ ferroFibre™ sutures secured with crimp collars are substantially equivalent to other sutures currently marketed in conjunction with suture anchors and present no substantial differences in design, material, intended use and function to previously approved products. Additionally, the subject device labeling is consistent both with FDA's guidance as well as current medical practice.
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the Core Essence
{3}------------------------------------------------
PONTiS™ ferroFibre stainless steel suture (K081060), the Pioneer Surgical SDB Cerclage System (K992616), the Ortheon Medical Tenofix tendon repair devices (K023594).
Brief Description of the Device, Suture Anchors
PONTiS™ Suture Anchors
The stainless steel PONTiS™ Suture Anchors are 3.0mm in diameter, are selftapping, and fully threaded. The anchors are each available with 4-0 and 3-0 multistrand ferroFibre™ stainless steel sutures. A disposable anchor driver holds the excess suture and delivers the anchor directly into the bone. The strands are then used to secure the soft tissue to bone.
Prior to installation, a hole is created in the bone using k-wire, drill or punch. The anchor is threaded into this hole. The anchor has self-tapping flutes.
Unlike the previously cleared titanium reNOVO™ Suture Anchor, PONTiS™ Suture Anchors are fully threaded allowing thread fixation in both cancellous and cortical bone.
The shorter overall length of the 3.0mm anchors, when compared to the self-drilling titanium reNOVO™ anchors, allows use in smaller/thinner bones where a lower profile is needed to achieve results without subsequent tissue injury.
The PONTiS™ threaded suture anchors are stainless steel to assure compatibility with the stainless steel in the ferroFibre™ stainless steel multi-strand sutures and the stainless steel crimp collars.
The size of the nonabsorbable stainless steel ferroFibre™ sutures used in the PONTiS™ Anchors is USP 4-0 and 3-0.
A single use driver and hand piece holds the excess suture and delivers the preloaded anchor into the bone. The suture strands are used to reapproximate and secure the soft tissue to bone.
The PONTiS™ Suture Anchors are provided sterile for single use applications.
The sizes and materials are designed to address the indications cited.
{4}------------------------------------------------
Indications for Use
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
Safety and Performance, Suture Anchors
The following safety and performance data has been provided to support substantial equivalence of the PONTiS™ Suture Anchors to reNOVO™ suture anchors.
Performance Testing: The reNOVO™ and the PONTiS™ Suture Anchors were compared in standardized foam bone model materials. The Anchor insertion torque, suture strength (knot pull), pull-out failure mechanism (pull-out or suture break strength) and pullout failure force were documented. In addition cyclic pull-out performance was characterized. The insertion torque was quantified in hard bone (bovine cortical bone) to assure the integrity of the anchor and anchor driver interface. The suture anchors were also functionally (vs. empirically) evaluated in animal and human cadaver bone by surgeons and engineers. The pullout strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
In all performance tests the PONTiS™ Sutures and Anchors secured by knots or crimps exhibited equivalent results to products currently marketed for the same indications as confirmed by a comparison to the literature documenting the comparative characteristics of suture anchors.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. The performance studies and biocompatibility reviews indicate that the PONTiS™ Sutures and Anchors are safe for their intended use.
Basis for Substantial Equivalence
The substantial equivalence of the PONTiS™ 3.0mm suture anchors are based on the equivalence in intended use, materials, operational principals, and indications to reNOVO™ and other suture anchors including the Core Essence Orthopaedics reNOVO™ (K071520) and the J&J DePuy Mitek Micro Anchor (K962511).
Relative to the hand, the PONTiS™ suture anchors have the same intended use (bone anchoring function), but different technological characteristics than the Mitek Micro-Anchors. (threads vs. barbs). The barbed predicate device from J&J Mitek incorporates the USP 4-0 suture sizes like the PONTiS™ suture anchors, and is indicated for soft tissue reattachment procedures in the hand.
The 3.0mm PONTiS™ suture anchors have threaded profiles similar to the 3.0mm reNOVO™ anchors. The PONTiS™ suture anchors are minor modifications of the
{5}------------------------------------------------
original titanium reNOVO™ anchors, and are indicated for the same procedures in the hand. The PONTiS™ suture anchors will be sold pre-assembled with stainless steel ferroFibre™ suture.
Conclusion
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the SBD Cerclage system and the Tenofix tendon repair system.
The Core Essence Orthopaedics PONTiS™ sutures used in the anchors are substantially equivalent to the previously cleared Core Essence ferroFibre™ sutures, and the Core Essence Orthopaedics PONTiS™ suture anchors are substantially equivalent to reNOVO™ and J&J Mitek Micro-Anchors currently marketed, and present no substantial differences in design, material, intended use and function.
END OF 510K SUMMARY
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized abstract symbol resembling a bird or a human figure with outstretched arms, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Core Essence Orthopaedics, Inc. % Mr. Jeff Miller Vice President Operations 575A Virginia Drive Fort Washington, Pennsylvania 19034
DEC - 2 2010
Re: K101126
Trade/Device Name: PONTis™ Sutures and Suture Anchors with Optional Crimps Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI, GAQ Dated: November 26, 2010 Received: November 29, 2010
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 - Mr. Jeff Miller
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vo
Sincerely yours,
Mark N. Mallekson
Mark I Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Statement of Indications for
・・・
DEC = 2 2010
510 (K) NUMBER IF KNOWN: K10176 (K) MANUFACTURER: Core Essence Orthopaedics, Inc. DEVICE NAME: PONTiS™ Sutures and Suture Anchors with Optional Crimps
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
. · .
- . Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons .
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- . Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons ●
Oputu for mxm
(Division Sign-Off) Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K101126
Prescription Use XX (Per 21 CFR 801 Subpart D) and/or
Over-the-Counter Use _ NO (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.