(224 days)
No
The device description focuses on the mechanical components and materials (sutures, anchors, crimps, drivers) and their physical properties and performance in securing soft tissue to bone or soft tissue. There is no mention of any software, algorithms, data processing, or learning capabilities that would indicate the use of AI or ML. The performance studies described are based on mechanical testing and biocompatibility, not algorithmic performance.
No
Explanation: A therapeutic device is used to treat a disease or condition for a therapeutic purpose. This device is indicated to secure soft tissue to soft tissue/bone in the hand. While it aids in the repair process, its primary function is mechanical fixation rather than active treatment of a disease or condition.
No
The device, PONTiS™ Sutures and Suture Anchors, is described as an implant used for securing soft tissue to soft tissue or bone reattachment. Its function is to provide physical support for damaged tissues, not to diagnose a condition or disease.
No
The device description clearly details physical components such as sutures, suture anchors, crimps, a crimping instrument, a tendon holder, and a disposable anchor driver, all made of stainless steel. These are hardware components, not software.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for securing soft tissue to soft tissue or bone reattachment in the hand. This is a surgical procedure performed directly on a patient.
- Device Description: The description details surgical implants (sutures, suture anchors, crimps) and instruments used for their placement. These are physical devices used within the body.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue samples) to provide information for diagnosis, monitoring, or screening.
IVD devices are used outside the body to analyze biological samples. This device is used inside the body for surgical repair.
N/A
Intended Use / Indications for Use
PONTiS™ Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons.
PONTiS™ Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons
Product codes (comma separated list FDA assigned to the subject device)
MBI, GAQ
Device Description
This premarket notification covers line extensions for the PONTiS™ product line. The ferroFibre™ stainless steel (SS) sutures were cleared for marketing in 2008, and are marketed in the PONTiS ™ product group. The PONTiS™ product family has been expanded to include stainless steel suture anchors geometrically similar to the reNOVO™ titanium product family (cleared in 2007), in combination with ferroFibre™ SS Sutures and optional suture crimps.
The PONTiS [1] implants incorporate non-absorbable ferroFibre [1] 316 L Stainless Steel Surgical sutures and are available in United States Pharmacopoeia (USP) sizes 4-0 and 3-0 in various lengths.
316L Stainless steel is used in a wide variety of medical devices including previously approved implants of this type. The ferroFibre™ Stainless Steel Sutures may be secured by either knot tying or crimping with a collar.
The ferroFibre™ multifilament sutures are supplied sterile, armed with cutting needles.
The implant components also include stainless steel collars specific to each suture size. The collars are used to secure multiple suture strands together simultaneously by crimping. For example, crimps are sized to accept between four (4) and eight (8) strands of each specific size of ferroFibre™ suture for crimping. A crimping instrument is supplied to crimp the collars to the suture strands. An adjustable tendon holder with suture tensioning slots is provided to facilitate suture cable tightening and tensioning. Excess suture is cut and removed.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
hand, PIP, DIP, MCP Joints
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
A collection of tests was conducted to characterize biocompatibility, diameter and tensile strength of the sutures in accordance with:
- ISO 10993 standards
- USP 32-NF 26 Monographs , ,
- Class II Special Control Guidance, Surgical Suture; Guidance for Industry and FDA, June 3, 2003
In addition, a crimped suture loop pull-strength test was developed and used to verify that multiple sutures secured together using a crimped multi-strand collar had sufficient strength to secure a tendon repair. In the performance tests, the PONTiS™ Sutures secured with Crimp Collars had sufficient strength and exhibited equivalent results to knot-secured suture products currently marketed for hand tendon repair as confirmed by a comparison to knotted sutures and to the literature.
The Crimp-Collar-secured multi-strand suture tensile strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. All tests were performed in accordance with Part-10993 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices). The performance studies and biocompatibility reviews indicate that the PONTiST™ Sutures, secured with crimped Collars are safe for their intended use.
Performance Testing for Suture Anchors: The reNOVO™ and the PONTiS™ Suture Anchors were compared in standardized foam bone model materials. The Anchor insertion torque, suture strength (knot pull), pull-out failure mechanism (pull-out or suture break strength) and pullout failure force were documented. In addition cyclic pull-out performance was characterized. The insertion torque was quantified in hard bone (bovine cortical bone) to assure the integrity of the anchor and anchor driver interface. The suture anchors were also functionally (vs. empirically) evaluated in animal and human cadaver bone by surgeons and engineers. The pullout strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
In all performance tests the PONTiS™ Sutures and Anchors secured by knots or crimps exhibited equivalent results to products currently marketed for the same indications as confirmed by a comparison to the literature documenting the comparative characteristics of suture anchors.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. The performance studies and biocompatibility reviews indicate that the PONTiS™ Sutures and Anchors are safe for their intended use.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.
0
Subject 510(k) Number: K101126
Sponsor
Core Essence Orthopaedics, Inc 575A Virginia Drive Fort Washington, PA 19034
FDA Establishment Registration Number 3004613836
Official Contact
Jeff Miller Vice President Operations Core Essence Orthopaedics, Inc. 575A Virginia Drive Fort Washington, PA 19034 Phone - (215) 310-9534 Fax - (215) 660-5015
Proprietary Name
1. PONTiSTM Sutures and Suture Anchors with Optional Crimps
Common Name
-
- Suture Anchor, Stainless Steel Suture
Regulatory Class & Device Product Codes, Suture Anchors
Class II 21 CFR § 888.3040 Smooth or threaded metallic bone fixation fastener Class II 21 CFR § 878.4495 Suture, Nonabsorbable, Steel, Monofilament and Multifilament, Sterile
Product Codes (Panel 87, Orthopedic):
PONTiS™ Suture Anchor with optional crimp
- MBI (fastener, fixation, nondegradable, soft tissue) .
Date Prepared: November 26, 2010
1
PONTiS™ Sutures with optional crimp
· GAQ
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the Core Essence PONTiS™ ferroFibre stainless steel suture (K081060), the Pioneer Surgical SDB Cerclage System (K992616), the Ortheon Medical Tenofix tendon repair devices (K023594).
The substantial equivalence of the PONTiS™ 3.0mm suture anchors are based on the equivalence in intended use, materials, operational principals, and indications to reNOVO™ and other suture anchors including the Core Essence Orthopaedics reNOVO™ (K071520) and the J&J DePuy Mitek Micro Anchor (K962511).
Brief Description of the Device, Sutures with Crimp Collars
This premarket notification covers line extensions for the PONTiS™ product line. The ferroFibre™ stainless steel (SS) sutures were cleared for marketing in 2008, and are marketed in the PONTiS ™ product group. The PONTiS™ product family has been expanded to include stainless steel suture anchors geometrically similar to the reNOVO™ titanium product family (cleared in 2007), in combination with ferroFibre™ SS Sutures and optional suture crimps.
The PONTiS [1] implants incorporate non-absorbable ferroFibre [1] 316 L Stainless Steel Surgical sutures and are available in United States Pharmacopoeia (USP) sizes 4-0 and 3-0 in various lengths.
316L Stainless steel is used in a wide variety of medical devices including previously approved implants of this type. The ferroFibre™ Stainless Steel Sutures may be secured by either knot tying or crimping with a collar.
The ferroFibre™ multifilament sutures are supplied sterile, armed with cutting needles.
The implant components also include stainless steel collars specific to each suture size. The collars are used to secure multiple suture strands together simultaneously by crimping. For example, crimps are sized to accept between four (4) and eight (8) strands of each specific size of ferroFibre™ suture for crimping. A crimping instrument is supplied to crimp the collars to the suture strands. An adjustable tendon holder with suture tensioning slots is provided to facilitate suture cable tightening and tensioning. Excess suture is cut and removed.
2
Indications for Use
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
Safety and Performance, PONTiSTM Anchors with ferroFibre™ Sutures and Crimp Collar Securing Method
A collection of tests was conducted to characterize biocompatibility, diameter and tensile strength of the sutures in accordance with:
- 0 · ISO 10993 standards
- 0 USP 32-NF 26 Monographs , ,
- Class II Special Control Guidance, Surgical Suture; Guidance for Industry and � FDA, June 3, 2003
In addition, a crimped suture loop pull-strength test was developed and used to verify that multiple sutures secured together using a crimped multi-strand collar had sufficient strength to secure a tendon repair. In the performance tests, the PONTiSTM Sutures secured with Crimp Collars had sufficient strength and exhibited equivalent results to knot-secured suture products currently marketed for hand tendon repair as confirmed by a comparison to knotted sutures and to the literature.
The Crimp-Collar-secured multi-strand suture tensile strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. All tests were performed in accordance with Part-10993 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices). The performance studies and biocompatibility reviews indicate that the PONTISTM Sutures, secured with crimped Collars are safe for their intended use.
Basis for Substantial Equivalence, Sutures with alternate Crimp Collar securing method:
Core Essence Orthopaedics PONTiS™ ferroFibre™ sutures secured with crimp collars are substantially equivalent to other sutures currently marketed in conjunction with suture anchors and present no substantial differences in design, material, intended use and function to previously approved products. Additionally, the subject device labeling is consistent both with FDA's guidance as well as current medical practice.
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the Core Essence
3
PONTiS™ ferroFibre stainless steel suture (K081060), the Pioneer Surgical SDB Cerclage System (K992616), the Ortheon Medical Tenofix tendon repair devices (K023594).
Brief Description of the Device, Suture Anchors
PONTiS™ Suture Anchors
The stainless steel PONTiS™ Suture Anchors are 3.0mm in diameter, are selftapping, and fully threaded. The anchors are each available with 4-0 and 3-0 multistrand ferroFibre™ stainless steel sutures. A disposable anchor driver holds the excess suture and delivers the anchor directly into the bone. The strands are then used to secure the soft tissue to bone.
Prior to installation, a hole is created in the bone using k-wire, drill or punch. The anchor is threaded into this hole. The anchor has self-tapping flutes.
Unlike the previously cleared titanium reNOVO™ Suture Anchor, PONTiS™ Suture Anchors are fully threaded allowing thread fixation in both cancellous and cortical bone.
The shorter overall length of the 3.0mm anchors, when compared to the self-drilling titanium reNOVO™ anchors, allows use in smaller/thinner bones where a lower profile is needed to achieve results without subsequent tissue injury.
The PONTiS™ threaded suture anchors are stainless steel to assure compatibility with the stainless steel in the ferroFibre™ stainless steel multi-strand sutures and the stainless steel crimp collars.
The size of the nonabsorbable stainless steel ferroFibre™ sutures used in the PONTiS™ Anchors is USP 4-0 and 3-0.
A single use driver and hand piece holds the excess suture and delivers the preloaded anchor into the bone. The suture strands are used to reapproximate and secure the soft tissue to bone.
The PONTiS™ Suture Anchors are provided sterile for single use applications.
The sizes and materials are designed to address the indications cited.
4
Indications for Use
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
Safety and Performance, Suture Anchors
The following safety and performance data has been provided to support substantial equivalence of the PONTiS™ Suture Anchors to reNOVO™ suture anchors.
Performance Testing: The reNOVO™ and the PONTiS™ Suture Anchors were compared in standardized foam bone model materials. The Anchor insertion torque, suture strength (knot pull), pull-out failure mechanism (pull-out or suture break strength) and pullout failure force were documented. In addition cyclic pull-out performance was characterized. The insertion torque was quantified in hard bone (bovine cortical bone) to assure the integrity of the anchor and anchor driver interface. The suture anchors were also functionally (vs. empirically) evaluated in animal and human cadaver bone by surgeons and engineers. The pullout strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
In all performance tests the PONTiS™ Sutures and Anchors secured by knots or crimps exhibited equivalent results to products currently marketed for the same indications as confirmed by a comparison to the literature documenting the comparative characteristics of suture anchors.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. The performance studies and biocompatibility reviews indicate that the PONTiS™ Sutures and Anchors are safe for their intended use.
Basis for Substantial Equivalence
The substantial equivalence of the PONTiS™ 3.0mm suture anchors are based on the equivalence in intended use, materials, operational principals, and indications to reNOVO™ and other suture anchors including the Core Essence Orthopaedics reNOVO™ (K071520) and the J&J DePuy Mitek Micro Anchor (K962511).
Relative to the hand, the PONTiS™ suture anchors have the same intended use (bone anchoring function), but different technological characteristics than the Mitek Micro-Anchors. (threads vs. barbs). The barbed predicate device from J&J Mitek incorporates the USP 4-0 suture sizes like the PONTiS™ suture anchors, and is indicated for soft tissue reattachment procedures in the hand.
The 3.0mm PONTiS™ suture anchors have threaded profiles similar to the 3.0mm reNOVO™ anchors. The PONTiS™ suture anchors are minor modifications of the
5
original titanium reNOVO™ anchors, and are indicated for the same procedures in the hand. The PONTiS™ suture anchors will be sold pre-assembled with stainless steel ferroFibre™ suture.
Conclusion
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the SBD Cerclage system and the Tenofix tendon repair system.
The Core Essence Orthopaedics PONTiS™ sutures used in the anchors are substantially equivalent to the previously cleared Core Essence ferroFibre™ sutures, and the Core Essence Orthopaedics PONTiS™ suture anchors are substantially equivalent to reNOVO™ and J&J Mitek Micro-Anchors currently marketed, and present no substantial differences in design, material, intended use and function.
END OF 510K SUMMARY
6
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized abstract symbol resembling a bird or a human figure with outstretched arms, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Core Essence Orthopaedics, Inc. % Mr. Jeff Miller Vice President Operations 575A Virginia Drive Fort Washington, Pennsylvania 19034
DEC - 2 2010
Re: K101126
Trade/Device Name: PONTis™ Sutures and Suture Anchors with Optional Crimps Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI, GAQ Dated: November 26, 2010 Received: November 29, 2010
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
7
Page 2 - Mr. Jeff Miller
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vo
Sincerely yours,
Mark N. Mallekson
Mark I Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
8
Statement of Indications for
・・・
DEC = 2 2010
510 (K) NUMBER IF KNOWN: K10176 (K) MANUFACTURER: Core Essence Orthopaedics, Inc. DEVICE NAME: PONTiS™ Sutures and Suture Anchors with Optional Crimps
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
. · .
- . Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons .
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- . Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons ●
Oputu for mxm
(Division Sign-Off) Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K101126
Prescription Use XX (Per 21 CFR 801 Subpart D) and/or
Over-the-Counter Use _ NO (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)