(77 days)
The ImaCor Zura system with ClariTEE probe is intended for use in the episodic assessment of cardiac function using transesophageal echocardiography. It is indicated for use in clinical settings, including long term settings such as the ICU. for an indwelling time period not to exceed 72 hours. The ImaCor Zura system with ClariTEE probe is not intended for pediatric use
The ImaCor Zura system with ClariTEE probe consists of three main components:
- Ultrasound Machine: A TEE predicate device optimized for use with ImaCor ClariTEE probe.
- Ultrasound Probe (ClariTEE formerly known as the Blue Probe in K080223): A miniaturized TEE probe optimized for longer dwell time relative to standard TEE probes enables use in longer term clinical settings such as the ICU. The probe distal tip is flexed upward transiently to obtain standard TEE images.
- Ultrasound Imaging Software: The software controls standard ultrasound machine functions such as imaging, recording and measuring. Continuous imaging is limited by a 20 minute software interlock should the operator mistakenly leave the machine in continuous imaging mode, thus limiting the potential unintentional exposure of the patient's mucosal tissue to acoustic energy. Maximum probe face temperature is limited according to FDA consensus standard IEC 60601-2-37. Latest software version enables two modes of operation; type B and color flow Doppler. The subsequent version cleared for marketing under K080223 enabled type B only.
The ImaCor Zura system with ClariTEE probe, a Transesophageal Echo Imaging System, received 510(k) clearance for software modification. The primary change was the addition of a color flow Doppler mode of operation.
1. Table of Acceptance Criteria and Reported Device Performance
Feature | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Intended Use | Identical to predicate device (ImaCor Zura K080223) | "The intended use of the ImaCor Zura with ClariTEE is identical to that of the ImaCor Zura (K080223)." |
Indications for Use | For episodic assessment of cardiac function using TEE; indicated for clinical settings including ICU for indwelling time not exceeding 72 hours; not for pediatric use. | "The ImaCor Zura system with ClariTEE probe is intended for use in the episodic assessment of cardiac function using transesophageal echocardiography. It is indicated for use in clinical settings, including long term settings such as the ICU, for an indwelling time period not to exceed 72 hours. The ImaCor Zura system with ClariTEE probe is not intended for pediatric use." (Verbatim from K100989) |
Technological Characteristics | Similar to predicate devices, with specified modifications related to the probe and software. | System consists of ultrasound machine (predicate optimized for ClariTEE), ClariTEE probe (miniaturized TEE probe for longer dwell time, flexed for images), and ultrasound imaging software (controls functions, 20-minute continuous imaging interlock, max probe face temp limited by IEC 60601-2-37). |
Software Modes | Addition of color flow Doppler mode to support cardiac function assessment. | "Latest software version enables two modes of operation; type B and color flow Doppler." |
Safety - Acoustic Energy | Limit potential unintentional exposure of patient's mucosal tissue to acoustic energy. | "Continuous imaging is limited by a 20 minute software interlock should the operator mistakenly leave the machine in continuous imaging mode, thus limiting the potential unintentional exposure of the patient's mucosal tissue to acoustic energy." |
Safety - Probe Temperature | Maximum probe face temperature limited by FDA consensus standard IEC 60601-2-37. | "Maximum probe face temperature is limited according to FDA consensus standard IEC 60601-2-37." |
Effectiveness (New Feature) | Color flow mode effectively demonstrates design modification for assessing cardiac function. | "Performance data using a flow phantom demonstrates the effectiveness of the design modification; the addition of a color flow mode of operation to support the assessment of cardiac function." |
Overall Safety & Effectiveness | As safe and effective as predicate devices. | "The ImaCor Zura System with ClariTEE probe is as safe and effective as the predicate devices." (Claim of Substantial Equivalence) |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "Performance data using a flow phantom demonstrates the effectiveness of the design modification; the addition of a color flow mode of operation to support the assessment of cardiac function."
- Test Set Sample Size: Not specified for the flow phantom. Flow phantom studies typically involve simulating various flow conditions but don't involve human subject sample sizes in the traditional sense.
- Data Provenance: The study was conducted on a "flow phantom," which is a controlled, laboratory-based setup designed to simulate fluid flow. This implies an in-vitro or ex-vivo experimental setting, not human subject data. The country of origin is not specified, but the submission is to the US FDA, so presumably the testing was done in a manner compliant with US regulatory expectations, likely in the US or a country with comparable standards. It is a prospective test specifically for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The performance data was gathered from a "flow phantom" for the new color flow mode. This type of testing typically relies on engineering specifications and physical measurements (e.g., flow velocity, direction accuracy) rather than expert human interpretation to establish ground truth for the device's technical function.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the performance data cited is from a "flow phantom," which means the "ground truth" for evaluating the function of the color flow mode would be based on the known parameters of the phantom and direct measurements from the device, not on expert adjudication of medical images.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The document only mentions performance data from a "flow phantom" and a claim of substantial equivalence to predicate devices based on this and other characteristics.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
Yes, a standalone performance assessment was done for the device's new feature (color flow mode) using a "flow phantom." This is an algorithm-only evaluation in the sense that the device's ability to accurately depict flow was tested independently of human interpretation of clinical images. The data pertains to the technical function of the device itself.
7. The Type of Ground Truth Used
For the specific performance data cited (color flow mode), the ground truth was known physical properties and measurements of the flow phantom. This is typically based on pre-established parameters of the phantom and/or reference measurements from other calibrated equipment, not on expert consensus, pathology, or outcomes data, as those relate more to clinical accuracy studies.
8. The Sample Size for the Training Set
Not applicable to this 510(k) submission. This submission is for modifications to an existing device (software change to add color flow Doppler), not the initial development of an AI algorithm that would require a distinct training set. The "performance data" refers to validation testing, not model training.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no mention of a training set or an AI algorithm being independently trained in this 510(k) summary. The submission focuses on modifications to an existing, cleared ultrasound system and probe.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.