K Number
K100984
Manufacturer
Date Cleared
2010-07-07

(90 days)

Product Code
Regulation Number
870.3470
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

XYLOS™ Vessel Guard is indicated as a cover for vessels during anterior vertebral surgery.

Device Description

XYLOS™ Vessel Guard is a flexible, non-resorbable, implantable sheet composed of microbialderived cellulose. The device is presented ready-to-use in a sterile double-pouched package. It is intended for one time use.

AI/ML Overview

The provided K100984 510(k) summary for the XYLOS™ Vessel Guard focuses on demonstrating substantial equivalence to predicate devices through material similarity and biomechanical testing, rather than clinical performance based on a study with acceptance criteria often associated with AI/software devices. Therefore, a direct response to some of the specific questions is not possible from the provided text, as they pertain to aspects that are not relevant to this type of medical device submission.

However, I can extract the relevant information and indicate where details regarding AI-specific criteria are not applicable.

Here's the breakdown based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/TestAcceptance CriteriaReported Device Performance
BiocompatibilityChemical CompositionIdentical to MTA™ Protective Sheet (K090778)Biologically and chemically identical to MTA™ Protective Sheet (K090778)
Biocompatibility PropertiesEquivalent to MTA™ Protective Sheet (K090778)Equivalent biocompatibility properties of the MTA™ Protective Sheet
Biomechanical PerformanceTensile StrengthEquivalent to Replication Medical Vessel Guard (K082782) and its predicatesDemonstrated biomechanical equivalence to Replication Medical Vessel Guard (K082782) and its predicates (PRECLUDE® IMA Sleeve, K960532; PRECLUDE® Vessel Guard, K061727)
Suture Pull-out TestingEquivalent to Replication Medical Vessel Guard (K082782) and its predicatesDemonstrated biomechanical equivalence to Replication Medical Vessel Guard (K082782) and its predicates

2. Sample Size Used for the Test Set and Data Provenance

This information is not applicable/not provided in the context of this 510(k) submission. The device is a physical implantable sheet, and its substantial equivalence was demonstrated through benchtop biomechanical testing and material comparison, not through a clinical study with a "test set" of patient data or clinical outcomes.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable/not provided. Ground truth, in the context of AI or diagnostic studies, refers to a definitive diagnosis or outcome, usually established by experts. For this device, "ground truth" would relate to the established material properties and biomechanical performance standards of the predicate devices. These are typically assessed by engineers and materials scientists in laboratory settings.

4. Adjudication Method for the Test Set

This information is not applicable/not provided. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies in expert opinions when establishing ground truth in clinical or image-based studies. Since this submission relies on benchtop testing and material comparisons, such adjudication is not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices or AI algorithms that assist human readers in interpreting data or images. The XYLOS™ Vessel Guard is an implantable surgical aid, and its effectiveness is demonstrated through its physical properties and biological compatibility, not through an assessment of human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone performance study was not done. This concept applies to AI algorithms that operate independently. The XYLOS™ Vessel Guard is a passive physical device.

7. The Type of Ground Truth Used

The "ground truth" in this context is based on:

  • Established Material Composition: The chemical and biological identity to the predicate MTA™ Protective Sheet.
  • Established Biomechanical Performance Standards: The tensile strength and suture pull-out performance as demonstrated by the predicate devices (Replication Medical Vessel Guard, PRECLUDE® IMA Sleeve, PRECLUDE® Vessel Guard).

Essentially, the ground truth for this submission is the well-characterized properties and performance of the legally marketed predicate devices.

8. The Sample Size for the Training Set

This information is not applicable/not provided. The concept of a "training set" applies to machine learning or AI models, where data is used to teach the algorithm. This device is a physical product, not an algorithm.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided for the same reasons as #8.

{0}------------------------------------------------

(വവു

XYLOS Corporation

510(k) Premarket Notification

JUL - 7 2010

Section 5: 510(k) Summary

The safety and effectiveness of XYLOS™ Vessel Guard is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate devices.

Sponsor:

Xylos Corporation 838 Town Center Drive Langhorne, PA 19047

Contact:

Joyce Elkins 838 Town Center Drive Langhorne, PA 19047 Phone: 215-867-0220 ext. 642 Facsimile: 215-741-4804

joyce.elkins@xyloscorp.com

Date of Submission: Proprietary Name: Common Name: Regulatory Class: Product Codes: Predicate Device(s):

April 6, 2010 XYLOS™ Vessel Guard

Vessel Guard

Class II

OMR

MTA™ Protective Sheet, K090778 Replication Medical Vessel Guard, K082782, and the predicates of Replication Medical Vessel Guard: PRECLUDE® Vessel Guard, K062161 PRECLUDE® IMA Sleeve, K960532

Device Description:

XYLOS™ Vessel Guard is a flexible, non-resorbable, implantable sheet composed of microbialderived cellulose. The device is presented ready-to-use in a sterile double-pouched package. It is intended for one time use.

Indications for Use:

XYLOS™ Vessel Guard is indicated as a cover for vessels during anterior vertebral, surgery.

CONFIDENTIAL

Page 14 of 34

{1}------------------------------------------------

Technological Characteristics and Substantial Equivalence:

XYLOS™ Vessel Guard is substantially equivalent to the previously cleared devices since it is biologically and chemically identical to MTA™ Protective Sheet (K090778), and therefore, has the equivalent biocompatibility properties of the MTA™ Protective Sheet.

Discussion of Performance Testing:

XYLOS™ Vessel Guard was subjected to biomechanical performance tests typical for vessel guard product (tensile strength and suture pull out testing demonstrated that XYLOS™ Vessel Guard is biomechanically equivalent to the predicate vessel guard device: Replication Medical Vessel Guard (K082782) and its predicates: PRECLUDE® IMA Sleeve, (K960532) and PRECLUDE® Vessel Guard (K061727). XYLOS™ Vessel Guard met the biomechanical performance test requirements, providing reasonable assurance of device performance for its intended use and supporting substantial equivalence.

Conclusion:

XYLOS™ Vessel Guard is substantially equivalent to the previously cleared devices since it is:

    1. Biologically and chemically identical to MTA™ Protective Sheet (K090778), and therefore. XYLQS™ Vessel Guard has the equivalent biocompatibility properties of the MTA™ Protective Sheet, and
    1. Biomechanically equivalent to Replication Medical Vessel Guard (K082782), as demonstrated via bench performance testing to its predicates.

CONFIDENTIAL

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, represented by three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL -- 7 2010

Xylos Corporation c/o Joyce Elkins Director, Quality Systems/Regulatory Affairs 838 Town Center Drive Langhorne, PA 19047

Re: K100984

Trade/Device Name: XYLOS™ Vessel Guard Regulation Number: 870.3470 Regulation Name: Intracardiac patch or pledget Regulatory Class: II Product Code: OMR Dated: April 06, 2010 Received: April 08, 2010

Dear Ms. Joyce Elkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for reducing the incidence, severity, and extent of post-operative adhesion formation have not been established.

{3}------------------------------------------------

Page 2 – Ms. Joyce Elkins

Furthermore, the indication for use as a cover for vessels following anterior vertebral surgery must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (301) 796-5540. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 301-796-6075. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

{4}------------------------------------------------

Page 3 – Ms. Joyce Elkins

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Christy Foreman

Christy Foreman Acting Director Office of Device Evaluation Center for Devices and Radiological Health Food and Drug Administration

Enclosure

{5}------------------------------------------------

510(k) Premarket Notification

XYLOS Corporation

Section 4: Indications for Use Statement

510(k) Number: To be assigned> K 1009 84

Device Name: XYLOS™ Vessel Guard

Indications for Use:

XYLOS™ Vessel Guard is indicated as a cover for vessels during anterior vertebral surgery.

Prescription Use X AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of __ 1_ Divin a. D. (Division Sign-Off) Division of Cardiovascular Devices CONFIDENTIAL Page 13 of 34 510(k) Number_K100984

§ 870.3470 Intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene.

(a)
Identification. An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.(b)
Classification. Class II (performance standards).