(85 days)
No
The description focuses on standard oscillometric blood pressure measurement technology and basic electronic features like LCD display, memory, and irregular heartbeat detection. There is no mention of AI, ML, or any advanced algorithms that would suggest their use.
No.
The device is a blood pressure monitor intended for measurement and monitoring, not for treating a condition.
Yes
The device is intended to measure diastolic and systolic blood pressures and pulse rate, which are physiological parameters used to assess a patient's health status and can aid in the diagnosis or monitoring of medical conditions.
No
The device description explicitly mentions physical components like an inflatable cuff, LCD, electronic interface module, touch key button, and silicon integrate pressure sensor technology, indicating it is a hardware device with integrated software, not a software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The KD-5965 series Fully Automatic Electronic Blood Pressure Monitor measures blood pressure and pulse rate by applying a cuff to the upper arm. This is a non-invasive measurement taken directly from the body, not from a sample in vitro (in glass or outside the body).
- Intended Use: The intended use clearly states it's a "non-invasive blood pressure measurement system."
Therefore, this device falls under the category of a non-invasive medical device, not an in vitro diagnostic.
N/A
Intended Use / Indications for Use
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Product codes
DXN
Device Description
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.
For KD-5965, the operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 60 times. If any irregular heartbeat is detected, it can be shown on the LCD. The LCD backlight, the touch key button and the voice function make it more convenient to use. The RCC (radio controlled clock) function can automatically receive the radio clock signal to adjust the clock accurately.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
upper arm
Indicated Patient Age Range
adult individual
Intended User / Care Setting
medical professionals or at home
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor will conform to the following standards before marketing:
- 60601-1, Medical Electrical Equipment Part 1: General ● EC Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
- · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
510(k) Summary
JUN 2 9 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
1.0 submitter's information
Name: Andon Health Co., Ltd. Address: No 3, Jinping Street, Ya An Road, Nankai District, Tianjin, P.R. China Phone number: 86-22-6052 6161 Fax number: 86-22-6052 6162 Contact: Liu Yi Date of Application: 03/16/2010
2.0 Device information
KD-5965 series Fully Automatic Electronic Blood Device name: Pressure Monitor
Model No: KD-5965XY(X =AZ, Y= blank or AZ)
The model in KD-5965 series are the modification to KD-5965, and the modification will rise no new 510(k) according to FDA's guidance document . 3.0 Classification
Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular
4.0 Predict device information
1 | Manufacturer: | Andon Health Co., Ltd. |
---|---|---|
Device: | KD-5905 Fully Automatic Electronic Blood Pressure | |
Monitor | ||
510(k) number: | K090770 | |
2 | Manufacturer: | Andon Health Co., Ltd. |
Device: | KD-5963 Fully Automatic Electronic Blood Pressure | |
Monitor |
1
510(k) number: K093528
5.0 Device description
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.
For KD-5965, the operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 60 times. If any irregular heartbeat is detected, it can be shown on the LCD. The LCD backlight, the touch key button and the voice function make it more convenient to use. The RCC (radio controlled clock) function can automatically receive the radio clock signal to adjust the clock accurately.
6.0 Intended use
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
The intended use and the indication for use of KD-5965 series, as described in its labeling are the same as the predict device KD-5905 or KD-5963.
7.0 Summary comparing technological characteristics with predicate device
Technological Characteristics | Comparison result |
---|---|
Design principle | Identical |
Appearance | Similar |
Patients contact Materials | Identical |
2
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor FDA 510(k) Files
100940 3/4
i. ে
Performance | Similar |
---|---|
Biocompatibility | Identical |
Mechanical safety | Identical |
Energy source | Identical |
Standards met | Identical |
Electrical safety | Identical |
EMC | Identical |
Function | Similar |
..
3
8.0 Performance summary
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor will conform to the following standards before marketing:
- 60601-1, Medical Electrical Equipment Part 1: General ● EC Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
- · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
9.0 Comparison to the predict device and the conclusion
Our device KD-5965 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-5905 whose 510(k) number is K090770 and the Fully Automatic Electronic Blood Pressure Monitor KD-5963 with the 510(k) number of K093528.
The KD-5965 is very similar with the predicted devices in the intended use, the design principle, the material, the energy source and the applicable standards. Only their appearance is different, moreover, the KD-5965 uses different MCU from the two predicted devices.
However, appropriate test will be conducted and specified acceptance criteria will be met before KD-5965 is marketed.
4
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 2 9 2010
Andon Health Co., Ltd. c/o Ms. Liu Yi President No. 3 JinPing Street, Ya An Road, NanKai District Tianjin 300190 CHINA
Re: K100940
Trade/Device Name: KD-5965 Series Fully Automatic Electronic Blood Pressure Monitor Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: 74 DXN Dated: May 27, 2010 Received: June 1, 2010
Dear Ms. Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
5
Page 2 - Ms. Liu Yi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
KI00940
Statement of Indications for Use
510(k) Number :
KD-5965 series Fully Automatic Electronic Blood Device name: Pressure Monitor
Indications for use:
KD-5965 series Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Prescription use Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR YES (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W.M. Wood
iovascular Devices
Page 1 of 1