K Number
K093528

Validate with FDA (Live)

Date Cleared
2009-12-04

(18 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KD-5963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

KD-7963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

Device Description

KD-5963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

KD-7963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.

The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability of KD-5963 and KD-7963 are both 60 times. If any irreqular heartbeat is detected, it can be shown on the LCD, KD-5963 has the voice function, KD-7963 has the function of touch button and LCD backlight.

AI/ML Overview

The provided text describes a 510(k) submission for two fully automatic electronic blood pressure monitors, KD-5963 and KD-7963. The submission focuses on demonstrating substantial equivalence to predicate devices (KD-5962 and KD-7962) rather than conducting a standalone clinical study to establish new performance criteria.

Therefore, the information required to fully answer the request, particularly regarding specific acceptance criteria, detailed study results (sample sizes, ground truth establishment, expert qualifications, MRMC studies, effect size, etc.), and standalone performance, is not present in the provided text. The document primarily highlights compliance with recognized standards.

However, I can extract the available information regarding compliance with standards, which can be interpreted as demonstrating that the device meets generally accepted performance criteria for blood pressure monitors.

Here's a breakdown based on the available information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Standards Met)Reported Device Performance
IEC 60601-1 (General Safety)Device conforms.
IEC 60601-1-2 (EMC)Device conforms.
AAMI SP10:2002 (Sphygmomanometers)Device conforms. This standard specifies performance requirements for automated sphygmomanometers, covering accuracy and other aspects. Meeting this standard indicates the device's accuracy and performance are within acceptable limits.
AAMI / ANSI SP10:2002/A1:2003Device conforms.
AAMI / ANSI SP10:2002/A2:2006Device conforms.

Study Proving Acceptance Criteria:

The document states that the devices (KD-5963 and KD-7963) "conform to the following standards" and that "the test in this submission provides demonstration that these small differences [from the predicate device] do not raise any new questions of safety and effectiveness." This implies that testing was conducted to verify compliance with the listed standards, particularly AAMI SP10, which governs the accuracy and performance of non-invasive blood pressure monitors.

Specific Details Not Available in the Provided Text:

The following requested information is not explicitly stated in the provided 510(k) summary:

  • 2. Sample sizes used for the test set and the data provenance: The document mentions "the test in this submission" but does not detail the sample size (number of subjects, measurements) or the origin (country, retrospective/prospective) of the data.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
  • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not mentioned, as a traditional ground truth adjudication process akin to image analysis wasn't described.
  • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a blood pressure monitor, not an AI-assisted diagnostic tool for Human-in-the-Loop interpretation.
  • 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The device itself is a "Fully Automatic Electronic Blood Pressure Monitor," implying standalone performance by design. However, the exact methodology and results of such a standalone performance test, beyond stating compliance with AAMI SP10, are not detailed.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For blood pressure monitors, "ground truth" for accuracy studies typically refers to reference measurements obtained using a validated method (e.g., a mercury sphygmomanometer performed by trained professionals). While compliance with AAMI SP10 strongly implies such a ground truth was used, the specific details are not provided.
  • 8. The sample size for the training set: Not applicable, as this is a device based on oscillometric technology and not a deep learning AI algorithm that requires a "training set" in the common sense.
  • 9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information

DEC - 4 2009

Name: Andon Health Co., Ltd. No 31, Changjiang Road, Nankai District, Tianjin, Address: P.R. China 86-22-6052 6161 Phone number: 86-22-6052 6162 Fax number: Contact: Liu Yi Date of Application: 11/13/2009

2.0 Device information

Trade name: Fully Automatic Electronic Blood Pressure Monitor Common name: Noninvasive blood pressure measurement system Noninvasive blood pressure measurement system Classification name:

3.0 Classification

Production code: DXN- Noninvasive blood pressure measurement system. Requlation number: 870.1130 Classification: II Panel: Cardiovascular

4.0 Predict device information

Andon Health Co., Ltd. Manufacturer: KD-5962 Fully Automatic Electronic Blood Pressure Monitor Device: KD-7962 Fully Automatic Electronic Blood Pressure Monitor K090771, K091997 510(k) number:

5.0 Device description

KD-5963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

5-1

10F4

{1}------------------------------------------------

KD-7963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.

The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability of KD-5963 and KD-7963 are both 60 times. If any irreqular heartbeat is detected, it can be shown on the LCD, KD-5963 has the voice function, KD-7963 has the function of touch button and LCD backlight.

6.0 Intended use

KD-5963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

KD-7963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

The intended use and the indication for use of KD-5963 and KD-7963, as described in its labeling are the same as the predict device KD-5962 and KD-7962.

7.0 Summary comparing technological characteristics with predicate device

PAGE 2 OF 4

{2}------------------------------------------------

: :

:

.

. . . . . . . .

.

:

: :

Technological CharacteristicsComparison result
Design principleIdentical
AppearanceSimilar
Patients contact MaterialsIdentical
PerformanceIdentical
BiocompatibilityIdentical
Mechanical safetyIdentical
Energy sourceIdentical
Standards metIdentical
Electrical safetyIdentical
EMCIdentical
FunctionSimilar

PAGE 3 OF 4

.

:

{3}------------------------------------------------

8.0 Performance summary

KD-5963 and KD-7963 Fully Automatic Electronic Blood Pressure Monitor conform to the following standards:

  • · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
  • · IEC 60601-1-2. Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
  • · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
  • · AAMI / ANSI SP10:2002/A1:2003 -- Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
  • · AAMI / ANSI SP10:2002/A2:2006 -- Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.

9.0 Comparison to the predict device and the conclusion

Our device KD-5963 and KD-7963 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-5962 and KD-7962 whose 510(k) number is K090771 and K091997.

KD-7963 and KD-7962 is very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Their appearance are different, they have different memory times, KD-7963 doesn't have the function of voice and averaging the last 3 times memory, the cuff circumference of KD-7963 is different from KD-7962.

KD-5963 and KD-5962 is very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Their appearance are different, KD-5963 also has a new cuff.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

PAGE 4 OF 4

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a human figure with outstretched arms, resembling a bird in flight.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

DEC - 4 2009

Andon Health Co., Ltd. c/o Mr. Liu Yi No. 31 Changjiang Road, Nankai District, Tianjin CHINA

Re: K093528

Trade/Device Name: KD-5963 Fully Automatic Electronic Blood Pressure Monitor; and, KD-7963 Fully Automatic Electronic Blood Pressure Monitor Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: November 13, 2009 Received: November 16, 2009

Dear Mr. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Liu Yi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K093528

Statement of Indications for Use

510(k) Number :

Device name:

KD-5963 and KD-7963 Fully Automatic Electronic Blood Pressure Monitor

Indications for use:

KD-5963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

KD-7963 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

Prescription use AND/OR Over-The-Counter Use YES Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

10(k) Number K07

Page 1 of

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).