K Number
K100845
Manufacturer
Date Cleared
2010-09-21

(180 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When intended for pedicle screw fixation from TI-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor, and failed previous fusion.

As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autograft or allograft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.

Device Description

The purpose of this 510(k) submission is to modernize the existing PathFinder Pedicle Screw System. A set of percutaneous insertion rods and new instrumentation have been included for minimally invasive procedures.

The Zimmer Spine PathFinder II system consists of polyaxial cannulated screws and rods and is intended to provide temporary stabilization following surgery to fuse the spine. A range of spinal rod lengths included in this system allows the surgeon to place polyaxial pedicle screws through an open or mini-open procedure. This system is intended only to provide stabilization during the development of a solid fusion with autograft or allograft. These implants are intended to be removed after the development of a solid fusion mass.

AI/ML Overview

The Zimmer Spine PathFinder® II Minimally Invasive Pedicle Screw System is indicated for pedicle screw fixation from T1-S1 for immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for various spinal instabilities or deformities. Additionally, for placement between L3 and S1, it is indicated for Grade 3 or Grade 4 spondylolisthesis.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Static and dynamic axial compression bending performance in accordance with ASTM F1717Performed as intended, demonstrating substantial equivalence to the predicate device.
Static torsion performance in accordance with ASTM F1717Performed as intended, demonstrating substantial equivalence to the predicate device.
Substantial equivalence to predicate device (Sequoia Pedicle Screw System) regarding intended use, indications for use, and no new safety/effectiveness issues.The proposed device was found to be as safe and effective and performs as well as or better than the predicate device.

2. Sample Size and Data Provenance:

  • Test Set Sample Size: Not explicitly stated in the provided text for the mechanical tests. The document mentions "the proposed PathFinder II system and predicate Sequoia Pedicle Screw system" as the subjects of the tests, implying full systems were tested, but the number of individual components or systems tested is not specified.
  • Data Provenance: The mechanical tests were performed by Zimmer Spine, Inc. The location of these tests and the nationality of the data are not specified. The tests are non-clinical, meaning they were likely conducted in a laboratory setting. This is a retrospective evaluation as it's part of a 510(k) submission for a device based on pre-existing technology and a predicate.

3. Number of Experts and Qualifications for Ground Truth:

  • Not applicable directly. This information is typically relevant for studies involving human interpretation (e.g., imaging devices, diagnostic algorithms). For this device, the "ground truth" for mechanical performance is established by standardized testing protocols (ASTM F1717) and engineering principles, rather than expert consensus on a data set. The FDA's review and determination of substantial equivalence act as the ultimate "expert" validation in this regulatory context.

4. Adjudication Method for the Test Set:

  • Not applicable directly. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human experts when establishing ground truth. In this case, the evaluation relies on objective mechanical test results compared against established standards. The FDA's review process itself can be seen as a form of adjudication in determining regulatory acceptance based on the submitted data.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed to assess the impact of a device or AI on human reader performance, especially in diagnostic or screening contexts. The Zimmer Spine PathFinder® II is a spinal implant, and its performance is evaluated through mechanical testing, not through human reader interpretation of images or data.

6. Standalone (Algorithm Only) Performance Study:

  • Yes, in a sense, a standalone performance was done for the mechanical aspects. The mechanical tests (static and dynamic axial compression bending, static torsion) were performed on the device itself, functioning independently of human interaction during the test. The "performance" here refers to the device's physical response to simulated forces as measured against a standard, not an algorithm's output.

7. Type of Ground Truth Used:

  • Standardized Mechanical Test Results/Engineering Principles. The "ground truth" for this device's performance is derived from the objective measurements obtained during mechanical testing in accordance with ASTM F1717 and the comparison of these results to those of the predicate device. The performance is deemed acceptable if it meets or exceeds the predicate's performance under these standardized conditions.

8. Sample Size for the Training Set:

  • Not applicable. This device is a physical medical implant, not an AI/machine learning algorithm that requires a "training set" of data. The "training" for such devices involves design, engineering, and manufacturing processes informed by established biomechanical principles and material science.

9. How Ground Truth for the Training Set Was Established:

  • Not applicable. As explained above, there is no "training set" in the context of an AI algorithm for this physical device. The design and validation of the device rely on engineering specifications, material testing, and the performance of previous or predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Zimmer Spine. The logo consists of a circle with a stylized "Z" inside it, above the words "zimmer" and "spine", with the word "spine" underlined. The text is in a bold, sans-serif font.

510(K) SUMMARY PathFinder® II Minimally Invasive Pedicle Screw System

Company/Address:Zimmer Spine, Inc.5301 Riata Park Court, Bldg FAustin, TX. 78727Telephone: (512) 533-1998Fax: (512) 258-0995SEP 21 2010
Contact Person:David Padgett, RAC (US)Senior Regulatory Affairs Specialist
Date Prepared:17 June 2010

Device Identification

Proprietary Name: Common Name: Classification:

PathFinder® II Minimally Invasive Pedicle Screw System Spinal Fixation System NKB - 21 CFR § 888.3070 - Spinal Interlaminal Fixation Orthosis

Predicate Device Information

Sequoia Pedicle Screw System, K082032 (S.E. 10/7/2008) PathFinder Minimally Invasive Pedicle Screw System, K030625 (S.E.3/28/2003) Depuy Viper II, K090648 (S.E. 6/10/09)

Device Description and Technological Characteristics:

The purpose of this 510(k) submission is to modernize the existing PathFinder Pedicle Screw System. A set of percutaneous insertion rods and new instrumentation have been included for minimally invasive procedures.

The Zimmer Spine PathFinder II system consists of polyaxial cannulated screws and rods and is intended to provide temporary stabilization following surgery to fuse the spine. A range of spinal rod lengths included in this system allows the surgeon to place polyaxial pedicle screws through an open or mini-open procedure. This system is intended only to provide stabilization during the development of a solid fusion with autograft or allograft. These implants are intended to be removed after the development of a solid fusion mass.

The predicate device, Sequoia Pedicle Screw System and the proposed PathFinder II device are fabricated from medical grade titanium alloy per ASTM F136 and commercially pure titanium per ASTM F67.

{1}------------------------------------------------

Intended Use / Indications for Use:

When intended for pedicle screw fixation from TI-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor, and failed previous fusion.

As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autograft or allograft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.

Performance Data:

For a determination of substantial equivalence, the following non-clinical mechanical tests were performed:

Static and dynamic axial compression bending and static torsion tests in accordance with ASTM F1717 were performed on the proposed PathFinder II system and predicate Sequoia Pedicle Screw system and the results compared. The proposed device functioned as intended and the observed test results demonstrate substantial equivalence to the aforementioned predicate device.

Substantial Equivalence:

The PathFinder II Minimally Invasive Pedicle Screw System has the same intended use and indications for use as the predicate Sequoia Pedicle Screw System. Additionally, no new issues of safety or effectiveness are raised. Based on the supporting documentation presented within this premarket notification, the proposed device is as safe and effective and performs as well as or better than the predicate device identified.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is smaller than the central image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002

Zimmer Spine Inc. % Mr. David Padgett Senior Regulatory Affairs Specialist 5301 Riata Park Court, Building F Austin, Texas 78727

SEP 21 2010

Re: K100845

Trade/Device Name: PathFinder® II Minimally Invasive Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: September 09, 2010 Received: September 10, 2010

Dear Mr. Padgett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. David Padgett

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark H. Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

K100845 SEP 2 1 2010

510(k) Number (if known): K100845

Device Name: PathFinder® II Minimally Invasive Pedicle Screw System

Indications for Use:

When intended for pedicle screw fixation from T1-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, and failed previous fusion.

As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autograft or allograft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division Sign-Off Division of Surgical, Orthopedic, and Kestorative Devices

KI00845 5 ! 0(k) Number_

Page 1 of 1

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.