(30 days)
The ergo Imaging System is intended to image the distribution of radionuclides in the body by means of a photon radiation detector. In so doing, the system produces images depicting the anatomical distribution of radioisotopes within the human body for interpretation by authorized medical personnel.
The proposed changes involve modifications to the 2020tc Imaging System to increase size of the detector head field of view (FOV) from 8"x 8" to 12"x 15". Modifications include mechanical and electrical design changes to support the large field-of-view (LFOV) detector head. The modified device (ergo Imaging System) incorporates Digirad's solid-state RIM detector design, modified with 3mm size pixels required for general purpose planar imaging. The 2020tc Imager detectors currently utilize the same 3mm pixel size.
The detectors used in the modified and predicate device utilize a pixelated, multi-crystal CsI scintillator detector with each pixel optically coupled to a low noise photodiode array. The charge detected from each gamma ray is amplified and processed using an amplifier circuit. The RIM detector design technology is currently used in the Digirad Cardius XPO imager systems with a 6mm pixel size for Cardiac SPECT imaging. The RIM detector has been modified to incorporate a 3mm size pixel required for general planar imaging. The RIM detector design includes electrical and mechanical configurations allowing for field replacement of detector modules and improved system performance (better energy resolution). The updated design of the RIM detector head assembly allows some of the current sub-systems to be moved into the detector head assembly (air dryer) and simplification of others (cooling and power distribution systems).
The modified device uses the 2020tc Imaging System SeeQuanta Acquisition software, with minor modifications required for use with the 3mm pixel size RIM detector modules.
The 2020tc Imager was initially marketed as the Digirad Notebook Imager (K961104), then re-branded as the 2020tc Imaging System (K982855) when it was used in conjunction, with the SPECTour Rotating Chair to obtain SPECT images in patients who are seated in an upright position. The modified device (ergo Imaging System) is a general purpose Nuclear Medicine Imaging device used for planar imaging, the same as the Notebook Imager/20201c Imager when imaging without the rotating chair.
The provided text describes a 510(k) premarket notification for the "ergo Imaging System," a scintillation (gamma) camera. The submission focuses on demonstrating substantial equivalence to predicate devices, rather than a standalone clinical study proving specific acceptance criteria in terms of diagnostic performance metrics like sensitivity or specificity.
Here's an analysis of the provided information based on your questions:
1. Table of acceptance criteria and the reported device performance
The document does not specify quantitative acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, accuracy) that are typically reported for AI/algorithm-based devices. Instead, it focuses on demonstrating functional equivalence and meeting design specifications compared to predicate devices. The acceptance criteria essentially revolve around the system performing as per its specifications, which are similar to the predicate device's functional specifications, and not raising new safety or effectiveness concerns.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Functions as per specifications | All tests passed with actual results substantially matching expected results. System meets design specifications. |
| Similar functional specifications to predicate devices | Design specifications are similar to predicate device functional specifications. |
| Equivalent efficacy to predicate devices (no new safety/effectiveness questions) | Digirad internal testing and phantom images demonstrated equivalent efficacy to predicate devices, and did not raise new questions regarding safety and effectiveness. |
| Substantially equivalent to predicate devices (intended use, technology, overall system performance) | Testing results demonstrate the ergo Imaging System meets specifications and is substantially equivalent based on comparisons of intended use, technology, and overall system performance. |
2. Sample size used for the test set and the data provenance
The document does not explicitly mention a "test set" in the context of diagnostic performance (e.g., a set of patient images for evaluation). The testing described is primarily internal verification and validation of the system's technical specifications and phantom images. Therefore, details like data provenance or sample size for a diagnostic test set are not provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The document describes technical testing and phantom images, not a clinical study involving experts establishing ground truth for diagnostic accuracy.
4. Adjudication method for the test set
Not applicable, as no diagnostic test set is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study is mentioned. This device is a gamma camera, a hardware imaging system, not an AI-assisted diagnostic software.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is a hardware device; the concept of "standalone performance" for an algorithm doesn't directly apply in the same way it would for AI-driven software. The "system meets design specifications" and "equivalent efficacy to predicate devices" with phantom images represent its standalone performance relative to its intended function.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document refers to "phantom images" and "Digirad internal testing." This implies the "ground truth" for the testing was based on known properties of the phantoms and the expected technical performance of the imaging system. There is no mention of clinical ground truth like pathology or outcomes data.
8. The sample size for the training set
Not applicable. This document describes a hardware device submission, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set is described.
{0}------------------------------------------------
Appendix 2: 510(k) Summary
A. Sponsor
Digirad Corporation 13950 Stowe Drive Poway, California 92064 Contact Person: Joel Tuckey Tel: (858) 726-1527 Fax: (858) 726-1546
APR 2 3 2010
B. Date Prepared: March 23, 2010
C. Device Name
| Trade Name: | ergo Imaging System |
|---|---|
| Common Name: | Camera, Scintillation (Gamma) |
| Classification Name: | Scintillation (gamma) camera |
| Device Class: | 21CFR 892.1100, Class I |
| Product Code: | IYX |
Cleared/Predicate Devices D.
The ergo Imaging System is substantially equivalent to the following cleared devices:
(1) Notebook Imager (Gamma Camera), cleared on May 28, 1997, under K9641104, and
(2) 2020tc SPECT Imaging System and the SPECTour Chair (SPECT Imaging System), cleared on November 9, 1998 under K982855, and
(3) Cardius-1, Cardius-2, Cardius-3, 2020tc SPECT Imaging System, cleared on July 13, 2005 under K051549, and
(4) Cardius XPO series and 2020tc SPECT Imaging Systems cleared on March 23, 2007 under K070542.
E. Device Description
The proposed changes involve modifications to the 2020tc Imaging System to increase size of the detector head field of view (FOV) from 8"x 8" to 12"x 15". Modifications include mechanical and electrical design changes to support the large field-of-view (LFOV) detector head. The modified device (ergo Imaging System) incorporates Digirad's solid-state RIM detector design, modified with 3mm size pixels required for general purpose planar imaging. The 2020tc Imager detectors currently utilize the same 3mm pixel size.
The detectors used in the modified and predicate device utilize a pixelated, multi-crystal CsI scintillator detector with each pixel optically coupled to a low noise photodiode array. The charge detected from each gamma ray is amplified and processed using an amplifier circuit. The RIM detector design technology is currently used in the Digirad Cardius XPO imager systems with a 6mm pixel size for Cardiac SPECT imaging. The RIM detector has been modified to incorporate a 3mm size pixel required for general planar imaging. The RIM detector design includes electrical and mechanical configurations allowing for field
{1}------------------------------------------------
replacement of detector modules and improved system performance (better energy resolution). The updated design of the RIM detector head assembly allows some of the current sub-systems to be moved into the detector head assembly (air dryer) and simplification of others (cooling and power distribution systems).
The modified device uses the 2020tc Imaging System SeeQuanta Acquisition software, with minor modifications required for use with the 3mm pixel size RIM detector modules.
The 2020tc Imager was initially marketed as the Digirad Notebook Imager (K961104), then re-branded as the 2020tc Imaging System (K982855) when it was used in conjunction, with the SPECTour Rotating Chair to obtain SPECT images in patients who are seated in an upright position. The modified device (ergo Imaging System) is a general purpose Nuclear Medicine Imaging device used for planar imaging, the same as the Notebook Imager/20201c Imager when imaging without the rotating chair.
F. Intended Use
The ergo imaging system is intended to image the distribution of radionuclides in the body by means of a photon radiation detector. In so doing, the system produces images depicting the anatomical distribution of radioisotopes within the human body for interpretation by authorized medical personnel.
The ergo imaging system will be used by trained medical personnel to perform nuclear medicine studies.
G. Technology
The ergo Imaging System utilizes Digirad's solid state detector technology used on other Digirad imaging systems. The LFOV (12" x 15") detector head assembly facilitates being able to image larger areas of the patient with a single acquisition. The changes do not alter the fundamental scientific technology of the predicate device.
H. Testing
Verification and Validation tests were conducted to demonstrate the ergo imaging system functioned as per its specifications. All tests passed with the actual results substantially matching the expected results. The testing shows the system meets the design specifications, which are similar to the predicate device functional specifications. Digirad internal testing and phantom images obtained with the ergo imaging system have demonstrated equivalent efficacy to the predicate devices, and did not raise new questions regarding safety and effectiveness.
I. Conclusion
Testing results demonstrate that the ergo Imaging System meets the specifications and is substantially equivalent to the predicate devices, based on comparisons of intended use and technology, and overall system performance.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The emblem is rendered in a blue color. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.
OCT 2 1 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Joel Tuckey Vice President QA/RA Digirad Corporation 13950 Stowe Drive POWAY CA 92064
Re: K100838
Trade/Device Name: ergo Imaging System Regulation Number: 21 CFR 892.1100 Regulation Name: Scintillation (gamma) Camera Regulatory Class: II Product Code: IYX Dated: March 23. 2010 Received: March 24, 2010
Dear Mr. Tuckey:
This letter corrects our substantially equivalent letter of April 23, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
David C. Brown, Ph.D.
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K100838
Device Name: ergo Imaging System
Indications for Use:
The ergo Imaging System is intended to image the distribution of radionuclides in the body by means of a photon radiation detector. In so doing, the system produces images depicting the anatomical distribution of radioisotopes within the human body for interpretation by authorized medical personnel.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Page 1 of
3
... は
Robert Lee Bell
(Division Sign-Off
Division of Radiological Devices
Office of In Vitro Diagnostic Device Eraluation and Safety
510K K100838
§ 892.1100 Scintillation (gamma) camera.
(a)
Identification. A scintillation (gamma) camera is a device intended to image the distribution of radionuclides in the body by means of a photon radiation detector. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class I (general controls).