K Number
K100818

Validate with FDA (Live)

Date Cleared
2010-07-09

(108 days)

Product Code
Regulation Number
866.2660
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The illumigene C. difficile DNA amplification assay, performed on the illumipro-10, is a qualitative in vitro diagnostic test for the direct detection of toxigenic C. difficile in human stool specients suspected of having Clostridium difficile associated disease (CDAD).

The illumigene C. difficile assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect the pathogenicity locus (PaLoc) of toxigenic Clostridium difficile PaLoc is a gene segment present in all known toxigenic C. difficile strains. The C. difficile PaLoc codes for both the Toxin A gene (tcdA) and the Toxin B gene (tcdB), has conserved border regions, and is found at the same site on the C. difficile genome for all toxigenic strains . The illumigene C. difficile assay detects the PaLoc by targeting a partial DNA fragment on the Toxin A gene. The tcdA target region was selected as an intact region remaining in all known A+B+ and A-B+ toxinotypes.

illumigene C. difficile is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.

Device Description

The illumigene Molecular Diagnostic Test System is comprised of the illuminene C. difficite DNA Ampification Test Kit. the illumigene C. difficile External Control Kit and the illumipro-10 Automated Isothermal Amplification and Detection System. The illumigene C. difficile DNA amplification assay utilizes loop-mediated isothermal ampification (LAMP) technology to detect the presence of toxigenic C. difficile in patients suspected of having C. difficile associated disease (CDAD). Each illumiaene C. difficile assay is combleted using an illuminene Sample Preparation Apparatus. Illumicene Reaction Buffer, illumigene C. difficile Test Device, Sample Collection Brush, and illumigene Extraction Tube. Samples are prepared using the Sample Collection Brush and the illumigene Sample Collection Apparatus, target DNA is heat extracted in the Extraction Tube and DNA amplification occurs in the illumigene C. difficile Test Device.

The illumipro-10 heats each illumigene C. difficile Test Device containing prepared samples, facilitation of target DNA. When toxigenic C. difficile is present in the patient specific sequence is amplified and Magnesium pyrophosphate is formed. Magnesium pyrophosphate in the reaction mixture. The Illumioro-10 detects the change in light transmission mixture created by the precipitating Magnesium pyrophosphate. Sample results are reported as Positive based on the detected change in transmission.

The illumigene C. difficile External Control Kit consists of a Positive Control Reagent. External Control reagents are provided to aid the user in deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors. The illumigene C. difficile External Control Kit is required for routine Quality Control.

AI/ML Overview

The provided document describes the illumigene™ C. difficile DNA Amplification Assay, illumipro-10™ Automated Isothermal Amplification and Detection System, and illumigene™ C. difficile External Control Kit. This system is designed for the direct detection of toxigenic C. difficile in human stool samples to diagnose Clostridium difficile-associated disease (CDAD).

Here's an analysis of the acceptance criteria and the study that proves the device meets these criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated in numerical thresholds (e.g., "sensitivity must be >90%"). Instead, the document presents the device's performance characteristics from a clinical study, implying these demonstrated values are acceptable for market clearance. The comparison is made against cytotoxic bacterial culture as the reference comparator.

MetricAcceptance Criteria (Implied)Reported Device Performance (Clinical Study)
SensitivityHigh95.2% (95% CI: 89.2% - 97.9%)
SpecificityHigh95.3% (95% CI: 93.2% - 96.7%)
CorrelationHigh95.3% (95% CI: 93.4% - 96.6%)
Invalid RateLow2.9% (Overall)
Analytical Sensitivity (LoD)Defined for various strains4-64 CFU/test for different strains
Non-interferenceDemonstrated for common substances & microorganismsNo interference from specified drugs, fecal components, white blood cells, whole blood, and a range of bacteria/viruses at tested concentrations.
ReproducibilityHigh agreement across sites and lots100% for negative and positive samples at various test levels, 91% for high negative.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Study (Test Set): 697 qualified patient samples. Of these, 677 samples were used for the overall performance data (Table 1), likely due to invalids or samples not meeting all inclusion criteria for the final analysis.
  • Data Provenance: Clinical trials were conducted in 2010 across four independent clinical test sites and the manufacturer located in the Midwestern and Southern regions of the United States. The study involved prospective collection of samples (patient samples collected from individuals suspected of having CDAD).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the clinical test set was established by cytotoxic bacterial culture, which is a laboratory method, not human expert consensus. Therefore, the concept of "number of experts" or their qualifications for establishing the ground truth is not applicable in this context.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by laboratory culture (cytotoxic bacterial culture), not by human interpretation requiring an adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a molecular diagnostic assay, not an imaging-based AI system that would involve human readers interpreting AI results. The comparison is between the automated system and a laboratory reference standard.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The clinical performance data (Sensitivity, Specificity, Correlation) presented in "Performance Comparison, Clinical Tests" and "Table 1. Overall performance data" represents the performance of the illumigene C. difficile assay (algorithm/device only) in detecting toxigenic C. difficile directly from patient stool samples, as compared to cytotoxic bacterial culture. The system is designed for automated amplification and detection.

7. The Type of Ground Truth Used

The primary ground truth used for the clinical performance evaluation was cytotoxic bacterial culture.

8. The Sample Size for the Training Set

The document does not specify a separate training set size for the illumigene C. difficile assay's development. This is common for molecular diagnostic assays that are designed based on known genetic targets and biological principles, rather than machine learning algorithms requiring explicit training data. The "Performance Comparison, Clinical Tests" section describes a clinical validation or test set.

9. How the Ground Truth for the Training Set was Established

As no explicit "training set" is mentioned in the context of machine learning, this question is not directly applicable. For the development and initial validation of the assay's ability to detect the pathogenicity locus (PaLoc), the ground truth for targeted sequences would have been established through well-characterized bacterial strains and genetic sequencing data specific to toxigenic C. difficile. The analytical sensitivity section describes testing against various known C. difficile strains with established toxinotypes and phenotypes to determine the limit of detection.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a stylized globe on the left, followed by the word "Meridian" in a bold, sans-serif font. A horizontal line underlines the word "Meridian", and below the line is the text "Bioscience, Inc." in a smaller, bold, sans-serif font.

510(k) Application illumigene C. difficile
Description:510(k) Summary illumigene C. difficile
Identification:Attachment 003. Revision 002
Date:June 21, 2010

JUL 0 9 2010

510(k) number:K100818
Date of preparation:June 21, 2010
Submitter:Meridian Bioscience, Inc.
Submitter's address:3471 River Hills DriveCincinnati, Ohio 45244
Contact:Michelle Smith
Contact number:(513) 271-3700
Device name:illumigene™ C. difficileillumipro-10™ Automated Isothermal Amplification and Detection Systemillumigene™ C. difficile External Control Kit
Common name:C. difficile DNA Amplification Assay
Classification name:C. difficile Nucleic AcidsOMN, CFR Section 866.2660
Predicate device:K091109: Cepheid® Xpert® C. difficileModel GXVDIFFICILE-10,900-0423, 900-0065, 900-0144, 900-0145, 900-0146,900-0381, 900-0391, 900-0392, 900-393, 950-0151
Reference comparator:Cytotoxic bacterial culture

Description of the device: 1

The illumigene Molecular Diagnostic Test System is comprised of the illuminene C. difficite DNA Ampification Test Kit. the illumigene C. difficile External Control Kit and the illumipro-10 Automated Isothermal Amplification and Detection System. The illumigene C. difficile DNA amplification assay utilizes loop-mediated isothermal ampification (LAMP) technology to detect the presence of toxigenic C. difficile in patients suspected of having C. difficile associated disease (CDAD). Each illumiaene C. difficile assay is combleted using an illuminene Sample Preparation Apparatus. Illumicene Reaction Buffer, illumigene C. difficile Test Device, Sample Collection Brush, and illumigene Extraction Tube. Samples are prepared using the Sample Collection Brush and the illumigene Sample Collection Apparatus, target DNA is heat extracted in the Extraction Tube and DNA amplification occurs in the illumigene C. difficile Test Device.

The illumipro-10 heats each illumigene C. difficile Test Device containing prepared samples, facilitation of target DNA. When toxigenic C. difficile is present in the patient specific sequence is amplified and Magnesium pyrophosphate is formed. Magnesium pyrophosphate in the reaction mixture. The Illumioro-10 detects the change in light transmission mixture created by the precipitating Magnesium pyrophosphate. Sample results are reported as Positive based on the detected change in transmission.

The illumigene C. difficile External Control Kit consists of a Positive Control Reagent. External Control reagents are provided to aid the user in deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors. The illumigene C. difficile External Control Kit is required for routine Quality Control.

{1}------------------------------------------------

MeridianBioscience, Inc.510(k) Application illumigene C. difficile
Description: 510(k) Summary illumigene C. difficile
Identification: Attachment 003. Revision 002
Date: June 21, 2010

Intended Use:

The illumigene C. difficile DNA amplification assay, performed on the illumipro-10, is a qualitative in vitro direct detection of toxigenic C. difficile in human stool spected of having Clostridium difficile-associated disease (CDAD).

The illumicene C. difficile assay utilizes loop-mediated issthermal DNA amplification (LAMP) technology to detect the pathogenicity locus (PaLoc) of toxigenic Clostridium difficile Paloc is a gene segment present in all known toxigenic C. difficile strains. The C. difficile Paloc codes for both the Toxin B gene (tcdA) and the Toxin B gene (tcdB), has conserved border regions, and is found at the same site on the C. difficile genome for all toxigenic strains . The illumigene C. difficile assay detects the PaLoc by targeting a partial DNA fragment on the Toxin A gene. The tcdA target region was selected as an intact region remaining in all known A+B+ and A-B+ toxinotypes.

Illumigene C. difficile is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.

Comparison to predicated device:

CharacteristicIllumigene™ C. difficileCepheid® Xpert® C. difficile
Test FormatDNA Amplification AssayDNA Amplification Assay
Intended Use
DNA Amplification TechnologyLoop-Mediated Isothermal Amplification (LAMP)Real-Time Polymerase Chain Reaction (PCR)
Target Sequences DetectedPartial DNA fragment on the Toxin A gene of thepathogenicity locus (PaLoc) found in all knownstrains for toxigenic C. difficile.Toxin B sequences
Qualitative/QuantitativeQualitativeQualitative
Screening, Diagnostic orIdentification TestDiagnosticDiagnostic
Specimen Types
Unformed Human StoolYesYes
Human Stool in Cary-Blair-basedMediaYesNo
illumigene Sample Preparation ApparatusXpert C. difficile Assay cartridges
illumigene Reaction BufferSample Reagent
Reagents/Componentsillumigene C. difficile Assay DeviceReagent 1
illumigene Extraction TubesReagent 2
Sample Collection Brushes
ExtractionManualSelf-contained and automated
AmplificationSelf-contained and automatedSelf-contained and automated
DetectionSelf-contained and automatedSelf-contained and automated
Testing TimeApproximately 60 minutesApproximately 45 minutes

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe graphic on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bolded, sans-serif font. A thin line is placed between the word "Meridian" and "Bioscience, Inc."

510(k) Application illumigene C. difficile
Description:510(k) Summary illumigene C. difficile
Identification:Attachment 003. Revision 002
Date:June 21, 2010
CalibrationNot requiredNot required
Comparison to predicated device:
CharacteristicIllumigene™ C. difficileCepheid® Xpert® C. difficile
Controls
Inhibition, AssayProvidedIllumigene Sample Preparation Apparatus:Staphylococcus aureusIllumigene C. difficile Assay Device:Staphylococcus aureus LAMP PrimersAdjunct ReagentsProvidedSample Processing Control (SPC): Bacillus globigiiProbe Check Control (PCC): Fluorescence emitting probes
ExternalIllumigene C. difficile External Control KitCatalog 279920User SuppliedKWIK-STIK™ from MicroBioLogics catalog 0329 (toxigenic C. difficile) as positive controlKWIK-STIK™ from MicroBioLogics catalog 0331 ( C. sordelli ) as negative control
ExtractionUser SuppliedUser Supplied
Equipment
Instrumentationillumipro-10™ Automated Isothermal Amplification and Detection SystemMicropipette 50 μL, 200 μLGeneXpert® Dx SystemVortex Mixer
General Laboratory EquipmentDry-bath with 12mm heat block, 95 CInterval TimerVortex Mixer
Reading MethodVisible Light TransmissionFluorescence
Results
C. difficile Toxinotypes Tested0 (A+/B+)III (A+/B+)V (A+/B+)VIII (A-/B+)X (A-/B+)XII (A+/B+)IX/XXIII (A+/B+)0 (A+/B+)III (A+/B+)V (A+/B+)VIII (A-/B+)XII (A+/B+)
Results InterpretationINVALIDPOSITIVENEGATIVEToxigenic C. difficile POSITIVEToxigenic C. difficile NEGATIVEINVALIDERRORNO RESULT

{3}------------------------------------------------

Image: Meridian Bioscience, Inc. logo510(k) Application illumigene C. difficile
Description: 510(k) Summary illumigene C. difficile
Identification: Attachment 003. Revision 002
Date: June 21, 2010

Performance Comparison, Non-clinical Tests:

Interference Testing

Selected drugs and other non-microbial substances that might be present in stool samples from healthy persons or patients suspected of having C. difficile associated disease were added to a natural negative and a contrived positive sample. The natural negative and contrived positive samples were prepared from donor samples and were confirmed negative by cytotoxic bacterial culture. The contrive sample was prepared by spiking a confirmed negative sample with toxinogenic C. difficile strain VPI 10463 to 18 CFU/test, slightly above the 16 CFU assay limit of delection for this organism. Potentially interfering substances were added at final concentrations of 5% V/V or greater. Dilution Controls for each sample were prepared by adding a phosphate-buffered saline solution in place of the potentially interfering substance. Each sample was tested in triplicate.

The following substances, at the specified saturated solventrations, do not interfere with illumigene C. difficile test results in the final concentrations listed: Barium sulfate (5 mg/mL), fecal fat (equivalent to 2.65 mg stearic plus 1.3 mg palmitic acids per mL), hemoglobin) (3.2 mg/mL), IgA (5 mg/mL), Inodium AD® (0.00667 mg/mL), Kaopectate® (0.87 mg/mL), Metronidazole (12.5 mg/mL) Mylanta® (4.2 mg/mL), Pepto-Bismol® (0.87 mg/mL), Prilosec® (0.5 mg/mL), TUMS® (0.5 mg/mL), Vancomycin (12.5 mg/mL), white blood cells (5%V/V), whole blood (5% V/V).

Cross-reactivity Study

Potentially cross-reactive microorganisms that might be present in stool samples from healthy persons or patients suspected of having C, difficile associated disease were added to a natural negalive sample. The natural negative and contrived positive samples were prepared from donor samples and were confirmed negative by cytotoxic bacterial culture. The contrived positive sample was prepared by spiking a confirmed negative sample with toxinogenic C. difficile strain VPI 10463 to 18 CFU assay limit of detection for this organism. Potentially cross-reactive microorganisms were added at concentrations of 1.2 x 10 fmL (bacteria and fungi) or 1 x 10539mL TCIDgs/mL (viruses). Dilution Controls for each sample were prepared by adding a phosphate-buffered saline solution in place of the potentially cross-reactive microorganisms. Each sample was tested in triplicate.

The following microorganisms, at the indicated concentrations, do not interfere with illumigene C. difficile test results: Aeromonas hydrophila. Bacteroides fragilis. Campylobacter fetus. Campylobacter ieiuni, Candida albicans. Citrobacter frendi. Clostridium perfringens, Enterobacter cloacae, Enterococcus faccalis, Escherichia coli, Escherichia coli 0157:H7, Escherichia hermannii, Helicobacter pylori, Klebsiella pneumoniae, Lactococcus lactis, Listeria monocytogenes, anaerobius, Plesiomonas shigelloides, Proteus vulgaris, Pseudomonas aeruginosa, Pseudomonas fluorescens, Salmonella Groups B-E, Serratia liquefaciens, Serratia marcescens, Shigella flexneri, Shigella sonnei, Staphylococcus aureus, Staphylococus epidermidis. Vibrio parahaemolyticus, Yersinia enterocolliica, Adenovirus, Echovirus, Echovirus, Rotavirus.

Performance Comparison, Clinical Tests:

Clinical trials for the illumigene C. difficile assay, including the illumipro-10 Automated Isothermal amplification and detection system, were conducted in 2010. Performance characteristics of the illumigene C. difficile assay were determined by comparison to cytotoxic culture. Four independent clinical test sites located in the Midwestern and Southern recions of the United States and the manufacturer evaluated a total of 697 qualified patient samples. Samples were collected from 274 (39.3%) males and 419 (60.1%) females. In the case of 4 (0.6%) of the patients, sex was not known. The age groups of patients range from 2 years of age to 96 years. No differences in test performance were observed based on patient age, sex, or geographical location. Overall Sensitivity was determined to be 95.2% (95% Cl: 89.2% - 97.9%). Overall Specificity was determined to be 95.3% - 96.7%). Subsequent tables show overall assay performance as well as performance by clinical site and patient age.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe graphic on the left, followed by the company name in bold, black font. The word "Meridian" is on the top line, and "Bioscience, Inc." is on the bottom line.

510(k) Application illumigene C. difficile
Description:510(k) Summary illumigene C. difficile
Identification:Attachment 003. Revision 002
Date:June 21, 2010

Table 1. Overall performance data

Cytotoxic bacterialcultureillumigene C. difficile
culturePositiveNegativeTotal
Positive995**104
Negative27*546573
Total126551677
95% CI
Sensitivity99/10495.2%89.2 - 97.9%
Specificity546/57395.3%93.2 - 96.7%
Correlation645/67795.3%93.4 - 96.6%

15/27 false positive results were positive by and the remaining 12 false positive results, 8 were positive results, 8 were positive by an FDA cleared assay for the detection of GDH.

** 2/5 false negative results were negative by another FDA cleared molecular assay

Table 2. Performance characteristics by site

SitePositive SamplesNegative Samples
illumigenelCytotoxicbacterialcultureSensitivity %95% CIillumigenelCytotoxicbacterialcultureSpecificity %95% CI
Total99/10495.2%89.2 - 97.9%546/57395.3%93.2 - 96.7%
Site 14/580.0%37.6 - 96.4%58/6097.6%88.6 - 99.1%
Site 212/12100%75.7 - 100%62/6792.5%83.7 - 96.8%
Site 320/20100%83.9 - 100%87/9294.6%87.9 - 97.7%
Site 48/8100%67.6 - 100%36/3992.3%79.7 - 97.3%
Site 555/5993.2%83.8 - 97.3%303/31596.2%93.5 - 97.8%

Table 3. Invalid rates by site

SiteClinical Site Evaluation
Total InvalidsAssayInvalidsInstrumentInvalidsInvalid Rate
Site 13303/68 (4.4%)
Site 21011/80 (1.3%)
Site 38178/120 (6.7%)
Site 41101/48 (2.1%)
Site 57617/381 (1.8%)
Total2011/697 (1.6%)***9/697 (1.3%)20/697 (2.9%)

*** 1 Specimen remained invalid after repeat testing from the original sample.

Table 4. Results by patient age

Positive SamplesNegative Samples
Patient ageillumigenelToxigeniccultureSensitivity %95% CIillumigenelToxigeniccultureSpecificity %95% CI
≥ 2 - 12 years10/1190.9%62.3 - 98.4%75/7994.9%87.7 - 98.0%
> 12 to 21 years5/5100%56.6 - 100%53/5694.6%85.4 - 98.2%
> 21 years83/8795.4%88.8 - 98.2%417/43795.4%93.0 - 97.0%
Age Unknown1/1100%20.7 - 100%1/1100%20.7 - 100%

{5}------------------------------------------------

14.000CAR A SEA CONSULERS OF CHILIPS OF CHILIP OF CHILIP OF CHILIP OF CHILIP OF OF CHILIP OF OTHER OF OF OF OF OF OF OF OF OF OF OF OF OF OF OF OFFICE OF OFFICE OF OFFICE OF OFFICE- A B & B & C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C------------------------------------------------------------------------------------------------------------------------------------------------------------------------------of the collection of the control of the continued and
. Be a series and the comments of the section of the section of the states of the section of the section of the states of the section of the section of the section of the sta
510(k) Application illumigene C. difficile
Description:510(k) Summary illumigene C. difficile
Identification:Attachment 003. Revision 002
Date:June 21, 2010

Analytical Sensitivity

The analytical sensitivity of this assay for C. difficile was based on 20 replicates for each measurand and with a stated probability (e.g., 95% or 19/20 positive replicates) of obtaining positive responses at the following levels of the measurands:

Strain IDToxinotypePhenotypeLoD/Test
VPI 104630A+/B+4 CFU/test
2007431III (NAP1)A+/B+32 CFU/test
CF1VIIIA-/B+64 CFU/test
2006240V (NAP7)A+/B+32 CFU/test
BI8IIIA+/B+64 CFU/test
2007858IX/XXIIIA+/B+32 CFU/test
8864XA-/B+64 CFU/test

Additional C. difficile stock cultures from different sources were tested and produced positive reactions at 64 CFU/test with illumicene C. difficile. Strains and toxinotypes tested were as follows: Type 0 Strains: 10463, 2004111, 2004205, 2005070, 2005257, 2008029, 2008162, 2008341, 2008351, 2009066, 2009099, B1, G1, J7, K12, Y1; Type III Strains: 2004052, 2004118, 2007431, Bl17, Bl8; Type V Strains: 2005325, 2006240, 2008188, 2009018, 2009065, BK6; Type VIII Strains: 43598, 2008016, CF1; Type XII Strains: 2007435; Type IX/XIII Strains: 2007858; Unknown Strains: 2009132, 2009155, 2009277.

Reproducibility

Blind coded panels of 10 samples were supplied to three independent laboratories for precision studies. Samples were randomly sorted within each panels identities. The panels included contrived samples manufactured at the assay limit of detection (n = 3) and just below the limit of blank (i.e., high negative sample, n = 3). The panels also included uncharacterized positive (n = 2) and negative (n = 2) samples. Testing was performed by different operators at each site on the same day (intra-assay variability) for five days (inter-assay variability). Three lots of illumigene C difficile were used in this study. The results are given in the table below:

Site 1Site 2Site 3Total
Sample TypePercent agreementPercent agreementPercent agreementPercent agreement
Negative20/20100%20/20100%19/19****100%59/59100%
High Negative25/3083%29/3097%28/3093%82/9091%
Low Positive30/30100%30/30100%30/30100%90/90100%
Positive20/20100%20/20100%20/20100%60/60100%

*** 1 specimen generated an instrument invalid test result.

Conclusions

The illumigene C. difficile assay used in coniunction with the illumioro-10 can be used to defficile in human stool samples. The test is diagnostic for toxigenic C. difficile infection.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three overlapping human profiles facing to the right, with the profiles gradually decreasing in size from front to back. The profiles are rendered in black, and the text is also in black against a white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center -- WO66-0609 Silver Spring, MD 20993-0002

Meridian Bioscience, Inc. c/o Michelle L. Smith Senior Director, Quality Systems 3471 River Hills Dr. Cincinnati, OH 45244

JUL 9 2010

Re: K100818

Trade/Device Name:Illumigene C. difficile Assay
Regulation Number:21 CFR §866.2660
Regulation Name:Microorganism differentiation and identification device
Regulatory Class:Class I
Product Code:OMN
Dated:June 21, 2010
Received:June 22, 2010

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 - Michelle L. Smith

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 115809.html 1580.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sal, aAr

Sally A. Hojvat, M.Sc., Ph.I Director

Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indication(s) for Use Form

510(k) Number (if known): K100818

Device Name: illumigene Molecular Diagnostic Test System (illumigene C. difficile DNA Amplification Assay, illumipro-10)

Indications for Use:

The illumigene C. difficile DNA amplification assay, performed on the illumipro-10, is a qualitative in vitro diagnostic test for the direct detection of toxigenic C. difficile in human stool specients suspected of having Clostridium difficile associated disease (CDAD).

The illumigene C. difficile assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect the pathogenicity locus (PaLoc) of toxigenic Clostridium difficile PaLoc is a gene segment present in all known toxigenic C. difficile strains. The C. difficile PaLoc codes for both the Toxin A gene (tcdA) and the Toxin B gene (tcdB), has conserved border regions, and is found at the same site on the C. difficile genome for all toxigenic strains . The illumigene C. difficile assay detects the PaLoc by targeting a partial DNA fragment on the Toxin A gene. The tcdA target region was selected as an intact region remaining in all known A+B+ and A-B+ toxinotypes.

illumigene C. difficile is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Kurdalin M. Fode

ion Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K10 0817

§ 866.2660 Microorganism differentiation and identification device.

(a)
Identification. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.