K Number
K100761

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-12-03

(261 days)

Product Code
Regulation Number
888.3390
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioPro Bipolar Head is intended for use in combination with a BioPro PSL femoral stem for primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
• Femoral neck and trochanteric fractures of the proximal femur
• Osteonecrosis of the femoral head
• Revision procedures where other devices or treatments for this indications have failed

Device Description

The BioPro Bipolar Head consist of a factory assembled Ultra High Molecular Weight Polyethylene (UHMWPE) liner in a cobalt chrome outer shell and UHMWPE retention ring with a Ti-6Al-4v spring. These bipolar heads include outer diameters ranging from 38-60mm, in 1mm increments, to properly fit the patient anatomy. The smaller bipolar heads (38-42 mm) have an inner diameter that mates with a 22mm diameter femoral head; the larger bipolar heads (43-60mm) have an inner diameter that mates with a 28mm diameter femoral head. The BioPro Bipolar Head may be used in conjunction with a BioPro PSL femoral stem (K872535 and K871462) for hemiarthroplasty.

AI/ML Overview

This is a 510(k) summary for a medical device, specifically a hemi-hip prosthesis. The information provided heavily emphasizes that the device is "substantially equivalent" to predicate devices, and therefore no new clinical studies were performed to establish its performance or meet acceptance criteria derived from clinical data.

Here's a breakdown based on your request, highlighting the relevant sections and the absence of clinical data:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a substantial equivalence submission relying on predicate devices, there aren't explicit "acceptance criteria" for a new study to meet, nor "reported device performance" in terms of clinical outcomes. Instead, the "acceptance criteria" are implied by the performance of the predicate devices and the successful completion of specific non-clinical tests.

Acceptance Criteria (Implied)Reported Device Performance
Mechanical Performance:
Push-out strength within range of legally marketed bipolar designs.Push-out and lever-out strengths were within the range of legally marketed bipolar designs.
Lever-out strength within range of legally marketed bipolar designs.(Addressed above)
Material Durability:
Locking ring spring should maintain function after heat exposure.Locking ring spring storage heat tolerance test (150 deg. F) was conducted and presumably passed, as the device was deemed substantially equivalent. (No specific performance metric provided, but implied successful completion).
Sterilization Efficacy:
ETO sterilization provides sterility assurance level.ETO sterilization validation 10^6 was performed and presumably passed, as the device was deemed substantially equivalent. (No specific performance metric provided, but implied successful completion).
Equivalence to Predicate Device:
Identical to predicate device (K082705) with specified modifications.The BioPro Bipolar Head is identical to the BioPro Bipolar head cleared in K082705 with the specified exceptions (modified UHMWPE retaining ring and added retaining spring). The device is also stated to be identical to the Omni Life Science Bipolar head 510k 100151 and manufactured to the same standards and technical drawings.
No change in Indications for Use from predicate.The indications have not changed from the predicate. (Listed in "Indications for Use" section: primary or revision hemiarthroplasty for femoral neck/trochanteric fractures, osteonecrosis, revision procedures where other treatments failed).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable in the context of clinical data. For the non-clinical tests (push-out, lever-out, heat tolerance, sterilization), the specific sample sizes are not provided in this summary.
  • Data Provenance: The "study" here is a series of non-clinical, benchtop tests. The data provenance is from laboratory testing conducted by or for the device manufacturer, BioPro, Inc. There is no clinical data provenance from human subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. No clinical test set with ground truth established by experts was used. The evaluation relies on engineering principles and comparison to predicate device specifications.

4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs. without AI assistance

Not applicable. This device is a physical implant (hemi-hip prosthesis), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC study involving human readers or AI was performed or is relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical implant, not an algorithm.

7. The Type of Ground Truth Used

Not applicable in the sense of clinical ground truth (e.g., pathology, outcomes data). For the non-clinical tests, the "ground truth" would be established engineering standards, material specifications, and the performance characteristics of the legally marketed predicate devices.

8. The Sample Size for the Training Set

Not applicable. There was no "training set" in the context of clinical data or machine learning for this device. The device design and testing are based on established engineering principles and comparison to predicate devices.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there was no training set, there was no ground truth for it to be established. The basis for the device's design and assessment is its substantial equivalence to previously cleared devices, supported by non-clinical testing demonstrating comparable mechanical and material properties.

{0}------------------------------------------------

510K Summary K100761

.

SubmitterBioPro, Inc.2929 Lapeer RdPort Huron, MIContactDavid MrakDirector of Product Development810-982-7777
Preparation Date:August 23, 2010DEC - 3 2010
Device Name
Common NameHemi-Hip prosthesis, uncemented
Trade NameBioPro Bipolar Head
Classification NameHip joint femoral (hemi-hip) metal/polymer cementedor uncemented prosthesis orthopedic (OR)
Regulatory ClassClass II per 21CFR 888.3390
Product CodeKWY
Legally MarketedPredicate Device(s)K082705 (BioPro Bipolar Head)K082468, K945793, K931655, K100151
Device DescriptionThe BioPro Bipolar Head consist of a factory assembled Ultra HighMolecular Weight Polyethylene (UHMWPE) liner in a cobalt chromeouter shell and UHMWPE retention ring with a Ti-6Al-4v spring. Thesebipolar heads include outer diameters ranging from 38-60mm, in 1mmincrements, to properly fit the patient anatomy. The smaller bipolarheads (38-42 mm) have an inner diameter that mates with a 22mmdiameter femoral head; the larger bipolar heads (43-60mm) have aninner diameter that mates with a 28mm diameter femoral head. TheBioPro Bipolar Head may be used in conjunction with a BioPro PSLfemoral stem (K872535 and K871462) for hemiarthroplasty.
Indications for UseThe indications have not changed from the predicate. The BioProBipolar head is intended for use in combination with the BioPro PSLfemoral stem for primary or revision hemiarthroplasty of the hip. Thisprosthesis may be used for the following conditions, as appropriate:• Femoral neck and trochanteric fractures of the proximal femur• Osteonecrosis of the femoral head• Revision procedures where other devices or treatments for thisindications have failed
Predicate DevicecomparisonThe BioPro Bipolar Head is identical to the BioPro Bipolar head clearedin K082705 with the following exceptions:• The UHMWPE (ASTM F-648) retaining ring in the acetabularcomponent was modified to increase the amount ofinterference/overlap between the retaining ring and the head

{1}------------------------------------------------

  • A retaining Spring (Ti-6Al-4V Eli) was added to the UHMWPE . retaining ring.
    Technological Characteristics

Non-Clinical Test Summary ...

The device is identical to the Omni Life Science Bipolar head 510k 100151 and is manufactured to the same standards and technical drawings

The following tests were conducted:

  • . Push-out and lever out testing. Push-out and lever out strengths were within the range of legally marketed bipolar designs.
  • Locking ring spring storage heat tolerance test (150 deg. F) .
  • ETO sterilization validation 10°6 ●

Clinical Test Summary

No Clinical studies were performed

Conclusions

The BioPro Bipolar head is a substantially equivalent device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 0 3 2010

BioPro Inc. % Mr. David Mrak Director of Product Development 2929 Lapeer Road Port Huron, Michigan 48060

Re: K100761 Trade/Device Name: BioPro Bipolar Heads Regulation Number: 21 CFR 888.3390 Regulation Name: Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis Regulatory Class: Class II Product Code: KWY Dated: September 28, 2010 Received: September 29, 2010

Dear Mr. Mrak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 – Mr. David Mrak

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Sincerely yours,

Mark McMullen

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K100761

Device Name: BioPro Bipolar Head

Indications for Use:

The BioPro Bipolar Head is intended for use in combination with a BioPro PSL femoral stem for primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:

  • Femoral neck and trochanteric fractures of the proximal femur; .
  • Osteonecrosis of the femoral head: .
  • Revision procedures where other devices or treatments for these . indications have failed.

Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Omitu for nxn
(Division Sign Off)

(Division Sign-G Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number K100761

§ 888.3390 Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis.

(a)
Identification. A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.(b)
Classification. Class II.