K Number
K100559
Date Cleared
2010-03-30

(29 days)

Product Code
Regulation Number
884.1690
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MyoSure™ Hysteroscopic Tissue Removal System is intended for hysteroscopic intrauterine procedures by a trained gynecologist to resect and remove tissue including submucous myomas and endometrial polyps.

Device Description

The modified MyoSure Hysteroscopic Tissue Removal System consists of the following procedural components which are identical to those found in the predicate MyoSure System:

  • MyoSure Control Unit O
  • MyoSure Tissue Removal Device o
  • MyoSure Foot Pedal o

The MyoSure Control Unit contains an electric motor and software controller that drives the MyoSure Tissue Removal Device. The Control Unit motor is activated and deactivated by the MyoSure Foot Pedal. The MyoSURE Tissue Removal Device is a tissue morcellator that is connected to the Control Unit via a flexible drive cable. The MyoSure Tissue Removal Device features a rotating/reciprocating (2.5 mm OD) cutter blade encased in a (3 mm OD) outer tube (i.e. morcellator). The morcellator's cutter-blade is controlled by a drive system that enables simultaneous rotation and reciprocation of the cutter. The cutter is also connected to a vacuum source which aspirates resected tissue through a side-facing cutting window in the device's outer tube. Distension fluid and resected tissue are transported from the MyoSure Tissue Removal Device to a tissue trap and vacuum canister via a tube protruding from the proximal end of the Tissue Removal Device. The MyoSure Hysteroscopic Tissue Removal System is compatible with commercially available fluid management systems and may be used with hysteroscopes that have a straight 3 mm working channel.

AI/ML Overview
  1. Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly list quantitative acceptance criteria in a table format. However, it indicates that the modified device's performance was evaluated against the predicate device. The conclusion states:

Acceptance Criteria CategoryReported Device Performance
Electrical SafetyMeets electrical safety standards.
EMC (Electromagnetic Compatibility)Meets EMC standards.
BiocompatibilityNew patient contact materials meet ISO 10993-1 requirements for Biological Evaluation of Medical Devices.
Performance (In-Vitro)Demonstrated to be equivalent to the predicate MyoSure device.

The study essentially aimed to prove equivalence to the predicate device (MyoSure Hysteroscopic Tissue Removal System, K091100) across these categories, rather than meeting specific numerical thresholds.

  1. Sample Size for Test Set and Data Provenance:

The document describes "in-vitro testing" for performance evaluation. It does not specify the sample size of the test set used for this in-vitro testing. Furthermore, there is no information regarding the data provenance (e.g., country of origin of the data, retrospective, or prospective), as this was an in-vitro study, not a clinical trial with human data.

  1. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This was an in-vitro study comparing a modified device to a predicate device. Ground truth, in the sense of expert human assessment of medical images or conditions, was not established for this type of test. The performance was likely measured against objective physical or mechanical parameters relevant to tissue removal.

  1. Adjudication Method for the Test Set:

Not applicable, as there were no human readers or expert panels involved in establishing ground truth for a test set in the traditional sense. The evaluation was based on in-vitro measurements and comparisons.

  1. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. The document describes an in-vitro study focused on demonstrating equivalence to a predicate device, not a comparative study involving human readers with and without AI assistance.

  1. Standalone (Algorithm Only) Performance:

Not applicable. The device described, the MyoSure™ Hysteroscopic Tissue Removal System, is a mechanical medical device for tissue removal, not an AI algorithm. Therefore, "standalone (algorithm only)" performance is not relevant to this submission. The performance assessment focused on the device's mechanical and material characteristics.

  1. Type of Ground Truth Used:

The "ground truth" for the in-vitro performance testing would have been objective measurements of physical parameters relevant to the device's function (e.g., cutting efficacy, tissue morcellation efficiency, fluid dynamics, material integrity under stress, etc.) compared against the performance of the predicate device under the same conditions. It was not based on expert consensus, pathology, or outcomes data in the context of human patients.

  1. Sample Size for the Training Set:

Not applicable. This device is a mechanical system, not an AI/machine learning model that requires a "training set."

  1. How the Ground Truth for the Training Set Was Established:

Not applicable, as there was no training set for an AI/machine learning model.

{0}------------------------------------------------

Interlace Medical Inc.

K100559

510(k) SUMMARY 5.

Submitter: 1.

Interlace Medical Inc. 135 Newbury St Framingham, MA 01701 Telephone: 508.875.1343, ext. 112 MAR 3 0 2010

Contact: John J. Vozella, VP Clinical & Regulatory Affairs Date Prepared: February 26, 2010

2. Device:

Trade Name: MyoSure™ Hysteroscopic Tissue Removal System Common Name: Hysteroscope and accessories Classification Name: Hysteroscope and accessories Class: II

Predicate Device: 3.

MyoSure™ Hysteroscopic Tissue Removal System (K091100)

Device Description: 4.

The modified MyoSure Hysteroscopic Tissue Removal System consists of the following procedural components which are identical to those found in the predicate MyoSure System:

  • MyoSure Control Unit O
  • MyoSure Tissue Removal Device o
  • MyoSure Foot Pedal o

The MyoSure Control Unit contains an electric motor and software controller that drives the MyoSure Tissue Removal Device. The Control Unit motor is activated and deactivated by the MyoSure Foot Pedal. The MyoSURE Tissue Removal Device is a tissue morcellator that is connected to the Control Unit via a flexible drive cable. The MyoSure Tissue Removal Device features a rotating/reciprocating (2.5 mm OD) cutter blade encased in a (3 mm OD) outer tube (i.e. morcellator). The morcellator's cutter-blade is controlled by a drive system that enables simultaneous rotation and reciprocation of the cutter. The cutter is also connected to a vacuum source which aspirates resected tissue through a side-facing cutting window in the device's outer tube. Distension fluid and resected tissue are transported from the MyoSure Tissue Removal Device to a tissue trap and vacuum canister via a tube protruding from the proximal end of the Tissue Removal Device. The MyoSure Hysteroscopic Tissue Removal System is compatible with commercially available fluid management systems and may be used with hysteroscopes that have a straight 3 mm working channel.

{1}------------------------------------------------

5. Intended Use:

The MyoSure™ Hysteroscopic Tissue Removal System is intended for hysteroscopic intrauterine procedures by a trained gynecologist to resect and remove tissue including submucous myomas and endometrial polyps.

Comparison of Characteristics: 6.

The modified MyoSure Hysteroscopic Tissue Removal System's intended use and indicated use are identical to that of the predicate MyoSure Hysteroscopic Tissue Removal System, K091100.

The principles of operation and primary functional specifications of the modified MyoSure Hysteroscopic Tissue Removal System are identical to those of the predicate MyoSure Hysteroscopic Tissue Removal System.

The primary functional components of the modified MyoSure Hysteroscopic Tissue Removal System are identical to those of the predicate MyoSure Hysteroscopic Tissue Removal System, i.e.:

  • each system employs a sterile, single use, Tissue Removal Device . (disposable straight surgical morcellator),
  • a Foot Pedal controls activation and deactivation of a motor in the . Control Unit which powers the morcellator in both systems,
  • the modified MyoSure cutter rotates and reciprocates at a fixed rate ● that is identical to the predicate device, and
  • the modified MyoSure cutter blade tip design is identical to that of the . predicate device, and

The modified MyoSure Hysteroscopic Tissue Removal System is different · from the predicate MyoSure Hysteroscopic Tissue Removal System as follows:

  • The modified MyoSure Control Unit's motor control software has been o changed to firmware and the controller component has been modified accordingly.
  • The maximum running current limit and current limit has been decreased o in the modified MyoSure device.
  • The modified MyoSure Tissue Removal Device morcellator's cutting tube O outer surface is now coated with a different material than was used in the predicate device.
  • An O ring in the predicate device's fluid path has been changed to a o different material in the modified MyoSure™ Hysteroscopic Tissue Removal System.

{2}------------------------------------------------

  • The modified MyoSure™ Hysteroscopic Tissue Removal System utilizes o a lubricant on the device's gear mechanism and O ring while the predicate MyoSure device had no gear lubrication.

Performance Testing: 7.

The modified MyoSure Hysteroscopic Tissue Removal System meets electrical safety and EMC standards. New patient contact materials in the modified MyoSure device meet the biocompatibility requirements of ISO 10993-1 Biological Evaluation of Medical Devices. In addition, in-vitro testing demonstrated that the modified MyoSure device's performance is equivalent to the predicate MyoSure device.

8. Conclusion:

Based on the intended use, descriptive information and performance evaluation provided in this submission, the modified MyoSure Hysteroscopic Tissue Removal System has been shown to be equivalent in technology, method of operation, functional performance and intended use to the predicate MyoSure Hysteroscopic Tissue Removal System.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. John Vozella V.P. Clinical and Regulatory Affairs Interlace Medical, Inc. 135 Newbury Street FRAMINGHAM MA 01701

MAR 3 0 2010

Re: K100559

Trade Name: MyoSure Hysteroscopic Tissue Removal System Regulation Number: 21 CFR §884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: February 26, 2010 Received: March 1, 2010

Dear Mr. Vozella:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device-Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Interlace Medical Inc.

4. INDICATIONS FOR USE STATEMENT

KI00559 510(k) Number (if known):

Device Name: MyoSURE™ Hysteroscopic Tissue Removal System

Indications For Use:

The MyoSURE™ Hysteroscopic Tissue Removal System is intended for hysteroscopic intrauterine procedures by a trained gynecologist to resect and remove tissue including submucous myomas and endometrial polyps.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Stan-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

CONFIDENTIAL

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.