K Number
K100556
Date Cleared
2010-08-02

(154 days)

Product Code
Regulation Number
890.5850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Flex Max by Stendertone is indicated for the improvement of abdominal muscle tone, for strengthening of the abdominal muscles and for the development of a firmer abdomen.

Device Description

Flex Max by Slendertone, Type 517-US is a two-channel battery operated muscle stimulation, system specifically designed to exercise the abdominal muscles. It offers 10 programs to the user and comprises of two main components, an electronic stimulator module which generates the required stimulation signals and an abdominal electrode belt, which connects the signals from the stimulator to the skin electrodes located on the inner surface of the belt. The product is supplied with a double set of double-sided adhesive electrodes, a double set of batteries, an extender belt and an instruction manual. Power is derived from 3 x 1.5 volt (LR03) batteries located in a compartment protected by a removable battery cover. Instructions are provided in the user instructions. For purposes of hygiene, the garment may be cleaned and instructions for device care are included in the user manual.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Flex Max by Slendertone, Type 517-US:

It's important to note that the provided documentation is a 510(k) Premarket Notification for a medical device. This type of submission focuses on demonstrating substantial equivalence to a previously legally marketed device (predicate device), rather than a full de novo clinical trial proving efficacy and safety from scratch.

Therefore, the "acceptance criteria" here are primarily about demonstrating that the new device is as safe and effective as the predicate device, not necessarily defining specific performance metrics for muscle tone improvement. The "study that proves the device meets the acceptance criteria" refers to the entire 510(k) submission process, which relies on demonstrating equivalence rather than a dedicated clinical trial for performance.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Demonstrated Equivalence to Predicate)Reported Device Performance (as demonstrated in submission)
Intended Use: Same intended use as predicate device"The Flex Max by Slendertone device has the same intended use and indications for use as the listed equivalent legally marketed device. It is intended for use by healthy persons to apply transcutaneous neuromuscular electrical stimulation ('NMES') through skin contact electrodes for the purpose of improving abdominal muscle tone, strength and firmness."
Indications for Use: Same indications as predicate"Flex Max by Stendertone is indicated for the improvement of abdominal muscle tone, for strengthening of the abdominal muscles and for the development of a firmer abdomen."
Technological Characteristics: Substantially equivalent to predicate"The Flex Max by Slendertone device is identical to the predicate device in design, materials, mode of operation, and methods of manufacture. It differs only in the addition of different program offerings to the user."
Safety: Compliance with relevant electrical safety and EMC standards"The Flex Max by Slendertone complies with international standards for electrical safety and electromagnetic compatibility."
Effectiveness: Inferred from equivalence to predicateImplied to be equivalent to the predicate device's effectiveness through the demonstration of identical design, materials, mode of operation, and methods of manufacture.

Explanation of "Acceptance Criteria" in this context: For a 510(k) submission, the primary "acceptance criterion" is to demonstrate substantial equivalence to a predicate device. This means showing that the new device has the same intended use, indications for use, and similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness. The device met this acceptance criterion, as indicated by the FDA's "substantial equivalence" determination.


2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not applicable in the context of a new clinical study. The submission primarily relies on the established safety and effectiveness of the predicate device (Slendertone Flex, Type 515, K030708). No new clinical "test set" data from human subjects is provided beyond proving the new device's equivalence to the predicate.
  • Data Provenance: Not applicable. No new human data is presented in this submission. The basis for equivalence is drawn from the predicate device’s prior clearance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not applicable. No "ground truth" was established for a new clinical test set in this submission.
  • Qualifications of Experts: Not applicable.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. No clinical test set requiring adjudication was performed as part of this 510(k) submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This is a muscle stimulator device, not an imaging/diagnostic device that would typically involve human "readers" or AI assistance in that sense.
  • Effect Size of Human Readers with/without AI: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. This device is a muscle stimulator, not an algorithm. Its operation inherently involves a human user applying it.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: The "ground truth" in this context is the FDA's prior determination of safety and effectiveness for the predicate device. The submission's "proof" relies on demonstrating that the new device is functionally identical or equivalent to this already cleared predicate device. There is no new clinical outcomes data or pathology presented for the Flex Max.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable.

Summary of the Study:

The "study" described in the provided document is a 510(k) Premarket Notification submission, not a clinical trial. The manufacturer's strategy was to demonstrate substantial equivalence to a previously cleared predicate device (Slendertone Flex, Type 515, K030708).

  • Key Argument: The device is "identical to the predicate device in design, materials, mode of operation, and methods of manufacture," differing only in "the addition of different program offerings to the user."
  • Clinical Testing: "No additional clinical studies were provided as part of this submission." This indicates that the manufacturer relied entirely on the established safety and effectiveness of the predicate device and the similarity of the new device to the predicate, along with compliance with international electrical safety and electromagnetic compatibility standards.
  • Ground Truth: The "ground truth" for the overall acceptance of the device is rooted in the FDA's previous clearance of the predicate device. By demonstrating substantial equivalence, the new device is deemed to carry the same safety and effectiveness profile as the predicate.

In essence, the "study" for this 510(k) submission primarily involved a technical comparison and regulatory compliance assessment to demonstrate that the Flex Max by Slendertone is sufficiently similar to an already approved device, making new clinical performance data unnecessary for its market clearance.

{0}------------------------------------------------

7AUG 02

IL100554

This summary is being submitted in accordance with the requirements of 21 CFR 807.92.

1. Contact Details

Name:Deirdre Barrow
Address:Bio-Medical Research Ltd.,Parkmore Business Park, WestGalway, Ireland
Telephone:+353 91 774300
Fax:+353 91 774301
E-Mail:dbarrow@bmr.ie
Alternate Contact:Michael O'Sullivan
Address:Bio-Medical Research Ltd.,Parkmore Business Park, WestGalway, Ireland
Telephone:+353 91 774300
Fax:+353 91 774301

Prepared:

.

25th May 2010

2. Device Name

Trade Name of Device:Flex Max by Slendertone, Type 517-US
Common Name:Muscle Stimulator
Classification Name:Stimulator, muscle, powered, for muscle conditioning
Regulation Number:21 CFR 890.5850
Product Code:NGX

3. Identification of Equivalent Legally Marketed Device

Name:Slendertone Flex, Type 515
Manufacturer:Bio-Medical Research Ltd.
510(k) No:K030708, June 2003

{1}------------------------------------------------

4. Description of Device

Flex Max by Slendertone, Type 517-US is a two-channel battery operated muscle stimulation, system specifically designed to exercise the abdominal muscles. It offers 10 programs to the user and comprises of two main components, an electronic stimulator module which generates the required stimulation signals and an abdominal electrode belt, which connects the signals from the stimulator to the skin electrodes located on the inner surface of the belt.

The product is supplied with a double set of double-sided adhesive electrodes, a double set of batteries, an extender belt and an instruction manual. Power is derived from 3 x 1.5 volt (LR03) batteries located in a compartment protected by a removable battery cover. Instructions are provided in the user instructions.

For purposes of hygiene, the garment may be cleaned and instructions for device care are included in the user manual.

5. Statement of Intended Use & Indications for Use

The Flex Max by Slendertone device has the same intended use and indications for use as the listed equivalent legally marketed device. It is intended for use by healthy persons to apply transcutaneous neuromuscular electrical stimulation ("NMES") through skin contact electrodes for the purpose of improving abdominal muscle tone, strength and firmness.

6. Technological Characteristics

The Flex Max by Slendertone device is identical to the predicate device in design, materials, mode of operation, and methods of manufacture. It differs only in the addition of different program offerings to the user.

7. Clinical and Non-Clinical Tests

The same technological platform and modus operandi is same as the predicate device, Slendertone Flex, Type 515. No additional clinical studies were provided as part of this The Flex Max by Slendertone complies with international standards for submission. electrical safety and electromagnetic compatibility.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Bio-Medical Research, Ltd. % Ms. Deirdre Barrow BMR House, Parkmore Business Park West Galway, Ireland

Re: K100556

Trade Name: Flex Max by Slendertone, Type 517-US Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: May 25, 2010 Received: June 4, 2010

AUG 02-2010

Dear Ms. Barrow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Deirdre Barrow

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A. Melkersen

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

AUG 02 2010

Indications for Use

K100556

510(k) Number (if known):

Device Name:

Flex Max by Slendertone, Type 517-US

Indications for Use:

Flex Max by Stendertone is indicated for the improvement of abdominal muscle tone, for strengthening of the abdominal muscles and for the development of a firmer abdomen.

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter. Use AND/OR (21 CFR 801 Subpart C) X

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Yurman W. Clary

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100556

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).