AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NDSsi ZeroWire Duo Wireless HD Video Transfer System is a paired transmitter and receiver, intended for delivery of video signals over a radio-frequency link to a video display during endoscopic and general surgical procedures. The ZeroWire wireless device is a non-sterile reusable device not intended for use in the sterile field. It is intended for use by qualified physicians having complete knowledge of these surgical procedures.

Device Description

The ZeroWire® wireless device is a wireless transmitter and receiver pair which allows delivery of a video signal over a radio frequency link to a video display such as Radiance and EndoVue LCD surgical monitors.

AI/ML Overview

The provided text is a 510(k) summary for the NDSsi ZeroWire Duo Wireless HD Video Transfer System. It states that "Clinical data is not needed for this type of wireless device 510(k) submission." Therefore, it does not include a study proving the device meets specific acceptance criteria based on clinical performance metrics such as sensitivity, specificity, accuracy, or reader performance.

Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (Stryker Vision Elect Wireless High Definition Television / K081995) through non-clinical tests and surgical-setting tests. The "acceptance criteria" can be inferred from the claims of equivalence and the characteristics described for the device, rather than explicit numerical performance targets.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Since clinical performance metrics (like sensitivity, specificity, or accuracy) are not provided, the "acceptance criteria" here relate to the functional and safety equivalence to the predicate device, and the "reported device performance" is a statement of that equivalence.

Acceptance Criteria (Implied)Reported Device Performance
Delivery of video signal over radio-frequency link"paired transmitter and receiver, intended for delivery of video signals over a radio-frequency link to a video display"
Compatible with video displays (e.g., Radiance and EndoVue LCD)"allows delivery of a video signal... to a video display such as Radiance and EndoVue LCD surgical monitors."
Full high-definition video"Delivers full high-definition video"
Low latency"less than 1 frame latency"
Resistant to interference"By utilizing the reserved UWB frequency spectrum, ZeroWire provides a wireless video link that is resistant to interference from other devices."
Non-clinical test conformity (medical safety, EMC)"Both systems conform to non-clinical tests including the medical safety and EMC standards defined by EN 60601-1 and EN 60601-1-2."
FCC Part 15 standards conformity (for wireless devices)"As wireless devices, both systems conform to relevant FCC Part 15 standards."
Data security (industry-standard encryption)"Both systems provide video data security by employing industry-standard data encryption techniques."
Performance, safety, and effectiveness equivalent to predicate"Based on non-clinical tests as well as surgical-setting tests, the ZeroWire wireless device provides performance, safety, and effectiveness that is equivalent to the predicate Stryker Vision Elect Wireless High Definition Television (Stryker VE VHDTV) / K081995." (Conclusion derived from "surgical-setting test results")
Not intended for sterile field / reusable / non-sterile"The Zero Wire wireless device is a non-sterile reusable device not intended for use in the sterile field."
Used by qualified physicians with knowledge of surgical procedures"It is intended for use by qualified physicians having complete knowledge of these surgical procedures."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The document mentions "surgical-setting test results" but does not detail the number of cases, participants, or specific parameters of these tests.
  • Data Provenance: Not explicitly stated. The document describes "surgical-setting test results" but does not specify the country of origin or whether the data was retrospective or prospective. Given the nature of a 510(k) summary focused on substantial equivalence through non-clinical and "surgical-setting tests" (likely functional testing in a surgical environment), it's improbable to be large-scale clinical trial data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • This information is not provided as the submission states, "Clinical data is not needed for this type of wireless device 510(k) submission." Therefore, there's no mention of experts establishing ground truth in the context of diagnostic accuracy or similar clinical outcomes. The "ground truth" for the non-clinical and surgical-setting tests would likely be objective measurements of video signal quality, latency, interference resistance, and adherence to technical standards.

4. Adjudication Method for the Test Set

  • This information is not provided for the same reasons as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states, "Clinical data is not needed for this type of wireless device 510(k) submission." The comparison is primarily technical and functional equivalence to a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • The device is a hardware system (transmitter and receiver for video transfer), not an algorithm that produces interpretations or diagnoses. Therefore, the concept of "standalone performance" in the context of an algorithm's diagnostic accuracy is not applicable to this device. The "standalone" performance here would refer to its ability to transmit video signals effectively, which is covered by the non-clinical and surgical-setting tests.

7. The Type of Ground Truth Used

  • As clinical data for diagnostic purposes was not required, the "ground truth" used would be based on technical specifications, objective measurements, and compliance with established standards (e.g., video resolution, latency measurements, signal strength, interference levels, adherence to EN 60601-1, EN 60601-1-2, and FCC Part 15).

8. The Sample Size for the Training Set

  • This device is a hardware system; it does not involve a "training set" in the context of machine learning or AI algorithms. The system's design and engineering are based on established electronics and wireless communication principles.

9. How the Ground Truth for the Training Set Was Established

  • As there is no training set for an AI/ML algorithm, this question is not applicable.

{0}------------------------------------------------

K.100195.

Page 1 of 2

510(k) SUMMARY

APR 2 1 2010

A. Manufacturer: NDS Surgical Imaging, LLC 5750 Heilyer Avenue San Jose, CA 95138 USA

B. Submitted By: Jim Leng Regulatory Engineer NDS Surgical Imaging, LLC

C. Date of Preparation: November 2, 2009 April 15, 2010 revision

Tel: 408-776-0085 408-776-9878 Fax:

E. Classification: Endoscope and Accessories

GCJ

F. Common Name: UWB Wireless Device

ZeroWire Duo Wireless HD Video Transfer System

H. Classification Number: 21 CFR 876.1500

  • I. Product Code:
    J. Substantial Equivalence:

D. Contact Information:

G. Proprietary Name:

Stryker Vision Elect Wireless High Definition Television (Stryker VE VHDTV) / K081995 Both the Stryker device and the NDS device contain a wireless, radio frequency transmitter that receives a digital video signal from a surgical camera. Both systems incorporate a wireless, radio frequency receiver that is exclusively linked to but only to the aforementioned transmitter. The wireless receiver of both systems delivers the received video signal to a suitable surgical monitor for use by qualified medical professionals. Both systems conform to non-clinical tests including the medical safety and EMC standards defined by EN 60601-1 and EN 60601-1-2. Both systems require a single transmitter to be linked to an individual receiver. As wireless devices, both systems conform to relevant FCC Part 15 standards. Both systems provide video data security by employing industry-standard data encryption techniques.

® RADIANCE

Image /page/0/Picture/18 description: The image shows the word "DOME" in bold, black letters. Above the word is a black semi-circle, which appears to be the top half of a dome shape. There is a horizontal line below the word "DOME". The background is white.

{1}------------------------------------------------

K100195

Page 2 of 2

The singular difference between the Stryker system and the NDS system is that in the Stryker system the receiver is embedded in the monitor, whereas in the NDS system, the receiver is a separate item which can be electrically connected by means of an appropriate cable to a suitable surgical monitor.

The ZeroWire® wireless device is a wireless transmitter K. Device Description: and receiver pair which allows delivery of a video signal over a radio frequency link to a video display such as Radiance and EndoVue LCD surgical monitors.

The NDSsi ZeroWire Duo Wireless HD Video Transfer System L. Intended Use: is a paired transmitter and receiver, intended for delivery of video signals over a radio-frequency link to a video display during endoscopic and general surgical procedures. The Zero Wire wireless device is a non-sterile reusable device not intended for use in the sterile field. It is intended for use by qualified physicians having complete knowledge of these surgical procedures.

M. Technological Characteristics: The ZeroWire Duo wireless device is a wireless video transfer solution for minimally invasive surgery and interventional procedures. It delivers full high-definition video with less than 1 frame latency. By utilizing the reserved UWB frequency spectrum, ZeroWire provides a wireless video link that is resistant to interference from other devices. ZeroWire improves clinical efficiency and safety in the OR by eliminating the need for video cables their cleaning, frequent replacement, and potential to be a tripping hazard. The proprietary memory-enabled pairing system makes installation quick and easy.

N. Clinical information: Clinical data is not needed for this type of wireless device 510(k) submission. However, this submission includes surgical-setting test results to demonstrate the equivalence of the ZeroWire device to the predicate Stryker device.

Based on non-clinical tests as well as surgical-setting tests, O. Conclusion: the ZeroWire wireless device provides performance, safety, and effectiveness that is equivalent to the predicate Stryker Vision Elect Wireless High Definition Television (Stryker VE VHDTV) / K081995.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR 21 2010

NDS Surgical Imaging % TUV Rheinland of North America, Inc. Tamas Borsai Manager, Medical Division 12 Commerce Road Newton, Connecticut 06470

Re: K100195

Trade/Device Name: ZeroWire Duo Wireless HD Video Transfer System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: April 02, 2010 Received: April 06, 2010

Dear Tamas Borsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Tamas Borsai

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

.fda.gov/MedicalDevices/ResourcestorYou/Industry/default.htm

Sincerely yours,

Mark A. Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for NDS Surgical Imaging. The logo consists of a grid of squares on the left, followed by the letters NDS in large, bold font. Below the letters is the text "SURGICAL · IMAGING" in a smaller font. The logo is black and white.

i00195

Indications for Use Statement

510 (k) Number (if Known):

Device Name: ZeroWire Duo Wireless HD Video Transfer System

Indications for Use:

The NDSsi ZeroWire Duo Wireless HD Video Transfer System is a paired transmitter and receiver, intended for delivery of video signals over a radio-frequency link to a video display during endoscopic and general surgical procedures. The ZeroWire wireless device is a non-sterile reusable device not intended for use in the sterile field. It is intended for use by qualified physicians having complete knowledge of these surgical procedures.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIE LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Phil R. Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100195

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.