(459 days)
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.
The Siemens IMMULITE® 2000 3gAllergy™ Specific IgE Assay is an in vitro diagnostic device for the quantitative measurement of allergen-specific IgE in human serum, assisting in the clinical diagnosis of IgE-mediated allergic disorders. The device met several acceptance criteria to demonstrate its performance.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for this diagnostic device are primarily defined by its precision, linearity, and specificity, with clinical performance (sensitivity, specificity, and agreement) validating its utility. The reported performance for these criteria is summarized below:
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Precision | Low Coefficients of Variation (CV%) for both within-run and total precision across different allergen concentrations. While no explicit thresholds are given, typical good laboratory practice would look for CVs generally below ~10-15%, especially for higher concentrations. Lower CVs indicate higher reproducibility. Some results (e.g., Parrot Feathers Positive #3 with 47.01% Total CV) suggest that higher variability is accepted at very low concentrations or for certain allergens, but overall, the majority of results show good precision. | Within-Run CV% Ranges:- Most allergens show within-run CVs below 10%. Some exceptions exist, e.g., Blomia tropicalis (7.69%), Loblolly Pine (5.86%), nDer p 2 (11.80%), Parrot Feathers (39.87% for one sample, indicating a potentially very low concentration or outlier).Total CV% Ranges:- Most allergens show total CVs below 10-15%. Similar to within-run, exceptions exist, e.g., Chicken Feathers (12.95%), nDer p 2 (12.60%), Parrot Feathers (47.01% for one sample, and 12.64% for another). Overall, the precision is considered acceptable, especially for clinically relevant concentrations, with acknowledgment of higher variability at very low concentrations as is common for many immunoassays. |
| Linearity | Regression analysis of observed vs. expected concentrations should yield slopes close to 1.0 and intercepts close to 0, with narrow 95% confidence intervals (CI) for both the slope and intercept, indicating accurate measurement across the assay's dynamic range. | Slope 95% CI: All allergens show a 95% CI for the slope that includes or is very close to 1.0 (e.g., Aureobasidium pullulans: 0.968 - 1.025, Blomia tropicalis: 0.967 - 1.008, Brazilian Peppertree: 0.939 - 1.024). Intercept 95% CI: All allergens show a 95% CI for the intercept that includes or is very close to 0 (e.g., Aureobasidium pullulans: -0.089 - 0.029, Blomia tropicalis: -0.063 - 0.119, Brazilian Peppertree: 0.070 - 0.360). These results demonstrate good linearity across the tested concentration ranges for all allergens. |
| Specificity | Competitive inhibition studies should show concentration-dependent inhibition by relevant extracts, with a target of 50% inhibition at the highest inhibitor concentration. Non-specific cross-reactivity with unrelated allergens should be minimized (e.g., below 7%). | Competitive Inhibition: The inhibition plots demonstrated that the allergens tested were inhibited by the relevant inhibitor extract in a concentration-dependent fashion. The target of 50% inhibition for the highest inhibitor concentration was met for all listed allergens. Inhibition Using Negative Controls (Cross-Reactivity): Results for most specific allergens when tested with unrelated allergen extracts were below 7%, except for Loblolly Pine. This indicates low cross-reactivity generally, supporting the specificity of the assays. |
| Clinical Performance (Agreement) | High overall agreement (and associated sensitivity and specificity) between IMMULITE® 2000 results and clinical data, indicating that the assay results align with actual patient allergic status. While no specific threshold is stated in the context, a high percentage is expected. | Overall Agreement: 90.5% (Lower Conf 89%, Upper Conf 92%). Sensitivity: 71.4% (Lower Conf 68%, Upper Conf 74%). Specificity: 98.3% (Lower Conf 98%, Upper Conf 99%). The high overall agreement, driven primarily by very high specificity, supports the utility of the assay as an aid in diagnosis, especially in confirming the absence of IgE-mediated allergy. The moderate sensitivity suggests it is a good indicator of allergy when positive, but a negative result needs to be interpreted carefully in conjunction with clinical findings. |
2. Sample Size and Data Provenance
- Test Set (Clinical Performance Study): The clinical performance study used a total of 3,140 serum samples.
- Data Provenance: The document does not explicitly state the country of origin. It is implied to be retrospective, using "clinically diagnosed atopic and non-atopic individuals," suggesting samples were collected from a clinical setting where diagnoses had already been made.
3. Number of Experts and Qualifications
- Role of Experts: The document does not describe the use of human experts in establishing the ground truth for the clinical test set in the way one might for an imaging AI device.
- Ground Truth Establishment: Clinical performance was evaluated by comparing IMMULITE® 2000 results "to clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." This implies a clinical diagnosis based on a combination of patient history, physical examination, and potentially other diagnostic tests performed by clinicians (e.g., allergists), rather than a specific panel of experts adjudicating each case.
4. Adjudication Method
- Not applicable in the context of this in vitro diagnostic assay. Adjudication methods like 2+1 or 3+1 are typically used for consensus reading of medical images by multiple human readers, not for establishing ground truth for laboratory tests based on clinical documentation.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not applicable. This is an in vitro diagnostic assay, not a medical imaging AI device designed to assist human readers. Therefore, an MRMC study and analysis of human reader improvement with AI assistance were not performed.
6. Standalone Performance
- Yes, the performance metrics (precision, linearity, specificity, and the clinical performance data for sensitivity, specificity, and agreement) represent the standalone performance of the IMMULITE® 2000 3gAllergy™ assay. This device is an automated immunoassay system, and its output (quantitative IgE levels) is the direct result of the algorithm/reagent system. There is no "human-in-the-loop" assistance being evaluated in the assay's core operational performance; the results are generated by the instrument.
7. Type of Ground Truth Used
- The ground truth for the clinical performance study was clinical documentation of presence or absence of signs, symptoms, and other diagnostic evidence of allergen sensitivity. This is akin to outcomes data or a composite clinical diagnosis rather than a single definitive gold standard like pathology.
8. Sample Size for the Training Set
- The document does not explicitly state the sample size used for a "training set" in the context of algorithm development. As a traditional immunoassay, rather than a machine learning/AI algorithm in the modern sense, the "training" would refer to the development and optimization of the assay reagents and parameters. This is not detailed in terms of a specific sample size used for that purpose. The studies presented (precision, linearity, specificity) are validation studies on the final product.
9. How the Ground Truth for the Training Set Was Established
- Since this is a traditional immunoassay rather than a machine learning model, the concept of establishing "ground truth for a training set" as it pertains to modern AI models is not directly applicable. The development of such assays involves iterative optimization based on known calibrators, controls, and characterized patient samples to ensure accurate and reliable measurement of the analyte (allergen-specific IgE). The documented studies (precision, linearity, specificity) evaluate the performance of the finalized assay itself using controlled samples and clinical samples, confirming its analytical and clinical validity.
{0}------------------------------------------------
SUMMARY OF SAFETY AND EFFECTIVENESS
Assigned 510(k) Number
The assigned 510(k) number is K093987
Sponsor Name and Address
Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles, CA 90045-6900 (310) 645-8200
MAR 2 8 2011
Contact
Donna Velasquez Regulatory Technical Specialist (310) 645-8200 x7403 (310) 645-9999 fax Donna.velasquez@siemens.com
Device Name
IMMULITE® 2000 3gAllergy™ Specific IgE Assay Trade name: Classification: Class II Classification Names: Radioallergosorbent (RAST) Immunological Test System Regulation Number: 866.5750 Product Code: DHB Catalog Numbers: L2KUN6 (600 tests)
Description of Device
IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format.12 (U.S. Patent No. 4.778.751) It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk.
The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.
Incubation Cycles: 2 × 30 minutes.
1 El Shami AS, Alaba O. Liquid-phase in vitro allergen-specific IgE assay with in situ immobilization. Adv Biosci 1989;74:191-201.
2 Alaba O. El Shami AS. Evaluation of non-specific IgE binding; comparison of two in vitro allergen assays. Adv Biosci 1989;74:203-14.
{1}------------------------------------------------
Indications for Use
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
Establishment Information
IMMULITE® 2000 3gAllergy Specific IgE assay is manufactured by Siemens Healthcare Diagnostics Inc. at the following locations:
Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles, CA 90045-6900 FDA Establishment #: 3005250747
・・
l
Predicate
The purpose of this 510(k) submission is for clearance of twenty one additional specific allergens, named in the table below, to be used with the IMMULITE® 2000 3gAllergy™ Specific IgE on the IMMULITE® 2000 analyzer.
| 1 | M12 – Aureobasidiumpullulans | 12 | A310 - nDer p 1 |
|---|---|---|---|
| 2 | D201 – Blomia tropicalis | 13 | A316 - nDer p 2 |
| 3 | T401 - Brazilian Peppertree | 14 | A345 - nFel d 1 |
| 4 | E78 - Budgerigar Feathers | 15 | F351 - nPen m 1 |
| 5 | E201 - Canary Feathers | 16 | E91 - Parrot Feathers |
| 6 | E85 - Chicken Feathers | 17 | M305 - Penicillium brevi-compactum |
| 7 | T43 - Loblolly Pine | 18 | E7 - Pigeon Droppings |
| 8 | A3050 - nAsp r 1 | 19 | W36 - Rabbit Bush |
| 9 | A174 - nCan f 1 | 20 | O201 - Tobacco |
| 10 | A295 - nDer f1 | 21 | E89 - Turkey Feathers |
| 11 | A302 - nDer f 2 |
FDA clearance was previously obtained for the assay kit and 196 specific allergens and allergen panels (K013134, K021206, K013135 and K021208).
Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corporation which was acquired by Siemens Medical Solutions Diagnostics in July 2006.
ﺮ
{2}------------------------------------------------
Siemens Medical Solutions Diagnostics was renamed Siemens Healthcare Diagnostics Inc. on January 1, 2008.
Precision
Precision studies were performed in accordance with Clinical Laboratory Standard Institute (CLSI) guidance: Evaluation of Precision Performance of Quantitative Methods; Approved Guideline-Second Edition. CLSI document EP5-A2 (ISBN 1-56238-542-9). CLSI. 940 West Valley Road. Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2004, assaying two aliquots of each test sample in two runs per day on 20 different days (Positives 1-3). Additional studies for Positives 4 were done assaying two aliquots of each test sample in two runs per day on 10 different days. Analysis of variance was used to estimate the within-run and total precision.
Three allergen lots were tested using three positive samples and one negative sample. Intraassay and inter-assay precision for the positive samples were evaluated by calculating the kU/L dose percent coefficients of variation (%CV) for each positive sample. Non-specific binding (NSB) was monitored by testing the negative control sample.
3
Representative precision claims for each allergen tested are presented below:
{3}------------------------------------------------
| Sample | MeankU/L | Within-Run | Total | |||
|---|---|---|---|---|---|---|
| SDkU/L | CV% | SDkU/L | CV% | |||
| Allergen = Aureobasidium pullulans, Lot 115 | ||||||
| Positive #1 | 0.72 | 0.027 | 3.75 | 0.041 | 5.69 | |
| Positive #2 | 0.96 | 0.037 | 3.85 | 0.041 | 4.27 | |
| Positive #3 | 12.19 | 0.529 | 4.34 | 0.788 | 6.46 | |
| Positive #4 | 0.54 | 0.072 | 13.28 | 0.081 | 14.90 | |
| Allergen = Blomia tropicalis, Lot 118 | ||||||
| Positive #1 | 36.21 | 1.403 | 3.87 | 1.645 | 4.54 | |
| Positive #2 | 1.73 | 0.133 | 7.69 | 0.169 | 9.77 | |
| Positive #3 | 2.55 | 0.079 | 3.10 | 0.121 | 4.75 | |
| Positive #4 | 0.42 | 0.023 | 5.36 | 0.031 | 7.20 | |
| Allergen = Brazilian Peppertree, Lot 115 | ||||||
| Positive #1 | 6.03 | 0.179 | 2.97 | 0.282 | 4.68 | |
| Positive #2 | 53.50 | 1.456 | 2.72 | 1.922 | 3.59 | |
| Positive #3 | 1.52 | 0.070 | 4.61 | 0.079 | 5.20 | |
| Allergen = Budgerigar Feathers, Lot 117 | ||||||
| Positive #1 | 0.45 | 0.026 | 5.78 | 0.029 | 6.44 | |
| Positive #2 | 1.35 | 0.070 | 5.19 | 0.095 | 7.04 | |
| Positive #3 | 1.89 | 0.094 | 4.97 | 0.159 | 8.41 | |
| Positive #4 | 4.34 | 0.144 | 3.32 | 0.165 | 3.81 | |
| Allergen = Canary Feathers, Lot 115 | ||||||
| Positive #1 | 0.74 | 0.028 | 3.78 | 0.038 | 5.14 | |
| Positive #2 | 0.99 | 0.033 | 3.33 | 0.048 | 4.85 | |
| Positive #3 | 6.80 | 0.271 | 3.99 | 0.312 | 4.59 | |
| Positive #4 | 0.41 | 0.05 | 3.76 | 0.027 | 6.71 | |
| Allergen = Chicken Feathers, Lot 117 | ||||||
| Positive #1 | 20.40 | 0.771 | 3.78 | 1.778 | 8.72 | |
| Positive #2 | 4.98 | 0.186 | 3.73 | 0.645 | 12.95 | |
| Positive #3 | 1.00 | 0.026 | 2.60 | 0.071 | 7.10 | |
| Allergen = Loblolly Pine, Lot 111 | ||||||
| Positive #1 | 1.05 | 0.036 | 3.43 | 0.052 | 4.95 | |
| Positive #2 | 0.87 | 0.051 | 5.86 | 0.055 | 6.32 | |
| Positive #3 | 6.00 | 0.295 | 4.92 | 0.381 | 6.35 | |
| Positive #4 | 0.31 | 0.016 | 5.25 | 0.021 | 6.71 | |
| Allergen = nAsp r 1, Lot 111 | ||||||
| Positive #1 | 1.78 | 0.082 | 4.61 | 0.111 | 6.24 | |
| Positive #2 | 62.51 | 2.541 | 4.06 | 3.704 | 5.93 | |
| Positive #3 | 0.46 | 0.026 | 5.65 | 0.033 | 7.17 | |
| Allergen = nCan f 1, Lot 112 | ||||||
| Positive #1 | 3.29 | 0.122 | 3.71 | 0.178 | 5.41 | |
| Positive #2 | 9.03 | 0.386 | 4.27 | 0.503 | 5.57 | |
| Positive #3 | 21.90 | 0.796 | 3.63 | 0.887 | 4.05 | |
| Allergen = nDer f 1, Lot 111 | ||||||
| Positive #1 | 3.22 | 0.142 | 4.41 | 0.179 | 5.56 | |
| Positive #2 | 9.16 | 0.394 | 4.30 | 0.486 | 5.31 | |
| Positive #3 | 65.82 | 2.478 | 3.76 | 3.046 | 4.63 | |
| Allergen = nDer f 2, Lot 111 | ||||||
| Positive #1 | 71.90 | 2.835 | 3.94 | 3.325 | 4.62 | |
| Positive #2 | 12.63 | 0.666 | 5.27 | 0.941 | 7.45 | |
| Positive #3 | 13.61 | 0.646 | 4.75 | 0.800 | 5.88 | |
| Positive #4 | 0.54 | 0.028 | 5.44 | 0.039 | 7.36 | |
| Allergen = nDer p 1, Lot 112 | ||||||
| Positive #1 | 50.96 | 2.143 | 4.21 | 2.450 | 4.81 | |
| Positive #2 | 7.38 | 0.328 | 4.44 | 0.427 | 5.79 | |
| Positive #3 | 38.26 | 1.791 | 4.68 | 2.087 | 5.45 | |
| Positive #4 | 0.45 | 0.018 | 4.11 | 0.027 | 5.96 | |
| Allergen = nDer p 2, Lot 111 | ||||||
| Positive #1 | 29.76 | 1.319 | 4.43 | 1.887 | 6.34 | |
| Positive #2 | 13.79 | 0.698 | 5.06 | 0.922 | 6.69 | |
| Positive #3 | 9.55 | 1.127 | 11.80 | 1.203 | 12.60 | |
| Positive #4 | 0.34 | 0.021 | 6.35 | 0.029 | 8.62 | |
| Allergen = nFel d 1, Lot 111 | ||||||
| Positive #1 | 5.08 | 0.249 | 4.90 | 0.353 | 6.95 | |
| Positive #2 | 20.83 | 1.016 | 4.88 | 1.171 | 5.62 | |
| Positive #3 | 1.75 | 0.071 | 4.06 | 0.093 | 5.31 | |
| Positive #1 | 3.21 | 0.146 | 4.55 | 5.26 | 5.26 | |
| Positive #2 | 2.25 | 0.094 | 4.18 | 6.71 | 6.76 | |
| Positive #3 | 30.79 | 1.409 | 4.58 | 5.14 | 5.26 | |
| Positive #4 | 0.42 | 0.020 | 4.64 | 0.025 | 5.88 | |
| Allergen = Parrot Feathers, Lot 116 | ||||||
| Positive #1 | 1.14 | 0.056 | 4.91 | 0.073 | 6.40 | |
| Positive #2 | 0.32 | 0.020 | 6.25 | 0.025 | 7.81 | |
| Positive #3 | 0.77 | 0.307 | 39.87 | 0.362 | 47.01 | |
| Positive #3 | 0.72 | 0.038 | 5.28 | 0.091 | 12.64 | |
| Positive #4 | 10.74 | 0.461 | 4.29 | 0.025 | 5.88 | |
| Allergen = Penicillium brevi-compactum, Lot 114 | ||||||
| Positive #1 | 1.87 | 0.109 | 5.83 | 0.127 | 6.79 | |
| Positive #2 | 1.37 | 0.053 | 3.87 | 0.070 | 5.11 | |
| Positive #3 | 4.87 | 0.190 | 3.90 | 0.226 | 4.64 | |
| Positive #4 | 0.43 | 0.013 | 3.05 | 0.018 | 4.15 | |
| Allergen = Pigeon Droppings, Lot 120 | ||||||
| Positive #1 | 0.48 | 0.033 | 6.88 | 0.042 | 8.75 | |
| Positive #2 | 6.22 | 0.401 | 6.45 | 0.497 | 7.99 | |
| Positive #3 | 5.33 | 0.232 | 4.35 | 0.261 | 4.90 | |
| Positive #4 | 2.27 | 0.111 | 4.88 | 0.178 | 7.85 | |
| Allergen = Rabbit Bush, Lot 111 | ||||||
| Positive #1 | 2.48 | 0.098 | 3.95 | 0.142 | 5.73 | |
| Positive #2 | 3.61 | 0.124 | 3.43 | 0.205 | 5.68 | |
| Positive #3 | 52.20 | 1.571 | 3.01 | 2.321 | 4.45 | |
| Positive #4 | 0.46 | 0.017 | 3.69 | 0.029 | 6.26 | |
| Allergen = Tobacco, Lot 111 | ||||||
| Positive #1 | 4.83 | 0.165 | 3.42 | 0.313 | 6.48 | |
| Positive #2 | 2.34 | 0.074 | 3.16 | 0.084 | 3.59 | |
| Positive #3 | 0.48 | 0.016 | 3.33 | 0.023 | 4.79 | |
| Allergen = Turkey Feathers, Lot 116 | ||||||
| Positive #1 | 0.54 | 0.021 | 3.89 | 0.027 | 5.00 | |
| Positive #2 | 0.75 | 0.027 | 3.60 | 0.035 | 4.67 | |
| Positive #3 | 30.23 | 1.004 | 3.32 | 1.345 | 4.45 |
、
.
:
.
ト!
:
{4}------------------------------------------------
: :
5
{5}------------------------------------------------
।
:
{6}------------------------------------------------
Linearity
For each allergen, two samples were diluted in 2-fold serial dilutions to 5 levels. The undiluted (neat) and diluted samples were tested with the specific allergen to demonstrate linearity at concentrations within the assay limits. Regression statistics for each allergen comparing observed to expected data are presented below.
| Allergen | Regression Equation | N | Slope95% CI | Intercept95% CI |
|---|---|---|---|---|
| M12 –Aureobasidium.Pullulans | Y = 0.9964x -0.0296 | 9 | 0.968 - 1.025 | -0.089 - 0.029 |
| D201 – Blomiatropicalis | Y = 0.9877x +0.0284 | 12 | 0.967 - 1.008 | -0.063 - 0.119 |
| T401 – BrazilianPeppertree | Y= 0.9818x +0.215 | 12 | 0.939 - 1.024 | 0.070 - 0.360 |
| E78 – BudgerigarFeathers | Y=0.939x +0.224 | 10 | 0.85 - 1.03 | 0.102 - 0.346 |
| E201 – CanaryFeathers | Y = 1.012x+0.2368 | 11 | 0.907 - 1.117 | -0.049 - 0.522 |
| E85 – ChickenFeathers | Y = 1.012x +0.2814 | 12 | 0.964 - 1.059 | -0.079 - 0.642 |
| T43 – LoblollyPine | Y = 1.009x -0.1098 | 12 | 0.994 - 1.023 | -0.177 - 0.043 |
| A3050 – nAsp r 1 | Y = 1.005x-0.0923 | 10 | 0.989 - 1.021 | -0.299 - 0.114 |
| A174 – nCan f 1 | Y=0.993x +0.0659 | 12 | 0.985 - 1.001 | 0.037 - 0.095 |
| A295 – nDer f 1 | Y = 1.002x +0.1762 | 12 | 0.979 - 1.024 | -0.018 - 0.370 |
| A302 – nDer f 2 | Y = 0.994x +0.0236 | 12 | 0.982 - 1.006 | -0.038 - 0.085 |
| A310 – nDer p 1 | Y = 0.996x -0.0213 | 12 | 0.978 - 1.104 | -0.122 - 0.079 |
| A316 – nDer p 2 | Y = 0.997x +0.0978 | 12 | 0.987 - 1.007 | 0.042 - 0.154 |
| A345 – nFel d 1 | Y = 0.997x +0.0484 | 12 | 0.964 - 1.030 | -0.203 - 0.299 |
| F351 – nPen m 1 | Y = 1.002x +0.1198 | 11 | 0.982 - 1.022 | 0.009 - 0.231 |
| E91 – ParrotFeathers | Y = 0.992x +0.4060 | 12 | 0.919 - 1.065 | 0.066 - 0.746 |
| M305 –Penicillium brevi-compactum | Y = 0.979x +0.2574 | 12 | 0.864 - 1.093 | -0.001 - 0.516 |
| E7 – PigeonDroppings | Y=0.959x +0.0758 | 10 | 0.893 - 1.025 | 0.0369 - 0.1147 |
| Allergen | Regression Equation | N | Slope95% CI | Intercept95% CI |
| W36 - RabbitBush | $Y = 0.982x +0.4142$ | 12 | 0.914 - 1.050 | 0.150 - 0.679 |
| 0201 - Tobacco | $Y = 0.964x +0.2661$ | 12 | 0.916 - 1.012 | 0.139-0.393 |
| E89 - TurkeyFeathers | $Y = 0.993x +0.2334$ | 12 | 0.932 - 1.053 | 0.058 - 0.409 |
Linearity
{7}------------------------------------------------
Specificity (Inhibition) Studies
Specificity of each allergen was verified through competitive inhibition testing using a single serum sample or pool of sera. A negative sample was used to measure the background response.
To initiate the inhibition experiment, 70 uL of undiluted and 3-4 levels of 5-fold serially diluted inhibitor extract were mixed with 250 uL of sample or pool to achieve final inhibitor concentrations of 218.75, 43.75, 8.75, 1.75, 0.35, 0.08, 0.07, 0.02, 0.01, 0.003 µg/mL. This mixture was incubated at room temperature (15-28 °C) for 1 hour allowing the immunological reaction to occur. Each sample mixture containing the inhibitor extract and the appropriate controls was assayed with 1 lot of each allergen. The percent (%) inhibition was calculated according to the following formula:
(Response of pos. control (pos. sample - neg. sample response with inhibitor extract) (Response of pos. control (pos. sample - neg. sample) )
X 100
The inhibition plots demonstrate that the allergens tested are inhibited by the relevant inhibitor extract in a concentration dependent fashion. Also, the target % inhibition of 50% for the highest inhibitor concentration tested was met. These results indicate specificity of Aureobasidium pullulans, Blomia tropicalis, Brazilian Peppertree, Budgerigar Feathers, Canary Feathers, Chicken Feathers, Loblolly Pine, nAsp r 1, nCan f 1, nDer f 1, nDer f 2, nDer p 1, nDer p 2, nFel d 1, nPen m 1, Parrot Feathers, Penicillium brevi-compactum, Pigeon Droppings, Rabbit Bush, Tobacco, and Turkey Feather allergens.
Inhibition Using Negative Controls
Additional inhibition studies were conducted to show that the specific allergens are not crossreacting to the unrelated allergens. Procedures were followed according to CLSI ILA20-A, Appendix-D. Testing was performed using one positive sample with three unrelated allergen extracts at 1 mg/ml. A negative sample was used to measure the background response. Results on the following specific allergen(s) were below 7% except for Loblolly Pine: Aureobasidium pullulans, Blomia tropicalis, Brazilian Peppertree, Budgerigar Feathers, Canary Feathers, Chicken Feathers, nAsp 1, nCan f 1, nDer f 2, nDer p 1, nDer p 1, nDer p 2, nFel d 1, nPen m 1, Parrot Feathers, Penicillium brevi-compactum, Pigeon Droppings, Rabbit Bush, Tobacco and Turkey Feathers.
{8}------------------------------------------------
Clinical Performance Studies
Clinical performance was demonstrated by testing serum samples against specific allergens from clinically diagnosed atopic and non-atopic individuals. Allergen-specific testing was obtained using the IMMULITE® 2000 3gAllergy™ assay.
Data summary agreement of the IMMULITE® 2000 3gAllergy results to clinical data is presented in the table below.
| IMMULITE® 2000 | Clinical Data | ||||
|---|---|---|---|---|---|
| Clinical | Normal | Total | |||
| Positive | 648 | 38 | 686 | ||
| Negative | 260 | 2,194 | 2,454 | ||
| Total | 908 | 2,232 | 3,140 | ||
| 71.4% | 98.3% | 90.5% | |||
| Sensitivity | Specificity | Agreement | |||
| Lower Conf | 68% | 98% | 89% | ||
| Upper Conf | 74% | 99% | 92% | ||
| Allergens included: Aureobasidium pullulans, Blomia tropicalis, Brazilian Peppertree, Budgerigar Feathers,Canary Feathers, Chicken Feathers, Loblolly Pine, nAsp r 1, nCan f 1, nDer f 1, nDer f 2, nDer p 1, nDer p 2,nFel d 1, nPen m 1, Parrot Feathers, Penicillium brevi-compactum, Pigeon Droppings, Rabbit Bush, Tobacco,and Turkey Feathers. |
IMMULITE® 2000 3gAllergy assay results for all allergens compare well with clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity.
Conclusions for all Studies
Allergens including : Aureobasidium pullulans. Blomia tropicalis, Brazilian Peppertree, Budgerigar Feathers, Canary Feathers, Chicken Feathers, Loblolly Pine, nAsp r 1, nCan f 1, nDer f 1, nDer f 2, nDer p 1, nDer p 2, nFel d 1, nPen m 1, Parrot Feathers, Penicillium brevi-compactum, Pigeon Droppings, Rabbit Bush, Tobacco, and Turkey Feathers for use with the IMMULITE® 2000 3gAllergy Specific IgE assay demonstrate acceptable analytical performance including precision, linearity and specificity. IMMULITE® 2000 assay results compare well with clinical documentation of presence of signs, symptoms and other diagnostic evidence of allergen sensitivity. Substantial equivalence was demonstrated to clinical data, supporting the following intended use of the IMMULITE® 2000 3gAllergys Specific IgE assay and the twenty seven previously listed allergens:
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with three curved lines representing the body and wings.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
MAR 2 8 2011
Siemens Healthcare Diagnostics Inc. c/o Ms. Donna Velasquez Regulatory Technical Specialist 5210 Pacific Concourse Drive Los Angeles, CA 90045
Re: K093987
Trade/Device Name: IMMULITE® 2000 3G Allergy™ Specific IgE Assay Kit Regulation Number: 21 CFR §866.5750 Regulation Name: Radioallergosorbent (RAST) immunological test system Regulatory Class: Class II Product Codes: DHB Dated: March 8, 2011 Received: March 11, 2011
Dear Ms. Velasquez :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{10}------------------------------------------------
Page 2 - Ms. Donna Velasquez
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Reema Philip
Maria M. Chan, Pl:D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
Indication for Use
510(k) Number (if known): K093987
Device Name: IMMULITE 3gAllergy™ Specific IgE Assay
Indication For Use:
For in vitro diagnostic use with the IMMULITE 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
Prescription Use ど (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Aiena Philip
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) 093987
§ 866.5750 Radioallergosorbent (RAST) immunological test system.
(a)
Identification. A radioallergosorbent immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the allergen antibodies (antibodies which cause an allergic reaction) specific for a given allergen. Measurement of specific allergen antibodies may aid in the diagnosis of asthma, allergies, and other pulmonary disorders.(b)
Classification. Class II (special controls). The device, when intended to detect any of the allergens included in Table 1 in this paragraph, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. Table 1—Class II Exempt Allergens Under § 866.5750—Radioallergosorbent (RAST) Immunological Test Systems
Allergen code Allergen product Source (taxonomical name) Grass Pollens g1 Sweet vernal grass Anthoxanthum odoratum. g3 Cocksfoot grass, Orchard grass Dactylis glomerata. g4 Meadow fescue Festuca elatior. g5 Rye-grass (perennial rye grass) Lolium perenne. g7 Common reed (common reed grass) Phragmites communis. g8 Meadow grass, Kentucky blue (June grass) Poa pratensis. g9 Redtop, Bentgrass Agrostis stolonifera ,Agrostis gigantea (Agrostis alba ). g11 Brome grass Bromus inermis. g12 Cultivated rye (cultivated rye grass) Secale cereale. g13 Velvet grass Holcus lanatus. g14 Cultivated oat (cultivated oat grass) Avena sativa. g15 Cultivated wheat (cultivated wheat grass) Triticum aestivum (Triticum spp.). g16 Meadow foxtail (meadow foxtail grass) Alopecurus pratensis. g17 Bahia grass Paspalum notatum. g24 Wheat grass, Western Agropyron smithii (Elymus smithii ). g30 Bluegrass, annual Poa annua. g70 Wild rye grass Elymus triticoides Elymus condensatus. g71 Canary grass Phalaris arundinacea. g201 Barley, cultivated Hordeum vulgare. g202 Maize, corn (cultivated corn) Zea mays. g203 Salt grass Distichlis spicata. g204 False oat-grass Arrhenatherum elatius. g216 Cyn d 1 Cynodon dactylon. g701 Phl p 1.0102, Phl p 5.0101 Phleum pratense. g702 Phl p 7.0101 Phleum pratense. g703 Phl p 12.0101 Phleum pratense. Weed Pollens w2 Western ragweed Ambrosia psilostachya. w4 False ragweed Ambrosia acanthicarpa (Franseria acanthicarpa ). w5 Wormwood Artemisia absinthium Artemisia annua. w6 Mugwort Artemisia vulgaris. w7 Marguerite, ox-eye daisy Chrysanthemum leucanthemum. w8 Dandelion Taraxacum vulgare ,Taraxacum officinale. w9 Plantain (English), Ribwort Plantago lanceolata. w10 Goosefoot, lamb's quarters Chenopodium album. w11 Saltwort (prickly), Russian thistle Salsola kali (Salsola pestifer ). w12 Goldenrod Solidago virgaurea (Solidago spp.). w13 Cocklebur, common Xanthium commune. w14 Common pigweed (rough pigweed) Amaranthus retroflexus. w15 Scale, Lenscale Atriplex lentiformis. w16 Rough marsh elder Iva ciliate ,Iva annua. w17 Firebush (Kochia) Kochia scoparia. w18 Sheep sorrel Rumex acetosella. w19 Wall pellitory Parietaria officinalis. w20 Nettle (Common stinging nettle) Urtica dioica. w21 Wall pellitory Parietaria judaica. w22 Japanese hop (careless weed) Humulus japonicas (Humulus scandens ). w23 Yellow dock, Yellow dockweed Rumex crispus. w24 Spiny pigweed Amaranthus spinosus. w27 Carnation Dianthus spp. w28 Rose Rosa rugosa. w33 Clover Trifolium pratense. w35 Mexican tea Chenopodium ambrosioides. w36 Rabbit bush Ambrosia deltoidea (Franseria deltoides ). w37 Salt bush, annual Atriplex wrightii. w39 Water hemp, Western Amaranthus rudis (Acnida tamariscina ). w41 Burrobrush Hymenoclea salsola. w42 Poverty weed Baccharis neglecta. w43 Common sagebrush Artemisia tridentata. w45 Alfalfa Medicago sativa. w46 Dog fennel Eupatorium capillifolium. w53 Geranium Geranium spp. w67 Groundsel bush Baccharis halimifolia. w69 Iodine bush Allenrolfea occidentalis. w70 Ragweed, slender Ambrosia confertiflora. w75 Wing scale (wingscale) Atriplex canescens. w82 Careless weed Amaranthus palmeri ,Amaranthus hybridus. w90 Japanese hop Humulus japonicas (Humulus scandens ). w203 Rape (rape pollen) Brassica napus. w204 Sunflower Helianthus annuus. w206 Camomile Matricaria chamomilla. w207 Lupin Lupinus spp. w210 Sugar-beet Beta vulgaris. w211 Par j 2.0101 Parietaria judaica. w231 Art v 1 Artemisia vulgaris (Mugwort). w232 Sal k 1 Salsola kali. w233 Art v 3 Artemisa vulgaris (LTP, Mugwort). w234 Pla l 1 Plantago lanceolata. w235 Che a 1.0101 Chenopodium album. w236 Mer a 1.0101 Mercurialis annua. a753 Art v 1 Artemisia vulgaris (Mugwort weed). Tree Pollens t1 Box-elder (Maple) Acer negundo ,Acer saccharum. t2 Gray alder, speckled alder (alder) Alnus incana. t4 Hazel, hazelnut Corylus avellana ,Corylus americana. t5 American beech (beech) Fagus grandifolia (Fagus americana ). t6 Mountain juniper, Mountain cedar Juniperus ashei (Juniperus sabinoides ). t8 Elm Ulmus americana. t9 Olive Olea europaea. t10 Walnut Juglans californica ,Juglans nigra. t11 Maple leaf sycamore, London plane, Plane tree Platanus acerifolia. t61 Sycamore Platanus occidentalis. t12 Willow Salix caprea ,Salix nigra. t14 Cottonwood (Eastern Cottonwood/Black Cottonwood) Populus deltoides. t15 White ash Fraxinus americana. t16 White pine Pinus strobus. t18 Eucalyptus, gum-tree Eucalyptus globulus (Eucalyptus spp.). t19/t26 Acacia Acacia longifolia (Acacia spp.). t20 Mesquite Prosopis glandulosa/ Prosopis juliflora. t21 Melaleuca, cajeput tree Melaleuca quinquenervia (Melaleuca leucadendron ). t22 Pecan, hickory Carya illinoinensis (Carya pecan ). t23 Italian/Mediterranean/funeral cypress Cupressus sempervirens. t24 Japanese cypress Chamaecyparis obtusa (Chamaecyparis spp.). t25 Ash Fraxinus excelsior. t27 Maple, red Acer rubrum. t29 Acacia Acacia spp. t30 Birch, white Betula populifolia. t32 Willow, black Salix nigra. t33 Ash, Arizona Fraxinus velutina. t35 Cedar, salt Tamarix gallica. t37 Bald cypress (white bald cypress) Taxodium distichum. t38 Elm, Chinese/Siberian Ulmus pumila. t40 Hazelnut tree Corylus americana. t41 White hickory Carya alba (Carya tomentosa ). t42 Oak, red Quercus rubra. t43 Loblolly pine Pinus taeda. t44 Hackberry Celtis occidentalis. t45 Cedar elm Ulmus crassifolia. t47 Juniper, one seed Juniperus monosperma. t48 Pine, lodgepole Pinus contorta. t49 Pine, ponderosa Pinus ponderosa. t50 Beech, European Fagus sylvatica. t51 Tree of Heaven Ailanthus altissima. t52 Western white pine Pinus monticola. t54 Russian olive Elaeagnus angustifolia. t55 Scotch broom Cytisus scoparius. t56 Bayberry Myrica cerifera. t57 Red cedar Juniperus virginiana. t60 Western juniper Juniperus occidentalis. t61 Sycamore Platanus occidentalis. t70 Mulberry (white mulberry) Morus alba. t71 Red mulberry Morus rubra. t72 Queen palm Arecastrum romanzoffiamon. t73 Australian pine Casuarina equisetifolia. t77 Oak mix (red, white, black) Quercus spp. t80 Japanese cypress Chamaecyparis obtusa. t81 Japanese alder Alnus japonica. t83 Mango tree Mangifera indica. t90 Walnut, black Juglans nigra. t96 Poplar, white (poplar) Populus alba. t103/t218 Virginia live oak (live oak) Quercus virginiana. t105 Pepper tree Schinus molle. t110 Orange tree Citrus sinensis. t201 Spruce, Norway spruce Picea abies (Picea excelsa ). t202 Alder, smooth Alnus incana spp.Rugosa (Alnus rugosa ). t203 Horse chestnut Aesculus hippocastanum. t205 Elder Sambucus nigra. t206 Chestnut Castanea sativa. t207 Douglas fir Pseudotsuga menziesii (Pseudotsuga taxifolia ). t208 Linden Tilia cordata. t209 Horn beam Carpinus betulus. t210 Privet Ligustrum vulgare. t211 Sweet gum Liquidambar styraciflua. t212 Cedar Libocedrus decurrens. t213 Pine Pinus radiata. t214 Date palm Phoenix canariensis. t215 Lilac Syringa vulgaris. t217 Pepper tree Schinus molle. t217 Red alder Alnus rubra. t218 Virginia live oak Quercus virginiana. t218 Bayberry (bayberry/sweet gale) Myrica gale. t219 Palo verde Cercidium floridum. t219 Red cedar Juniperus virginiana. t220 Bet v 4 Betula verrucosa (Birch). t221 Bet v 2.0101, Bet v 4 Betula verrucosa (Birch). t222 Cypress (Arizona cypress) Cupressus arizonica. t223 Oil palm Elaeis guineensis. t224 Ole e 1 Olea europaea. t225 Bet v 6 Betula verrucosa (Birch). t226 Cup a 1 Cupressus arizonica. t227 Ole e 7 Olea Europaea. t228 Aspen, quaking Populus tremuloides. t229 Eastern hemlock Tsuga canadensis. t230 Redwood (sequoia) Sequoia sempervirens. t232 Pussy willow Salix discolor. t240 Ole e 9.0101 Olea Europaea. t241 Pla a 1.0101 Platanus acerifolia. t242 Pla a 2 Platanus acerifolia. t243 Pla a 3.0101 Platanus acerifolia. t244 Cor a 1.0103 Corylus avellana. t245 Aln g 1.0101 Alnus glutinosa. t246 Cry j 1 Cryptomeria japonica. t280 Locust tree Robinia pseudoacacia. t401 Brazilian peppertree Schinus terebinthifolius. t402 Mastic tree Pistacia lentiscus. t404 Tree of heaven Ailanthus altissima. t406 Date palm Phoenix dactylifera. a482 Ole e 1 Olea europaea (Olive Oil). Mites d207 Blo t 5.0101 Blomia tropicalis. d208 Lep d 2.0101 Lepidoglyphus destructor. Microorganisms, Molds, Yeast m1 Penicillium chrysogenum (Penicillium notatum ) Penicillium chrysogenum (Penicillium notatum ). m2 Cladosporium herbarum (Hormodendrum ) Cladosporium herbarum (Hormodendrum ). m3 Aspergillus fumigatus Aspergillus fumigatus. m4 Mucor racemosus Mucor racemosus. m5 Candida albicans Candida albicans. m7 Botrytis cinerea Botrytis cinerea. m8 Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes, Helminthosporium interseminatum ) Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes. m9 Fusarium moniliforme (Fusarium proliferatum ) Fusarium moniliforme (Fusarium proliferatum ). m10 Stemphylium botryosum Stemphylium herbarum (Stemphylium botryosum ). m11 Rhizopus nigricans Rhizopus nigricans. m12 Aureobasidium pullulans Aureobasidium pullulans. m13 Phoma betae Phoma betae. m14 Epicoccum purpurascens Epicoccum purpurascens (Epicoccum nigrum ). m15 Trichoderma viride Trichoderma viride. m16 Curvularia lunata Curvularia lunata ,Curvularia specifera ([K923044](https://510k.innolitics.com/device/[K923044](https://510k.innolitics.com/device/K923044)) ). m17 Cladosporium fulvum Cladosporium fulvum. m18 Fusarium culmorum Fusarium culmorum. m19 Aspergillus versicolor Aspergillus versicolor. m20 Mucor mucedo Mucor mucedo. m21 Aspergillus clavatus Aspergillus clavatus. m22 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m23 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m24 Stachybotrys atra Stachybotrys chartarum (Stachybotrys atra ). m24 Paecilomyces spp Paecilomyces spp. m25 Aspergillus versicolor Aspergillus versicolor. m25 Penicillium brevicompactum Penicillium brevicompactum. m26 Cladosporium cladosporioides Cladosporium cladosporioides. m26 Penicillium citrinum Penicillium citrinum. m27 Penicillium spp Penicillium spp. m29 Aspergillus repens Aspergillus repens. m30 Penicillium roqueforti Penicillium roqueforti. m32 Cladosporium cladosporioides Cladosporium cladosporioides. m34 Serpula lacrymans Serpula lacrymans. m36 Aspergillus terreus Aspergillus terreus. m37 Trichophyton mentagrophytes Trichophyton mentagrophytes. m40 Aspergillus amstelodami Aspergillus amstelodami. m43 Saccharomyces Carlsberg Saccharomyces carlsbergensis. m44 Saccharomyces cerevisiae Saccharomyces cerevisiae. m45 Hormodendrum hordei Hormodendrum hordei. m46 Bipolaris spicifera Bipolaris spicifera. m47 Aspergillus nidulans Aspergillus nidulans. m48 Aspergillus oryzae Aspergillus oryzae. m49 Fusarium oxysporum Fusarium oxysporum. m50 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m51 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m53 Microspora canis Microsporum canis (Microspora canis ). m54 Aspergillus flavus Aspergillus flavus. m63 Helminthosporium intersemin Helminthosporium intersemin. m66 Mucor plumbeus Mucor plumbeus. m67 Mycogone Mycogone perniciosa. m68 Nigrospora oryzae Nigrospora oryzae. m69 Rhodotorula Rhodotorula rubra (Rhodotorula mucilaginosa ). m70 Malassezia furfur (Pityrosporum orbiculare ) Malassezia furfur (Pityrosporum orbiculare ). m71 Spondylocladium Spondylocladium spp. m72 Epidermophyton Epidermophyton floccosum. m73 Epicoccum nigrum Epicoccum purpurascens (Epicoccum nigrum ). m80 Staphylococcal enterotoxin A (Sta a SEA ) Staphylococcus aureus. m80 Helminthosporium spp Helminthosporium spp. m81 Staphylococcal enterotoxin B (Sta a SEB ) Staphylococcus aureus. m88 Stemphylium solani Stemphylium solani. m93 Gliocladium fimbriatum Gliocladium fimbriatum. m94 Phycomyces blakesleeanus Phycomyces blakesleeanus. m201 Tilletia tritici (Ustilago nuda, Ustilago tritici ) (Barley smut) Tilletia tritici (Ustilago nuda, Ustilago tritici ). m202 Acremonium kiliense (Cephalosporium acremonium ) Acremonium kiliense (Cephalosporium acremonium ). m203 Trichosporon pullulans Trichosporon pullulans. m204 Ulocladium chartarum Ulocladium chartarum. m205 Trichophyton rubrum Trichophyton rubrum. m207 Aspergillus niger Aspergillus niger. m208 Chaetomium globosum Chaetomium globosum. m209 Penicillium frequentans Penicillium glabrum (Penicillium frequentans ). m209 Stachybotrys chartarum Stachybotrys chartarum (Stachybotrys atra ). m210 Trichophyton mentagrophytes var. goetzii Trichophyton mentagrophytes var. goetzii. m211 Trichophyton mentagrophytes var. interdigitale Trichophyton mentagrophytes var. interdigitale. m211 Oat smut Ustilago avenae. m212 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m212 Geotrichum candidum Geotrichum candidum. m213 Bermuda grass smut Ustilago cynodontis. m214 Johnson grass smut Sphacelotheca cruenta. m215 Corn smut Ustilago maydis. m218 Asp f 1.0101 Aspergillus fumigatus. a3050 Asp r 1 Aspergillus restrictus. m219 Asp f 2 Aspergillus fumigatus. m220 Asp f 3.0101 Aspergillus fumigatus. m221 Asp f 4 Aspergillus fumigatus. m222 Asp f 6.0101 Aspergillus fumigatus. m223 Staphylococcal enterotoxin C (Sta a SEC ) Staphylococcus aureus. m224 Staphylococcal enterotoxin D (Sta a SED ) Staphylococcus aureus. m226 Staphylococcal enterotoxin TSST (Sta a TSST ) Staphylococcus aureus. m227 Malassezia spp. (Pityrosporum spp.) Malassezia spp. (Pityrosporum spp.). m228 Aspergillus flavus. m229 Alt a 1.0101 Alternaria alternata (Alternaria tenuis ). m230 Alt a 6.0101 Alternaria alternata (Alternaria tenuis ). m231 Cla h 8.0101 Cladosporium herbarum (Hormodendrum ). m300 Eurotium spp Eurotium spp. m304 Aspergillus oryzae Aspergillus oryzae. m305 Penicillium brevicompactum Penicillium brevicompactum. m309 Aspergillus terreus Aspergillus terreus. m310 Aspergillus nidulans Aspergillus nidulans. m311 Aspergillus flavus Aspergillus flavus. m312 Aspergillus clavatus Aspergillus clavatus. Epidermal & Animal e6 Guinea pig epithelium Cavia porcellus. e7 Pigeon droppings Columba palumbus ,Columba livia. e25 Chicken serum Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e26 Parrot serum Psittacoidea spp. e62 Camel Camelus dromedaries. e70 Goose feathers Anser anser. e71 Mouse epithelium Mus musculus (Mus spp.). e73 Rat epithelium Rattus norvegicus. e74 Rat urine proteins Rattus norvegicus ,Rattus rattus. e75 Rat serum proteins Rattus norvegicus ,Rattus rattus. e76 Mouse serum proteins Mus musculus (Mus spp.). e77 Budgerigar droppings Melopsittacus undulatus. e78 Budgerigar feathers Melopsittacus undulatus. e79 Budgerigar serum proteins Melopsittacus undulatus. e80 Goat epithelium Capra hircus. e81 Sheep epithelium Ovis aries (Ovis spp.). e82 Rabbit epithelium Oryctolagus cuniculus (Oryctolagus spp.). e83 Swine epithelium Sus scrofa (Sus scrofa domesticus; Sus spp.). e84 Hamster epithelium Cricetus cricetus, Mesocricetus auratus, and Phodopus sungorus. e85 Chicken feathers Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e86 Duck feathers Anas platyrhynchos. e87 Rat epithelium, serum proteins, and urine proteins Rattus norvegicus Rattus rattus. e88 Mouse epithelium, serum proteins, and urine proteins (mouse) Mus musculus (Mus spp.). e89 Turkey feathers Meleagris gallopavo. e90 Budgerigar serum proteins, feathers, and droppings Melopsittacus undulatus. e91 Pigeon serum proteins, feathers, and droppings Streptopelia roseogrisea ,Psittacidae spp. e92 Parrot serum proteins, feathers, and droppings Ara spp. e93 Pigeon serum proteins Streptopelia roseogrisea. e94 Fel d 1.0101 Felis domesticus. a345 Fel d 1 Felis domesticus. e98 Parrot droppings Psittacoidea spp. e101 Can f 1.0101 Canis familiaris (Canis domesticus ). a174 Can f 1 Canis familiaris (Canis domesticus ). e102 Can f 2.0101 Canis familiaris (Canis domesticus ). e196 Parakeet feathers Nymphicus hollandicus. e197 Parakeet droppings Nymphicus hollandicus. e198 Parakeet serum Nymphicus hollandicus. e199 Canary bird serum Serinus canarius. e200 Canary bird droppings Serinus canarius. e201 Canary bird feathers (Canary feathers) Serinus canarius. e202 Reindeer epithelium Rangifer tarandus. e203 Mink epithelium Mustela spp. e204 Bos d 6 Bos domesticus (Bos taurus; Bos spp.). e205 Horse, serum proteins Equus caballus (Equus spp.). e206 Rabbit, serum proteins Oryctolagus cuniculus (Oryctolagus spp.). e208 Chinchilla epithelium Chinchilla laniger. e209 Gerbil epithelium Meriones unguiculatus. e210 Fox epithelium Vulpes vulpes. e211 Rabbit, urine proteins Oryctolagus cuniculus (Oryctolagus spp.). e212 Swine, urine proteins Sus scrofa (Sus scrofa domesticus; Sus spp.). e213 Parrot feathers Ara spp. e214 Finch feathers Lonchura domestica. e215 Pigeon feathers Streptopelia roseogrisea (Streptopelia spp.),Columbia spp. e216 Deer epithelium Dama dama. e217 Ferret epithelium Mustela putorius. e218 Chicken droppings Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e219 Chicken, serum proteins Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e220 Fel d 2, Cat serum albumin Felis domesticus. e221 Can f 3 Canis familiaris (Canis domesticus ) (Dog serum albumin ). e222 Swine serum albumin (Sus s PSA) Sus scrofa (Sus scrofa domesticus; Sus spp.). e225 Lovebird feathers Psittacoidea agapomis. e226 Can f 5.0101 Canis familiaris. e227 Equ c 1.0101 Equus caballus. e228 Fel d 4.0101 Felis domesticus. e230 Equ c 3 Equus caballus. e231 Mus m 1 Mus musculus. Food f9 Rice Oryza sativa. f12 Pea (green pea) Pisum sativum. f15 White bean Phaseolus vulgaris. f19 Cayenne pepper Capsicum frutescens (Capsicum annum ). f21 Sugar cane Saccharum officinarum. f22 Raspberry Rubus idaeus. f26 Pork Sus scrofa (Sus scrofa domesticus; Sus spp.). f29 Watermelon Citrullus lanatus (Citrullus vulgaris ). f31 Carrot Daucus carota. f32 Oyster mushroom Pleurotus ostreatus. f33 Orange Citrus sinensis. f35 Potato Solanum tuberosum. f43 Mother's milk Homo sapiens. f44 Strawberry Fragaria vesca (Fragaria spp.). f45 Yeast, baker's Saccharomyces cerevisiae. f46 Pepper, Red Capsicum annuum. f47 Garlic Allium sativum. f48 Onion Allium cepa. f49 Apple Malus x domestica (Malus spp.). f51 Bamboo shoot Phyllostachys pubescens. f52 Cacao/chocolate Theobroma cacao. f54 Sweet potato Ipomoea batatas. f55 Common millet Panicum miliaceum. f56 Foxtail millet Setaria italica. f57 Japanese millet Echinochloa crus-galli. f58 Pacific squid Todarodes pacificus. f59 Octopus Octopus vulgaris (Octopus spp.). f63 Kefir NA. f67 Parmesan cheese NA. f81 Cheese, cheddar type NA. f82 Cheese, mold type NA. f83 Chicken Gallus domesticus (Gallus gallus domesticus; Gallus spp.). f86 Parsley Petroselinum crispum. f87 Melon Cucumis melo Cucumis melo + Citrullus lanatus. f88 Mutton (lamb) Ovis aries (Ovis spp.). f90 Malt Hordeum vulgare. f92 Banana Musa spp. f93 Cacao Theobroma cacao. f94 Pear Pyrus communis (Pyrus spp.). f97 Yam Dioscorea spp.Dioscorea opposita. f97 Chamomile tea Matricaria chamomilla. f98 Gliadin Triticum aestivum (Triticum spp.). f102 Cantaloupe Cucumis melo var. cantalupensis. f105 Chocolate Theobroma cacao. f109 Cottonseed Gossypium hirsutum. f110 Giant radish Raphanus sativus. f118 Zucchini Cucurbita pepo. f119 Radish Raphanus sativus. f120 Venison Capreolus capeolus. f121 Pinto bean Phaseolus vulgaris. f122 Cheese, American NA. f127 Black-eyed pea Vigna unguiculata. f131 Black Olive Olea europaea. f136 Red beet Beta vulgaris var. conditiva. f139 Goat's Cheese Capra aegagrus. f140 Bran NA. f141 Corn (vegetables) Zea mays. f152 Green bell pepper Capsicum annuum. f155 Brewer's yeast Saccharomyces carlsbergensis. f157 Duck Anas domesticus. f158 Goose Anser anser. f160 Camembert cheese NA. f162 Nectarine Prunus persica var. nucipersica. f163 Kohlrabi Brassica oleracea var. gongylodes. f65 Perch f166 Leek Allium porrum. f170 Cheese (Switzerland) (Swiss cheese) NA. f174 Fig Ficus carica. f177 Cranberry Vaccinium macrocarpon. f179 Raisin Vitis spp. f182 Lima bean Phaseolus lunatus. f198 Flaxseed (bruised grain) Linum usitatissimum. f199 Untreated native milk Bos domesticus (Bos taurus; Bos spp.). f208 Lemon Citrus limon. f209 Grapefruit Citrus paradisi. f210 Pineapple Ananas comosus. f211 Blackberry Rubus fruticosus. f212 Mushroom (champignon) Agaricus hortensis (Agaricus spp.). f213 Rabbit Oryctolagus cuniculus (Oryctolagus spp.). f214 Spinach Spinacia oleracea. f215 Lettuce Lactuca sativa. f216 Cabbage Brassica oleracea var. capitata. f217 Brussels sprouts Brassica oleracea var. gem. f218 Paprika, sweet pepper Capsicum annuum. f219 Fennel seed Foeniculum vulgare. f219 Sage Salvia officinalis. f220 Cinnamon Cinnamomum spp. f221 Coffee Coffea spp. f222 Tea Camellia sinensis. f223 Green olive Olea europaea. f225 Summer squash, pumpkin Cucurbita pepo. f225 Pumpkin Cucurbita maxima. f226 Pumpkin seed Cucurbita pepo. f227 Sugar-beet seed Beta vulgaris. f229 Safflower Seed Carthamus tinctorius. f231 Milk, boiled Bos domesticus (Bos taurus; Bos spp.). f234 Vanilla Vanilla planifolia. f237 Apricot Prunus armeniaca. f241 Gouda cheese NA. f242 Cherry Prunus avium. f244 Cucumber Cucumis sativus. f246 Guar, guar gum Cyamopsis tetragonoloba. f247 Honey NA. f248 Rosemary Rosmarinus officinalis. f254 Plaice Pleuronectes platessa. f255 Plum Prunus domestica ,Prunus americana. f258 Squid Loligo spp. f259 Grape Vitis vinifera (Vitis spp.). f260 Broccoli Brassica oleracea var. italica (Brassica oleracea var. cultivar ). f261 Asparagus Asparagus officinalis. f262 Aubergine, eggplant Solanum melongena. f263 Green pepper Piper nigrum ,Capsicum annuum. f264 Eel Anguilla anguilla. f265 Caraway Carum carvi. f265 Cumin Cuminum cyminum. f266 Mace Myristica fragrans. f267 Cardamon Elettaria cardamomum. f268 Clove Syzygium aromaticum. f269 Basil Ocimum basilicum. f270 Ginger Zingiber officinale. f271 Anise Pimpinella anisum. f272 Tarragon Artemisia dracunculus. f273 Thyme Thymus vulgaris. f274 Marjoram Origanum majorana. f275 Lovage Levisticum officinale. f276 Fennel, fresh Foeniculum vulgare. f277 Dill Anethum graveolens. f278 Bay leaf Laurus nobilis. f279 Chili pepper Capsicum frutescens. f280 Black pepper Piper nigrum. f281 Curry (Santa Maria) NA. f282 Nutmeg Myristica fragrans. f283 Oregano Origanum vulgare. f284 Turkey meat Meleagris gallopavo. f285 Elk/moose meat Alces spp. f286 Mare's milk Equus caballus (Equus spp.). f287 Red kidney bean Phaseolus vulgaris. f288 Blueberry Vaccinium myrtillus (Vaccinium spp.). f289 Date Phoenix dactylifera. f291 Cauliflower Brassica oleracea var. botrytis. f292 Guava Psidium guajava. f293 Papaya Carica papaya. f294 Passion fruit, Maracuja Passiflora edulis (Passiflora spp.). f295 Carambola Averrhoa carambola. f296 Carob Ceratonia siliqua. f297 Gum Arabic Acacia senegal (Acacia spp.). f298 Tragacanth Astragalus spp. f299 Sweet chestnut (chestnut) Castanea sativa. f300 Pinto bean Phaseolus spp. f301 Persimmon (kaki fruit, sharon) Diospyros kaki. f302 Mandarin (tangerine, clementine, satsumas) Citrus reticulata. f305 Fenugreek Trigonella foenum-graecum. f306 Lime Citrus aurantifolia. f307 Hake Merluccius merluccius. f308 Sardine (pilchard) Sardina pilchardus. f310 Blue vetch Lathyrus sativus. f311 Megrim Lepidorhombus whiffiagonis. f315 Green bean Phaseolus vulgaris. f316 Rape seed Brassica napus. f317 Coriander Coriandrum sativum. f318 Jack fruit Artocarpus heterophyllus. f319 Beetroot Beta vulgaris. f320 Crayfish Astacus astacus. f321 Horse meat Equus caballus (Equus spp.). f322 Red currant Ribes sylvestre. f324 Hop (fruit cone) Humulus lupulus. f325 Saffron Colchicum autumnale. f328 Fig Ficus carica. f329 Watermelon Citrullus lanatus. f330 Rose hip Rosa spp. f331 Saffron Crocus sativus. f332 Mint Mentha piperita. f333 Linseed Linum usitatissimum. f336 Jujube Ziziphus jujuba. f336 Wine vinegar Vitis vinifera (Vitis spp.). f337 Sole Solea solea. f337 English sole Parophrys vetulus. f338 Wine, white Vitis vinifera (Vitis spp.). f339 Allspice Pimenta dioica. f339 Wine, red Vitis vinifera (Vitis spp.). f341 Cranberry Vaccinium oxycoccus ,Vaccinium macrocarpon. f342 Olive (black, fresh) Olea europaea. f343 Raspberry Rubus idaeus. f344 Sage Salvia officinalis. f346 Chives Allium schoenoprasum. f347 Quinoa Chenopodium quinoa. f348 Litchi Litchi chinensis. f349 Chum salmon roe Oncorhynchus keta. f358 Artichoke Cynara scolymus. f360 Yogurt NA. f368 Black bass Micropterus dolomieu (Micropterus dolomieui ). f374 Karaya gum Sterculia urens. f375 Horseradish Armoracia rusticana. f377 Maple syrup NA. f379 Okra Abelmoschus esculentus. f382 Beet, sugar Beta vulgaris var. altissima. f401 Loquat Eriobotrya japonica. f402 Fig Ficus carica. f403 Brewer's yeast Saccharomyces cerevisiae. f405 Mint Mentha spp. f406 Arugula Eruca vesicaria. House Dust h1 Greer Labs., Inc NA. h2 Hollister-Stier Labs NA. h6 Japan NA. Venoms & Insects i7 Midge Chironomus yoshimatsui. i8 Moth Bombyx mori ,Heterocera spp. i47 Water flea Daphnia spp. i49 Deer fly Chrysops spp. i51 Black ant Camponotus pennsylvanicus. i54 Flea mix (dog/cat), common flea Ctenocephalides spp. i71 Mosquito Aedes communis ,Aedes spp.and Culex spp. i72 Green nimitti Cladotanytarsus lewisi. i73 Blood worm Chironomus thummi ,Chironomusri parius ,Chironomus spp. i75 European hornet Vespa crabro. i76 Berlin beetle Trogoderma angustum. i77 European paper wasp Polistes dominulus. i78 Fly Musca domestica. i80 Bumblebee Bombus pennsylvanicus. i201 Horse bot fly Gasterophilus intestinalis. i202 Grain weevil Sitophilus granarius. i203 Mediterranean flour moth Ephestia kuehniella (Anagasta kuehniella ). i204 Horse fly Tabanus spp. i205 Bumblebee Bombus terrestris. i208 Api m 1.0101 Apis mellifera. a45 Api m 1 Apis mellifera. i209 Ves v 5.0101 Vespula vulgaris. a670 Ves v 5 Vespula vulgaris. i210 Pol d 5.0101 Polistes dominulus. i211 Ves v 1.0101 Vespula vulgaris. i213 Api m 4 Apis mellifera. i214 Api m 2 Apis mellifera. i215 Api m 3 Apis mellifera. i216 Api m 5 Apis mellifera. i217 Api m 10 Apis mellifera. i220 Bla g 1.0101 Blattella germanica. i221 Bla g 2.0101 Blattella germanica. i222 Bla g 5.0101 Blattella germanica. i223 Bla g 7 Blattella germanica. a46 Api m 2 Apis mellifera. Miscellaneous o1 Cotton, crude fibers Gossypium spp. o3 Cotton (treated) Gossypium spp. o70 Seminal fluid Homo sapiens. o71 Staphylococcus aureus Staphylococcus aureus. o72 Pichia pastoris crude extract customer specific Pichia pastoris. o72 Sperm-sediment Homo sapiens. o73 Pichia pastoris crude extr. vector customer specific Pichia pastoris. o74 Pichia pastoris with vector customer specific Pichia pastoris. o201 Tobacco leaf, tobacco dust Nicotiana tabacum. o202 Artemia salina, fish feed Artemia salina. o203 Tetramin, fish feed NA. o207 Daphnia, fish feed Daphnia spp. o211 Mealworm Tenebrio molitor. o212 Streptavidin Streptomyces avidini. o213 MBP (maltose binding protein) Escherichia coli. o214 CCD; MUXF3 from bromelain Ananas comosus. o72 Enterotoxin A (Sta a SEA) Staphylococcus aureus. o73 Enterotoxin B (Sta a SEB) Staphylococcus aureus. Parasites p1 Ascaris Ascaris suum. p2 Echinococcus Echinococcus granulosus. p3 Schistosoma Schistosoma mansoni. p4 Anisakis (Herring Worm) Anisakis simplex (Anisakis spp.). p5 Toxocara canis Toxocara canis. p10 Ani s 3.0101 Anisakis simplex (Anisakis spp.). p11 Ani s 1 Anisakis simplex (Anisakis spp.). Occupational k4 Threshing dust NA. k5 Flax NA. k7 Hay Dust NA. k8 Hop (hops) Humulus lupulus. k12 Grain mill dust NA. k14 Kapok NA. k20 Sheep's wool (treated) (wool) Ovis aries (Ovis spp.). k21 Sheep's wool (Untreated) Ovis aries (Ovis spp.). k23 Straw Dust NA. k33 Oak NA. k70 Green coffee bean Coffea spp. k71 Castor bean Ricinus communis. k72 Ispaghula Plantago psyllium/Plantago ovata. k73 Silk waste NA. k74 Silk Bombyx mori. k75 Isocyanate TDI (Toluene diisocyanate) NA. k76 Isocyanate MDI (Diphenylmethane diisocyanate) NA. k77 Isocyanate HDI (Hexamethylen diisocyanate) NA. k78 Ethylene oxide NA. k79 Phthalic anhydride NA. k80 Formaldehyde/Formalin NA. k81 Ficus Ficus benjamina (Ficus spp.). k83 Cotton seed Gossypium hirsutum. k84 Sunflower seed Helianthus annuus. k85 Chloramin T NA. k86 Trimellitic anhydride, TMA NA. k87 Asp o 21, alpha-amylase Aspergillus oryzae. k89 Orris root Iris florentina. k99 HSA (Human Serum Albumin) (Hom s HSA) Homo sapiens. k201 Car p 1, Papain Carica papaya. k202 Ana c 2, Bromelain Ananas comosus. k204 Maxatase Bacillus licheniformis. k205 Alcalase Bacillus spp. k206 Savinase, Protease 1 (Bac l Subtilisin) Bacillus spp. k208 Gal d 4, Lysozyme Gallus domesticus (Gallus gallus domesticus; Gallus spp.). k209 Hexahydrophtalic anhydrid NA. k210 Maleic anhydride NA. k211 Methyltetrahydrophtalic anhydrid NA. k212 Abachi wood dust Triplochiton scleroxylon. k213 Pepsin (Sus s Pepsin) Sus scrofa (Sus scrofa domesticus; Sus spp.). k213 TCPA NA. k214 Bougainvillea Bougainvillea spp. k225 Horse radish peroxidase (Arm r HRP) Armoracia rusticana. k226 Ascorbate oxidase (Cuc p ascorbate oxidase) Cucurbita pepo. k301 Flour dust Triticum spp. k501 Savinase customer specific Proprietary knowledge of customer. k502 Lipolase customer specific Proprietary knowledge of customer. k503 Termamyl customer specific Proprietary knowledge of customer. k504 Clazinase customer specific Proprietary knowledge of customer.