K Number
K093834
Manufacturer
Date Cleared
2010-09-29

(289 days)

Product Code
Regulation Number
866.3110
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Crit-Line Anemia Management software is a database application used to record, track and trend patient data as pertains to the management of anemia, and provide a dosage recommendation.

Device Description

The CLAM software is a database application that records, tracks, trends, and analyzes patient data (Hematocrit, Oxygen Saturation, and Hemoglobin) along with current ESA dose in order to effectively manage anemia. During dialysis treatment, the CLAM downloads initial, real-time Hematocrit (HCT), Oxygen Saturation (SaO2), and Hemoglobin (Hb) data from the Crit Line blood monitor. The CLAM software trends this patient data along with current Erythropoiesis Stimulating Agent (ESA) dosage to determine the patient's Hb levels, the need (if any) for ESA dosing, and to track patient's response to treatment. The CLAM software uses a proprietary algorithm to calculate a recommendation for ESA dosage based on patient data trends and current ESA dose in order to establish and maintain predetermined patient Hb levels. The CLAM proprietary algorithm used to determine ESA dosing for anemia treatment is based on the current manual algorithm. In addition, the CLAM software generates and prints patient reports for review and record keeping. The CLAM software resides on a host computer separate from the Crit Line units and acquires real-time data from the clinic's Crit Line monitors via wireless radio signal or by a serial cable.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Crit Line Anemia Management (CLAM) software:

The provided 510(k) summary focuses primarily on establishing substantial equivalence to predicate devices and describes a performance comparison of its ESA dosage recommendation algorithm to manual calculations. It does not provide explicit "acceptance criteria" in the typical quantitative sense (e.g., target specificity, sensitivity, or accuracy thresholds) for the device's performance, nor does it conduct a comprehensive clinical study to prove these against a definitive ground truth. Instead, the study aims to show agreement with current manual practice.

However, based on the information provided, we can infer the implicit "acceptance criteria" as the calculated ESA dosage recommendation being "exactly the same" as manually calculated doses.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
CLAM software calculated recommended ESA dose = Manually calculated ESA doseThe CLAM software calculated recommended dose and the manually calculated dose were exactly the same.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 62 dialysis patients
  • Data Provenance:
    • Country of Origin: New England (United States, as per typical use of "New England" in this context)
    • Retrospective or Prospective: Retrospective data using "the last two weeks of a patient's Hb values" being recorded for analysis.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The ground truth was established by "manually calculated ESA dosage calculations," implying that medical personnel (e.g., clinicians, nurses, physicians as described in Intended User Population) performed these calculations.
  • The document does not specify the number of experts, their specific roles (e.g., physician, nurse), or their qualifications (e.g., years of experience).

4. Adjudication Method for the Test Set

  • The document does not describe an adjudication method. The comparison was directly between the CLAM software's calculation and the "manually calculated dose," which presumably represents the established practice. There's no mention of a consensus process among multiple manual calculators or a tie-breaking mechanism.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • No, an MRMC comparative effectiveness study was not explicitly done. The study's focus was on the agreement between the CLAM software's recommendation and manual calculations, not on comparing human reader performance with and without AI assistance or quantifying an "effect size" of improvement.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, a standalone performance evaluation was done. The "performance test" directly compared the CLAM software's calculated ESA dosage recommendations (algorithm only) to manually calculated recommendations. The CLAM software generated its recommendation based on inputted patient data (Hb values and current ESA dose), independently of human input for that specific calculation step.

7. The Type of Ground Truth Used

  • Expert Consensus / Expert Determination (via manual calculation): The ground truth for this specific performance test was the "manually calculated ESA dosage calculations." This represents a determination made by medical professionals based on established protocols. It is not pathology, imaging, or direct outcomes data, but rather a calculation based on clinical parameters and existing practice.

8. The Sample Size for the Training Set

  • The document does not explicitly state the sample size used for training the CLAM software's proprietary algorithm. It mentions the algorithm is "based on the current manual algorithm" but provides no details on how it was developed or trained on specific datasets.

9. How the Ground Truth for the Training Set Was Established

  • The document does not specify how ground truth for any potential training set was established. It states the CLAM algorithm "is based on the current manual algorithm," which implies that the logic and principles of existing manual ESA dosing protocols formed the basis of the algorithm's design. It does not detail if or how a separate, labeled dataset with established ground truth was used to train or refine the algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for HEMA METRICS. The word "HEMA" is on the top left, and the word "METRICS" is on the bottom right. A curved line goes from the bottom left to the top right, connecting the two words.

510(k) Summary

K093834
SEP 29 2010

As Required by 21 CFR 807.92

510(k) Number: __

1.Submitter Information
Submitter Name:Hema Metrics, LLC695 North 900 WestKaysville, UT 84037-4118Tel: 801-451-9000Fax: 801-451-9007
Establishment Reg.:1721979
Contact Person:Douglas L. CoxDirector of Quality Assurance/Regulatory Affairs
Date Prepared:June 24, 2010
2.Device InformationTrade Name:Crit Line Anemia Management (CLAM)Hemodialysis system monitor accessory
Product Code:MSQ, JQP
Classification Reg.876.5820, 862.2100
Class:Class II

3. Predicate Device Information

Predicate 1 - Crit Line

Trade NameCrit Line IIITQA Access Management
Common NameNon-invasive Hematocrit, Blood Volume,Oxygen Saturation, Recirculation andAccess Blood Flow
Classification NameCrit Line Monitor III with TQA (CLM IIITQA)
510(k)K001763
ClassificationRegulations876.5820
Product CodeMQS

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for HEMA METRICS. The word "HEMA" is on the top line, and the word "METRICS" is on the bottom line. A curved line goes from the bottom left of the image to the top right, passing over the words.

Predicate 2 - Vasc-Alert

Trade NameVasc-Alert
Common NameHemodialysis access site patencymonitoring software
Classification NameSystem, hemodialysis, AccessRecirculation Monitoring
510(k)K042566
ClassificationRegulations876.5820
Product CodeMQS

4. Intended Use

The CLAM software is a database application used to record, track and trend patient data pertaining to the management of anemia, and to provide a dosage recommendation. The CLAM software is intended to manage anemia in adult end stage renal failure patients.

5. CLAM Software Description

The CLAM software is a database application that records, tracks, trends, and analyzes patient data (Hematocrit, Oxygen Saturation, and Hemoglobin) along with current ESA dose in order to effectively manage anemia.

During dialysis treatment, the CLAM downloads initial, real-time Hematocrit (HCT), Oxygen Saturation (SaO2), and Hemoglobin (Hb) data from the Crit Line blood monitor. The CLAM software trends this patient data along with current Erythropoiesis Stimulating Agent (ESA) dosage to determine the patient's Hb levels, the need (if any) for ESA dosing, and to track patient's response to treatment. The CLAM software uses a proprietary algorithm to calculate a recommendation for ESA dosage based on patient data trends and current ESA dose in order to establish and maintain predetermined patient Hb levels. The The CLAM proprietary algorithm used to determine ESA dosing for anemia treatment is based on the current manual algorithm. In addition, the CLAM software generates and prints patient reports for review and record keeping.

The CLAM software resides on a host computer separate from the Crit Line units and acquires real-time data from the clinic's Crit Line monitors via wireless radio signal or by a serial cable.

A. Crit Line IIITQA Blood Volume Monitor and CLAM Accessory

The Crit Line uses optical methods to measure HCT, SaO2, and then calculates Hb from HCT for patients undergoing hemodialysis treatment. The Crit Line system consists of a blood monitor, optical clip and blood chamber.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the text "HEMA METRICS" in a bold, sans-serif font. The word "HEMA" is positioned above the word "METRICS". A curved line spans from below the left side of "HEMA" to above the right side of "METRICS", visually connecting the two words. The letters "TM" are in superscript to the right of the word "METRICS".

The blood chamber is attached to the dialysis tubing prior to treatment and the optical clip is clipped to the blood chamber, allowing optical access to the blood flow. The Crit Line monitor collects baseline (pre-treatment) readings of HCT, SaO2, and Hb levels and subsequent ongoing readings during treatment. When Crit Line units are used with the CLAM software accessory, baseline readings are transferred to the CLAM application, which establishes a baseline and subsequently performs near real-time monitoring as described above.

B. Intended Patient Population

The CLAM software currently is intended for use in dialysis clinics which care for adult renal failure patients :(see Section 21 - Other for the study: "Use of 12x/month haemoglobin monitoring with a computer algorithm reduces haemoglobin variability").

C. Intended User Population and Use Environment

The CLAM software is intended for use by medical personnel such as clinicians, nurses, and physicians in dialysis clinics or where ever dialysis treatment is provided and/or Crit Line is used.

D. CLAM Software User Interface

The CLAM software interface is designed to centralize and automate many of the routine tasks performed by medical staff when collecting and handling patient data used in the management of anemia. Features of the CLAM software enables the user to collect, record, track, trend, report, analyze, and print data for review and record keeping. An additional feature of the CLAM software is the ability to calculate a recommended ESA dosage based on trending of patient Hb and current ESA dose.

6. · Technological Comparisons

A. Similarities to Predicate Devices 1 & 2

  • The CLAM software, Crit Line (predicate 1), and Vasc-Alert (predicate 2) devices provide a (pre-treatment) baseline of patient HCT readings.
  • 『 The CLAM software and predicate 2 devices are software database applications designed to record, track, trend patient data.
  • ® The CLAM software and Vasc-Alert software device perform trend analysis on patient data routinely collected during dialysis treatment
  • 0 All three allow the medical staff to set thresholds or limits for each patient, which when not met signal an alert.
  • © All three enable medical staff to record and track patient data with each dialysis treatment providing continual monitoring of patient status verses the standard once a month measurement protocol.
  • © All three devices are intended to be a reference for patient status and when considering course of treatment. For comparison of each predicate to new device see. Table 1 Similarities and Table 2 Similarities:

{3}------------------------------------------------

HEMA METR . Sin

SIMILARITIES to CRIT LINE (predicate 1)
ItemNew DevicePredicate Device
Indication for UseManagement of Hematocrit based Hemoglobin levels indialysis patientsSame
Principle of OperationNon-invasive means of recording and tracking patient data:Hematocrit, Hemoglobin, Oxygen Saturation which can beused to determine the need and/or course of treatment.Same
TechnologyAlgorithm used to trend patient data collected during eachdialysis treatment.Same
Patient DemographicsDialysis & CRRT patientsSame
Intended UserPhysician/Clinicians/NursesSame
Data StorageData is stored on computer media or networkSame
Data ManagementReports and graphs are available for view or printSame
Manual Data EntryGUI interface allows for patient data to be entered andedited manually.Same
Safeguards/AlertsSystem flags patients whose Hematocrit based Hemoglobinlevels fall outside set limits.Same
SIMILARITIES to Vasc-Alert (predicate 2)
ItemNew DevicePredicate Device
Principle of OperationSoftware database application which records, tracks andtrends patient data used as a reference when determiningtreatment.Same
Data AnalysisTrends and analyses patient data and sets thresholds orlimitsSame
TechnologyAlgorithm used record, track, and establish a baseline ofpatient data collected during initial phase of dialysistreatmentSame
Patient DemographicsDialysis & CRRT patientsSame
Intended UserPhysician/Clinicians/NursesSame
Data StorageData is stored on computer media or on a networkSame
Data ManagementReports and graphs are available for view or printSame
Manual Data EntryGUI interface allows for patient data to be entered andedited manually via keyboard and/or mouse interface.Same
Safeguards/AlertsSystem flags patients who exceed set limits.Same

B. Differences

Differences between the CLAM software and predicate devices 1 & 2:

  • © The CLAM software offers an ESA dosage recommendation based on trending of patient's hematocrit based hemoglobin levels and the current ESA dose.
  • ් The CLAM software algorithm is for the purpose of managing anemia following hemodialysis therapy, refer to Table 3 Differences and Table 4 Differences below:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for HEMA METRICS. The word "HEMA" is on the top line, and the word "METRICS" is on the bottom line. A curved line goes from the bottom left of the word "HEMA" to the top right of the word "METRICS".

DIFFERENCES - CRIT LINE (predicate 1)
ItemNew DevicePredicate Device
Technology/AlgorithmAlgorithm used to trend Hb levels and current ESAdose and based on trending of data calculate arecommended ESA dosageAlgorithm used to plot HCT,SaO2, and Hb data and reportresults in real time
Data AnalysisProvides an improved recommended ESA dosagebased on significantly more data points than currentonce per month laboratory blood draws thereforetrending of historical patient Hemoglobin levels andESA dosage keeps ESA dosing in better complianceto ESA instructionsProvides results of collectedand analyzed data only
HardwareCLAM is a software database which resides on ahost computer and communicates with Crit Linemonitor via radio and/or serial linkCLM III Monitor is a standalone device whichcommunicates with CLAM viaradio and/or serial link

Table 3 Differences

DIFFERENCES - VASC-ALERT (predicate 2)
ItemNew DevicePredicate Device
Indication for UseManagement of Hematocrit basedHemoglobin levels in dialysis patientsMonitors access flow patency offistulas and grafts inhemodialysis patients
Technology /AlgorithmAlgorithm used to trend Hb levels and currentESA dose and based on trending of datacalculate a recommended ESA dosageVAPR algorithm used to trendhematocrit and pressure flowrate to determine access flowrate.
Data AnalysisProvides an improved recommended ESAdosage based on significantly more datapoints than current once per month laboratoryblood draws therefore trending of historicalpatient Hemoglobin levels and ESA dosagekeeps ESA dosing in better compliance toESA instructionsPlaces patient on report forreferral to an interventionalist.

Table 4 Differences

C. Performance Data

A performance test was completed in September of 2009 at a dialysis center in New England where the CLAM software ESA dosage calculations were compared to ESA dosage calculations performed manually.

The test consisted of taking data from sixty-two dialysis patients for analysis. The last two weeks of a patient's Hb values were recorded and the average of Week 1 was subtracted from the average of Week 2. If the difference between the two weeks is positive then the trend is falling, and if the difference was negative the trend was rising. The average of Week 1, the absolute value of the difference, and the rising or falling trend is then used to derive the next recommended dose. The results showed that the CLAM software calculated recommended dose and the manually calculated dose were exactly the same.

{5}------------------------------------------------

HEMA METRICS ...

D. Conclusion

The CLAM software is designed to work in tandem with the predicate device (Crit Line) to provide a more comprehensive and accurate approach to managing anemia treatment following hemodialysis.

The Crit Line uses an optical method to measure HCT and SaO2 and then calculate Hb from HCT for patients undergoing hemodialysis. Crit Line takes a baseline reading just prior to the start of hemodialysis and then continues to take measurements during the duration of treatment. Upon completion of the hemodialysis run the CLAM software downloads these baseline readings to the CLAM database where it is stored for tracking and trending.

The current protocol used by most clinics to assess Hb levels in a dialysis patient is to draw blood one time per month from which the physician prescribes (if needed) ESA dosage for the following month or until the next blood draw. This method provides little data and does not account for the time and changes experienced by the patient between blood draws. One change experienced by the patient is the removal of excess fluid during hemodialysis. On average a dialysis patient will underqo ten to twelve dialysis treatments between the once a month blood draw. The changes to HCT in relation to total blood volume are changed with each hemodialysis treatment affecting Hb levels as well.

The CLAM software provides the means to record, track, and trend these critical changes along with current ESA dose with each hemodialysis treatment. This method provides up to twelve points of data per month to measure patient status compared to the one point currently used. Based on trending of Hb levels and current ESA dose the CLAM software gives an ESA dosage recommendation for the physician's consideration.

The CLAM software device, like the Crit Line monitor, is for use only under supervision of a healthcare provider and is not intended to replace the knowledge and experience of the healthcare provider when prescribing treatment

Additional safeguards are incorporated in the CLAM software interface and algorithm to mitigate or prevent harm to the patient by limiting the maximum dosage recommendation of the CLAM software to 20,000 IU per treatment, to flag those patients whose Hb levels fall outside the established limits, and to prevent interchange of Hb and HCT values when entered manually.

In conclusion, the CLAM software like the predicate devices is designed to record, track, and trend patient data for the purpose of providing accurate readings, ongoing analyses of patient data, and as a reference for physicians when determining an effective course of treatment. For these reasons the CLAM software is considered equivalent to the predicate devices and does not pose any additional safety or effectiveness issues.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows a logo for the Department of Health & Human Services-USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is a symbol that looks like three stylized waves or lines. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Hema Metrics, LLC C/O Mr. Terry Lanier 136 W. 2nd Street Ogden, Utah 84404

SEP 2 9 2010

Re: K093834

Trade/Device Name: Crit-Line Anemia Management Software Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis System Monitor Accessory Regulatory Class: II Product Code: MSQ, JQP Dated: September 13, 2010 Received: September 15, 2010

Dear Mr. Lanier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2- Mr. Lanier

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

h fac

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for HEMA Metrics. The word "HEMA" is on the top line, and the word "METRICS" is on the bottom line. A curved line goes over the top of the word "METRICS" and under the word "HEMA".

510(k) Number:

Indication for Use Statement

K093834

SEP 2 9 2010

T

Crit-Line Anemia Management Software Device Name:

Indications for Use: The Crit-Line Anemia Management software is a database application used to record, track and trend patient data as pertains to the management of anemia, and provide a dosage recommendation.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rith C. Chape 9/27/10

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antonio Dental Devices

510(k) Number: K093834

§ 866.3110

Campylobacter fetus serological reagents.(a)
Identification. Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identifyCampylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases.Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.(b)
Classification. Class I (general controls).