K Number
K093684

Validate with FDA (Live)

Device Name
EASYCARE ONLINE
Manufacturer
Date Cleared
2010-02-26

(88 days)

Product Code
Regulation Number
868.5905
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EasyCare Online is intended to augment the standard follow-up care of patients diagnosed with obstructive sleep apnea by displaying usage and therapeutic information that has been transmitted remotely from the patient's flow generator located in the home to the care giver. EasyCare Online also provides remote settings capabilities.

It is intended to be used by Clinicians in conjunction with ResMed compatible flow generators.

Device Description

The performance and functional characteristics of EasyCare Online includes all the user features of the predicate device, ResTraxx Online (K083816).

EasyCare Online is designed to function with ResMed OSA treatment systems for the transfer, storage, retrieval and display of stored information from the flow generator to a centralized server. This information can then be accessed by the clinician where the information can be displayed on the clinician's PC using a web browser application. Flow Generator settings changes, via wireless transmission can also be made from the clinicians PC. Access to the data is limited to subscribers of the system. Patients cannot access the system.

EasyCare Online comprises two distinct components, the wireless transmitter/receiver located on the flow generator and the Server System. Data taken from the flow generator is transmitted via a wireless network and stored on the EasyCare Online database.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ResMed EasyCare Online device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text states: "All tests confirmed that EasCare Online met the predetermined acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria. Instead, the study focuses on demonstrating "Substantial Equivalence" to predicate devices through functional testing.

Acceptance Criteria (Not explicitly stated quantitatively, inferred from text)Reported Device Performance (as stated in the document)
Successful transfer of patient and treatment data from flow generator to EasyCare Online system."End-to-End testing that shows patient and treatment data can be successfully transferred from the flow generator to the EasyCare Online system." "All tests confirmed that EasCare Online met the predetermined acceptance criteria."
Successful sending of appropriate commands from PC to flow generator to update treatment parameters."If the user decides to update treatment parameters in the flow generator, this can be done by sending appropriate commands from the PC to the flow generator via the server and GSM network. The flow generator, via the GSM network, will send confirmation back to the user once these new settings have been implemented."
Performance in a new operating platform and software operating system (Windows 2000 NT to Windows 2008)."Areas of difference relate to new operating platform and software operating system (Windows 2000 NT to Windows 2008) which provides improved performance in the connectivity and transfer of data... along with improved (GUI) which provides better usability."
Compliance with FDA guidance documents for software in medical devices, off-the-shelf software, and cybersecurity."The new device complies with the applicable requirements referenced in the FDA quidance documents: [lists 3 FDA guidance documents]."
Functionality and user features comparable to the predicate device, ResTraxx Online (K083816)."The performance and functional characteristics of EasyCare Online includes all the user features of the predicate device, ResTraxx Online (K083816)."
Secure interface for accessing data."provide a secure interface into the system."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document refers to "End-to-End testing" and "System testing" but does not quantify the number of test cases, patient data sets, or flow generators used in these tests.
  • Data Provenance: The testing was "bench testing." It is not specified if any real patient data was used for testing, or if simulated data was utilized. The document mentions the flow generators are "located in the patient's home," but the testing itself appears to be simulated environment testing ("bench testing"), not involving actual patients at home for the purpose of the 510(k) submission. Therefore, it's not strictly "retrospective or prospective" in terms of patient data collection for the study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable/not stated. The ground truth for this device's performance was established through functional verification against defined requirements and specifications, not through expert consensus on medical diagnoses or interpretations. The "bench testing" would have been evaluated against technical specifications and expected system behavior.
  • Qualifications of Experts: Not applicable/not stated, as the ground truth was not expert consensus.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable/none. The testing involved verifying the system's functional performance against predetermined technical requirements, not subjective interpretation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, an MRMC comparative effectiveness study was not done.
  • Effect Size: Not applicable, as no MRMC study was performed. The device is a data management and remote setting system, not an AI diagnostic tool.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Standalone Study: Yes, in essence, the "End-to-End bench testing" represents the standalone performance of the system. The device's primary function is to transfer, store, retrieve, display, and remotely set parameters, which are algorithmic and system-level functions. The testing focused on whether the system performed these functions correctly and reliably without human intervention during the data flow.
    • The document explicitly states: "No clinical trials were needed to demonstrate Safety and Efficacy. Based upon the Risk Analysis, End-to-End bench testing alone is sufficient as no new features were added that could raise concerns regarding Safety or Efficacy." This reinforces that only the standalone system performance was evaluated.

7. Type of Ground Truth Used

  • Ground Truth Type: The ground truth used was functional specification and predetermined acceptance criteria for system operations, data transfer, storage, display, and remote control. It was based on engineering and software design requirements, rather than clinical outcomes, pathology, or expert medical consensus.

8. Sample Size for the Training Set

  • Sample Size: Not applicable/not stated. This device is a software system for data management and remote control, not an AI/machine learning algorithm that requires a "training set" in the conventional sense. The software development lifecycle and verification activities would involve testing against requirements, but not a distinct "training set" for model learning.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable. As there is no training set for an AI/ML model, there is no corresponding ground truth to be established for it. The ground truth for the overall system's development and testing was its design specifications and functional requirements.

{0}------------------------------------------------

KC093684

RESMED

EasyCare Online Traditional 510(k) Premarket Notification

510(k) Summary - EasyCare Online

Date Prepared1st Feb, 2010FEB 2 6 2010
Official ContactDr Lionel KingV.P., Quality Assurance & Regulatory AffairsResMed Ltd1 Elizabeth Macarthur DriveBella Vista, NSW 2153AustraliaTel: +61 (2) 8884 1000Fax: +61 (2) 8884 2021
Classification Reference21 CFR 868.5905
Product Code73 BZD
Common/Usual NameNoncontinuous ventilator (IPPB).
Proprietary NameEasyCare Online
Predicate Device(s)ResTraxx Online (K083816)ResScan Pro (K082983)
Reason for submissionNew Device

1গ February, 2010

{1}------------------------------------------------

Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

  • Similar intended use >
  • Same operating principle A
  • Similar technologies 入
  • Same manufacturing process

EasyCare Online remains largely the same as the predicate devices ResTraxx Online (K083816) for web based browser and patient data management and ResScan Pro (K082983) for Graphics User Interface (GUI) functionality, including the use of ResMed's proprietary communications protocol. Areas : of difference relate to new operating platform and software operating system (Windows 2000 NT to Windows 2008) which provides improved performance in the connectivity and transfer of data from the flow generator to the clinician along with improved (GUI) which provides better usability.

Design and Verification activities such as software development life cycle according to FDA guidance documents were followed as a result of the risk analysis and design requirements. Demonstrating Substantial Equivalence between the predicate device ResTraxx Online (K083816) and the new device (EasyCare Online) involved End to-End testing that shows patient and treatment data can be successfully transferred from the flow generator to the EasyCare Online system.

System testing includes a PC (GU1) with a Web Browser which is used to access the EasyCare Online server. Based upon the user's instructions, the PC sends a request via the internet to the server (EasyCare Online) to retrieve patient and machine data from the flow generator located in the patient's home via a secure telemetry network (GSM network).

If the user decides to update treatment parameters in the flow generator, this can be done by sending appropriate commands from the PC to the flow generator via the server and GSM network. The flow generator, via the GSM network, will send confirmation back to the user once these new settings have been implemented.

All tests confirmed that EasCare Online met the predetermined acceptance criteria. ResMed has determined that the new device is Substantially Equivalent to the predicate devices – ResTraxx Online (K083816) for web based wireless transfer and functionality of patient and machine information and ResScan Pro (K082983) for (GUI). The new device complies with the applicable requirements referenced in the FDA quidance documents:

  • FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices > (May 11, 2005)
  • FDA Off-the-Shelf Software Use in Medical Devices (September 9,1999) A
  • FDA Guidance for Industry Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software (January 14, 2005)

No clinical trials were needed to demonstrate Safety and Efficacy. Based upon the Risk Analysis, Endto-End bench testing alone is sufficient as no new features were added that could raise concerns regarding Safety or Efficacy.

Intended Use

EasyCare Online is intended to augment the standard follow-up care of patients diagnosed with obstructive sleep apnea by displaying usage and therapeutic information that has been transmitted from the patient's flow generator located in the home to the care giver. EasyCare Online also provides remote settings capabilities.

It is intended to be used by Clinicians in conjunction with ResMed compatible flow generators.

1st February, 2010

{2}------------------------------------------------

Device Description

General

The performance and functional characteristics of EasyCare Online includes all the user features of the predicate device, ResTraxx Online (K083816).

EasyCare Online is designed to function with ResMed OSA treatment systems for the transfer, storage, retrieval and display of stored information from the flow generator to a centralized server. This information can then be accessed by the clinician where the information can be displayed on the clinician's PC using a web browser application. Flow Generator settings changes, via wireless transmission can also be made from the clinicians PC. Access to the data is limited to subscribers of the system. Patients cannot access the system.

EasyCare Online comprises two distinct components, the wireless transmitter/receiver located on the flow generator and the Server System. Data taken from the flow generator is transmitted wa a wireless network and stored on the EasyCare Online database.

Wireless Module

The Wireless Module is similar in size to a mobile phone, it contains a GSM device and some interface hardware and firmware that functions in a similar manner as a mobile phone by transferring data to and from the server and the flow generator. It is designed to attach to a compatible ResMed flow generator using a docking mechanism and is powered via the electrical connection from the flow generator. This mechanism allows the device to be electrically connected via the existing expansion port located at the rear of the flow generator. When attached, the wireless modules can automatically collect patient and machine information stored within the flow generator's memory. The wireless module sends information utilizing existing wireless networks providing coverage to large portions of the US population.

All flow generators capable of being connected to the GSM wireless module and therefore able to transfer patient and machine data to the server are classified as 73 BZD devices.

Server System

The Server System consists of several functional software modules that are designed to retrieve information from ResMed flow generators through the wireless network, store the information in a database and provide a secure interface into the system. The system allows users to schedule information retrieval and view patient and treatment information, including the ability to retrieve existing settings from and apply new settings to the flow generator remotely from the Clinical reviewer's PC.

Conclusion

Based upon the Risk Analysis and results obtained from End-to-End bench testing as detailed in this submission, EasyCare Online is as safe, as effective, and performs as well as the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

ResMed Limited C/O Mr. David D'Cruz Vice President, Clinical & Regulatory Affairs ResMed Corporation 9001 Spectrum Center Boulevard San Diego, California 92123

FEB 2 6 2010

Re: K093684 Trade/Device Name: EasyCare Online Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: November 20, 2009 Received: November 30, 2009

Dear Mr. D'Cruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRI4's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

for
Arthur D. Watson, B.S., M.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number (if known):

Device Name:

EasyCare Online

Indication for Use

EasyCare Online is intended to augment the standard follow-up care of patients diagnosed with obstructive sleep apnea by displaying usage and therapeutic information that has been transmitted remotely from the patient's flow generator located in the home to the care giver. EasyCare Online also provides remote settings capabilities.

It is intended to be used by Clinicians in conjunction with ResMed compatible flow generators.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-T he-Counter Use_

(Part 21 CFR 801 Subpart D)

(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONT NUE ON ANOT HER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

L. Schult

Page 1 of _ 1

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K093684

20吨 Nov, 2009

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).