K Number
K093679

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-06-10

(195 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AtriClip LAA Exclusion System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.

Device Description

The AtriClip LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier, and a Selection Guide to aid in appropriate Clip size selection. The frame assembly of the implantable Clip consists of two springs connecting two opposing tubes which are covered with pressure pads. This assembly is covered with a knit fabric. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure occlusion of the left atrial appendage (LAA). The Clip is available in the following lengths to accommodate different sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm. The Clip Applier is a disposable device with a handle, shaft, and an end effector which contains the Clip. Both sides of the Clip are free to move within the end effector when activated. Pulling the thumb lever proximally opens the Clip. The bias of the Clip allows both the clip and the lever to return to the closed position when the lever is released.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information provided in the document for the AtriClip LAA Exclusion System:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in a quantitative format (e.g., "efficacy must be X% or greater"). Instead, it presents the results of the clinical study and concludes that these results demonstrate excellent performance and support substantial equivalence.

Based on the descriptive statements, we can infer the following:

Acceptance Criterion (Inferred)Reported Device Performance
Efficacy: Complete and consistent LAA exclusion.- Acute and Long-term Exclusion: Demonstrated complete appendage exclusion both acutely and long-term (up to 90 days) in all animals in non-clinical testing. - Clinical Efficacy: 95.1% composite exclusion rate (intra-operative and 3 month). - Clinical Efficacy (3 Months): 98% exclusion rate at 3-month follow-up.
Safety: No device or procedure-related adverse events.- Animal Safety: No adverse events attributed to the Clip device in non-clinical testing. - Clinical Safety: 0% event rate for the primary safety endpoint. Independent physician adjudication found no events linked to the device or procedure.
Biocompatibility: Materials suitable for intended use.- All materials suitable for intended use, tested per ISO 10993-1.
Functional Performance (e.g., proper LAA occlusion).- Non-clinical testing demonstrated complete appendage exclusion. - Post-implant verification confirmed no communication between LAA and left atrium in all animals (non-clinical).
Substantial Equivalence to Predicate Devices- Concluded to be "as safe, as effective and performs as well as or better than the predicate device," based on technological characteristics, non-clinical, and clinical data.

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study Test Set Sample Size: 70 subjects
  • Data Provenance: The document explicitly states "a controlled clinical study," implying prospective data collection in humans. The country of origin is not specified, but the submission to the FDA for a US market device implies it would be relevant to the US regulatory context.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document states: "Independent physician adjudication has demonstrated that there were no events that were in any way linked to the device or the procedure."
  • Number of Experts: Not specified.
  • Qualifications of Experts: Described as "Independent physician adjudication," but specific qualifications (e.g., specialty, years of experience) are not provided.

4. Adjudication Method for the Test Set

  • The document mentions "Independent physician adjudication" for safety events. The specific method (e.g., 2+1, 3+1, none) is not detailed. It implies a review process by independent physicians to determine the linkage of reported events to the device or procedure.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done.
  • This device is an implantable medical device (a clip for LAA exclusion), not an AI-powered diagnostic or assistive tool. Therefore, the concept of human readers improving with AI assistance does not apply to this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, this is not applicable. As mentioned above, this is an implantable medical device, not an algorithm. The performance is the direct result of the device's physical interaction with tissue, not an independent algorithm.

7. The Type of Ground Truth Used

  • Clinical Efficacy/Exclusion: Visual confirmation (intra-operative) and likely imaging modalities (e.g., transesophageal echocardiography) at 3-month follow-up to confirm no communication between the LAA and the left atrium.
  • Clinical Safety: Adverse event reporting and subsequent "Independent physician adjudication" to determine causality.
  • Non-Clinical (Animal) Efficacy/Exclusion: Direct observation, post-implant verification for communication, and likely pathological examination.

8. The Sample Size for the Training Set

  • The document describes "non-clinical testing" on "acute and chronic models ranging from 7 days to 90 days." It refers to "all animals" being tested. However, a specific number for the animal "training" or "pre-clinical" sample size is not explicitly stated beyond "all animals."
  • There's no mention of a separate training set for algorithm development, as this is not an AI device.

9. How the Ground Truth for the Training Set Was Established

  • For the non-clinical (animal) testing, the ground truth for LAA exclusion and safety was established through:
    • Direct observation during surgical procedures.
    • Post-implant verification (likely imaging and/or direct inspection) to confirm no communication between the LAA and left atrium.
    • Observation for adverse events.
    • The document implies that the "training set" (animal data) served as foundational evidence before human clinical trials.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "AtriCure" in a stylized font. The letters "A," "t," "r," and "i" are in a bold, sans-serif font, while the "Cure" portion of the word is outlined. There is a solid black circle inside the "C" of "Cure." The "e" in "Cure" has a small trademark symbol next to it.

KC093679

ﯿﻨﮧ ﮨ

510(k) Summary

JUN 1 0 2010

General Information

ClassificationClass 2
Trade nameAtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip
Common nameImplantable Clip
Classification NameClip, Implantable (21 CFR 878.4300, Product Code FZP)
ManufacturerAtriCure, Inc.6217 Centre Park Dr.West Chester, OH 45069P: 513-755-4100F: 513-755-4108
ContactJames Lucky, RACVice President Quality Assurance and Regulatory Affairs
Date of SubmissionNovember 27, 2009

Intended Use

The AtriClip LAA Exclusion System is indicated for the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.

Predicate Devices

The predicate devices for the AtriClip LAA Exclusion System are the Weck Hem-O-Lok® Ligating Clip and Clip Applier (K030311), Medtronic VNUS U-Clip and Applier (K031623), Tyco AutoSuture TA and GIA Staplers (K032696), Power Medical SurgASSIST® Straight Linear 4 Row No Knife DLUs with Reloads (K040398), Demetech Braided Nonabsorbable Polyester Suture and needle driver (K023030) and Gore SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material (K043056).

Device Description

The AtriClip LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier, and a Selection Guide to aid in appropriate Clip size selection. The frame assembly of the implantable Clip consists of two springs connecting two opposing tubes which are covered with pressure pads. This assembly is covered with a knit fabric. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure occlusion of the left atrial appendage (LAA).

{1}------------------------------------------------

The Clip is available in the following lengths to accommodate different sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm.

The Clip Applier is a disposable device with a handle, shaft, and an end effector which contains the Clip. Both sides of the Clip are free to move within the end effector when activated. Pulling the thumb lever proximally opens the Clip. The bias of the Clip allows both the clip and the lever to return to the closed position when the lever is released.

Materials

All materials in the AtriClip LAA Exclusion System are suitable for their intended use and have been used in previously cleared products. Testing was conducted in accordance with ISO 10993-1 to ensure appropriate biocompatibility of all materials.

Predicate Comparison

The technological characteristics of the device are the same as the predicate devices. The same materials are used in the AtriClip and predicate devices. The sizes of the devices are equivalent and are provided in similar ranges. The method of LAA closure, permanent approximation of the sides of the LAA, is the same for the AtriClip and the predicates. All of the devices may be repositioned on the LAA prior to deployment and once deployed exert similar pressures on the tissue.

Non-Clinical Testing

The AtriClip was tested on in both acute and chronic models ranging from 7 days to 90 days Testing presented in the 510(k) document demonstrated complete appendage duration. exclusion both acutely and long-term in all animals with no adverse events attributed to the Clip device. Post-implant verification of efficacy confirmed there was no communication between the LAA and left atrium in all animals.

Clinical Testing

The AtriClip has been evaluated for safety and efficacy in a controlled clinical study of seventy subjects. The results of the study demonstrate the excellent performance of the product in its ability to exclude the LAA in a safe and efficient manner. Efficacy has been demonstrated at a very high rate intra-procedurally and this rate has been shown to improve as further healing was achieved. The observed exclusion rate of 95.1% composite (both intra-operative and 3 month) and 98% at 3 month follow-up support that this procedure can achieve excellent efficacy results.

The safety of the device and procedure has been well established through a careful scrutiny of all events reported in the study. Independent physician adjudication has demonstrated that there were no events that were in any way linked to the device or the procedure. The primary safety endpoint event rate in this trial is 0%. These results confirm the safety of the product and the procedure. The results further support the technical comparability of the product to other currently marketed options for LAA exclusion and therefore provides the clinical support to demonstrate a substantial equivalence of the AtriClip to the identified predicate devices.

{2}------------------------------------------------

AtriCure®

Summary of Substantial Equivalence

As demonstrated by the predicate comparison of technological characteristics, data provided in the non-clinical studies and clinical testing, the AtriClip LAA Exclusion System is equivalent to the predicate products. The data presented in K093679 demonstrate that the device is as safe, as effective and performs as well as or better than the predicate device.

.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like bird with its wings spread, positioned to the right of a circular inscription. The inscription reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement around the bird. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

JUN 1 0 2010

AtriCure, Inc. c/o Mr. James Lucky Vice President of Quality Assurance and Regulatory Affairs 6217 Centre Park Drive West Chester, OH 45069

Re: K093679

AtriClip™ LAA Exclusion System w/Pre-loaded Gillnov-Cosgrove™ Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: II Product Code: FZP Dated: June 1, 2010 Received: June 2, 2010

Dear Mr. Lucky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. James Lucky

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

STATEMENT: INDICATION FOR USE

K093679 510 (k) Number:

Device Name: AtriClip LAA Exclusion System

Indications for Use:

The AtriClip LAA Exclusion System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.

Prescription Use X Over the Counter Use (Part 21 CFR 801 Subpart D) and/or (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGES IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

W.W.A.

ardiovascular Devices

510(k) Number K093679

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.