K Number
K093523
Date Cleared
2010-03-04

(111 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This is a medical glove to be worn on the hands of health care and similar personnel to prevent contamination between health care personnel and the patient's body, fluids, waste or environment, and tested for use with chemotherapy drugs.

Device Description

Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs meet all of the current specifications of ASTM D6319-00a(2005)e1, Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

The provided text describes the acceptance criteria and performance of "Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs" as part of a 510(k) submission (K093523).

Here's an analysis based on the provided information:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
General Glove Standards
DimensionsASTM D 6319-00a(2005)e1Meets
Physical PropertiesASTM D 6319-00a(2005)e1Meets
Freedom from HolesASTM D 6319-00a(2005)e1 & ASTM D 5151-06Meets
Powder-FreeASTM D 6124-06≤2 mg per glove
Biocompatibility
Primary Skin IrritationAnimal StudyPasses
Dermal Sensitization AssayAnimal StudyPasses
Chemotherapy Drug Permeation (ASTM D6978-05)
CarmustineNot explicitly stated as a minimum, but breakthrough time recorded12.4 minutes
CyclophosphamideNot explicitly stated as a minimum, but breakthrough time recorded>240 minutes
Doxorubicin HydrochlorideNot explicitly stated as a minimum, but breakthrough time recorded>240 minutes
Etoposide (Toposar)Not explicitly stated as a minimum, but breakthrough time recorded>240 minutes
5-FluorouracilNot explicitly stated as a minimum, but breakthrough time recorded>240 minutes
Paclitaxel (Toxol)Not explicitly stated as a minimum, but breakthrough time recorded>240 minutes
ThioTEPANot explicitly stated as a minimum, but breakthrough time recorded19.6 minutes
CisplatinNot explicitly stated as a minimum, but breakthrough time recorded>240 minutes
DacarbazineNot explicitly stated as a minimum, but breakthrough time recorded>240 minutes

Note on Chemotherapy Drug Permeation Acceptance Criteria: While specific minimum breakthrough times are not explicitly listed as "acceptance criteria" in the table within the document, the "Indications for Use" section presents these values as the performance data for which the device was "tested for use." The underlying acceptance criterion for these drugs would be defined by the ASTM D6978-05 standard for acceptable permeation, likely implying a desired "resistant" or "protective" duration, which the ">240 minutes" results aim to demonstrate. The WARNING regarding Carmustine and ThioTEPA with their lower breakthrough times indicates that these specific drugs do not meet an acceptable level of protection, hence the cautionary instruction.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes (e.g., number of gloves tested) for each of the performance tests (Dimensions, Physical Properties, Freedom from Holes, Powder-Free, Biocompatibility, and Chemotherapy Drug Permeation). It only reports the performance results based on standard test methods.

The data provenance is not explicitly stated in terms of country of origin, nor is it classified as retrospective or prospective. However, given that this is a 510(k) submission for a new device, the testing would generally be considered prospective in nature, performed specifically for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

There is no mention of experts being used to establish a "ground truth" in the context of interpretation or diagnosis for this device. This product is a physical device (examination gloves), and its performance is evaluated through standardized laboratory tests, not through expert review of clinical images or data. Therefore, the concept of qualified experts establishing ground truth for a test set is not applicable here.

4. Adjudication Method for the Test Set

Not applicable. As described above, the performance of the gloves is based on objective laboratory measurements and adherence to ASTM standards, not on subjective assessment requiring adjudication among experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This is not an imaging or diagnostic device that would typically involve human readers or MRMC studies. The device is a medical glove, and its effectiveness is determined by its physical and chemical barrier properties.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in a sense. The testing performed for the gloves is entirely "standalone" in that it evaluates the intrinsic properties of the device itself (e.g., integrity, dimensions, chemical resistance) without human interaction or a "human-in-the-loop" component for its functional assessment. There is no algorithm involved.

7. The Type of Ground Truth Used

The "ground truth" for this device is based on:

  • Standardized Test Methods and Specifications: Adherence to defined ASTM standards (e.g., ASTM D 6319-00a(2005)e1 for dimensions and physical properties, ASTM D 5151-06 for freedom from holes, ASTM D 6124-06 for powder residue, and ASTM D6978-05 for chemotherapy drug permeation).
  • Established Biological Safety Standards: Demonstrated biocompatibility through "Animal Study" results (Primary Skin Irritation and Dermal Sensitization Assay).

These standards and test results serve as the objective "truth" against which the device's performance is measured.

8. The Sample Size for the Training Set

Not applicable. This device does not involve a "training set" as it is not an AI/ML algorithm or a diagnostic system that learns from data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of medical device.

{0}------------------------------------------------

K093523

Image /page/0/Picture/1 description: The image shows the word "Ansell" in a bold, sans-serif font. The word is underlined with a thick, black line. The font is a simple, blocky style, and the overall design is clean and straightforward. The image is likely a logo or branding element for the company Ansell.

MAR ~ 4 2010

1635 Industrial Road Dothan, AL 36303 Tel: (334) 615-2563 Fax: (334) 615-2574

510(k) SUMMARY

  • February 4, 2010 1) Date:
    1. Submitter:

February 4, 2010

  • Ansell Healthcare Products LLC 1635 Industrial Road Dothan, AL 36303
  • Cynthia A. Ingram, Regulatory Affairs Manager, Americas 3) Contact Information: (334) 615-2563 Fax: (334) 615-2568 Telephone:
    1. Name of Device:
Trade Name:Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs
Common Name:Patient Examination Gloves
Classification Name:Glove, Patient Examination, Nitrile
  1. Legally Marketed Device to Which Equivalency is being Claimed:

Device Name: Non-Sterile Powder-Free Blue Color Nitrile Examination Gloves 510(k) Number: K031580

    1. Identification of the Device:
      Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs meet all of the requirements of ASTM D 6319-00a(2005)e1.
    1. Description of the Device:
      Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs meet all of the current specifications of ASTM D6319-00a(2005)e1, Standard Specification for Nitrile Examination Gloves for Medical Application.
    1. Intended Use of the Device:
      This is a medical glove to be worn on the hands of health care and similar personnel to prevent contamination between health care personnel and the patient's body, fluids, waste or environment, and tested for use with chemotherapy drugs.

Chemotherapy Drug Permeation (average breakthrough detection time in minutes) (ASTM D6978-05)

*Carmustine12.4
Cyclophosphamide>240
Doxorubicin Hydrochloride>240
Etoposide (Toposar)>240
5-Fluorouracil>240
Paclitaxel (Toxol)>240
*ThioTEPA19.6
Cisplatin>240
Dacarbazine>240

*WARNING: Do not use with Carmustine or ThioTEPA.

1

{1}------------------------------------------------

K093523

    1. Summary of Technological Characteristics of the Device:
      Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs are summarized with the following technological characteristics compared to ASTM or equivalent standards.
CharacteristicsStandardDevice Performance
DimensionsASTM D 6319-00a(2005)e1Meets
Physical PropertiesASTM D 6319-00a(2005)e1Meets
Freedom fromHolesASTM D 6319-00a(2005)e1ASTM D 5151-06Meets
Powder-FreeASTM D 6124-06≤2 mg per glove
BiocompatibilityPrimary Skin Irritation(Animal Study)Dermal Sensitization Assay(Animal Study)PassesPasses
  1. Substantial Equivalence Based on Assessment of Non-Clinical Performance Data:

The performance test data of the non-clinical tests are the same as mentioned immediately above.

  1. Substantial Equivalence Based on Assessment of Clinical Performance Data:

Clinical data is not needed for medical gloves or for most devices cleared by the 510(k) process.

  1. Conclusion:

It is concluded that Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs are as safe, as effective, and perform as well as the glove performance standards referenced in Section 8 above and therefore meet:

ASTM listed standards, FDA hole requirements, and labeling claims for the product.

This device is substantially equivalent to currently marketed devices.

This summary will include any other information reasonably deemed necessary by the FDA.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three wing segments, representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Cynthia A. Ingram Regulatory Affairs Manager Ansell Healthcare Products LLC 1635 Industrial Road Dothan, Alabama 36303

MAR - 4 2010

Re: K093523

Trade/Device Name: Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs Regulation Number: 21CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: February 4, 2010 Received: February 12, 2010

Dear Ms. Ingram:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Ingram

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reporta Problem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Susan Kimes

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K093523

3.0 Indications for Use Statement:

INDICATIONS FOR USE

510(k) Number (if known):

Device Name: Micro-Touch® Nitrile Powder-Free Blue Examination Gloves, Tested for Use with Chemotherapy Drugs

Indications For Use:

This is a medical glove to be worn on the hands of health care and similar personnel to prevent contamination between health care personnel and the patient's body, fluids, waste or environment, and tested for use with chemotherapy drugs.

Chemotherapy Drug Permeation (average breakthrough detection time in minutes) (ASTM D6978-05)

*Carmustine12.4
Cyclophosphamide>240
Doxorubicin Hydrochloride>240
Etoposide (Toposar)>240
5-Fluorouracil>240
Paclitaxel (Toxol)>240
*ThioTEPA19.6
Cisplatin>240
Dacarbazine>240

* WARNING: Do not use with Carmustine or ThioTEPA.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Elizabet F. Hamin-Wells

(Division Sign-Off) (Division Sign-Om)
Division of Anesthesiology, General Hospital Infection Control. Dental Devices

510(k) Number: K093523

.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.