The device description and performance studies focus on the physical properties and vacuum performance of a rigid plastic cannula, with no mention of software, algorithms, or data processing that would indicate AI/ML.
No
The device, Rigid Cannulae, is used for uterine aspiration/evacuation, which is a procedure rather than for therapy. While the procedure might be therapeutic in effect, the device itself acts as a surgical instrument.
No
The device is described as a tool for uterine aspiration/evacuation, performing a therapeutic or interventional function, not a diagnostic one.
No
The device description explicitly states it is an injection molded plastic device (Rigid Cannulae), which is a physical hardware component, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for "uterine aspiration/uterine evacuation" and "rapid transcervical aspiration of the uterine cavity". This is a surgical or procedural intervention performed directly on the patient's body.
- Device Description: The device is a "Rigid Cannulae" which is a physical instrument used for aspiration.
- Lack of IVD Characteristics: IVD devices are used to examine specimens derived from the human body (like blood, urine, tissue) in vitro (outside the body) to provide information for diagnosis, monitoring, or screening. This device does not perform any such analysis of specimens.
The device is a surgical instrument used for a medical procedure, not a diagnostic test performed on a sample.
N/A
Intended Use / Indications for Use
For uterine aspiration/uterine evacuation in obstetric and gynecologic patients. Indications for use are rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy.
Product codes (comma separated list FDA assigned to the subject device)
HGH
Device Description
The Rigid Cannulae are injection molded plastic devices approximately 190mm in length, made from a styrenic copolymer manufactured in sizes 6mm. 7mm. 8mm. 9mm. 10mm, 11mm, and 12mm (outer diameter), straight and curved.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
uterine cavity
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench Testing was performed to compare the physical strength of the materials (three point bend and compression testing) as well as vacuum testing (integrity and performance).
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 884.5070 Vacuum abortion system.
(a)
Identification. A vacuum abortion system is a device designed to aspirate transcervically the products of conception or menstruation from the uterus by using a cannula connected to a suction source. This device is used for pregnancy termination or menstrual regulation. This type of device may include aspiration cannula, vacuum source, and vacuum controller.(b)
Classification. Class II (performance standards).
0
KO93508
- 510(k) Summary
Submitters Name: Address:
Doranne Frano WomanCare Global 300 Market Street Suite 134 Chapel Hill, NC 27516 919 442-2621 Susanne Parks November 6, 2009
Phone Number: Contact Person: Date Summary Prepared:
Device Name: Rigid Uterine Cannulae Common Name: Uterine Cannulae Classification Name: Vacuum Abortion System, Regulation Number (884.5070) Product Code: Cannulae Suction Uterine (HGH), Class II device
Establishment Registration Number: 3008007615
Identification of Substantially Equivalent Device: Berkeley VC10-Vacuum Curettage System and Accessories (K030935) and Synevac Vacuum Curettage System 10 (K813282). These 510(k) applications include vacuum pumps as well as the uterine suction cannulae and accessories.
Description of Device: The Rigid Cannulae are injection molded plastic devices approximately 190mm in length, made from a styrenic copolymer manufactured in sizes 6mm. 7mm. 8mm. 9mm. 10mm, 11mm, and 12mm (outer diameter), straight and curved.
Intended Use: For uterine aspiration/uterine evacuation in obstetric and gynecologic patients. Indications for use are rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy.
Comparison to Predicate: The Rigid Cannulae and the predicate device are both plastic injection molded devices. They share the same design. They both have a single scoop end and similar dimensional features. The raw materials used to produce the parts are similar. Both are attached to a vacuum source to perform uterine aspiration procedures. Bench Testing was performed to compare the physical strength of the materials (three point bend and compression testing) as well as vacuum testing (integrity and performance).
MAR - 4 2010
1
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
MAR - 4 2010
:--.
Ms. Susanne Parks Director Regulatory and Logistics WomanCare Global 300 Market Street, Suite 134 CHAPEL HILL NC 27516
Re: K093508
Trade/Device Name: Rigid Uterine Cannulae, Curved and Straight Regulation Number: 21 CFR §884.5070 Regulation Name: Vacuum abortion system Regulatory Class: II Product Code: HGH Dated: February 1, 2010 Received: February 3, 2010
Dear Ms. Parks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devise-Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
2
Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
3
4. Indications for Use Statement
Device Name:
Rigid Uterine Cannulae
KO93508
Indications for Use: For rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy.
Prescription Use X ___________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please do not write below this line-continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation(ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, and
Radiological Devices
510(k) Number K093508