K Number
K093493
Manufacturer
Date Cleared
2010-07-27

(259 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aesculap Sterilcontainer S is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and disposable polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container has been validated with stainless steel lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the STERRAD 200, STERRAD NX (Standard cycle and Advanced cycle), and STERRAD 100NX (Standard cycle and Flex cycle). The Sterilcontainer S System includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

The Sterilcontainer S is recommended for surface and lumens:

STERRAD 200, stainless steel lumens ≥ 3mm I.D. x ≤ 400mm L STERRAD NX standard cycle, stainless steel lumens ≥ 2mm I.D. x ≤ 400mm L STERRAD NX advanced cycle, stainless steel lumens ≥ 1mm I.D. x ≤ 500mm L STERRAD 100NX standard cycle, stainless steel lumens ≥ 0.7mm I.D. x ≤ 500mm L STERRAD 100NX flex cycle, porous lumens (flexible endoscope) ≥ 1mm I.D. x ≤ 850mm L

30 days of shelf life testing has been conducted for the Sterilicontainer S.

The attached tables identify which products may be sterilized in which STERRAD sterilization cycle.

Device Description

The Aesculap Sterilcontainer S is designed as a container system that will allow for sterilization and storage of other medical devices. This container is designed to be compatible for use with the STERRAD sterilization processes [200, NX (Standard cycle and Advanced cycle), and 100NX (Standard cycle and Flex cycle). The container is made from non-anodized Aluminum and utilizes a disposable (single use) polypropylene filter. The Sterilcontainer S System includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Aesculap Sterilcontainer S, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated or Implied)Reported Device Performance
Sterilization Compatibility (STERRAD 200): Maintain sterility for enclosed devices and allow sterilization for lumens: ≥ 3mm I.D. x ≤ 400mm L (stainless steel).The SterilContainer S has been "fully validated" for STERRAD 200. Validated for full, ¾, ½, and ¼ size containers with 5 stainless steel lumens per system. Validated for Extra Long Mini container with 2 stainless steel lumens per system.
Sterilization Compatibility (STERRAD NX Standard Cycle): Maintain sterility for enclosed devices and allow sterilization for lumens: ≥ 2mm I.D. x ≤ 400mm L (stainless steel).The SterilContainer S has been "fully validated" for STERRAD NX (Standard cycle). Validated for full, ¾, ½, and ¼ size containers with 5 stainless steel lumens per system. Validated for Extra Long Mini container with 2 stainless steel lumens per system.
Sterilization Compatibility (STERRAD NX Advanced Cycle): Maintain sterility for enclosed devices and allow sterilization for lumens: ≥ 1mm I.D. x ≤ 500mm L (stainless steel).The SterilContainer S has been "fully validated" for STERRAD NX (Advanced cycle). Validated for full, ¾, ½, and ¼ size containers with 5 stainless steel lumens per system. Validated for Extra Long Mini container with 2 stainless steel lumens per system.
Sterilization Compatibility (STERRAD 100NX Standard Cycle): Maintain sterility for enclosed devices and allow sterilization for lumens: ≥ 0.7mm I.D. x ≤ 500mm L (stainless steel).The SterilContainer S has been "fully validated" for STERRAD 100NX (Standard cycle). Validated for full, ¾, ½, and ¼ size containers with 5 stainless steel lumens per system. Validated for Extra Long Mini container with 2 stainless steel lumens per system. (Note in text seems to have a typo for Extra Long Mini, but it's likely intended to be 2 lumens like other cycles)
Sterilization Compatibility (STERRAD 100NX Flex Cycle): Maintain sterility for enclosed devices and allow sterilization for porous lumens (flexible endoscope): ≥ 1mm I.D. x ≤ 850mm L.The SterilContainer S has been "fully validated" for STERRAD 100NX (Flex cycle). Validated for full, ¾, and ½ size containers with 1 PTFE/PE lumen per system.
Sterility Maintenance (Shelf Life): Maintain sterility of the enclosed device until used for a specified duration."30 days of shelf life testing has been conducted for the Sterilicontainer S." (No explicit pass/fail criteria and result provided beyond simply stating it was conducted).
Compatibility with various instruments: Validate with stainless steel lumens, hinged, and knurled instruments."This container has been validated with stainless steel lumens, hinged, and knurled instruments."
Overall Performance: Meet "all required testing" per "Draft Guidance for the Preparation of Premarket Notifications 510(k)'s" for sterilization containers."All required testing...was completed." "This validation was conducted in accordance with FDA guidance and available AAMI standards by a qualified testing laboratory."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state a numerical sample size for the "test set" in terms of how many containers or sterilization cycles were performed. Instead, it refers to "validation" of various container sizes with a specific number of lumens per container system (e.g., "5 stainless steel lumens per container system" for full, ¾, ½, and ¼ sizes, and "2 stainless steel lumens per container system" for Extra Long Mini). For the 100NX flex cycle, it states "1 PTFE/PE lumen per container system." This suggests that the "sample size" is the combination of container types and the specified number of lumens.
  • Data Provenance: The document explicitly states the validation was conducted by "a qualified testing laboratory." The country of origin for the data is not specified. The study is implicitly prospective as it's a validation study for a new device to demonstrate its performance against predefined criteria.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the document. The validation appears to rely on objective measures of sterilization efficacy rather than expert subjective assessment.

4. Adjudication Method for the Test Set

  • This information is not applicable as the nature of the study (sterilization container validation) does not involve human expert adjudication of results but rather objective measurements of sterility and physical compatibility.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This information is not applicable. The device is a sterilization container, not an AI-powered diagnostic tool, so an MRMC study related to human reader performance with or without AI is irrelevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

  • This information is not applicable. The device is a physical sterilization container, not an algorithm.

7. The Type of Ground Truth Used

  • The ground truth in this context would be objective laboratory measurements of sterility (e.g., absence of microbial growth on biological indicators, successful chemical indicator changes, maintenance of barrier properties, and lumen patency) and physical compatibility (e.g., fitness for use within the specified STERRAD sterilizers). The document implies validation against "FDA guidance and available AAMI standards," which would define the methods and criteria for establishing sterility.

8. The Sample Size for the Training Set

  • This information is not applicable. The Aesculap Sterilcontainer S is a physical device, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

KCA343

Page 1 of 5

B.510(k) SUMMARY (as required by 21 CFR 807.92)
Aesculap Sterilcontainer S forSTERRAD 200, STERRAD NX, and STERRAD NX100July 27, 2010
COMPANY:Aesculap®, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714JUL 2-7 2010
CONTACT:Kathy A. Racosky610-984-9291 (phone)610-791-6882 (fax)
TRADE NAME:Aesculap Sterilcontainer 'S
COMMON NAME:Sterilization Container Wrap
CLASSIFICATION NAME:Wrap, Sterilization
REGULATION NUMBER:880.6850
PRODUCT CODE:FRG

SUBSTANTIAL EQUIVALENCE

Aesculap®, Inc. believes that the SterilContainer S is substantially equivalent to:

  • Aesculap STERRAD 100S Compatible Sterilcontainer (K040865), .
  • SteriTite Universal Container System with MediTray Products for STERRAD 200, . STERRAD NX, Ozone 125L, Sterilization, and Prevacuum Steam Flash Sterilization (K080558).

DEVICE DESCRIPTION

The Aesculap Sterilcontainer S is designed as a container system that will allow for sterilization and storage of other medical devices. This container is designed to be compatible for use with the STERRAD sterilization processes [200, NX (Standard cycle and Advanced cycle), and 100NX (Standard cycle and Flex cycle). The container is made from non-anodized Aluminum and utilizes a disposable (single use) polypropylene filter. The Sterilcontainer S System includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

{1}------------------------------------------------

INDICATIONS FOR USE:

The Aesculap Sterilcontainer S is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and disposable polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container has been validated with stainless steel lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the STERRAD 200, STERRAD NX (Standard cycle and Advanced cycle), and STERRAD 100NX (Standard cycle and Flex cycle). The Sterilcontainer S System includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

The Sterilcontainer S is recommended for surface and lumens:

STERRAD 200, stainless steel lumens ≥ 3mm I.D. x ≤ 400mm L STERRAD NX standard cycle, stainless steel lumens ≥ 2mm I.D. x ≤ 400mm L STERRAD NX advanced cycle, stainless steel lumens ≥ 1mm I.D. x ≤ 500mm L STERRAD 100NX standard cycle, stainless steel lumens ≥ 0.7mm I.D. x ≤ 500mm L STERRAD 100NX flex cycle, porous lumens (flexible endoscope) ≥ 1mm I.D. x ≤ 850mm L

30 days of shelf life testing has been conducted for the Sterilicontainer S.

The attached tables identify which products may be sterilized in which STERRAD sterilization cycle.

{2}------------------------------------------------

Page 3 of 5

Item #DescriptionTotal loadedcontainerweight (lbs)
JM440Full Size Perforated Bottom 90mm (4 ¼")21.46
JM441Full Size Perforated Bottom 120mm (5 ½")21.46
JM442Full Size Perforated Bottom 135mm (6")21.46
JM444Full Size Perforated Bottom 187mm (8")21.46
JM740¾ Size Perforated Bottom 90mm(4 ¼")14.42
JM741¾ Size Perforated Bottom 120mm (5 ½")14.42
JM742¾ Size Perforated Bottom 135mm (6")14.42
JM340½ Size Perforated Bottom 90mm (4 ¼")14.42
JM341½ Size Perforated Bottom 120mm (5 ½")14.42
JM342½ Size Perforated Bottom 135mm (6")14.42
JM344½ Size Perforated Bottom 187mm (8")14.42
JM094¼ Size Perforated Bottom with Lid 65mm (2 ½")14.42
JM096¼ Size Perforated Bottom with Lid 130mm (5 1/8")14.42
JM021Extra Long Mini Bottom 73mm (3")7.64
JM489Full Size Lid
JM789¾ Size Lid
JM389½ Size Lid
See JM094 & JM096¼ Size Lid
JM020Extra Long Mini Lid

TABLE 1 ilContainer & Comp tible Containers in STERBAD 200

Note: Full, ¾, ½, and ¼ size containers have been validated with 5 stainless steel lumens per container system. The Extra Long Mini container has been validated with 2 stainless steel lumens per container system.

TABLE 2 SterilContainer S Compatible Containers in STERRAD NX Standard and Advanced Cvcle

Item #DescriptionTotal loadedcontainer weight(lbs)
JM440Full Size Perforated Bottom 90mm (4 ¼")21.46
JM441Full Size Perforated Bottom 120mm (5 ½")21.46
JM740¾ Size Perforated Bottom 90mm (4 ¼")13.85
JM741¾ Size Perforated Bottom 120mm (5 ½")13.85
JM742¾ Size Perforated Bottom 135mm (6")13.85
JM340½ Size Perforated Bottom 90mm (4 ¼")13.85
JM341½ Size Perforated Bottom 120mm (5 ½")13.85
JM094¼ Size Perforated Bottom with Lid 65mm (2 ½")13.85
JM096½ Size Perforated Bottom with Lid 130mm (5 1/8")13.85
JM021Extra Long Mini Bottom 73mm (3")7.64
JM489Full Size Lid
JM7893/4 Size Lid
JM389½ Size Lid
See JM094 & JM096¼ Size Lid
JM020Extra Long Mini Lid

Note: Full, ¾, ½, and ¼ size containers have been validated with 5 stainless steel lumens per container system. The Extra Long Mini container has been validated with 2 stainless steel lumens per container system.

{3}------------------------------------------------

Page 4 of 5

TABLE 3
SterilContainer S Compatible Containers in STERRAD 100NX
Standard Cycle

.

Item #DescriptionTotal loadedcontainerweight (lbs)
JM440Full Size Perforated Bottom 90mm (4 ¼")21.46
JM441Full Size Perforated Bottom 120mm (5 ½")21.46
JM442Full Size Perforated Bottom 135mm (6")21.46
JM444Full Size Perforated Bottom 187mm (8")21.46
JM740¾ Size Perforated Bottom 90mm(4 ¼")13.85
JM741¾ Size Perforated Bottom 120mm (5 ½")13.85
JM742¾ Size Perforated Bottom 135mm (6")13.85
JM340½ Size Perforated Bottom 90mm (4 ¼")13.85
JM341½ Size Perforated Bottom 120mm (5 ½")13.85
JM342½ Size Perforated Bottom 135mm (6")13.85
JM344½ Size Perforated Bottom 187mm (8")13.85
JM094¼ Size Perforated Bottom with Lid 65mm (2 ½")13.85
JM096¼ Size Perforated Bottom with Lid 130mm (5 1/8")13.85
JM021Extra Long Mini Bottom 73mm (3")7.64
JM489Full Size Lid
JM789¾ Size Lid
JM389½ Size Lid
See JM094 & JM096¼ Size Lid
JM020Extra Long Mini Lid

Note: Full, ¾, ½, and ¼ size containers have been validated with 5 stainless steel lumens per container system. The Extra Long Mini container has been validated with 2 stainless steel lumens per container system.

TABLE 4 SterilContainer S Compatible Containers in STERRAD 100NX Flex Cycle

Item #DescriptionTotal loadedcontainerweight (lbs)
JM440Full Size Perforated Bottom 90mm (4 ¼")9.67
JM441Full Size Perforated Bottom 120mm (5 ½")9.67
JM442Full Size Perforated Bottom 135mm (6")9.67
JM444Full Size Perforated Bottom 187mm (8")9.67
JM740¾ Size Perforated Bottom 90mm(4 ¼")9.67
JM741¾ Size Perforated Bottom 120mm (5 ½")9.67
JM742¾ Size Perforated Bottom 135mm (6")9.67
JM340½ Size Perforated Bottom 90mm (4 ¼")9.67
JM341½ Size Perforated Bottom 120mm (5 ½")9.67
JM342½ Size Perforated Bottom 135mm (6")9.67
JM344½ Size Perforated Bottom 187mm (8")9.67
JM489Full Size Lid
JM789¾ Size Lid
JM389½ Size Lid

Note: Full, ¾, and ½ size containers have been validated with 1 PTFE/PE lumen per container system.

{4}------------------------------------------------

TECHNOLIGICAL CHARACTERISTICS(compared to predicate(s)}

The Aesculap Sterilcontainer S is compatible for use in STERRAD 200, STERRAD NX (Standard cycle and Advanced cycle), and STERRAD 100NX (Standard cycle and Flex cycle). The Sterilcontainer is offered in similar shapes and sizes as the predicate devices. The material used for the STERRAD Compatible Aesculap Sterilcontainer S is the same as that used to manufacture the predicate devices.

PERFORMANCE DATA

All required testing per "Draft Guidance for the Preparation of Premarket Notifications 510(k)'s" for Aesculap Sterilcontainer S was completed. The Aesculap Sterilcontainer S was fully validated for STERRAD 200, STERRAD NX (Standard cycle and Advanced cycle), and STERRAD 100NX (Standard cycle and Flex cycle). Sterilization Systems. This validation was conducted in accordance with FDA guidance and available AAMI standards by a qualified testing laboratory.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized emblem resembling a caduceus, with three wave-like lines representing the department's mission of promoting health and well-being. The emblem is accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Kathy Racosky Regulatory Affairs Specialist AESCULAP, Incorporated 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K093493

Trade/Device Name: AESCULAP Sterilcontainer S Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: June 29, 2010 Received: June 30, 2010

JUL 2-7 2010

Dear Ms. Racosky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Ms. Racosky

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Susan Luong

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE STATEMENT

JUL 27 2010

K093493

510(k) Number:

Device Name: Aesculap Sterilcontainer S

Indications for Use:

The Aesculap Sterilcontainer S is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and disposable polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This confainer has been validated with stainless steel lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the STERRAD 200, STERRAD NX (Standard cycle and Advanced cycle), and STERRAD 100NX (Standard cycle and Flex cycle). The Steriicontainer S System includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

The Stericontainer S is recommended for surface and lumens:

STERRAD 200, stainless steel lumens ≥ 3mm I.D. x ≤ 400mm L STERRAD NX standard cycle, stainless steel lumens ≥ 2mm 1.D. x ≤ 400mm L STERRAD NX advanced cycle, stainless steel lumens ≥ 1mm 1.D. x ≤ 500mm L STERRAD 100NX standard cycle, stainless steel lumens ≥ 0.7mm |.D. x ≤ 500mm L STERRAD 100NX flex cycle, porous lumens (flexible endoscope) ≥ 1mm I.D. x ≤ 850mm L

30 days of shelf life testing has been conducted for the Sterilicontainer S.

The attached tables identify which products may be sterilized in which STERRAD sterilization cycle.

Prescription Use

and/or . Over-the-Counter Use __

(per 21 CFR 801 Subpart D)

(per 21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 3

(Division Sign-Uff) (Division Sign-On)
Division of Anesthesiology, General Hospital

{8}------------------------------------------------

Item #DescriptionTotal loadedcontainerweight (lbs)
JM440Full Size Perforated Bottom 90mm (4 ¼")21.46
JM441Full Size Perforated Bottom 120mm (5 ½")21.46
JM442Full Size Perforated Bottom 135mm (6")21.46
JM444Full Size Perforated Bottom 187mm (8")21.46
JM740¾ Size Perforated Bottom 90mm(4 ¼")14.42
JM741¾ Size Perforated Bottom 120mm (5 ½")14.42
JM742¾ Size Perforated Bottom 135mm (6")14.42
JM340½ Size Perforated Bottom 90mm (4 ¼")14.42
JM341½ Size Perforated Bottom 120mm (5 ½")14.42
JM342½ Size Perforated Bottom 135mm (6")14.42
JM344½ Size Perforated Bottom 187mm (8")14.42
JM094¼ Size Perforated Bottom with Lid 65mm (2 ½")14.42
JM096¼ Size Perforated Bottom with Lid 130mm (5 1/8")14.42
JM021Extra Long Mini Bottom 73mm (3")7.64
JM489Full Size Lid
JM789¾ Size Lid
JM389½ Size Lid
See JM094 & JM096¼ Size Lid
JM020Extra Long Mini Lid

TABLE 1 SterilContainer S Compatible Containers in STERRAD 200

Note: Full, ¾, ½, and ¼ size containers have been validated with 5 stainless steel lumens per container system. The Extra Long Mini container has been validated with 2 stainless steel lumens per container system.

TABLE 2 SterilContainer S Compatible Containers in STERRAD NX Standard and Advanced Cycle

Item #DescriptionTotal loadedcontainer weight(lbs)
JM440Full Size Perforated Bottom 90mm (4 1/4")21,46
JM441Full Size Perforated Bottom 120mm (5 1/2")21.46
JM7403/4 Size Perforated Bottom 90mm (4 1/4")13.85
JM7413/4 Size Perforated Bottom 120mm (5 1/2")13.85
JM7423/4 Size Perforated Bottom 135mm (6")13.85
JM3401/2 Size Perforated Bottom 90mm (4 1/4")13.85
JM3411/2 Size Perforated Bottom 120mm (5 1/2")13.85
JM0941/4 Size Perforated Bottom with Lid 65mm (2 1/2")13.85
JM0961/4 Size Perforated Bottom with Lid 130mm (5 1/8")13.85
JM021Extra Long Mini Bottom 73mm (3")7.64
JM489Full Size Lid
JM7893/4 Size Lid
JM389½ Size Lid
See JM094 & JM0961/4 Size Lid
JM020Extra Long Mini Lid

Note: Full, ¼, ½, and ¼ size containers have been validated with 5 stainless steel lumens per container system. The Extra Long Mini container has been validated with 2 stainless steel lumens per container system.

{9}------------------------------------------------

TABLE 3 SterilContainer S Compatible Containers in STERRAD 100NX Standard Cycle

Item #DescriptionTotal loadedcontainerweight (lbs)
JM440Full Size Perforated Bottom 90mm (4 1/4")21.46
JM441Full Size Perforated Bottom 120mm (5 1/2")21.46
JM442Full Size Perforated Bottom 135mm (6")21.46
JM444Full Size Perforated Bottom 187mm (8")21.46
JM7403/4 Size Perforated Bottom 90mm(4 1/4")13.85
JM7413/4 Size Perforated Bottom 120mm (5 1/2")13.85
JM7423/4 Size Perforated Bottom 135mm (6")13.85
JM3401/2 Size Perforated Bottom 90mm (4 1/4")13.85
JM3411/2 Size Perforated Bottom 120mm (5 1/2")13.85
JM3421/2 Size Perforated Bottom 135mm (6")13.85
JM3441/2 Size Perforated Bottom 187mm (8")13.85
JM0941/4 Size Perforated Bottom with Lid 65mm (2 1/2")13.85
JM0961/4 Size Perforated Bottom with Lid 130mm (5 1/8")13.85
JM021Extra Long Mini Bottom 73mm (3")7.64
JM489Full Size Lid
JM7893/4 Size Lid
JM3891/2 Size Lid
See JM094 & JM0961/4 Size Lid
JM020Extra Long Mini Lid

Note: Full, ¾, ½, and ¼ size containers have been validated with 5 stainless steel lumens per Note: I dir, %, 72, and 14 oize oonlainer has been validated with 2 stainless steel lumens per container system.

TABLE 4 SterilContainer S Compatible Containers in STERRAD 100NX Flex Cycle

Item #DescriptionTotal loadedcontainerweight (lbs)
JM440Full Size Perforated Bottom 90mm (4 ½")9.67
JM441Full Size Perforated Bottom 120mm (5 1/2")9.67
JM442Full Size Perforated Bottom 135mm (6")9.67
JM444Full Size Perforated Bottom 187mm (8")9.67
JM7403/4 Size Perforated Bottom 90mm(4 ½")9.67
JM7413/4 Size Perforated Bottom 120mm (5 1/2")9.67
JM7423/4 Size Perforated Bottom 135mm (6")9.67
JM340½ Size Perforated Bottom 90mm (4 ¼")9.67
JM341½ Size Perforated Bottom 120mm (5 ½")9.67
JM342½ Size Perforated Bottom 135mm (6")9.67
JM344½ Size Perforated Bottom 187mm (8")9.67
JM489Full Size Lid.
JM7893/4 Size Lid
JM389½ Size Lid

Note: Full, ¾, and ½ size containers have been validated with 1 PTFE/PE lumen per container system.

Page 3 ol 3

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).