(217 days)
Not Found
No
The summary describes a standard hysteroscopic endoscope with a camera and light source, and there is no mention of AI, ML, or image processing beyond basic video capture.
No.
The device is used for diagnostic purposes by viewing the cervical canal and uterine cavity, not for therapeutic intervention.
Yes
Explanation: The "Intended Use / Indications for Use" section explicitly states, "The FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures." Additionally, the "Device Description" section reiterates that it is "intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures."
No
The device description clearly states it is a "handheld, low-voltage camera device" consisting of a "hysteroscopic endoscope" with a "rigid shaft," "camera," "integrated light source," and a "battery cartridge." These are all hardware components, not software.
Based on the provided information, the FemEye Two™ Hysteroscope is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health. These tests are performed outside of the body.
- FemEye Two™ Function: The FemEye Two™ Hysteroscope is used to directly visualize the cervical canal and uterine cavity inside the patient's body. It's an endoscopic device used for diagnostic procedures through direct observation, not by analyzing samples taken from the body.
The device description and intended use clearly indicate it's an instrument for internal examination and visualization, which falls under the category of an endoscopic device, not an IVD.
N/A
Intended Use / Indications for Use
The FemSuite, LLC, FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:
- Abnormal Uterine Bleeding .
- Infertility and Pregnancy Wastage .
- Evaluation of Abnormal Hysterosalpingogram, ●
- Intrauterine Foreign Body .
- Amenorrhea ●
- Pelvic Pain .
Product codes (comma separated list FDA assigned to the subject device)
HIH
Device Description
The FemEye Two™ Hysteroscope is designed as a handheld, low-voltage camera device suitable for use during both in-office as well as in-hospital diagnostic procedures. The FemEye Two is a continuous flow hysteroscope intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures.
The device consists of a hysteroscopic endoscope and is provided sterile for single patient use. It consists of a small diameter, rigid shaft and a handle. The shaft contains a camera and integrated light source as well as an irrigation/working channel port. The nondisposable portion of the FemEye Two is the battery cartridge which powers the device.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
cervical canal and uterine cavity
Indicated Patient Age Range
Not Found
Intended User / Care Setting
in-office as well as in-hospital diagnostic procedures
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench testing has demonstrated that the device is in compliance with ISO 8600-3 and ISO 8600-5. In addition the bench testing included the Tensile Strength test, Torque test, Fatigue test and Flow test. The Tensile Strength Test demonstrated that joints can withstand a 5.0 Ib tensile load, the Torque Test demonstrated that the device can withstand a 1 lb- inch torque. Fatigue test demonstrated that the rotation knob can withstand multiple actuations without affecting the image functionality of the device and the Flow test verified the flow of 120 ml per minute without any leaks.
A clinical usability study was performed with FemEye Two. The results of this usability study indicate that the FemEye Two is safe and effective when operated by the intended user.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 884.1690 Hysteroscope and accessories.
(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
0
K093218
pg. 1 of 3
Section V-510(k) Summary
Device Information 5.1.
MAY 1 8 2010
Category | Comments |
---|---|
Sponsor: | Company: FemSuite, LLC |
Address: The Presidio | |
16A Funston Avenue | |
San Francisco, Ca, 94129, USA | |
Correspondence Contact Information: | Dr. Pavan Sethi, Ph.D. |
Vice President, Regulatory, Quality & Scientific | |
Affairs | |
Tel: (408) 513-7529 | |
Pavan@femsuite.com | |
Device Common Name: | Hysteroscope |
Device Classification and Code: | Class II, HIH |
Device Classification Name: | Hysteroscope and accessories |
Device Proprietary Name: | FemEye TwoTM |
5.2. Predicate Device Information
| Predicate Device | Manufacturer | Predicate
Device
Common Name | Predicate Device
Classification
and Code | Predicate
Device
Classification |
|-------------------------------------------------------------|------------------------|------------------------------------|------------------------------------------------|---------------------------------------|
| MicroSpan
Hysteroscope | Imagyn
Medical, Inc | Hysteroscope | Class II, HIH | 21CFR884.16
90 |
| Gyrus ACMI ®
Invisio ® Digital
Hysteroscope
System | Gyrus ACMI,
Inc | Hysteroscope | Class II, HIH | 21CFR884.16
90 |
5.3. Date Summary Prepared
Date Submitted: October 09, 2009
5.4. Description of Device:
The FemEye Two™ Hysteroscope is designed as a handheld, low-voltage camera device suitable for use during both in-office as well as in-hospital diagnostic procedures. The FemEye Two is a continuous flow hysteroscope intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures.
The device consists of a hysteroscopic endoscope and is provided sterile for single patient use. It consists of a small diameter, rigid shaft and a handle. The shaft contains a camera
Section V-510(k) Summary or 510(k) StatementPage
Company Confidential
V-1
1
K093218
Pg 2 of 3
and integrated light source as well as an irrigation/working channel port. The nondisposable portion of the FemEye Two is the battery cartridge which powers the device.
Intended Use 5.5.
The FemSuite, LLC, FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:
- Abnormal Uterine Bleeding .
- Infertility and Pregnancy Wastage .
- Evaluation of Abnormal Hysterosalpingogram, ●
- Intrauterine Foreign Body .
- Amenorrhea ●
- Pelvic Pain .
5.6. Comparison to Predicate Device:
Fem Suite's FemEye Two is substantially equivalent in intended use and operation to following predicate devices:
K961688 Imagyn Medical Microspan Hysteroscope
K092278 Gyrus ACMI Invisio Digital HysteroscopeSystem
The FemEye Two is identical to Imagyn Medical Microspan Hysteroscope in method of operation and intended use. The FemEye Two provides video images similar in resolution to the predicate device. Both systems provide light adequate to allow physicians to visualize the cervical and uterine cavity. Both devices have one inflow and one outflow channel for saline instillation for irrigation and distention.
The FemEye Two is different in design from Imagyn Medical Microspan Hysteroscope . The FemEye Two provides illumination of the cervical and uterine cavity using an integrated white-light LED and a fiber optic light guide, while the predicate Microspan Hysteroscope provides illumination by means of fiber optic bundles that deliver light from an external OES Xenon light source. The FemEye Two uses embedded CMOS (complimentary medical oxide semiconductor) technology to generate an image, whereas the predicate device uses fiber optic technology to generate an image. The FemEye Two™ is a sterile, single patient use, disposable device except for the battery which is provided non sterile and is reusable. The Microspan Hysteroscope is supplied non sterile and must be sterilized prior to each use. The device is reusable but is intended to be used with a single use sterile sheath. The FemEye Two is designed to be used for diagnostic hysteroscopy, where as Microspan Hysteroscope is designed to be used for operative hysteroscopy in addition to diagnostic hysteroscopy.
In addition to method of operation and intended use for diagnostic hysteroscopy, the FemEye Two is identical to Gyrus ACMI Invisio Digital Hysteroscope System in video imaging technology. Both devices are endoscopic, and transmit images through a video camera to a video monitor and use a light source for illumination. Both
Section V-510(k) Summary or 510(k) StatementPage
Company Confidential
V-2
2
K093218
Pg. 3 of 3
devices have an embedded video camera, an embedded light source, and an integrated cable.
However, the devices have different design for video imaging technology. FemEye Two has integrated LEDs in the shaft where as Gyrus ACMI Invisio Digital Hysteroscope has integrated LEDs in the handle. FemEye Two has an integrated cable to connect to external video monitor, where as in Gyrus ACMI Invisio Digital Hysteroscope System, an integrated cable connects the endoscope to the camera control unit (CCU).
These differences do not impact the intended use or optical performance of the FemEye Two. The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy.
FemSuite concludes that the devices are substantially equivalent.
5.7. Summary of Supporting Data:
Biocompatibility data demonstrates that the device is in compliance with ISO 10993.
Electrical Safety and Electromagnetic Compatibility data demonstrates that the device is in compliance with IEC 60601-2-18 IEC, 60601-1 and IEC 60601-1-2.
Bench testing has demonstrated that the device is in compliance with ISO 8600-3 and ISO 8600-5. In addition the bench testing included the Tensile Strength test, Torque test, Fatigue test and Flow test. The Tensile Strength Test demonstrated that joints can withstand a 5.0 Ib tensile load, the Torque Test demonstrated that the device can withstand a 1 lb- inch torque. Fatigue test demonstrated that the rotation knob can withstand multiple actuations without affecting the image functionality of the device and the Flow test verified the flow of 120 ml per minute without any leaks.
A clinical usability study was performed with FemEye Two. The results of this usability study indicate that the FemEye Two is safe and effective when operated by the intended user.
Company Confidential
3
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Pavan Sethi. M.D., Ph.D. Vice President, Regulatory, Quality & Scientific Affairs FemSuite, LLC 19991 7th Street East SONOMA CA 95476
Re: K093218 MAY 1 8 2010
Trade/Device Name: FemEye Two™ Regulation Number: 21 CFR §884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: May 10, 2010 Received: May 12, 2010
Dear Dr. Sethi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
4
Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/SafetyReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Section IV-Indications for Use Statement
KD93218 510(k) Number:
FemSuite, LLC, FemEye Two™ Device Name:
Indications for Use:
The FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:
- Abnormal Uterine Bleeding .
- Infertility and Pregnancy Wastage .
- Evaluation of Abnormal Hysterosalpingogram, ●
- Intrauterine Foreign Body .
- Amenorrhea
- Pelvic Pain
Prescription Use X (Per 21 CFR §801 subpart D) OR
Over-The-Counter Use (Per 21 CFR §801 subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sheldon Adelson
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number