K Number
K093218

Validate with FDA (Live)

Device Name
FEMEYE TWO
Manufacturer
Date Cleared
2010-05-18

(217 days)

Product Code
Regulation Number
884.1690
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:

  • Abnormal Uterine Bleeding .
  • Infertility and Pregnancy Wastage .
  • Evaluation of Abnormal Hysterosalpingogram, ●
  • Intrauterine Foreign Body .
  • Amenorrhea ●
  • Pelvic Pain .
Device Description

The FemEye Two™ Hysteroscope is designed as a handheld, low-voltage camera device suitable for use during both in-office as well as in-hospital diagnostic procedures. The FemEye Two is a continuous flow hysteroscope intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures.
The device consists of a hysteroscopic endoscope and is provided sterile for single patient use. It consists of a small diameter, rigid shaft and a handle. The shaft contains a camera and integrated light source as well as an irrigation/working channel port. The nondisposable portion of the FemEye Two is the battery cartridge which powers the device.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the FemEye Two™ Hysteroscope. It details the device, its intended use, comparison to predicate devices, and supporting data. However, the document does not contain information about acceptance criteria or a study that specifically proves the device meets those criteria in the way typically expected for an AI/algorithm-based medical device.

The studies mentioned are for general medical device safety and functionality, not for performance metrics related to diagnostic accuracy or AI performance. Therefore, I cannot populate the table or answer all of your specific questions related to AI/algorithm performance.

Here's a breakdown of what is available and what is not:


Acceptance Criteria and Reported Device Performance

This information is not provided in the document. The document describes a traditional medical device (hysteroscope), not an AI device with performance metrics like accuracy, sensitivity, or specificity. There are no acceptance criteria for diagnostic performance or AI algorithm output listed.

Acceptance CriterionReported Device Performance
Not applicable - no performance acceptance criteria for an AI/algorithm device are specified.Not applicable - no performance metrics for an AI/algorithm device are reported.

Study Details (as inferable from the document, with limitations for AI specifics)

The document refers to a "clinical usability study" but does not provide details on its methodology, sample size, or specific performance metrics that would be relevant to an AI model's output. The other studies are for physical device characteristics and safety.

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified for any study, particularly not for a diagnostic performance test set.
    • Data Provenance: Not specified.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not specified.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, this type of study is not mentioned as having been performed. The device is a hysteroscope for direct viewing, not an AI-assisted diagnostic tool for image analysis.
    • Effect size of human readers with/without AI assistance: Not applicable, as no AI component or related study is described.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable, as no AI algorithm is described.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not specified, as no diagnostic performance study is detailed.
  7. The sample size for the training set:

    • Not applicable, as no AI model or training process is described.
  8. How the ground truth for the training set was established:

    • Not applicable, as no AI model or training process is described.

Summary of Supporting Data (as provided in the document):

The document lists the following studies and compliance:

  • Biocompatibility Data: Demonstrates compliance with ISO 10993.
  • Electrical Safety and Electromagnetic Compatibility Data: Demonstrates compliance with IEC 60601-2-18, IEC 60601-1, and IEC 60601-1-2.
  • Bench Testing: Demonstrates compliance with ISO 8600-3 and ISO 8600-5. This included:
    • Tensile Strength test: Joints can withstand a 5.0 lb tensile load.
    • Torque test: Device can withstand a 1 lb-inch torque.
    • Fatigue test: Rotation knob can withstand multiple actuations without affecting image functionality.
    • Flow test: Verified flow of 120 ml per minute without leaks.
  • Clinical Usability Study: Results indicate the FemEye Two is "safe and effective when operated by the intended user." No further details on this study (e.g., sample size, specific outcomes, metrics of "effectiveness") are provided.

Conclusion: The provided text describes a traditional hysteroscopic device that is cleared based on mechanical, electrical, biocompatibility, and general usability testing, along with substantial equivalence to predicate devices. It does not involve artificial intelligence or algorithms requiring diagnostic performance metrics or studies as outlined in your questions for AI-based devices.

{0}------------------------------------------------

K093218
pg. 1 of 3

Section V-510(k) Summary

Device Information 5.1.

MAY 1 8 2010

CategoryComments
Sponsor:Company: FemSuite, LLCAddress: The Presidio16A Funston AvenueSan Francisco, Ca, 94129, USA
Correspondence Contact Information:Dr. Pavan Sethi, Ph.D.Vice President, Regulatory, Quality & ScientificAffairsTel: (408) 513-7529Pavan@femsuite.com
Device Common Name:Hysteroscope
Device Classification and Code:Class II, HIH
Device Classification Name:Hysteroscope and accessories
Device Proprietary Name:FemEye TwoTM

5.2. Predicate Device Information

Predicate DeviceManufacturerPredicateDeviceCommon NamePredicate DeviceClassificationand CodePredicateDeviceClassification
MicroSpanHysteroscopeImagynMedical, IncHysteroscopeClass II, HIH21CFR884.1690
Gyrus ACMI ®Invisio ® DigitalHysteroscopeSystemGyrus ACMI,IncHysteroscopeClass II, HIH21CFR884.1690

5.3. Date Summary Prepared

Date Submitted: October 09, 2009

5.4. Description of Device:

The FemEye Two™ Hysteroscope is designed as a handheld, low-voltage camera device suitable for use during both in-office as well as in-hospital diagnostic procedures. The FemEye Two is a continuous flow hysteroscope intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures.

The device consists of a hysteroscopic endoscope and is provided sterile for single patient use. It consists of a small diameter, rigid shaft and a handle. The shaft contains a camera

Section V-510(k) Summary or 510(k) StatementPage

Company Confidential

V-1

{1}------------------------------------------------

K093218
Pg 2 of 3

and integrated light source as well as an irrigation/working channel port. The nondisposable portion of the FemEye Two is the battery cartridge which powers the device.

Intended Use 5.5.

The FemSuite, LLC, FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:

  • Abnormal Uterine Bleeding .
  • Infertility and Pregnancy Wastage .
  • Evaluation of Abnormal Hysterosalpingogram, ●
  • Intrauterine Foreign Body .
  • Amenorrhea ●
  • Pelvic Pain .

5.6. Comparison to Predicate Device:

Fem Suite's FemEye Two is substantially equivalent in intended use and operation to following predicate devices:

K961688 Imagyn Medical Microspan Hysteroscope

K092278 Gyrus ACMI Invisio Digital HysteroscopeSystem

The FemEye Two is identical to Imagyn Medical Microspan Hysteroscope in method of operation and intended use. The FemEye Two provides video images similar in resolution to the predicate device. Both systems provide light adequate to allow physicians to visualize the cervical and uterine cavity. Both devices have one inflow and one outflow channel for saline instillation for irrigation and distention.

The FemEye Two is different in design from Imagyn Medical Microspan Hysteroscope . The FemEye Two provides illumination of the cervical and uterine cavity using an integrated white-light LED and a fiber optic light guide, while the predicate Microspan Hysteroscope provides illumination by means of fiber optic bundles that deliver light from an external OES Xenon light source. The FemEye Two uses embedded CMOS (complimentary medical oxide semiconductor) technology to generate an image, whereas the predicate device uses fiber optic technology to generate an image. The FemEye Two™ is a sterile, single patient use, disposable device except for the battery which is provided non sterile and is reusable. The Microspan Hysteroscope is supplied non sterile and must be sterilized prior to each use. The device is reusable but is intended to be used with a single use sterile sheath. The FemEye Two is designed to be used for diagnostic hysteroscopy, where as Microspan Hysteroscope is designed to be used for operative hysteroscopy in addition to diagnostic hysteroscopy.

In addition to method of operation and intended use for diagnostic hysteroscopy, the FemEye Two is identical to Gyrus ACMI Invisio Digital Hysteroscope System in video imaging technology. Both devices are endoscopic, and transmit images through a video camera to a video monitor and use a light source for illumination. Both

Section V-510(k) Summary or 510(k) StatementPage

Company Confidential

V-2

{2}------------------------------------------------

K093218
Pg. 3 of 3

devices have an embedded video camera, an embedded light source, and an integrated cable.

However, the devices have different design for video imaging technology. FemEye Two has integrated LEDs in the shaft where as Gyrus ACMI Invisio Digital Hysteroscope has integrated LEDs in the handle. FemEye Two has an integrated cable to connect to external video monitor, where as in Gyrus ACMI Invisio Digital Hysteroscope System, an integrated cable connects the endoscope to the camera control unit (CCU).

These differences do not impact the intended use or optical performance of the FemEye Two. The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy.

FemSuite concludes that the devices are substantially equivalent.

5.7. Summary of Supporting Data:

Biocompatibility data demonstrates that the device is in compliance with ISO 10993.

Electrical Safety and Electromagnetic Compatibility data demonstrates that the device is in compliance with IEC 60601-2-18 IEC, 60601-1 and IEC 60601-1-2.

Bench testing has demonstrated that the device is in compliance with ISO 8600-3 and ISO 8600-5. In addition the bench testing included the Tensile Strength test, Torque test, Fatigue test and Flow test. The Tensile Strength Test demonstrated that joints can withstand a 5.0 Ib tensile load, the Torque Test demonstrated that the device can withstand a 1 lb- inch torque. Fatigue test demonstrated that the rotation knob can withstand multiple actuations without affecting the image functionality of the device and the Flow test verified the flow of 120 ml per minute without any leaks.

A clinical usability study was performed with FemEye Two. The results of this usability study indicate that the FemEye Two is safe and effective when operated by the intended user.

Company Confidential

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Pavan Sethi. M.D., Ph.D. Vice President, Regulatory, Quality & Scientific Affairs FemSuite, LLC 19991 7th Street East SONOMA CA 95476

Re: K093218 MAY 1 8 2010

Trade/Device Name: FemEye Two™ Regulation Number: 21 CFR §884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: May 10, 2010 Received: May 12, 2010

Dear Dr. Sethi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/SafetyReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section IV-Indications for Use Statement

KD93218 510(k) Number:

FemSuite, LLC, FemEye Two™ Device Name:

Indications for Use:

The FemEye Two™ Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:

  • Abnormal Uterine Bleeding .
  • Infertility and Pregnancy Wastage .
  • Evaluation of Abnormal Hysterosalpingogram, ●
  • Intrauterine Foreign Body .
  • Amenorrhea
  • Pelvic Pain

Prescription Use X (Per 21 CFR §801 subpart D) OR

Over-The-Counter Use (Per 21 CFR §801 subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheldon Adelson

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.